Aims. Revision hip arthroplasty for femoral stem loosening remains challenging due to significant bone loss and deformities requiring specialized
Aims. A number of methods have been described to remove a well-fixed humeral implant as part of revision shoulder arthroplasty. These include the use of cortical windows and humeral osteotomies. The router bit extraction technique uses a high-speed router bit to disrupt the bone-implant interface. The implant is then struck in a retrograde fashion with a square-tip impactor and mallet. The purpose of this study was to determine the characteristics and frequency of the different techniques needed for the removal of a well-fixed humeral
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral
Near 70% of failures of knee arthroplastys due to septic or aseptic mobilization are complicated because of massive loss of bone stock. In these cases surgeon have to perform a salvage procedure to restore legamentous balance, articular plane direction and axes of lower limb, finally to fill bone defects. Today intrinsic biomechanical stability of revision implant is entrust to sophisticated design and materials of custom made and modular implants. Endomedullary stem has to assolve specific functions: mechanical stability contrasting stress due to the boneloss, offering support for omoplastic or spongious bone innests in femoralor tibial defects. There are paucity of study in literature about dimension and morphology of endomedullary canal, probably because of variability between periostal andendostal anatomy in each patients, specially age related. This date has conditioned production of several number of
Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. . We review the risks and causes of fracture of tapered titanium
modular
Introduction: Revision surgery for stem loosening and peri-implant bone loss of variable extent is a major challenge for orthopedic surgeons. Our strategy has been to use cementless straight stems of rectangular cross-section for revisions. To ascertain the value of this implant we analyzed the results at a minimum follow-up of 10 years. Material and Method: Between October 1991 and end 1998, 125 patients (134 hips) underwent revision surgery. Of these, 39 were males and 86 were females. Sixty-seven of the original implants were cemented and as many were cementless. Cementless
We describe the clinical and radiological results of 120 consecutive revision hip replacements in 107 patients, using the JRI Furlong hydroxyapatite-ceramic-coated femoral component. The mean age of the patients at operation was 71 years (36 to 92) and the mean length of follow-up 8.0 years (5.0 to 12.4). We included patients on whom previous revision hip surgery had taken place. The patients were independently reviewed and scored using the Harris hip score, the Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and the Charnley modification of the Merle d’Aubigné and Postel score. Radiographs were assessed by three reviewers for the formation of new bone, osteolysis, osseointegration and radiolucent lines in each Gruen zone. The mean Harris hip score was 85.8 (42 to 100) at the latest post-operative review. The mean WOMAC and Merle d’Aubigné and Postel scores were 34.5 and 14.8, respectively. The mean visual analogue score for pain (possible range 0 to 10) was 1.2 overall, but 0.5 specifically for mid-thigh pain. There were no revisions of the femoral component for aseptic loosening. There were four re-revisions, three for infection and one for recurrent dislocation. Radiological review of all the femoral components, including the four re-revisions showed stable bony ingrowth and no new radiolucent lines in any zone. Using revision or impending revision for aseptic loosening as an end-point, the cumulative survival of the femoral component at ten years was 100% (95% confidence interval 94 to 100). We present excellent medium- to long-term clinical, radiological and survivorship results with the fully hydroxyapatite-ceramic-coated femoral component in revision hip surgery.
We aimed to assess whether using long stem femoral components, with cemented distal fixation and proximal impaction grafting allows early patient mobilisation, reconstitution of the proximal femur and long term stability of fixation in patients with aseptic loosening and proximal femoral analysis. Over the past ten years 239 patients have been treated with an Elite Plus cemented long stem femoral implant, 33% with concomitant proximal impaction bone grafting. Many of the patients had co-morbidities. The average age at revision was 72 years (range 48 to 91). There was a slight female predominance. Fourteen percent of hips had been previously revised. Forty-eight patients were deceased and 22 were not available for follow-up; this left a cohort of 169 patients who were available for radiological and questionnaire review at an average of 4.5 years. According to the Paprosky grading for pre-operative bone loss 40% had moderate to severe bone loss (grade IIIb or IV). The Barrack grading was used to assess the cement mantle post-operatively with 65% showing good cementation. The Harris and O'Neill grades were used at final review to assess probability of loosening with only 8% being probably or definitely loose. The average Oxford Hip Score was 29. Mesh ± cables were required in a third of cases to allow adequate containment and pressurisation. It was generally felt that the long stem needed to be at least one third longer than the initial component. The re-revision rate was 1.2 with a 10 year survivorship analysis of 94%. The complication rate of almost 6% included periprosthetic fractures, dislocations, infection and mesh breakage. A long stem cemented femoral implant can be useful in bypassing proximal femoral deficiency in the appropriate patient.
