Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 180 - 180
1 Sep 2012
Banks D Boden R Mehan R Fehily M
Full Access

Background

Magnetic resonance arthrography is the current method of choice for investigating patients with a clinical diagnosis of femoroacetabular impingement prior to performing hip arthroscopy. The aim of our study was to assess the efficacy of this investigation by comparing the findings of MR arthrogram with those found at arthroscopy, with reference to labral tears and chondral damage.

Methods

A prospective trial to investigate the sensitivity, specificity, accuracy and predictive value of MRA for diagnosis of labral tears and chondral defects. Over a 25-month period 69 hips undergoing hip arthroscopy were investigated with MRA prior to the definitive operative procedure. MRA findings were compared to the intraoperative findings.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 10 - 10
23 Feb 2023
Hardwick-Morris M Twiggs J Miles B Jones E Bruce WJM Walter WL
Full Access

Dislocation is one of the most common complications in total hip arthroplasty (THA) and is primarily driven by bony or prosthetic impingement. The aim of this study was two-fold. First, to develop a simulation that incorporates the functional position of the femur and pelvis and instantaneously determines range of motion (ROM) limits. Second, to assess the number of patients for whom their functional bony alignment escalates impingement risk. 468 patients underwent a preoperative THA planning protocol that included functional x-rays and a lower limb CT scan. The CT scan was segmented and landmarked, and the x-rays were measured for pelvic tilt, femoral rotation, and preoperative leg length discrepancy (LLD). All patients received 3D templating with the same implant combination (Depuy; Corail/Pinnacle). Implants were positioned according to standardised criteria. Each patient was simulated in a novel ROM simulation that instantaneously calculates bony and prosthetic impingement limits in functional movements. Simulated motions included flexion and standing-external rotation (ER). Each patient's ROM was simulated with their bones oriented in both functional and neutral positions. 13% patients suffered a ROM impingement for functional but not neutral extension-ER. As a result, 48% patients who failed the functional-ER simulation would not be detected without consideration of the functional bony alignment. 16% patients suffered a ROM impingement for functional but not neutral flexion. As a result, 65% patients who failed the flexion simulation would not be detected without consideration of the functional bony alignment. We have developed a ROM simulation for use with preoperative planning for THA surgery that can solve bony and prosthetic impingement limits instantaneously. The advantage of our ROM simulation over previous simulations is instantaneous impingement detection, not requiring implant geometries to be analysed prior to use, and addressing the functional position of both the femur and pelvis


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 11 - 11
23 Feb 2023
Hardwick-Morris M Twiggs J Miles B Walter WL
Full Access

Iliopsoas tendonitis occurs in up to 30% of patients after hip resurfacing arthroplasty (HRA) and is a common reason for revision. The primary purpose of this study was to validate our novel computational model for quantifying iliopsoas impingement in HRA patients using a case-controlled investigation. Secondary purpose was to compare these results with previously measured THA patients. We conducted a retrospective search in an experienced surgeon's database for HRA patients with iliopsoas tendonitis, confirmed via the active hip flexion test in supine, and control patients without iliopsoas tendonitis, resulting in two cohorts of 12 patients. The CT scans were segmented, landmarked, and used to simulate the iliopsoas impingement in supine and standing pelvic positions. Three discrete impingement values were output for each pelvic position, and the mean and maximum of these values were reported. Cup prominence was measured using a novel, nearest-neighbour algorithm. The mean cup prominence for the symptomatic cohort was 10.7mm and 5.1mm for the asymptomatic cohort (p << 0.01). The average standing mean impingement for the symptomatic cohort was 0.1mm and 0.0mm for the asymptomatic cohort (p << 0.01). The average standing maximum impingement for the symptomatic cohort was 0.2mm and 0.0mm for the asymptomatic cohort (p << 0.01). Impingement significantly predicted the probability of pain in logistic regression models and the simulation had a sensitivity of 92%, specificity of 91%, and an AUC ROC curve of 0.95. Using a case-controlled investigation, we demonstrated that our novel simulation could detect iliopsoas impingement and differentiate between the symptomatic and asymptomatic cohorts. Interestingly, the HRA patients demonstrated less impingement than the THA patients, despite greater cup prominence. In conclusion, this tool has the potential to be used preoperatively, to guide decisions about optimal cup placement, and postoperatively, to assist in the diagnosis of iliopsoas tendonitis


