Whilst there is a great deal of research on hip implants, few studies have looked at implant orientation and the subsequent effect upon the wear performance of a hip resurfacing. This study aimed to measure implantation angles through radiographic analysis and linear wear for retrieved acetabular cups in order to investigate possible causal links between wear and implant orientation. Seventy Birmingham Hip Resurfacing (Smith & Nephew, UK) cups with known time in vivo were analysed. Linear wear of retrieved cups were assessed using a Talyrond 290 roundness machine. Deviations from the characteristic manufactured profile, was identified as a region of wear. Polar measurements across the wear region were taken to determine wear. The linear wear rate (LWR) of a component was defined as the linear wear (μm) divided by the duration of the implant life in vivo (years). Cups which showed the wear crossing over the edge of the cup were classified as edge loaded (EL). For all non-edge loaded (NEL) cups, the wear area was within the bearing surface. Cup orientation angles were conducted for 31 cups. This was determined by superimposing BHR models of appropriate size, generated by CAD ProEngineer Wildfire 4, onto anterior-posterior x-rays. Anatomical landmarks and specific features of the BHR were used as points of reference to determine cup version and inclination angles.INTRODUCTION
MATERIALS & METHODS
Analysis of retrieved ceramic components have shown areas of localized ‘stripe wear’, which have been attributed to joint laxity and/or impingement resulting in subluxation of the head, causing wear on the edge of the cup. Studies have been conducted into the effects of mild subluxation, however few in vitro tests have looked at severe subluxation. The aim of this study was to develop a more clinically relevant subluxation protocol. Seven (Subluxation n=4; standard test n=3) of 36mm Biolox Forte (R3, Smith & Nephew) ceramic devices were tested for 0.5m cycles (mc). Two of the subluxed joints were further tested to 1 Mc. The devices were subjected to subluxation under standard testing conditions. The flex/ext was 30° and 15° respectively, with internal/external rotation of ±10°. The force was Paul type stance phase loading with a maximum load of 3 kN, and a standard ISO swing phase load of 0.3 kN at 1 Hz. The test was conducted on a ProSim hip joint wear simulator (SimSol, UK). The simulator is equipped with a novel mechanism to achieve translation of the head, to achieve subluxation. During the ISO swing phase load of 0.3kN, a controlled lateral force required for the translation of the head is applied by a cam mechanism, head retraction then occurs during heel strike. The lubricant used was new born calf serum diluted with de-ionised water to achieve average protein concentration of 20 g/l, with 0.2 wt % concentration NaN3, and changed every 250k cycles. Measurements have been taken at 0.5 & 1 mc stages.INTRODUCTION
MATERIALS & METHODS
The in vivo abduction angle and version angle of the cup were determined by superimposing the BHR models onto the radiographs (ProEngineer Wildfire 4 with ISDX II extension software) using anatomical references and specific features of the BHR.
Cup orientation: The abduction angles of the two non-edge loaded cups were 31° and 39°, and their version angles were 12 and 16° respectively. These angles were within recommended orientation for the BHR. In contrast, the adduction angles and/or version angles of all edge loaded devices were outside the recommended orientation. Their abduction angle varied from 40° to 66° and version angle from 5° to 46°. The edge loaded devices with higher inclination angles and/or higher version angels generally had higher linear wear. There is strong correlation between the cup orientation and the linear wear of the implant.
Most friction studies are conducted in a bovine serum- carboxymethylcellulose (bs-cmc) medium. Following implantation however, the joint is bathed in blood which contains macromolecules and cells. The effect of these on friction is not fully understood. A progressive radiolucent line (fig 1) observed in some low clearance resurfacings raises the concern that increased friction may be affecting component fixation. The purpose of this investigation was to study the effect of clearance on friction for a given bearing diameter in the presence of blood as lubricant.
The results of this study suggest that reduced clearance bearings have the potential to generate higher friction when blood is the lubricant. this higher friction in the low clearance bearings may produce micromotion in the early postoperative period and hamper bony ingrowth resulting in impaired fixation with long-term implications for survival.
Hip Resurfacing has always been an attractive concept for the treatment of hip arthritis in younger patients. Introduction of modern metal-on-metal hip resurfacing in 1991 in Birmingham, UK made this concept a reality. In the early years, resurfacings were used only by a few experienced surgeons. From 1997, Birmingham Hip Resurfacings (BHRs) are being widely used in younger and more active patients. A breakdown of the ages at operation in the regional NHS hospital in Birmingham during the period April 1999 to March 2004 show that the mean age of metal-metal resurfacings is 51 years and the mean age of total hip replacements is 70 years. At a 3.7 to 10.8 year follow-up (mean follow-up 5.8 years), the cumulative survival rate of metal-metal resurfacing in young active patients with osteoarthritis is 99.8%. In the long term, none of these patients were constrained to change their occupational or leisure activities as a result of the procedure. The overall revision rate of BHRs in all ages and all diagnoses is also very low (19 out of 2167 [0.88%] with a maximum follow-up of 7.5 years). Improvements in instrumentation and a minimally invasive approach developed by the senior author have made this successful device even more attractive. Although objective evidence does not support the fact that the longer approach was any more invasive than the minimal route, patient feedback shows that it is very popular with them. While minimal approach is indeed appealing, it has a steep learning curve. In the early phase of this curve, care should be taken to avoid the potential risk of suboptimal component placement which can adversely affect long-term outcome. It is true that metal-metal bearings are associated with elevated metal ion levels. In vitro studies of BHRs show that they have a period of early run-in wear. This is not sustained in the longer term. These findings are found to hold true in vivo as well, in our studies of 24- hour cobalt output and whole blood metal ion levels. Epidemiological studies show that historic metal-metal bearings are not associated with carcinogenic effects in the long-term. Metal ion levels in patients with BHRs are in the same range as the levels found in those with historic metal-metal total hip replacements.