Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 32 - 32
1 Nov 2015
O'Neill C Molloy D Patterson C Beverland D
Full Access

Introduction. Operative inclination (OI) is defined as the angle between the acetabular axis and the sagittal plane. With the patient in the true lateral decubitus position, this corresponds to the angle formed between the handle of the acetabular component inserter and the theatre floor intra-operatively. Patients/Materials & Methods. The primary study aim was to determine which method of acetabular component insertion most accurately allows the surgeon to obtain a target OI of 35o. 270 consecutive patients undergoing cementless THA were randomised to one of three possible methods for acetabular component implantation:. 1. Freehand,. 2. 35o mechanical alignment guide (MAG), or. 3. Digital inclinometer assisted. Two surgeons participated. Target OI was 35o in all cases. OI was measured using a digital inclinometer. For the freehand and MAG cases, the surgeon was blinded to inclinometer readings intra-operatively. Results. Freehand: OI 25.2 – 43.2o. Mean deviation from target 2.92o. 35o MAG: OI 29.3 – 39.3o. Mean deviation from target 1.83o. Inclinometer assisted: OI 27.5 – 37.5o. Mean deviation from target 1.28o. Overall, when comparing mean deviation from target OI, a statistically significant difference between both the Freehand/Inclinometer group and Freehand/MAG group was demonstrated (p<0.001). A statistically significant difference between the Inclinometer/MAG groups was also demonstrated (p<0.023). Discussion. Both the 35o MAG and digital inclinometer assisted methods provided a considerably narrower OI range when compared to the freehand method. Though the range was similar for both the 35o MAG and digital inclinometer assisted methods, the SD was smaller in the inclinometer assisted group. Conclusion. The novel method of using a digital inclinometer to control operative inclination appears to be more accurate than both the freehand and mechanical alignment guide methods and may help further optimise acetabular component orientation


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 12 - 12
1 Jun 2016
Meermans G Peeters W Van Doorn W Kats J
Full Access

Introduction. In total hip arthroplasty (THA), a high radiographic inclination angle (RI) of the acetabular component has been linked to an increased dislocation rate, liner fracture, and increased wear. In contrast to version, we have more proven boundaries when it comes to a safe zone for angles of RI. Although intuitively it seems easier to achieve a target RI, most studies demonstrate a lack of accuracy and the trend towards a high RI with all surgical approaches when using a freehand technique or a mechanical guide. This is due to pelvic motion during surgery, which can be highly variable. The current study had two primary aims, each with a different primary outcome. The first aim was to determine how accurate a surgeon could obtain the target operative inclination (OI) during THA when using a cementless cup using a digital protractor. The second aim was to determine how accurate a surgeon can estimate the target OI to obtain a RI of 40° based on the patient's hip circumference as demonstrated in a previous study. Methods. In this prospective study, we included 200 consecutive patients undergoing uncemented primary THA in the lateral decubitus position using a posterior approach. Preoperatively, the surgeon determined the target OI based on the patient's hip circumference (22.5°, 25°, 27.5° or 30°). Intraoperatively, the effective OI was measured with the aid of a digital inclinometer after seating of the acetabular component. Six weeks postoperatively anteroposterior pelvic radiographs were made and two evaluators, blinded to the effective OI, measured the RI of the acetabular component. The safe zone for inclination was defined as 30°-45° of inclination. Results. The mean difference between the target OI and the effective OI of the acetabular component was −0.7° SD 1.4 (95% CI −0.9° to −0.5°). The difference between the target and effective OI was less than 1° in 108 patients (54%), less than 2° in 160 patients (80%) and less than 3° in 186 patients (93%). In 14 patients (7%) the difference was 3°-5°. The mean RI was 37.9° SD 4.7 (95% CI 37.2° to 38.5°). The mean difference between the RI and effective OI was 11.5° SD 4.7 (95% CI 10.8° to 12.1°). Overall, 188 cups (94%) were within the inclination safe zone. When analysing the RI outliers, 1 could have be avoided if a better target OI was chosen and 2 could have been avoided if the difference between the target and effective OI would have been smaller. For the remaining 9 outliers (75%) the difference between the RI and effective OI was in the upper and lower 7. th. percentile, indicating more or less than average motion of the pelvis in these patients. Discussion and Conclusions. When using a digital protractor, the mean difference between the target OI and the effective OI of the acetabular component was less than 3° in 93% and less than 5° in all patients. The use of a digital protractor allows surgeons to accurately implant the acetabular component in the desired OI in a cheap and easy way. By adjusting the target OI based on the patient's hip circumference, 94% of the acetabular components were placed within an inclination safe zone of 30°-45°. Most outliers were caused by more of less than average intraoperative pelvic motion


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 862 - 866
1 Jul 2018
Darrith B Bell JA Culvern C Della Valle CJ

Aims

Accurate placement of the acetabular component is essential in total hip arthroplasty (THA). The purpose of this study was to determine if the ability to achieve inclination of the acetabular component within the ‘safe-zone’ of 30° to 50° could be improved with the use of an inclinometer.