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 12 - 12
23 Feb 2023
Hardwick-Morris M Twiggs J Miles B Balakumar J Walter WL
Full Access

Iliopsoas impingement occurs in between 5–30% of patients after hip arthroplasty and has been thought to only be caused by an oversized cup, cup malpositioning, or the depth of the psoas valley. However, no study has associated the relationship between preoperative measurements with the risk of impingement. This study sought to assess impingement between the iliopsoas and acetabular cup using a novel validated model to determine the risk factors for iliopsoas impingement. 413 patients received lower limb CT scans and lateral x-rays that were segmented, landmarked, and measured using a validated preoperative planning protocol. Implants were positioned according to the preference of ten experienced surgeons. The segmented bones were transformed to the standing reference frame and simulated with a novel computational model that detects impingement between the iliopsoas and acetabular cup. Definitions of patients at-risk and not at-risk of impingement were defined from a previous validation study of the simulation. At-risk patients were propensity score matched to not at-risk patients. 21% of patients were assessed as being at-risk of iliopsoas impingement. Significant differences between at-risk patients and not at-risk patients were observed in standing pelvic tilt (p << 0.01), standing femoral internal rotation (p << 0.01), medio-lateral centre-of-rotation (COR) change (p << 0.01), supine cup anteversion (p << 0.01), pre- to postoperative cup offset change (p << 0.001), postoperative gross offset (p = 0.009), and supero-inferior COR change (p = 0.02). Impingement between the iliopsoas and acetabular cup is under-studied and may be more common than is published in the literature. Previously it has been thought to only be related to cup size or positioning. However, we have observed significant differences between at-risk and not at-risk patients in additional measurements. This indicates that its occurrence is more complex than simply being related to cup position


Abstract. Optimal acetabular component position in Total Hip Arthroplasty is vital for avoiding complications such as dislocation and impingement, Transverse acetabular ligament (TAL) have been shown to be a reliable landmark to guide optimum acetabular cup position. Reports of iliopsoas impingement caused by acetabular components exist. The Psoas fossa (PF) is not a well-regarded landmark for Component positioning. Our aim was to assess the relationship of the TAL and PF in relation to Acetabular Component positioning. A total of 22 cadavers were implanted on 4 occasions with the an uncemented acetabular component. Measurements were taken between the inner edge of TAL and the base of the acetabular component and the distance between the lower end of the PF and the most medial end of TAL. The distance between the edge of the acetabular component and TAL was a mean of 1.6cm (range 1.4–18cm). The distance between the medial end of TAL and the lowest part of PF was a mean of 1.cm (range 1,3–1.8cm) It was evident that the edge of PF was not aligned with TAL. Optimal acetabular component position is vital to the longevity and outcome following THA. TAL provides a landmark to guide acetabular component position. However we feel the PF is a better landmark to allow appropriate positioning of the acetabular component inside edge of the acetabulum inside the bone without exposure of the component rim and thus preventing iliopsoas impingement at the psoas notch and resultant groin pain


