header advert
Results 1 - 3 of 3
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 137 - 137
1 Jan 2016
Laende E Richardson G Biddulph M Dunbar M
Full Access

Introduction

Debate over appropriate alignment in total knee arthroplasty has become a topical subject as technology allows planned alignments that differ from a neutral mechanical axis. These surgical techniques employ patient-specific cutting blocks derived from 3D reconstructions of pre-operative imaging, commonly MRI or CT. The patient-specific OtisMed system uses a detailed MRI scan of the knee for 3D reconstruction to estimate the kinematic axis, dictating the cutting planes in the custom-fit cutting blocks machined for each patient [1, 2].

The purpose of this study was to evaluate the correlation between post-operative limb alignment and implant migration in subjects receiving shape match derived kinematic alignment.

Methods

In a randomized controlled trial comparing patient-specific cutting blocks to navigated surgery, seventeen subjects in the patient specific group had complete 1 year data. They received cruciate retaining cemented total knee replacements (Triathlon, Stryker) using patient-specific cutting blocks (OtisMed custom-fit blocks, Stryker). Intra-operatively, 6–8 tantalum markers (1 mm diameter) were inserted in the proximal tibia. Radiostereometric analysis (RSA) [3, 4] exams were performed with subjects supine on post-operative day 1 and at 6 week, 3, 6, and 12 month follow-ups with dual overhead tubes (Rad 92, Varian Medical Systems, Inc., Palo Alto, CA, USA), digital detectors (CXDI-55C, Canon Inc., Tokyo, Japan), and a uniplanar calibration box (Halifax Biomedical Inc., Mabou, NS, Canada). RSA exams were analyzed in Model-based RSA (Version 3.32, RSAcore, Leiden, The Netherlands. Post-operative limb alignment was evaluated from weight-bearing long-leg films.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 136 - 136
1 Jan 2016
Laende E Richardson G Biddulph M Dunbar M
Full Access

Introduction

Surgical techniques for implant alignment in total knee arthroplasty (TKA) is a expanding field as manufacturers introduce patient-specific cutting blocks derived from 3D reconstructions of pre-operative imaging, commonly MRI or CT. The patient-specific OtisMed system uses a detailed MRI scan of the knee for 3D reconstruction to estimate the kinematic axis, dictating the cutting planes in the custom-fit cutting blocks machined for each patient. The resulting planned alignment can vary greatly from a neutral mechanical axis. The purpose of this study was to evaluate the early fixation of components in subjects randomized to receive shape match derived kinematic alignment or conventional alignment using computer navigation. A subset of subjects were evaluated with gait analysis.

Methods

Fifty-one patients were randomized to receive a cruciate retaining cemented total knees (Triathlon, Stryker) using computer navigation aiming for neutral mechanical axis (standard of care) or patient-specific cutting blocks (OtisMed custom-fit blocks, Stryker). Pre-operatively, all subjects had MRI scans for cutting block construction to maintain blinding. RSA exams and health outcome questionnaires were performed post-operatively at 6 week, 3, 6, and 12 month follow-ups. A subset (9 subjects) of the patient-specific group underwent gait analysis (Optotrak TM 3020, AMTI force platforms) one-year post-TKA, capturing three dimensional (3D) knee joint angles and kinematics. Principal component analysis (PCA) was applied to the 3D gait angles and moments of the patient-specific group, a case-matched control group, and 60 previously collected asymptomatic subjects.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 138 - 138
1 Jan 2016
Laende E Dunbar M Richardson G Biddulph M
Full Access

Introduction

The dual mobility design concept for acetabular liners is intended to reduce the risk of dislocation and increase range of motion, but the wear pattern of this design is unclear and may have implications in implant fixation. Additionally, the solid back cups do not have the option for supplementary screw fixation, providing an additional smooth articulating surface for the liner to move against. The objective of this study was to assess cup fixation by measuring implant migration. A secondary objective was to evaluate the mobile bearing motion after rotating the hip.

Methods

Thirty subjects were recruited in a consecutive series prospective study and received Anatomic Dual Mobility (Stryker Orthopedics) uncemented acetabular components with mobile bearing polyethylene liners through a direct lateral approach. Femoral stems were cemented (Exeter) or uncemented (Accolade, Stryker Orthopedics). The femur, acetabulum, and non-articulating surface of the polyethylene liner were marked with tantalum beads. Radiostereometric analysis (RSA) exams were performed post-operatively and at 6 weeks, 3, 6, months, and at 1 year. At the 1 year exam, a frog leg RSA exam was performed to assess the mobility of the cup compared to its position during a supine exam.