Patients and Methods

We reviewed 167 primary THAs performed by a single surgeon over a period of 14 months. Procedures were performed at two institutions: an inpatient hospital, where an inclinometer was used (inclinometer group); and an ambulatory centre, where an inclinometer was not used as it could not be adequately sterilized (control group). We excluded 47 patients with a body mass index (BMI) of > 40 kg/m2, age of > 68 years, or a surgical indication other than osteoarthritis whose treatment could not be undertaken in the ambulatory centre. There were thus 120 patients in the study, 68 in the inclinometer group and 52 in the control group. The inclination angles of the acetabular component were measured from de-identified plain radiographs by two blinded investigators who were not involved in the surgery. The effect of the use of the inclinometer on the inclination angle was determined using multivariate regression analysis.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1010 - 1017
1 Aug 2018
Jain S Magra M Dube B Veysi VT Whitwell GS Aderinto JB Emerton ME Stone MH Pandit HG

Aims

This study aimed to evaluate implant survival of reverse hybrid total hip arthroplasty (THA) at medium-term follow-up.

Patients and Methods

A consecutive series of 1082 THAs in 982 patients with mean follow-up of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21 to 94). Of these, 194 (17.9%) were in patients under 60 years, 663 (61.3%) in female patients and 348 (32.2%) performed by a trainee. Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%). Survival analysis was performed and subgroups compared using log rank tests.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1589 - 1596
1 Dec 2016
Magill P Blaney J Hill JC Bonnin MP Beverland DE

Aims

Our aim was to report survivorship data and lessons learned with the Corail/Pinnacle cementless total hip arthroplasty (THA) system.

Patients and Methods

Between August 2005 and March 2015, a total of 4802 primary cementless Corail/Pinnacle THAs were performed in 4309 patients. In March 2016, we reviewed these hips from a prospectively maintained database.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 312 - 318
1 Mar 2014
Meermans G Van Doorn WJ Koenraadt K Kats J

The orientation of the acetabular component can influence both the short- and long-term outcomes of total hip replacement (THR). We performed a prospective, randomised, controlled trial of two groups, comprising of 40 patients each, in order to compare freehand introduction of the component with introduction using the transverse acetabular ligament (TAL) as a reference for anteversion. Anteversion and inclination were measured on pelvic radiographs.

With respect to anteversion, in the freehand group 22.5% of the components were outside the safe zone versus 0% in the transverse acetabular ligament group (p = 0.002). The mean angle of anteversion in the freehand group was 21° (2° to 35°) which was significantly higher compared with 17° (2° to 25°) in the TAL group (p = 0.004). There was a significant difference comparing the variations of both groups (p = 0.008).

With respect to inclination, in the freehand group 37.5% of the components were outside the safe zone versus 20% in the TAL group (p = 0.14). There was no significant difference regarding the accuracy or variation of the angle of inclination when comparing the two groups.

The transverse acetabular ligament may be used to obtain the appropriate anteversion when introducing the acetabular component during THR, but not acetabular component inclination.

Cite this article: Bone Joint J 2014;96-B:312–18.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 603 - 610
1 May 2015
Meermans G Goetheer-Smits I Lim RF Van Doorn WJ Kats J

A high radiographic inclination angle (RI) contributes to accelerated wear and has been associated with dislocation after total hip arthroplasty (THA). With freehand positioning of the acetabular component there is a lack of accuracy, with a trend towards a high radiographic inclination angle. The aim of this study was to investigate whether the use of a digital protractor to measure the operative inclination angle (OI) could improve the positioning of the acetabular component in relation to a ‘safe zone’.

We measured the radiographic inclination angles of 200 consecutive uncemented primary THAs. In the first 100 the component was introduced freehand and in the second 100 a digital protractor was used to measure the operative inclination angle.

The mean difference between the operative and the radiographic inclination angles (∆RI–OI) in the second cohort was 12.3° (3.8° to 19.8°). There was a strong correlation between the circumference of the hip and ∆RI–OI. The number of RI outliers was significantly reduced in the protractor group (p = 0.002).

Adjusting the OI, using a digital protractor and taking into account the circumference of the patient’s hip, improves the RI significantly (p < 0.001) and does not require additional operating time.

Cite this article: Bone Joint J 2015; 97-B:603–610.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 163 - 167
1 Feb 2006
Kalteis T Handel M Bäthis H Perlick L Tingart M Grifka J

In a prospective randomised clinical study acetabular components were implanted either freehand (n = 30) or using CT-based (n = 30) or imageless navigation (n = 30). The position of the component was determined post-operatively on CT scans of the pelvis.