Long femoral nails for neck of femur fractures and prophylactic fixation have a risk of anterior cortex perforation. Previous studies have demonstrated the radius of curvature (ROC) of a femoral nail influencing the finishing point of a nail and the risk of anterior cortex perforation. This study aims to calculate a patients femoral ROC using preoperative XR and CT and therefore nail finishing position. We conducted a retrospective study review of patients with long femoral cephalomedullary nailing for proximal femur fractures (OTA/AO 31(A) and OTA/AO 32) or impending pathological fractures at a level 1 trauma centre between January 1, 2015 and December 31, 2020 with both full length lateral X-ray and CT imaging. Femoral ROC was calculated on both imaging modalities. Outcomes measured including nail finishing position, anterior cortex encroachment and impingement. The mean femoral ROC was 1026mm on CT and 1244mm on XR. CT femoral ROC strongly correlated with nail finishing point with a spearmans coefficient of 0.77. Additionally, femurs with a ROC <1000mm were associated with a higher risk of anterior encroachment (OR 6.12) and femurs with a ROC <900mm were associated with a higher risk of anterior cortex impingement (OR 6.47). To our knowledge this is the first study to compare a measured femoral ROC to nail finishing position. The use of CT to measure femoral ROC and to a lesser extent XR was able to predict both nail finishing position and risk of anterior cortex encroachment. Preoperative XRs and CTs were able to identify patients with a small femoral ROC. This predicted patients at risk of anterior cortex impingement, anterior cortex encroachment and nail finishing position. We may be able to select femoral nails that resemble the native femoral ROC and mitigate the risk of anterior cortex perforation


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_15 | Pages 23 - 23
7 Nov 2023
Mulaudzi NP Mzayiya N Rachuene P
Full Access

Os acromiale is a developmental defect caused by failure of fusion of the anterior epiphysis of the acromion between the ages of 22 and 25. The prevalence of os acromiale in the general population ranges from 1.4% to 15%. Os acromiale has been reported as a contributory factor to shoulder impingement symptoms and rotator cuff injuries, despite being a common incidental observation. In this retrospective study, we examined the prevalence of os acromiale in black African patients with shoulder pain. We retrospectively reviewed the clinical records, radiographs, and magnetic resonance imaging (MRI) scans of 119 patients who presented with atraumatic and minor traumatic shoulder pain at a single institution over a one-year period. Anteroposterior, scapula Y-view, and axillary view plain images were initially evaluated for the presence of os acromiale, and this was corroborated with axial MRI image findings. Patients with verified os acromiale had their medical records reviewed to determine their first complaint and the results of their clinical examination and imaging examination. Radiographs and MRI on 24 patients (20%) revealed an osacromiale. This cohort had a mean age of 59.2 years, and there were significantly more females (65%) than males. Meso-acromion was identified as the most prevalent type (n=11), followed by pre-os acromion (n=7). All patients underwent bilateral shoulder x-rays, and 45 percent of patients were found to have bilateral meso- acromion. Most patients (70%) were reported to have unstable os acromiale with subacromial impingement symptoms, and nine patients (36%) had confirmed rotator cuff tears based on clinical and Mri findings. Surgery was necessary for 47% of the 24 patients with confirmed Os acromiale (arthroscopic surgery, n=7; open surgery, n=1) in order to treat their symptoms. The prevalence of os acromiale in our African patients with atraumatic shoulder symptoms is greater than that reported in the general population. Os acromiale is a rare condition that should always be considered when evaluating shoulder pain patients


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 11 - 11
10 Feb 2023
Boyle A George C MacLean S
Full Access

A larger radial tuberosity, and therefore a smaller radioulnar space, may cause mechanical impingement of the DBT predisposing to tear. We sought to investigate anatomic factors associated with partial DBT tears by retrospectively reviewing 3-T MRI scans of elbows with partial DBT tears and a normal elbow comparison group. 3-T MRI scans of elbows with partial DBT tears and elbows with no known pathology were reviewed retrospectively by two independent observers. Basic demographic data were collected and measurements of radial tuberosity length, radial tuberosity thickness, radio-ulnar space, and radial tuberosity-ulnar space were made using simultaneous tracker lines and a standardised technique. The presence or absence of enthesophytes and the presence of a single or double DBT were noted. 26 3-T MRI scans of 26 elbows with partial DBT tears and 30 3-T MRI scans of 30 elbows without pathology were included. Basic demographic data was comparable between the two groups. The tear group showed statistically significant larger mean measurements for radial tuberosity length (24.3mm vs 21.3mm, p=0.002), and radial tuberosity thickness (5.5mm vs 3.7mm, p=<0.0001. The tear group also showed statistically significant smaller measurements for radio-ulnar space (8.2mm vs 10.0mm, p=0.010), and radial tuberosity-ulnar space (7.2mm vs 9.1mm, p=0.013). There was a statistically significant positive correlation between partial DBT tears and presence of enthesophytes (p=0.007) as well as between partial DBT tears and having two discrete DBTs rather than a single or interdigitating tendon (p=<0.0001). Larger radial tuberosities, and smaller radio-ulnar and radial tuberosity-ulnar spaces are associated with partial DBT tears. This may be due to chronic impingement, tendon delamination and consequent weakness which ultimately leads to tears. Enthesophytes may be associated with tears for the same reason. Having two discrete DBTs that do not interdigitate prior to insertion is also associated with partial tears


