header advert
Results 1 - 1 of 1
Results per page:
Applied filters
Hip

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 50 - 50
1 Oct 2020
Kraay MJ Bigach SD Rimnac CM Moore RD Kolevar MP Adavi P
Full Access

Introduction

The purpose of this study was to evaluate the long term changes in bone mineral density (BMD) following implantation of a low-modulus composite femoral component designed to closely match the stiffness of the proximal femur and minimize stress shielding. Specifically, we asked: 1) How does BMD in the proximal femur change with time and with Gruen zone location; 2) Does BMD in the proximal femur stabilize after two years of implantation?

Methods

We retrospectively reviewed a subgroup of sixteen patients who had preoperative and postoperative DEXA scans in an FDA multi-center prospective trial of this composite stem. Five of these sixteen patients returned for long-term DEXA scans at a mean 22.0 years post-op (range 21.2–22.6 years). BMD in the 7 Gruen zones at final follow-up was compared to immediate post-operative and 2-year follow-up values. Percentage change was calculated and change in BMD was plotted against time from immediate post-operative measurements to each subsequent follow-up.