Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 97 - 97
1 May 2016
Dai Y Angibaud L Harris B Gulbransen S Begin D
Full Access

Introduction

Evaluations of Computer-assisted orthopaedic surgery (CAOS) systems generally overlooked the intrinsic accuracy of the systems themselves, and have been largely focused on the final implant position and alignment in the reconstructed knee [1]. Although accuracy at the system-level has been assessed [2], the study method was system-specific, required a custom test bench, and the results were clinically irrelevant. As such, clinical interpolation/comparison of the results across CAOS systems or multiple studies is challenging. This study quantified and compared the system-level accuracy in the intraoperative measurements of resection alignment between two CAOS systems.

Materials and Methods

Computer-assisted TKAs were performed on 10 neutral leg assemblies (MITA knee insert and trainer leg, Medial Models, Bristol, UK) using System I (5 legs, ExactechGPS®, Blue-Ortho, Grenoble, FR) and System II (5 legs, globally established manufacturer). The surgeries referenced a set of pre-defined anatomical landmarks on the inserts (small dimples). Post bone cut, the alignment parameters were collected by the CAOS systems (CAOS measured alignment). The pre- and post- operative leg surfaces were scanned, digitized, and registered (Comet L3D, Steinbichler, Plymouth, MI, USA; Geomagic, Lakewood, CO, USA; and Unigraphics NX version 7.5, Siemens PLM Software, Plano, TX, USA). The alignment parameters were measured virtually, referencing the same pre-defined anatomical landmarks (baseline). The signed and unsigned measurement errors between the baseline and CAOS measured alignment were compared between the two CAOS systems (significance defined as p<0.05), representing the magnitude of measurement errors and bias of the measurement error generated by the CAOS systems, respectively.