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 103 - 103
23 Feb 2023
Gupta V Van Niekerk M Hirner M
Full Access

Scapular notching is a common problem following reverse shoulder arthroplasty (RSA). This is due to impingement between the humeral polyethylene cup and scapular neck in adduction and external rotation. Various glenoid component strategies have been described to combat scapular notching and enhance impingement-free range of motion (ROM). There is limited data available detailing optimal glenosphere position in RSA with an onlay configuration. The purpose of this study was to determine which glenosphere configurations would maximise impingement free ROM using an onlay RSA prosthesis. A three-dimensional (3D) computed tomography (CT) scan of a shoulder with Walch A1, Favard E0 glenoid morphology was segmented using validated software. An onlay RSA prosthesis was implanted and a computer model simulated external rotation and adduction motion of the virtual RSA prosthesis. Four glenosphere parameters were tested; diameter (36mm, 41mm), lateralization (0mm, 3mm, 6mm), inferior tilt (neutral, 5 degrees, 10 degrees), and inferior eccentric positioning (0.5mm, 1.5mm. 2.5mm, 3.5mm, 4.5mm). Eighty-four combinations were simulated. For each simulation, the humeral neck-shaft angle was 147 degrees and retroversion was 30 degrees. The largest increase in impingement-free range of motion resulted from increasing inferior eccentric positioning, gaining 15.0 degrees for external rotation and 18.8 degrees for adduction. Glenosphere lateralization increased external rotation motion by 13. 6 degrees and adduction by 4.3 degrees. Implanting larger diameter glenospheres increased external rotation and adduction by 9.4 and 10.1 degrees respectively. Glenosphere tilt had a negligible effect on impingement-free ROM. Maximizing inferior glenosphere eccentricity, lateralizing the glenosphere, and implanting larger glenosphere diameters improves impingement-free range of motion, in particular external rotation, of an onlay RSA prosthesis. Surgeons’ awareness of these trends can help optimize glenoid component position to maximise impingement-free ROM for RSA. Further studies are required to validate these findings in the context of scapulothoracic motion and soft tissue constraints


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_7 | Pages 30 - 30
1 Jul 2020
Faizan A Zhang J Scholl L
Full Access

