Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 80 - 80
1 May 2016
Nebergall A Freiberg A Greene M Malchau H Muratoglu O Rowell S Zumbrunn T Varadarajan K
Full Access

Introduction

The large diameter mobile polyethylene liner of the dual mobility implant provides increased resistance to hip dislocation. However, a problem specific to the dual mobility system is intra-prosthetic dislocation (IPD), secondary to loss of the retentive rim, causing the inner head to dissociate from the polyethylene liner. We hypothesized that impingement of the polyethylene liner with the surrounding soft-tissue inhibits liner motion, thereby facilitating load transfer from the femoral neck to the liner and leading to loss of retentive rim over time. This mechanism of soft-tissue impingement with the liner was evaluated via cadaver experiments, and retrievals were used to assess polyethylene rim damage.

Methods

Total hip arthroplasty was performed on 10 cadaver hips using 3D printed dual mobility components. A metal wire was sutured to the posterior surface (underside) of the iliopsoas, and metal wires were embedded into grooves on the outer surface of the liner and inner head to identify these structures under fluoroscopy. Tension was applied to the iliopsoas to move the femur from maximum hyperextension to 90° of flexion for the purpose of visualizing the iliopsoas and capsule interaction with the mobile liner. The interaction of the mobile liner with the iliopsoas was studied using fluoroscopy and direct visual observation. Fifteen retrieved dual mobility liners were assessed for rim edge and rim chamfer damage. Rim edge damage was defined as any evidence of contact, and rim chamfer damage was classified into six categories: impact ribs on the chamfer surface, loss of machining marks, scratching or pitting, rim deformation causing a raised lip, a rounded rim edge, or embedded metal debris.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 51 - 51
23 Feb 2023
Full Access

Dual mobility is a French concept that appeared in the 1970s and was initially intended to reduce dislocation rates. In recent years, this concept has evolved with new HA titanium spray coatings, new external macrostructures, and better-quality polyethylene. This has allowed to extend the indications to younger and therefore active populations. The objective of our work is to analyze at least 10 years a homogeneous and continuous series of 170 primary total hip replacements associating a latest generation Novae Sunfit. ®. dual mobility cup with a straight femoral stem. Only primary arthroplasties for osteoarthritis or necrosis were included. Total hip arthroplasty was always performed through a posterolateral approach. All patients had regular clinical and radiological follow-up. The average follow-up in our series was 11.5 years. The average age of the population is 71 years. At the last follow-up, there were 17 deaths, 6 losses to follow up and 9 adverse events, including 1 cup change for psoas impingement and 1 dislocation. The low rate of dislocation at 11 years confirms the high stability of the dual mobility, which should be recommended for primary procedure for patients at high risk of postoperative instability. The lack of intraprosthetic dislocation due to wear at 11 years of follow-up highlights the good quality of the latest generation of polyethylene, and the need to combine high-polished surfaces and a refined femoral neck with a dual mobility cup. Finally, the lack of aseptic loosening confirms the quality of the secondary fixation of these implants and justifies their wider use in all patients


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 52 - 52
23 Feb 2023
Full Access

THA in patients with acetabular bone defects is associated with a high risk of dislocation. Dual mobility (DM) cups are known to prevent and treat chronic instability. The aim of this study was to evaluate the dislocation rate and survival of jumbo DM cups. This was a retrospective, continuous, multicenter study of all the cases of jumbo DM cup implantation between 2010 and 2017 in patients with acetabular bone loss (Paprosky 2A: 46%, 2B: 32%, 2C: 15% and 3A: 6%). The indications for implantation were revisions for aseptic loosening of the cup (n=45), aseptic loosening of the femoral stem (n=3), bipolar loosening (n=11), septic loosening (n=10), periprosthetic fracture (n=5), chronic dislocation (n=4), intraprosthetic dislocation (n=2), cup impingement (n=1), primary posttraumatic arthroplasty (n=8), and acetabular dysplasia (n=4). The jumbo cups used were COPTOS TH (SERF), which combines press-fit fixation with supplemental fixation (acetabular hook, two superior flanges with one to four screws, two acetabular pegs). A bone graft was added in 74 cases (80%). The clinical assessment consisted of the Harris hip score. The primary endpoint was surgical revision for aseptic acetabular loosening or the occurrence of a dislocation episode. In all, 93 patients were reviewed at a mean follow-up of 5.3 ± 2.3 years [0, 10]. As of the last follow-up, the acetabular cup had been changed in five cases: three due to aseptic loosening (3.2%) and two due to infection (2.1%). The survivorship free of aseptic loosening was 96.8%. Three patients (3%) suffered a dislocation. At the last follow-up visit, the mean HSS scores were 72.15, (p < 0.05). Use of a jumbo DM cup in cases of acetabular bone defects leads to satisfactory medium-term results with low dislocation and loosening rates


