Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 31 - 31
1 Nov 2016
Morellato J Louati H Bodrogi A Stewart A Papp S Liew A Gofton W
Full Access

Traditional screw fixation of the syndesmosis can be prone to malreduction. Suture button fixation however, has recently shown potential in securing the fibula back into the incisura even with intentional malreduction. Yet, if there is sufficient motion to aid reduction, the question arises of whether or not this construct is stable enough to maintain reduction under loaded conditions. To date, there have been no studies assessing the optimal biomechanical tension of these constructs. The purpose of this study was to assess optimal tensioning of suture button fixation and its ability to maintain reduction under loaded conditions using a novel stress CT model.

Ten cadaveric lower limbs disarticulated at the knee were used. The limbs were placed in a modified external fixator frame that allows for the application of sustained torsional (5 Nm), axial (500 N) and combined torsional/axial (5Nm/500N) loads. Baseline CT scans of the intact ankle under unloaded and loaded conditions were obtaining. The syndesmosis and the deltoid ligament complex were then sectioned. The limbs were then randomised to receive a suture button construct tightened at 4 kg force (loose), 8 kg (standard), or 12 kg (maximal) of tension and CT scans under loaded and unloaded conditions were again obtained. Eight previously described measurements were taken from axial slices 10 mm above the tibiotalar joint to assess the joint morphology under the intact and repair states, and the three loading conditions: a measure of posterolateral translation (a, b), medial/lateral translation (c, g), a measure of anterior-posterior translation (f), a ratio of anterior-posterior translation (d/e), an angle (Angle 1) created by a line parallel to the incisura and the axis of the fibula, and an angle (Angle 2) created between the medial surfaces of two malleoli. These measurements have all been previously described. Each measurement was taken at baseline and compared with the three loading scenarios. A repeated measures ANOVA with a Bonferroni correction for multiple comparisons was used to test for significance.

Significant lateral (g, maximum 5.26 mm), posterior (f, maximum 6.42 mm), and external rotation (angle 2, maximum 11.71°) was noted with the 4 kg repair when compared to the intact, loaded state. Significant posterior translation was also seen with the both the 8 kg and 12 kg repairs, however the incidence and magnitude was less than with the 4 kg repair. Significant overcompression (g, 1.69 mm) was noted with the 12 kg repair.

Suture button constructs must be appropriately tensioned to maintain reduction and re-approximate the degree of physiological motion at the distal tibiofibular joint. If inserted too loosely, these constructs allow for supraphysiologic motion which may have negative implications on ligament healing. These constructs also demonstrate overcompression of the syndesmosis when inserted at maximal tension however the clinical effect of this remains to be determined.