Iliopsoas tendonitis after total hip arthroplasty (THA) can be a considerable cause of pain and patient dissatisfaction. The optimal cup position to avoid iliopsoas tendonitis has not been clearly established. Implant designs have also been developed with an anterior recess to avoid iliopsoas impingement. The purpose of this cadaveric study was to determine the effect of cup position and implant design on iliopsoas impingement. Bilateral THA was performed on three fresh frozen cadavers using oversized (jumbo) offset head center revision acetabular cups with an anterior recess (60, 62 and 66 mm diameter) and tapered wedge primary stems through a posterior approach. A 2mm diameter flexible stainless steel cable was inserted into the psoas tendon sheath between the muscle and the surrounding membrane to identify the location of the psoas muscle radiographically. CT scans of each cadaver were imported in an imaging software. The acetabular shells, cables as well as pelvis were segmented to create separate solid models of each. The offset head center shell was virtually replaced with an equivalent diameter hemispherical shell by overlaying the outer shell surfaces of both designs and keeping the faces of shells parallel. The shortest distance between each shell and cable was measured. To determine the influence of cup inclination and anteversion on psoas impingement, we virtually varied the inclination (30°/40°/50°) and anteversion (10°/20°/30°) angles for both shell designs. The CT analysis revealed that the original orientation (inclination/anteversion) of the shells implanted in 3 cadavers were as follows: Left1: 44.7°/23.3°, Right1: 41.7°/33.8°, Left2: 40/17, Right2: 31.7/23.5, Left3: 33/2908, Right3: 46.7/6.3. For the offset center shells, the shell to cable distance in all the above cases were positive indicating that there was clearance between the shells and psoas. For the hemispherical shells, in 3 out of 6 cases, the distance was negative indicating impingement of psoas. With the virtual implantation of both shell designs at orientations 40°/10°, 40°/20°, 40°/30° we found that greater anteversion helped decrease psoas impingement in both shell designs. When we analyzed the influence of inclination angle on psoas impingement by comparing wire distances for three orientations (30°/20°, 40°/20°, 50°/20°), we found that the effect was less pronounced. Further analysis comparing the offset head center shell to the conventional hemispherical shell revealed that the offset design was favored (greater clearance between the shell and the wire) in 17 out of 18 cases when the effect of anteversion was considered and in 15 out of 18 cases when the effect of inclinations was considered. Our results indicate that psoas impingement is related to both cup position and implant geometry. For an oversized jumbo cup, psoas impingement is reduced by greater anteversion while cup inclination has little effect. An offset head center cup with an anterior recess was effective in reducing psoas impingement in comparison to a conventional hemispherical geometry. In conclusion, adequate anteversion is important to avoid psoas impingement with jumbo acetabular shells and an implant with an anterior recess may further mitigate the risk of psoas impingement


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 86 - 86
1 Feb 2020
Dennis D Pierrepont J Bare J
Full Access

Introduction. Instability continues to be the number one reason for revision in primary total hip arthroplasty (THA). Commonly, impingement precedes dislocation, inducing a levering out the prosthetic head from the liner. Impingement can be prosthetic, bony or soft tissue, depending on component positioning and anatomy. The aim of this virtual study was to investigate whether bony or prosthetic impingement occurred first in well positioned THAs, with the hip placed in deep flexion and hyperextension. Methods. Twenty-three patients requiring THA were planned for a TriFit/Trinity ceramic-on-poly cementless construct using the OPS. TM. dynamic planning software (Corin, UK). The cups were sized to best fit the anatomy, medialised to sit on the acetabular fossa and orientated at 45° inclination and 25° anteversion when standing. Femoral components and head lengths were then positioned to reproduce the native anteversion and match the contralateral leg length and offset. The planned constructs were flexed and internally rotated until anterior impingement occurred in deep flexion [Fig. 1]. The type (bony or prosthetic), and location, of impingement was then recorded. Similarly, the hips were extended and externally rotated until posterior impingement occurred, and the type and location of impingement recorded [Fig. 2]. Patients with minimal pre-operative osteophyte were selected as a best-case scenario for bony impingement. Results. 6/23 (26%) patients were planned with only a 32mm articulation (<50mm cup size), with the remaining 17 patients all planned with both 32mm and 36mm articulations (≥50mm cup size). Anterior impingement was 26% prosthetic and 74% bony with the 32mm articulations, and 100% bony with the 36mm articulations. Bony impingement in deep flexion was exclusively anterior neck on anterior inferior iliac spine. Posterior impingement was 57% prosthetic and 43% bony with the 32mm articulations, and 41% prosthetic and 59% bony with the 36mm articulations. Bony impingement in hyperextension was exclusively lesser trochanter (LT) on ischium. Of the patients planned with both 32mm and 36mm articulations, there was a 14% increase in prosthetic impingement when a 32mm head was planned (35% and 21% respectively). Discussion. Impingement in THA usually precedes dislocation and should be avoided with appropriate component positioning. We found that in hyperextension, prosthetic and bony impingement were equally common. In deep flexion, impingement was almost exclusively bony. Further studies should investigate the effects of stem version, cup orientation, liner design, cup depth, native offset and retained osteophytes on the type of impingement in THA. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 28 - 28
1 Feb 2020
Kamada K Takahashi Y Tateiwa T Shishido T Masaoka T Pezzotti G Yamamoto K
Full Access

