Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_19 | Pages 16 - 16
1 Apr 2013
Shah A Badge R Joshi Y Choudhary N Sochart D
Full Access

Introduction

Negative remodelling of the femoral cortex in the form of calcar resorption due to stress shielding and cortical hypertrophy at the level of the tip of the implant, due to distal load transfer, is frequently noted following cemented total hip replacement, most commonly with composite beam implants, but also with polished double tapers. The C-stem polished femoral component was designed with a third taper running from lateral to medial across and along the entire length of the implant, with the aim of achieving more proximal and therefore more natural loading of the femur. The hoop stresses generated in the cement mantle are transferred to the proximal bone starting at the calcar, which should theoretically minimise stress-shielding and calcar resorption, as well as reducing distal load transfer, as signified by the development of distal femoral cortical hypertrophy.

Materials/Methods

We present the results of a consecutive series of 500 total hip replacements performed between March 2000 and December 2005 at a single institution, using a standard surgical technique and third generation cementing with Palacos-R antibiotic loaded cement. Data was collected prospectively and the patients remain under annual follow-up.

500 arthroplasties were performed on 455 patients with an average age of 68.3 years (23–92). 77 patients have died (73 arthroplasties) and the average duration of follow-up for the entire series is 81 months (52–124).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_19 | Pages 13 - 13
1 Apr 2013
Badge R Shah A Joshi Y Choudhary N Sochart D
Full Access

Introduction

Traditionally the use of small diameter femoral head (22mm) with the posterior approach has been perceived as an increased risk of dislocation. We present this prospective study of 400 consecutive total hip replacements performed using a 22mm femoral head and the posterior approach.

Materials/Methods

Between March 2000 and November 2005 364 patients underwent 400 total hip replacements with a small diameter 22mm head under the care of four different consultants, using a standard posterior approach. All of the femoral implants were cemented using modular C-stems (Depuy Ltd.) and all of the acetabular components were cemented flanged monobloc all-polyethylene components with long posterior wall with a third generation cementing technique. A standard posterior approach was used in all cases, with direct repair of the capsule and short external rotators.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_III | Pages 32 - 32
1 Feb 2012
Al-Shawi A Badge R Bunker T
Full Access

Ultrasound imaging has become an essential adjunct to clinical examination when assessing a patient with suspected rotator cuff pathology. With the new high-resolution portable machines it has become feasible for the shoulder surgeon to perform the scans himself in the clinic and save a great deal of time. This study was conducted to examine the accuracy of the ultrasound scans performed by a single surgeon over a period of four years (2001-2004).

The ultrasound findings were uniformly documented and collected prospectively. Out of a total of 364 scanned patients we selected 143 who ultimately received an operation and we compared the surgical findings with the ultrasound reports. The intra-operative findings included 77 full thickness supraspinatus tears, 24 partial thickness tears and 42 normal cuffs. Three full thickness tears were missed on ultrasound and reported as normal/ partially torn. Four normal/ partially torn cuffs were thought to have a full thickness tear. This presents 96.3% sensitivity and 94.3% specificity for full thickness tears. Three partial thickness tears were reported normal on ultrasound and eight normal cuffs were thought to have partial thickness tears. This presents 89% sensitivity and 93.7% specificity for partial thickness tears. The size estimation of full thickness tears was more accurate for large/massive tears (96%) than moderate (82%) and small/pinhole tears (75%). The tear sizes were more often underestimated which may partly reflect disease progression during the unavoidable time lag between scan and surgery.

We conclude that shoulder ultrasound performed by a sufficiently trained orthopaedic surgeon is a safe and reliable practice to identify rotator cuff tears.