Charcot neuro-osteoarthropathy (CN) of the midfoot
presents a major reconstructive challenge for the foot and ankle
surgeon. The Synthes 6 mm Midfoot Fusion Bolt is both designed and
recommended for patients who have a deformity of the medial column
of the foot due to CN. We present the results from the first nine
patients (ten feet) on which we attempted to perform fusion of the
medial column using this bolt. Six feet had concurrent hindfoot fusion
using a retrograde nail. Satisfactory correction of deformity of
the medial column was achieved in all patients. The mean correction
of calcaneal pitch was from 6°
(-15° to +18°) pre-operatively to 16° (7° to 23°) post-operatively;
the mean Meary angle from 26° (3° to 46°) to 1° (1° to 2°); and
the mean talometatarsal angle on dorsoplantar radiographs from 27°
(1° to 48°) to 1° (1° to 3°). However, in all but two feet, at least one joint failed to fuse.
The bolt migrated in six feet, all of which showed progressive radiographic
osteolysis, which was considered to indicate loosening. Four of
these feet have undergone a revision procedure, with good radiological
evidence of fusion. The medial column bolt provided satisfactory correction
of the deformity but failed to provide adequate fixation for fusion
in CN deformities in the foot. In its present form, we cannot recommend the routine use of this
bolt. Cite this article:
We report the outcomes of 20 patients (12 men,
8 women, 21 feet) with Charcot neuro-arthropathy who underwent correction
of deformities of the ankle and hindfoot using retrograde intramedullary
nail arthrodesis. The mean age of the patients was 62.6 years (46
to 83); their mean BMI was 32.7 (15 to 47) and their median American
Society of Anaesthetists score was 3 (2 to 4). All presented with
severe deformities and 15 had chronic ulceration. All were treated
with reconstructive surgery and seven underwent simultaneous midfoot
fusion using a bolt, locking plate or a combination of both. At
a mean follow-up of 26 months (8 to 54), limb salvage was achieved
in all patients and 12 patients (80%) with ulceration achieved healing
and all but one patient regained independent mobilisation. There was
failure of fixation with a broken nail requiring revision surgery
in one patient. Migration of distal locking screws occurred only
when standard screws had been used but not with hydroxyapatite-coated
screws. The mean American Academy of Orthopaedic Surgeons Foot and
Ankle (AAOS-FAO) score improved from 50.7 (17 to 88) to 65.2 (22
to 88), (p = 0.015). The mean Short Form (SF)-36 Health Survey Physical
Component Score improved from 25.2 (16.4 to 42.8) to 29.8 (17.7
to 44.2), (p = 0.003) and the mean Euroqol EQ‑5D‑5L score improved
from 0.63 (0.51 to 0.78) to 0.67 (0.57 to 0.84), (p = 0.012). Single-stage correction of deformity using an intramedullary
hindfoot arthrodesis nail is a good form of treatment for patients
with severe Charcot hindfoot deformity, ulceration and instability
provided a multidisciplinary care plan is delivered. Cite this article:
Charcot arthropathy is a complex condition affecting diabetic patients with neuropathy. Diagnosis of acute Charcot arthropathy particularly in absence of any perceptible trauma is very challenging as clinically it can mimic osteomyelitis and cellulitis. Delay in recognition of Charcot arthropathy can result in gross instability of foot and ankle. Early diagnosis can provide an opportunity to halt the progression of disease. We report the role of SPECT /CT in the early diagnosis and elucidation of the natural progression of the disease. Our multidisciplinary team analysed the scans of neuropathic patients presented with acute red, hot, swollen foot with normal radiological findings (Eichenholtz stage 0), attending the diabetic foot clinic from 2009–2013. The patients were selected from our database, clinic and nuclear medicine records. Initial workup included the assessment of peripheral neuropathy, temperature difference, between the feet, serum inflammatory markers and weight bearing dorsoplantar, lateral and oblique x-rays. All patients had three dimensional triple Phase Bone Scan using 800Mbq 99mTc HDP followed by CT scan. Those patients with obvious radiological findings and signs of infection were excluded.Introduction:
Methods: