Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_18 | Pages 42 - 42
14 Nov 2024
Kato K Hayashi S
Full Access

Purpose. To compare postoperative clinical outcomes between posterior cruciate ligament (PCL) retaining and resecting total knee arthroplasty (TKA) using same cruciate-substituting (CS) inserts, and to elucidate the clinical relevance of the residual PCL in cruciate-retaining TKA, considering intraoperative influence factors, such as the posterior tibial slope, posterior condylar offset, joint gap, joint balance, and joint laxity. Methods. A total of 64 consecutive knees (44 patients) were enrolled in this study and divided into following two groups: 39 knees underwent PCL-retaining TKA group (CR group), and 25 underwent PCL-resecting TKA group (CS group). Preoperative patients’ demographic data and one-year postoperative clinical outcomes including range of motion, the Knee Injury and Osteoarthritis Outcome Score (KOOS), the Japanese Orthopaedic Association (JOA) score, and Forgotten Joint Score-12 (FJS-12) were compared between two groups. Results. Regarding range of motion, the average preoperative ROM was -14.3/120.0 degrees in the CR group and improved to -2.4/118.9 degrees postoperatively. In the CS group, the average preoperative ROM was -7.5/130 degrees and changed to -2.2/122.4 degrees postoperatively. There was no significant difference in the postoperative ROM between the groups (P=0.16). The KOOS (from 47.1 to 69.5 in CR group; from 41.1 to 70.8 in CS group) and JOA scores (from 59.2 to 76.9 in CR group; from 55.6 to 80.8 in CS group) were significantly improved postoperatively in both groups (P < 0.01). However, there was no significance in these postoperative scores between two groups (P = 0.09). There was also no significance in FJS-12 between two groups (70.3 in CR group and 66.9 in CS group; P=0.53). Conclusions. Residual PCL in TKA with a CS insert would not impact one-year postoperative clinical outcomes including KOOS, JOA, and FJS-12


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 1 | Pages 173 - 179
1 Jan 1998
Race A Amis AA

We compared the ability of three different posterior cruciate ligament (PCL) reconstructions to restore normal anteroposterior laxity to the knee from 0 to 130° of knee flexion. Cadaver knees were tested intact, after PCL rupture or after bone-patellar tendon-bone grafting. Grafts were performed isometrically or with a single bundle representing the anatomical anterior PCL fibre bulk (aPC) or with a double bundle that added the posterior PCL fibre bulk (pPC). The grafts were tensioned to restore normal knee laxity at 60° of flexion, except for the pPC which was tensioned at 130°. The isometric graft led to overconstraint as the knee extended resulting in high graft tension in extension and excess laxity in flexion. The aPC graft matched normal laxity from 0 to 60° of flexion but was lax from 90 to 130° of flexion. Only the double-bundled graft could restore normal knee laxity across the full range of flexion


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 5 | Pages 765 - 773
1 Jul 2003
Gupte CM Bull AMJ Thomas RD Amis AA

We have tested the hypothesis that the meniscofemoral ligaments make a significant contribution to resisting anteroposterior and rotatory laxity of the posterior-cruciate-ligament-deficient knee. Eight cadaver human knees were tested for anteroposterior and rotatory laxity in a materials-testing machine. The posterior cruciate ligament (PCL) was then divided, followed by division of the meniscofemoral ligaments (MFLs). Laxity results were obtained for intact, PCL-deficient, and PCL-MFL-deficient knees. Division of the MFLs in the PCL-deficient knee increased posterior laxity between 15° and 90° of flexion. Force-displacement measurements showed that the MFLs contributed 28% to the total force resisting posterior drawer at 90° of flexion in the intact knee, and 70.1% in the PCL-deficient knee. There was no effect on rotatory laxity. This is the first study which shows a function for the MFLs as secondary restraints to posterior tibial translation. The integrity of these structures should be assessed during both imaging and arthroscopic studies of PCL-injured knees since this may affect the diagnosis and management of such injuries


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 450 - 456
1 Apr 2004
Nakagawa S Johal P Pinskerova V Komatsu T Sosna A Williams A Freeman MAR

The posterior cruciate ligament (PCL) was imaged by MRI throughout flexion in neutral tibial rotation in six cadaver knees, which were also dissected, and in 20 unloaded and 13 loaded living (squatting) knees. The appearance of the ligament was the same in all three groups. In extension the ligament is curved concave-forwards. It is straight, fully out-to-length and approaching vertical from 60° to 120°, and curves convex-forwards over the roof of the intercondylar notch in full flexion. Throughout flexion the length of the ligament does not change, but the separations of its attachments do. We conclude that the PCL is not loaded in the unloaded cadaver knee and therefore, since its appearance in all three groups is the same, that it is also unloaded in the living knee during flexion. The posterior fibres may be an exception in hyperextension, probably being loaded either because of posterior femoral lift-off or because of the forward curvature of the PCL. These conclusions relate only to everyday life: none may be drawn with regard to more strenuous activities such as sport or in trauma


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_2 | Pages 109 - 109
1 Jan 2017
Beukes G Patnaik S Sivarasu S
Full Access

The function of the knee joint is to allow for locomotion and is comprised of various bodily structures including the four major ligaments; medial collateral ligament (MCL), lateral collateral ligament (LCL), anterior cruciate ligament (ACL) and posterior cruciate ligament (PCL). The primary function of the ligaments are to provide stability to the joint. The knee is prone to injury as a result of osteoarthritis as well as ligamentous and meniscal lesions. Furthermore, compromised joint integrity due to ligamentous injury may be a result of direct and indirect trauma, illness, occupational hazard as well as lifestyle. A device capable of non-invasively determining the condition of the ligaments in the knee joint would be a useful tool to assist the clinician in making a more informed diagnosis and prognosis of the injury. Furthermore, the device would potentially reduce the probability of a misdiagnosis, timely diagnosis and avoidable surgeries. The existing Laxmeter prototype (UK IPN: GB2520046) is a Stress Radiography Device currently limited to measuring the laxity of the MCL and LCL at multiple fixed degrees of knee flexion. Laxity refers to the measure of a ligament's elasticity and stiffness i.e. the condition of the ligament, by applying a known load (200N) to various aspects of the proximal tibial and thereby inducing tibial translation. The extent of translation would indicate the condition of the ligament. The Laxmeter does not feature a load applying component as of yet, however, it allows for the patient to be in the most comfortable and ideal position during radiographic laxity measurement testing. The entire structure is radiolucent and attempts to address the limitations of existing laxity measurement devices, which includes: excessive radiation exposure to the radiographic assistant, little consideration for patient ergonomics and restrictions to cruciate or collateral ligament laxity measurements. The study focusses on further developing and modifying the Laxmeter to allow for: the laxity measurement of all four major ligaments of the knee joint, foldability for improved storage and increased structural integrity. Additionally, a load applicator has been designed as an add-on to the system thereby making the Laxmeter a complete Stress Radiography Device. Various materials including Nylon, Polycarbonate, Ultra High Molecular Weight Polyethylene (UHMWPE) – PE 1000, and Acetal/ POM were tested, using the Low Dose X-ray (Lodox) scanner, to determine their radiolucency. All materials were found to be radiolucent enough for the manufacture of the Laxmeter structure as well as the load applicator in order to identify and measure the translation of the tibia with respect to the stationary femur. The Laxmeter allows for the measurement of the laxity of the MCL and LCL at multiple fixed degrees of flexion by providing the ideal patient position for testing. The next iteration of the device will present an affordable and complete Stress Radiography Device capable of measuring the laxity of all four major ligaments of the knee joint at multiple fixed degrees of flexion. Future work would include aesthetic considerations as well as an investigation into carbon-fibre-reinforced plastics


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_2 | Pages 20 - 20
1 Jan 2017
Pai S Li J Wang Y Lin C Kuo M Lu T
Full Access

Knee ligament injury is one of the most frequent sport injuries and ligament reconstruction has been used to restore the structural stability of the joint. Cycling exercises have been shown to be safe for anterior cruciate ligament (ACL) reconstruction and are thus often prescribed in the rehabilitation of patients after ligament reconstruction. However, whether it is safe for posterior cruciate ligament (PCL) reconstruction remains unclear. Considering the structural roles of the PCL, backward cycling may be more suitable for rehabilitation in PCL reconstruction. However, no study has documented the differences in the effects on the knee kinematics between forward and backward pedaling. Therefore, the current study aimed to measure and compare the arthrokinematics of the tibiofemoral joint between forward and backward pedaling using a biplane fluoroscope-to- computed tomography (CT) registration method. Eight healthy young adults participated in the current study with informed written consent. Each subject performed forward and backward pedaling with an average resistance of 20 Nm, while the motion of the left knee was monitored simultaneously by a biplane fluoroscope (ALLURA XPER FD, Philips) at 30 fps and a 14-camera stereophotogrammetry system (Vicon, OMG, UK) at 120 Hz. Before the motion experiment, the knee was CT and magnetic resonance scanned, which enabled the reconstruction of the bones and articular cartilage. The bone models were registered to the fluoroscopic images using a volumetric model-based fluoroscopy-to-CT registration method, giving the 3-D poses of the bones. The bone poses were then used to calculate the rigid-body kinematics of the joint and the arthrokinematics of the articular cartilage. In this study, the top dead center of the crank was defined as 0° so forward pedaling sequence would begin from 0° to 360°. Compared with forward pedaling, for crank angles from 0° to 180°, backward pedaling showed significantly more tibial external rotation. Moreover, both the joint center and contact positions in the lateral compartment were more anterior while the contact positions in the medial compartment was more posterior, during backward pedaling. For crank angles from 180° to 360°, the above-observed phenomena were generally reversed, except for the anterior-posterior component of the contact positions in the medial compartment. Forward and backward pedaling displayed significant differences in the internal/external rotations while the rotations in the sagittal and frontal planes were similar. Compared with forward cycling, the greater tibial external rotation for crank angles from 0° to 180° during backward pedaling appeared to be the main reason for the more anterior contact positions in the lateral compartment and more posterior contact positions in the medial compartment. Even though knee angular motions during forward and backward pedaling were largely similar in the sagittal and frontal planes, significant differences existed in the other components with different contact patterns. The current results suggest that different pedaling direction may be used in rehabilitation programs for better treatment outcome in future clinical applications


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 1 - 1
1 Jul 2014
Gao B Angibaud L
Full Access

Summary Statement. Femorotibial constraint is a key property of a total knee arthroplasty (TKA) prosthesis and should reflect the intended function of the device. With a validated simulation methodology, this study evaluated the constraint of two TKA prostheses designed for different intentions. Introduction. TKA prostheses are semi-constrained artificial joints. Femorotibial constraint level is a major property of a prosthesis and should be designed to match the device's intended function. Cruciate Retaining (CR) prostheses are usually indicated for patients with a functioning posterior cruciate ligament (PCL). For patients without a fully functioning PCL, CR-Constrained (CRC) prostheses with additional built-in constraint may be indicated. A CRC prosthesis usually consists of a CR femoral component and a tibial insert which has a more conforming sagittal profile to offer an increased femorotibial constraint. This study evaluated the anterior-posterior (AP) constraint behavior of two lines of prostheses (CR and CRC) from a same TKA product family. Using a validated computer simulation approach, multiple sizes of each product line were evaluated. Methods. Both the CR and CRC prostheses are from the same TKA product family (Optetrak Logic, Exactech, FL, USA) and share identical femoral components and tibial baseplates. The CRC tibial inserts have a more conforming sagittal profile than the CR tibial inserts, especially in the anterior aspect. Three sizes (sizes 1, 3, and 5) from each product line were included in this study. Computer simulations using finite element analysis (FEA) were performed to evaluate the femorotibial constraint of each prosthesis per ASTM F1223 standard [1]. The simulation has been validated by comparison with physical testing (more details submitted in a separate paper to CORS 2013). Briefly, FEA models were created using 10-node tetrahedral elements with all materials considered linear elastic. The tibial baseplate was distally fixed and a constant compressive force (710 N) was applied to the femoral component. Nonlinear Surface-Surface-Contact was established at the articulating surfaces, as well as between the tibial insert and the tibial baseplate. A coefficient of friction of 0.1 was assumed for all articulations [2]. The femoral component was driven under a displacement-controlled scheme to slide along AP direction on the tibial insert. Constraint force occurring at the articulation was derived from the reaction force at the distal fixation; thus, the force-displacement curve can be plotted to characterise the constraint behavior of the prosthesis. A nonlinear FEA solver (NX Nastran SOL601, Siemens, TX, USA) was used to solve the simulations. Results. The force-displacement curves predicted by the simulation exhibited the hysteresis loop appearance for both CR and CRC prostheses. The profile of the curves was generally consistent across different sizes for both product lines. The anterior constraint of the CRC prosthesis was significantly greater than the CR prosthesis. The posterior constraint of the CRC prosthesis was also slightly greater. Larger sizes exhibited reduced constraint compared to smaller sizes. Discussion/Conclusion. The increased constraint of the CRC prosthesis revealed in the study is consistent with the geometrical characteristics and the functional intent of the device. The CRC tibial insert is expected to provide significantly greater anterior constraint than the CR prosthesis to prevent paradoxical femoral translation when the patient's PCL is not fully functioning. The CRC tibial insert is also expected to provide slightly increased posterior constraint due to its elevated posterior lip. The observed hysteresis loop appearance is consistent with physical testing and the existence of friction. The reduced constraint on larger sizes is functionally desirable to offer proportional translation freedom. This study demonstrated the effectiveness of the simulation approach in quantifying the constraint behavior of different TKA prosthesis designs


Bone & Joint Research
Vol. 8, Issue 3 | Pages 136 - 145
1 Mar 2019
Cerquiglini A Henckel J Hothi H Allen P Lewis J Eskelinen A Skinner J Hirschmann MT Hart AJ

Objectives

The Attune total knee arthroplasty (TKA) has been used in over 600 000 patients worldwide. Registry data show good clinical outcome; however, concerns over the cement-tibial interface have been reported. We used retrieval analysis to give further insight into this controversial topic.

Methods

We examined 12 titanium (Ti) PFC Sigma implants, eight cobalt-chromium (CoCr) PFC Sigma implants, eight cobalt-chromium PFC Sigma rotating platform (RP) implants, and 11 Attune implants. We used a peer-reviewed digital imaging method to quantify the amount of cement attached to the backside of each tibial tray. We then measured: 1) the size of tibial tray thickness, tray projections, peripheral lips, and undercuts; and 2) surface roughness (Ra) on the backside and keel of the trays. Statistical analyses were performed to investigate differences between the two designs.


Objectives

Preservation of both anterior and posterior cruciate ligaments in total knee arthroplasty (TKA) can lead to near-normal post-operative joint mechanics and improved knee function. We hypothesised that a patient-specific bicruciate-retaining prosthesis preserves near-normal kinematics better than standard off-the-shelf posterior cruciate-retaining and bicruciate-retaining prostheses in TKA.

Methods

We developed the validated models to evaluate the post-operative kinematics in patient-specific bicruciate-retaining, standard off-the-shelf bicruciate-retaining and posterior cruciate-retaining TKA under gait and deep knee bend loading conditions using numerical simulation.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 830 - 834
1 Jun 2009
Pinskerova V Samuelson KM Stammers J Maruthainar K Sosna A Freeman MAR

There has been only one limited report dating from 1941 using dissection which has described the tibiofemoral joint between 120° and 160° of flexion despite the relevance of this arc to total knee replacement. We now provide a full description having examined one living and eight cadaver knees using MRI, dissection and previously published cryosections in one knee.