Following conventional freehand placement of the acetabular component, only 14 of the 30 were within the safe zone as defined by Lewinnek et al (40° inclination sd 10°; 15° anteversion sd 10°). After computer-assisted navigation 25 of 30 acetabular components (CT-based) and 28 of 30 components (imageless) were positioned within this limit (overall p < 0.001). No significant differences were observed between CT-based and imageless navigation (p = 0.23); both showed a significant reduction in variation of the position of the acetabular component compared with conventional freehand arthroplasty (p < 0.001). The duration of the operation was increased by eight minutes with imageless and by 17 minutes with CT-based navigation.

Imageless navigation proved as reliable as that using CT in positioning the acetabular component.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1021 - 1026
1 Aug 2011
Kalteis T Sendtner E Beverland D Archbold PA Hube R Schuster T Renkawitz T Grifka J

Orientation of the native acetabular plane as defined by the transverse acetabular ligament (TAL) and the posterior labrum was measured intra-operatively using computer-assisted navigation in 39 hips. In order to assess the influence of alignment on impingement, the range of movement was calculated for that defined by the TAL and the posterior labrum and compared with a standard acetabular component position (abduction 45°/anteversion 15°).

With respect to the registration of the plane defined by the TAL and the posterior labrum, there was moderate interobserver agreement (r = 0.64, p < 0.001) and intra-observer reproducibility (r = 0.73, p < 0.001). The mean acetabular component orientation achieved was abduction of 41° (32° to 51°) and anteversion of 18° (−1° to 36°). With respect to the Lewinnek safe zone (abduction 40° ±10°, anteversion 15° ±10°), 35 of the 39 acetabular components were within this zone. However, there was no improvement in the range of movement (p = 0.94) and no significant difference in impingement (p = 0.085).

Alignment of the acetabular component with the TAL and the posterior labrum might reduce the variability of acetabular component placement in total hip replacement. However, there is only a moderate interobserver agreement and intra-observer reliability in the alignment of the acetabular component using the TAL and the posterior labrum. No reduction in impingement was found when the acetabular component was aligned with the TAL and the posterior labrum, compared with a standard acetabular component position.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1158 - 1164
1 Sep 2007
Lusty PJ Watson A Tuke MA Walter WL Walter WK Zicat B

We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm3/yr for the femoral heads, and 0.04 mm3/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of < 15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm3/yr, compared with 0 mm3/yr for those with an anteversion of ≥15° (15 femoral heads, p < 0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 762 - 769
1 Jun 2005
Biedermann R Tonin A Krismer M Rachbauer F Eibl G Stöckl B

Malposition of the acetabular component is a risk factor for post-operative dislocation after total hip replacement (THR). We have investigated the influence of the orientation of the acetabular component on the probability of dislocation. Radiological anteversion and abduction of the component of 127 hips which dislocated post-operatively were measured by Einzel-Bild-Röentgen-Analysis and compared with those in a control group of 342 patients.

In the control group, the mean value of anteversion was 15° and of abduction 44°. Patients with anterior dislocation after primary THR showed significant differences in the mean angle of anteversion (17°), and abduction (48°) as did patients with posterior dislocation (anteversion 11°, abduction 42°). After revision patients with posterior dislocation showed significant differences in anteversion (12°) and abduction (40°).

Our results demonstrate the importance of accurate positioning of the acetabular component in order to reduce the frequency of subsequent dislocations. Radiological anteversion of 15° and abduction of 45° are the lowest at-risk values for dislocation.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1072 - 1078
1 Aug 2010
Grammatopoulos G Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gill HS Murray DW

Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by edge loading due to poor orientation of the acetabular component. Our aim was to determine the optimal acetabular orientation to minimise the risk of pseudotumour formation.

We matched 31 hip resurfacings revised for pseudotumour formation with 58 controls who had a satisfactory outcome from this procedure. The radiographic inclination and anteversion angles of the acetabular component were measured on anteroposterior radiographs of the pelvis using Einzel-Bild-Roentgen-Analyse software. The mean inclination angle (47°, 10° to 81°) and anteversion angle (14°, 4° to 34°) of the pseudotumour cases were the same (p = 0.8, p = 0.2) as the controls, 46° (29° to 60°) and 16° (4° to 30°) respectively, but the variation was greater. Assuming an accuracy of implantation of ± 10° about a target position, the optimal radiographic position was found to be approximately 45° of inclination and 20° of anteversion. The incidence of pseudotumours inside the zone was four times lower (p = 0.007) than outside the zone.

In order to minimise the risk of pseudotumour formation we recommend that surgeons implant the acetabular component at an inclination of 45° (± 10) and anteversion of 20° (± 10) on post-operative radiographs. Because of differences between the radiographic and the operative angles, this may be best achieved by aiming for an inclination of 40° and an anteversion of 25°.