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_10 | Pages 25 - 25
1 Jun 2018
Della Valle C
Full Access

Dislocation remains among the most common complications of, and reasons for, revision of both primary and revision total hip arthroplasties in the United States. We have advocated identifying the primary cause of instability to plan appropriate treatment (Wera, Della Valle, et al., JOA 2012). Once implant position, leg length, and offset have been optimised and sources of impingement have been removed, the surgeon can opt for a large femoral head, a dual mobility articulation or a constrained liner. Given the limitations of constrained liners, we have looked to dual mobility articulations as an alternative, including its use in patients with abductor deficiency. We retrospectively compared a consecutive series of revision THA that were at high risk for instability and treated with either a constrained liner or a dual mobility articulation. At a minimum of two years, there were ten dislocations in the constrained group (10/43 or 23.3%) compared to three in the dual-mobility group (3/36 or 8.3%; p = 0.06). With repeat revision for instability as an endpoint, the failure rate was 23% for the constrained group and 5.5% for the dual mobility group (p = 0.03). We have also performed a systematic review of the published literature on the use of dual mobility in revision THA. Of the 3,088 hips reviewed, the dislocation rate was 2.2%, the risk of intraprosthetic dislocation was 0.3% and overall survivorship was 96.6% at 5 years. Dual mobility articulations offer anatomic sized femoral heads that greatly increase jump distance, without many of the negatives of a constrained liner. While dual mobility is associated with its own concerns and problems (including intraprosthetic dislocation and wear) our initial results suggest that they are a viable alternative to a constrained liner, even in the most challenging situations


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_22 | Pages 45 - 45
1 Dec 2016
Lachiewicz P
Full Access

Dual mobility components for total hip arthroplasty provide for an additional articular surface, with the goals of improving range of motion, jump distance, and overall stability of the prosthetic hip joint. A large polyethylene head articulates with a polished metal acetabular component, and an additional smaller metal head is snap-fit into the large polyethylene. The first such device was introduced for primary total hip arthroplasty by Bousquet in the 1970s, thus, the “French connection”. Dual mobility components have been released for use in North America over the past five years. In some European centers, these components are routinely used for primary total hip arthroplasty. However, their greatest utility may be to manage recurrent dislocation in the setting of revision total hip arthroplasty. Several retrospective series and the Swedish hip registry have shown satisfactory results for this indication at short- to medium-term follow-up times. However, there are important concerns with polyethylene wear, late intraprosthetic dislocation, and the lack of long-term follow-up data. These components are an important option in the treatment of recurrent dislocation in younger patients, revision of failed metal-metal resurfacing, and salvage of failed constrained liners. There are more recent concerns of possible iliopsoas tendinitis, elevated metal levels with one design, and acute early intraprosthetic dislocation following attempted closed reduction. However, a dual mobility component may now be the preferred solution in revision surgery for recurrent hip dislocation


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 25 - 25
1 Feb 2021
Cascardo C Gehrke C Moore D Karadsheh M Flierl M Baker E
Full Access