Introduction. Highly crosslinked, ultra-high molecular weight polyethylene (HXLPE) acetabular liners inherently have a risk of fatigue failure associated with femoral neck impingement. One of the potential reasons for liner failure was reported as crosslinking formulations of polyethylene, increasing the brittleness and structural rigidity. In addition, the acetabular component designs greatly affect the mechanical loading scenario, such as the offset (lateralized) liners with protruded rim above the metal shells, which commonly induce a weak resistance to rim impingement. The purpose of the present study was to compare the influence of the liner offset length on the impingement resistance in the annealed (first generation) and vitamin E-blended (second-generation) HXLPE liners with a commercial design. Materials and Methods. The materials tested were the 95-kGy irradiated annealed GUR1020, and the 300-kGy irradiated vitamin E-blended GUR1050 HXLPE offset liners, which were referred to as “20_95” and “50E_300”, respectively. These liners had 2, 3, 4-mm rim offset, 2.45-mm rim thickness, and 36-mm internal diameter. Their rims were protruded above the metal rim at 2, 3, 4mm. Rim impingement testing was performed using an electrodynamic axial-torsional machine. The cyclic impingement load of 25–250N was applied on the rims through the necks of the femoral stems at 1Hz. The rotational torque was simultaneously generated by swinging the stem necks on the rims at 1Hz and its rotational angle was set at the range of 0–10˚. The percent crystallinity was analyzed on the as-received (intact) and impinged HXLPE acetabular rims by confocal Raman microspectroscopy. Results. The number of cycles to failure was dependent on the offset length (2, 3, 4-mm) in 20_95 and 50E_300 liners. Our results showed that the shorter the rim offset, the shorter the number of cycles to failure. In both HXLPEs, accumulation of impingement damages significantly decreased crystallinity in their near-surfaces, indicating the occurrence of crystallographic breakdown. In each offset length tested, the fracture always occurred much earlier in 50E_300 than 20_95. However, the magnitudes of the microstructural changes at the time of failure were much less in 50E_300 than 20_95. Conclusions. Although it is known that vitamin E blend into HXLPE can improve the fatigue resistance of HXLPE, the impingement resistance of 50E_300 was lower than vitamin-E free 20_95, indicating a larger negative contribution of high-dose radiation (300kGy) over a positive contribution of the vitamin E blend in 50E_300. Our results implied that the reduction of the protruded rim length in the offset liners may increase the neck-rim contact stresses at the time of impingement, causing a decrease in the fatigue durability. Therefore, if HXLPE offset liner is used, surgeons should take special care in maximizing the volume of the protruded lip section


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 95 - 95
10 Feb 2023
Mowbray J Frampton C Maxwell R Hooper G
Full Access

Cementless fixation is an alternative to cemented unicompartmental knee replacement (UKR), with several advantages over cementation. This study reports on the 15-year survival and 10-year clinical outcomes of the cementless Oxford unicompartmental knee replacement (OUKR). This prospective study describes the clinical outcomes and survival of first 693 consecutive cementless medial OUKRs implanted in New Zealand. The sixteen-year survival was 89.2%, with forty-six knees being revised. The commonest reason for revision was progression of arthritis, which occurred in twenty-three knees, followed by primary dislocation of the bearing, which occurred in nine knees. There were two bearing dislocations secondary to trauma and a ruptured ACL, and two tibial plateau fractures. There were four revisions for polyethylene wear. There were four revisions for aseptic tibial loosening, and one revision for impingement secondary to overhang of the tibial component. There was only one revision for deep infection and one revision where the indication was not stated. The mean OKS improved from 23.3 (7.4 SD) to 40.59 (SD 6.8) at a mean follow-up of sixteen years. In conclusion, the cementless OUKR is a safe and reproducible procedure with excellent sixteen-year survival and clinical outcomes