We have reviewed the intermediate term results of 56 out of 61 consecutive Wagner
Short-stem total hip arthroplasty (THA) may have bone sparing properties, which could be advantageous in a younger population with high risk of future revision surgery. We used data from the AOANJRR, LROI and SAR to compare survival rates of primary THA, stems used in the first-time revision procedures as well as the overall survival of first-time revisions between a cohort of short-stem and standard-stem THA. Short-stem THAs (designed as a short stem with mainly metaphyseal fixation) between 2007 and 2021 were identified (n=16,258). A propensity score matched cohort (1:2) with standard THAs in each register was identified (n=32,515). The cohorts were merged into a research dataset. Overall survival at 12 years follow-up was calculated using Kaplan-Meier survival analyses.
We report the medium term results using an extensively porous coated cementless femoral stem for revision hip arthroplasty in 129 cases. 166 femoral revisions were performed using the Solution cementless stem between 1991 and 1997 in 4 hospitals within our region. 30 patients had died and 7 were lost to follow up leaving 122 patients ( 129 hips ) available for assessment. All were independently reviewed, questioned about thigh pain, and scored using the Charnley modification of the Postel-D’Aubigne Hip Score. Post operative complications and need for further surgery were noted. Radiographs were assessed to identify component subsidence, osteolysis and stress shielding. At mean 5 year follow up (range 2–8 yrs), 9 stems (7%) had been revised and a further 4 stems (3%) were subsiding. Of the remaining components, 8 stems (7%) showed fibrous union and 108 stems (92%) bone ingrowth. Mild to moderate stress shielding was common but did not seem to affect fixation. Our findings indicate that satisfactory medium term stability can be achieved using diaphyseal fixation in the mechanically or biologically proximally deficient femur. Failure due to subsidence occurs due to undersizing, occurs early and progresses.
Introduction.
We revised 24 consecutive hips with loosening of the femoral stem using impaction allograft and a cemented stem with an unpolished proximal surface. Repeated radiostereometric examinations for up to two years showed a slow rate of subsidence with a mean of 0.32 mm (−2.0 to +0.31). Fifteen cases followed for a further year showed the same mean subsidence after three years, indicating stabilisation. A tendency to retroversion of the stems was noted between the operation and the last follow-up. Retroversion was also recorded when displacement of the stem was studied in ten of the patients after two years. Repeated determination of bone mineral density showed an initial loss after six months, followed by recovery to the postoperative level at two years. Defects in the cement mantle and malalignment of the stem were often noted on postoperative radiographs, but did not correlate with the degrees of migration or displacement. After one year, increasing frequency of trabecular remodelling or resorption of the graft was observed in the greater trochanter and distal to the tip of the stem. Cortical repair was noted distally and medially (Gruen regions 3, 5 and 6). Migration of the stems was the lowest reported to date, which we attribute to the improved grafting technique and to the hardness of the graft.
We report our experience of revision total hip
replacement (THR) using the Revitan curved modular titanium fluted revision
stem in patients with a full spectrum of proximal femoral defects.
A total of 112 patients (116 revisions) with a mean age of 73.4
years (39 to 90) were included in the study. The mean follow-up
was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died
but their data were included in the survival analysis, and four
patients (4 hips) were lost to follow-up. The clinical outcome,
proximal bone regeneration and subsidence were assessed for 101
hips. The mean Harris Hip Score was 88.2 (45.8 to 100) after five years
and there was an increase of the mean Barnett and Nordin-Score,
a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7).
Five stems had to be revised (4.3%), three (2.9%) showed subsidence,
five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%)
a periprosthetic infection. . At the latest follow-up, the survival with revision of the stem
as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%)
and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic
fractures were not observed. We report excellent results with respect to subsidence, the risk
of fracture, and loosening after femoral revision using a modular
curved
Femoral impaction grafting with cancellous bone and cement is an important technique in reconstituting deficient bone stock in revision hip arthroplasty. We report the medium to long term results of 75 consecutive patients using a collarless, polished, tapered femoral stem with an average age of 68 (±11.4) years and a mean follow up of 10.5 (±2.4) years (range 6.3 to 14.1 years). The median Endoklinik pre-operative bone defect score was 3 (IQR: 2–3) with a median subsidence at 1 year of 2mm (IQR: 1–3mm). At the most recent follow-up (mean 10.5±2.4 years), the median Harris Hip Score (HHS) was 80.6 (IQR: 67.6–88.9) and median subsidence 2mm (IQR: 1–4mm). Ten-year survivorship with any further femoral operation as an endpoint was 92%. Four prostheses required further revision. Subsidence of the Exeter stem continued, albeit at a slower rate after the first year and was related to the Endoklinik pre-operative bone loss (p=0.037). The degree of subsidence at 1 year was a strong predictor of long term subsidence (p<0.001). Neither subsidence nor bone stock were related to long term outcome (HHS). There was a correlation between previous revision surgery and a poor Harris Hip Score (p=0.028) and those who had undergone previous revision surgery for infection had a higher risk of complications (p=0.048). The good long term results of this technique commend its use in revision hip arthroplasty for patients with poor femoral bone stock.