In the range of flexion from 120° to 160° the flexion facet centre of the medial femoral condyle moves back 5 mm and rises up on to the posterior horn of the medial meniscus. At 160° the posterior horn is compressed in a synovial recess between the femoral cortex and the tibia. This limits flexion. The lateral femoral condyle also rolls back with the posterior horn of the lateral meniscus moving with the condyle. Both move down over the posterior tibia at 160° of flexion.

Neither the events between 120° and 160° nor the anatomy at 160° could result from a continuation of the kinematics up to 120°. Therefore hyperflexion is a separate arc. The anatomical and functional features of this arc suggest that it would be difficult to design an implant for total knee replacement giving physiological movement from 0° to 160°.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1460 - 1465
1 Oct 2010
Rauh PB Clancy WG Jasper LE Curl LA Belkoff S Moorman CT

We evaluated two reconstruction techniques for a simulated posterolateral corner injury on ten pairs of cadaver knees. Specimens were mounted at 30° and 90° of knee flexion to record external rotation and varus movement. Instability was created by transversely sectioning the lateral collateral ligament at its midpoint and the popliteus tendon was released at the lateral femoral condyle. The left knee was randomly assigned for reconstruction using either a combined or fibula-based treatment with the right knee receiving the other. After sectioning, laxity increased in all the specimens. Each technique restored external rotatory and varus stability at both flexion angles to levels similar to the intact condition. For the fibula-based reconstruction method, varus laxity at 30° of knee flexion did not differ from the intact state, but was significantly less than after the combined method.

Both the fibula-based and combined posterolateral reconstruction techniques are equally effective in restoring stability following the simulated injury.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1528 - 1533
1 Nov 2007
Jeffcote B Nicholls R Schirm A Kuster MS

Achieving deep flexion after total knee replacement remains a challenge. In this study we compared the soft-tissue tension and tibiofemoral force in a mobile-bearing posterior cruciate ligament-sacrificing total knee replacement, using equal flexion and extension gaps, and with the gaps increased by 2 mm each. The tests were conducted during passive movement in five cadaver knees, and measurements of strain were made simultaneously in the collateral ligaments. The tibiofemoral force was measured using a customised mini-force plate in the tibial tray. Measurements of collateral ligament strain were not very sensitive to changes in the gap ratio, but tibiofemoral force measurements were. Tibiofemoral force was decreased by a mean of 40% (sd 10.7) after 90° of knee flexion when the flexion gap was increased by 2 mm. Increasing the extension gap by 2 mm affected the force only in full extension. Because increasing the range of flexion after total knee replacement beyond 110° is a widely-held goal, small increases in the flexion gap warrant further investigation.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 557 - 564
1 Apr 2009
Rumian AP Draper ERC Wallace AL Goodship AE

An understanding of the remodelling of tendon is crucial for the development of scientific methods of treatment and rehabilitation. This study tested the hypothesis that tendon adapts structurally in response to changes in functional loading. A novel model allowed manipulation of the mechanical environment of the patellar tendon in the presence of normal joint movement via the application of an adjustable external fixator mechanism between the patella and the tibia in sheep, while avoiding exposure of the patellar tendon itself. Stress shielding caused a significant reduction in the structural and material properties of stiffness (79%), ultimate load (69%), energy absorbed (61%), elastic modulus (76%) and ultimate stress (72%) of the tendon compared with controls. Compared with the material properties the structural properties exhibited better recovery after re-stressing with stiffness 97%, ultimate load 92%, energy absorbed 96%, elastic modulus 79% and ultimate stress 80%. The cross-sectional area of the re-stressed tendons was significantly greater than that of stress-shielded tendons.

The remodelling phenomena exhibited in this study are consistent with a putative feedback mechanism under strain control. This study provides a basis from which to explore the interactions of tendon remodelling and mechanical environment.