Introduction. Dual mobility (DM) total hip arthroplasty (THA) prostheses are designed to increase stability. In the setting of primary and revision THA, DM THA are used most frequently for dysplasia and instability diagnoses, respectively. As the use of DM THA continues to increase, with 8,031 cases logged in the American Joint Replacement Registry from 2012–2018, characterizing in vivo damage and clinical failure modes are important to report. Methods. Under IRB-approved implant retrieval protocol, 43 DM THA systems from 41 patients were included. Each DM THA component was macroscopically examined for standard damage modes. Clinically-relevant data, including patient demographics and surgical elements, were collected from medical records. Fretting and corrosion damage grading is planned, according to the Goldberg et al. classification system. Results. In this 43-retrieved implant series, there were 23 female and 17 male patients (n=1, unknown), with an average body mass index of 29 (range, 19–49), and average ages at index and revision of 63 years (range, 34–80) and 64 years (range, 38–88), respectively. The average duration of implantation was 12.9 months (range, 0.1–72.0). Reasons for revision included infection (n=11, 26%), mechanical complication (n=10, 23%), intraprosthetic dislocation (n=6, 14%), periprosthetic fracture (n=5, 12%), pain (n=4, 9%), acetabular-associated loosening (n=3, 7%), unknown (n=3, 7%), hematoma (n=2, 5%), leg length discrepancy (n=1, 2%), and inflammatory reaction (n=1, 2%); some cases included multiple reasons for revision. On articular surfaces, scratching was the most commonly observed damage mode on all components, with more than 40% of acetabular cup and femoral heads showing scratching damage (Figure 1A). Abrasion, burnishing, and pitting damage were also observed in more than 10% of acetabular cup and acetabular liner components; further, approximately 20% of polyethylene acetabular liners exhibited edge deformation damage. On backside surfaces, polyethylene acetabular liners showed the greatest damage, with more than 60% of components exhibiting abrasion, scratching, or pitting damage (Figure 1B). Conclusion. This series showed various reasons for revision as well as in vivo damage of retrieved DM systems following short-to-midterm implantation. Damage was observed on both articular and backside surfaces of the five components of DM THA. Modularity of DM THA prostheses may amplify rates of in vivo damage. Future studies are needed to confirm these results and clinical significance. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_10 | Pages 30 - 30
1 Jun 2018
Taunton M
Full Access

Patients with neuromuscular disease and imbalance present a particularly challenging clinical situation for the orthopaedic hip surgeon. The cause of the neuromuscular imbalance may be intrinsic or extrinsic. Intrinsic disorders include those in which the hip is in development, such as cerebral palsy, polio, CVA, and other spinal cord injuries and disease. This can result in subluxation and dislocation of the hip in growing children, and subsequent pain, and difficulty in sitting and perineal care. Extrinsic factors involve previously stable hips and play a secondary role in the development of osteoarthritis and contractures in later life. Examples of extrinsic factors are Parkinson's disease, dyskinesis, athetosis, and multiple sclerosis. Goals of treatment in adults with pain and dysfunction in the setting of neuromuscular imbalance are to treat contractures and to perform salvage procedures to improve function and eliminate pain. Treatment of patients with neuromuscular imbalance may include resection arthroplasty (Girdlestone), arthrodesis, or total hip arthroplasty. Resection arthroplasty is typically reserved for patients that are non-ambulatory, or hips that are felt to be so unstable that arthroplasty would definitely fail due to instability. In modern times arthrodesis has limited use as it negatively impacts function and self-care in patients with neuromuscular disorders. Total hip arthroplasty has the ability to treat pain, relieve contractures, and provide improved function. Due to the increased risk of instability, special considerations must be made during primary total hip arthroplasty in this patient cohort. Risk of instability may be addressed by surgical approach, head size, or use of alternative bearing constructs. Posterior approach may have increased risk of posterior dislocation in this patient group, particularly if a posterior capsular repair is not possible due to the flexion contractures and sitting position in many patients. Surgeons familiar with the approaches may utilise the anterolateral or direct anterior approach judicially. Release of the adductors may be performed in conjunction with primary total hip arthroplasty to help with post-operative range of motion and to decrease risk of instability. In a standard bearing, the selected head size should be the largest that can be utilised for the particular cup size. Rigorous testing of intra-operative impingement, component rotation, and instability is required. If instability cannot be adequately addressed by a standard bearing, the next option is a dual mobility bearing. Multiple studies have shown improved stability with the use of these bearings, but they are also at risk for instability, intraprosthetic dislocation, and fretting and corrosion of the modular connections. Another option is a constrained liner. However, this results in reduced range of motion, and an increased risk for mechanical complications of the construct. The use of a constrained liner in a primary situation should be limited to the most severe instability cases, and the patient should be counseled with the associated risks. If total hip arthroplasty results in repeated instability, revision surgery or Girdlestone arthroplasty may be considered


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_1 | Pages 5 - 5
1 Feb 2015
Su E
Full Access

Dislocation is a particular problem after total hip replacement in femoral neck fractures and elderly especially female patients. The increased rate of dislocation in this population is probably due to significant ligamentous laxity in these patients and poor coordination and proprioception. Another population of patients with increased propensity for dislocation is the revision hip replacement patient. Current dislocation rates in these patients can approach 10% with conventional implant systems. The Dual Mobility total hip system is composed of a cobalt chrome acetabular shell with a grit blasted, beaded and/or hydroxyapatite coating to improve bone ingrowth. The polyethylene liner is highly cross-linked polyethylene and fits congruently into the cobalt chrome shell and acts like a large femoral head (usually >40mm). The femoral head attached to the trunnion is usually 28mm or 32mm. The femoral head snaps into the polyethylene liner to acts as a second protection against dislocation. Indications for the Dual Mobility socket are in the high risk for dislocation patient and particularly in elderly female patients. One hundred fifty-six patients with an average age of 79 have been performed to date with a maximum follow up to 4.2 years. To date there have been no mechanical or septic failures and no dislocations. Pain relief has been no different than conventional hip replacement and range of motion is unchanged as well. There have been reported cases of intraprosthetic dislocation but these have not occurred to date


