23 Jun 2023
Engh CA Bhal V Hopper RH
Full Access

When the Birmingham Hip Resurfacing (BHR) metal-on-metal implant system was approved by the United States Food and Drug Administration in 2006, a multicenter, prospective, post-approval study (PAS) was required. This study uses data from the PAS to investigate metal level and glomerular filtration rate (GFR) trends over the first decade in vivo.

Between October 2006 and March 2011, 290 primary BHR procedures were performed among 262 patients at 5 sites. Whole blood samples were sent to a single specialized laboratory to determine GFR, cobalt (Co) and chromium (Cr) levels. The population for this study consists of 117 unrevised unilateral patients with a mean age at surgery of 51.3±6.5 years who had pre-operative, 1-year, 4-year, 5-year and 10-year laboratory data. The mean follow-up for these patients that included 36 females was 10.1±0.2 years.

Median metal levels at 1-year increased relative to pre-operative values for Co (by a factor of 9.7 from 0.13 to 1.26 ppb, p<0.001) and Cr (by a factor of 2.5 from 0. 60 to 1.50 ppb, p<0.001). Metal levels subsequently remained relatively constant over time with a median 10-year value of 1.12 ppb for Co and 1.29 ppb for Cr. Based on 585 blood samples from all 117 patients, there was no relationship between GFR and Co (=−0.06, p=0.14) or Cr (=0.05, p=0.27) levels. However, lower pre-operative GFR values were associated with larger increases in Co at 1-year relative to the pre-operative level (=−0.26, p=0.005). There was no relationship between pre-operative GFR values and changes in Cr at 1 year (=−0.13, p=0.15).

Through the first decade in vivo, elevated whole blood metal levels for unilateral BHR patients do not appear to adversely affect GFR. However, patients with lower pre-operative GFR values tend to have larger increases in their Co level at 1-year.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 20 - 20
1 Oct 2020
Engh CA Ho H Bhal V Housman LR Masonis JL Noble JW Hopper RH Su EP
Full Access

Introduction

The BIRMINGHAM HIP Resurfacing is a metal-on-metal (MOM) hip implant system approved by the US FDA in 2006. The approval required a multicenter, prospective, post-approval study (PAS). Our purpose is to report the current minimum 10-year results.

Methods

253 patients (280 hips) had surgery between October 2006 and December 2009 at one of 5 sites. We report revisions, survivorship, EQ-5D, Harris Hip Score (HHS), radiographic findings, and metal levels including cobalt (Co) and chromium (Cr). The mean age at surgery was 51 years, 74% male, BMI 28, osteoarthritis 95%. 243 (87%) of hips have known outcome or 10-year minimum follow-up (fup). Prior to 10 years, 5 patients died, 20 hips were revised, and 37 hips did not complete 10-year fup.