Vol. 98-B, Issue SUPP_2 | Pages 19 - 19
1 Jan 2016
Hachiya Y Watanabe H Taniguchi S Muramatsu K Tanaka K Yoshioka A Ando K
Full Access

(Introduction)

Metal on Metal (MoM) THA bearing as the second generation Total Hip Arthroplasty, widely came into use in Japan since 2000s. And it was expected to be able to ensure long-term use, the range of motion in the joints after surgeries, and the resistance to the dislocation, because MoM THAhad been considered to be able to use the large diameter head for its lower wear. However, there are some reports that the metal wear debris from MoM bearing surface generates the elevated serum ion level in the blood and the incidence of pseudotumor may occur subsequent to the metal wear debris. We performed MoM THA for 291 joints in 234 patients in our hospital since May 2005. But now, we refrain to use this. In this time, we will describe two cases of pseudotumor after implantation of MoM THA.

(Case1) Age 64 years, female

This case showed the hip osteoarthritis on the right side, and we performed MoM THA in 2005. After that, she had slipped just prior to the three-year postoperative check-up, andstarted swelling up, hip joint pain, and the rise in CRP. Then, we performed to remove pesudotumor in 2008 and her CRP turned into negative. However, she was repeating hip dislocation several times and an MRI showed recur of pseudotumor. Finally, we changed her metal liner into a polyethylene liner in 2012.