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 31 - 31
1 May 2016
Barlow B Mclawhorn A Westrich G
Full Access

Introduction. Postoperative dislocation remains a vexing problem for patients and surgeons following total hip arthroplasty (THA). It is the commonest reason for revision THA in the US. Dual mobility (DM) THA implants markedly decrease the risk of THA instability. However, DM implants are more expensive than those used for conventional THA. The purpose of this study was to perform a cost-effectiveness analysis of DM implants compared to conventional bearing couples for unilateral primary THA using a computer model-based evaluation. Methods. A state-transition Markov computer simulation model was developed to compare the cost-utility of dual mobility versus conventional THA for hip osteoarthritis from a societal perspective (Figure 1). The model was populated with health outcomes and probabilities from registry and published data. Health outcomes were expressed as quality-adjusted life years (QALYs). Direct costs were derived from the literature and from administrative claims data, and indirect costs reflected estimated lost wages. All costs were expressed in 2013 US dollars. Health and cost outcomes were discounted by 3% annually. The base case modeled a 65-year-old patient undergoing THA for unilateral hip osteoarthritis. A lifetime time horizon was analyzed. The primary outcome was the incremental cost-effectiveness ratio (ICER). The willingness-to-pay threshold was set at $100,000/QALY. Threshold, one-way, two-way, and probabilistic sensitivity analyses were performed to assess model uncertainty. Results. DM THA exhibited absolute dominance over conventional THA with lower accrued costs (US$45,960 versus $47,103) and higher accrued utility (12.08 QALY versus 11.84 QALY). The ICER was -$4,771/QALY, suggesting that DM THA is cost-saving compared to conventional THA. The cost threshold at which dual mobility implants were cost-ineffective was $25,000 (Figure 2), and the threshold at which DM implants ceased being cost-saving was $12,845. Sensitivity analyses demonstrated that the probability of intraprosthetic dislocation, primary THA utility, and age at index THA most influenced model results. In the probabilistic sensitivity analysis, 90% of model iterations resulted in cost savings for DM THA (Figure 3). Discussion. Dual mobility components showed clear cost-utility advantages over conventional THA components, and DM implants are cost-saving for primary unilateral THA from a societal perspecitve. Model results suggest that DM THA need not be limited to only high-risk patients, although patient age and risk of dislocation are important determinants of its cost-utility


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_12 | Pages 89 - 89
1 Jul 2014
Della Valle C
Full Access

Dislocation remains among the most common complications of, and reasons for, revision of both primary and revision total hip arthroplasties in the United States. Hence, there is great interest in maximising stability to prevent this complication. Highly crosslinked polyethylene has allowed us to increase femoral head size, without a clinically important increase in wear. As femoral head size increases, stability is augmented, secondary to a decrease in component-to-component impingement, which is theoretically eliminated at head sizes greater than 36mm in diameter (however osseous impingement can still occur). Larger heads sizes also greatly increase the “jump distance” required for the head to dislocate (in an appropriately positioned cup) and eliminate the need for skirts. Hence, large heads have become the mainstay for preventing and treating instability in contemporary practice. Large heads, however, have been shown to have poor performance in patients with abductor insufficiency. Constrained liners are a tantalising solution to both prevent and treat instability, as they markedly increase the force needed for a dislocation to occur. They have, however, several important negatives that the surgeon must consider before entertaining their use including: . –. Increased stresses at the implant bone interface which can increase the risk of loosening or cause catastrophic failure in the early post-operative period. –. Decreased range of motion with a greater risk of impingement. –. Usually require an open reduction if they dislocate or otherwise fail. Given the limitations of constrained liners, we have moved to dual mobility articulations in most situations where we would have used a constrained liner in the past, including patients with abductor deficiency. These articulations offer anatomic sized femoral heads that greatly increase the jump distance, without many of the negatives of a constrained liner. While dual mobility is associated with its own concerns and problems (including intraprosthetic dislocation and wear) our initial results suggest that they are a viable alternative to a constrained liner, even in the most challenging situations