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998. The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification. The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively. Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p <
0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97).
Severe bone loss creates a challenge for fixation in femoral revision. The goal of the study was to assess reproducibility of fixation and clinical outcomes of femoral revision with bone loss using a modular, fluted, tapered distally fixing stem. 92 consecutive patients (96 hips) underwent hip revision surgery using the same design of a modular, fluted, tapered titanium stem between 1998 and 2005. Fourteen patients with 16 hips died before a 2-year follow-up. Eighty hips were followed for an average of 11.3 years (range of 8 to 13.5 years). Bone loss was classified as per Paprosky's classification, osseointegration assessed according to a modified system of Engh et al, and Harris Hip Score was used to document pain and function. Serial radiographs were reviewed by an independent observer to assess subsidence, osseointegration and bony reconstitution.Introduction:
Methods:
Periprosthetic fractures (PPF) of the femur following total hip arthroplasty represent a significant complication with a rising incidence. The commonest subtype is Vancouver B2 type, for which revision to a long uncemented tapered fluted stem is a widely accepted management. In this study we compare this procedure to the less commonly performed cement-in-cement revision. All patients undergoing surgical intervention for a Vancouver B2 femoral PPF in a cemented stem from 2008 – 2018 were identified. We collated patient age, gender, ASA score, BMI, operative time, blood transfusion requirement, change in haemoglobin (Hb) level, length of hospital stay and last Oxford Hip Score (OHS). Radiographic analysis was performed to assess time to fracture union and leg length discrepancy. Complications and survivorship of implant and patients were recorded. 43 uncemented and 29 cement-in-cement revisions were identified. There was no difference in patient demographics between groups. A significantly shorter operative time was found in the cement-in-cement group, but there was no difference in transfusion requirement, Hb change, or length of hospital stay. OHS was comparable between groups. A non-significant increase in overall complication rates was found in the revision uncemented group, with a significantly higher dislocation rate. Time of union was comparable and there were no non-unions in the cement-in-cement group. A greater degree of stem subsidence was found in the uncemented group. There was no difference in any revision surgery required in either group. Three patients in the uncemented group died in the perioperative period, compared to none in the cement-in-cement group. With appropriate patient selection, both cement-in-cement and long uncemented tapered
In relation to the conduct of this study, one or more of the authors is in receipt of a research grant from a non-commercial source.
The early revision rate in elective Total Hip Arthroplasty (THA) three years after surgery in elderly patients over 80 years is significantly lower for cemented stems in the German Arthroplasty Register (EPRD): cemented 3,1% (3.0 – 3.2) vs. uncemented 4.2% (4.1 – 4.3; p < 0.001). However, the mortality rate in elderly patients is elevated for cemented fixation. This study presents a detailed analysis of the influence of stem type and fixation on revision and mortality rate in this patient cohort. Elective primary THA cases for primary Coxarthrosis using uncemented cups from the EPRD data base were analysed (n0= 37,183). Four stem type groups were compared: cementless, cementless with collar, cementless short, and cemented. Stems with at least 300 cases at risk three years after surgery were analysed individually. The reference stem was determined as the stem with the lowest revision rate and at least 1000 cases under surveillance 3 years after surgery (n3 = 28,637). The revision rate for cemented stems (2.5% [2.2–1.81] was lower than for uncemented (4.5% [4.2–4.9]; p<0.001) and uncemented short stems (4.2% [3.1–5.7]; p=0.002). The revision rate of uncemented collared stems (2.3% [1.5–3.6]) was similar to cemented stems (p=0.89) and lower than for uncemented stems (p=0.02). One year mortality showed no sig. differences between the groups (p>0.17): cemented 3.2% [2.9–3.6], uncemented 3.4% [3.1–3.7], uncemented short 3.5% [2.5–4.9], uncemented collar 2.0% [1.2–3.2]. “Cementless” and “cementless short” stems should not be used in patients over 80 years due to the higher revision risk. If cementing should be avoided, “cementless collared” stems seem to be a good alternative combined with a tendency for a lower one year mortality rate.