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 51 - 51
23 Feb 2023
Full Access

Dual mobility is a French concept that appeared in the 1970s and was initially intended to reduce dislocation rates. In recent years, this concept has evolved with new HA titanium spray coatings, new external macrostructures, and better-quality polyethylene. This has allowed to extend the indications to younger and therefore active populations. The objective of our work is to analyze at least 10 years a homogeneous and continuous series of 170 primary total hip replacements associating a latest generation Novae Sunfit. ®. dual mobility cup with a straight femoral stem. Only primary arthroplasties for osteoarthritis or necrosis were included. Total hip arthroplasty was always performed through a posterolateral approach. All patients had regular clinical and radiological follow-up. The average follow-up in our series was 11.5 years. The average age of the population is 71 years. At the last follow-up, there were 17 deaths, 6 losses to follow up and 9 adverse events, including 1 cup change for psoas impingement and 1 dislocation. The low rate of dislocation at 11 years confirms the high stability of the dual mobility, which should be recommended for primary procedure for patients at high risk of postoperative instability. The lack of intraprosthetic dislocation due to wear at 11 years of follow-up highlights the good quality of the latest generation of polyethylene, and the need to combine high-polished surfaces and a refined femoral neck with a dual mobility cup. Finally, the lack of aseptic loosening confirms the quality of the secondary fixation of these implants and justifies their wider use in all patients


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 52 - 52
23 Feb 2023
Full Access

THA in patients with acetabular bone defects is associated with a high risk of dislocation. Dual mobility (DM) cups are known to prevent and treat chronic instability. The aim of this study was to evaluate the dislocation rate and survival of jumbo DM cups. This was a retrospective, continuous, multicenter study of all the cases of jumbo DM cup implantation between 2010 and 2017 in patients with acetabular bone loss (Paprosky 2A: 46%, 2B: 32%, 2C: 15% and 3A: 6%). The indications for implantation were revisions for aseptic loosening of the cup (n=45), aseptic loosening of the femoral stem (n=3), bipolar loosening (n=11), septic loosening (n=10), periprosthetic fracture (n=5), chronic dislocation (n=4), intraprosthetic dislocation (n=2), cup impingement (n=1), primary posttraumatic arthroplasty (n=8), and acetabular dysplasia (n=4). The jumbo cups used were COPTOS TH (SERF), which combines press-fit fixation with supplemental fixation (acetabular hook, two superior flanges with one to four screws, two acetabular pegs). A bone graft was added in 74 cases (80%). The clinical assessment consisted of the Harris hip score. The primary endpoint was surgical revision for aseptic acetabular loosening or the occurrence of a dislocation episode. In all, 93 patients were reviewed at a mean follow-up of 5.3 ± 2.3 years [0, 10]. As of the last follow-up, the acetabular cup had been changed in five cases: three due to aseptic loosening (3.2%) and two due to infection (2.1%). The survivorship free of aseptic loosening was 96.8%. Three patients (3%) suffered a dislocation. At the last follow-up visit, the mean HSS scores were 72.15, (p < 0.05). Use of a jumbo DM cup in cases of acetabular bone defects leads to satisfactory medium-term results with low dislocation and loosening rates


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 50 - 50
10 Feb 2023
Eagles A Erian C Kermeci S Lovell D Weinrauch P
Full Access

Arthroscopic hip surgery is increasingly common in Australia. Hip arthroscopy is indicated for a range of diagnostic and therapeutic purposes, including labral tears, capsular laxity and femoral-acetabular impingement (FAI). Despite this, previous cohort studies aiming to characterise hip pathology seen on arthroscopic examination are mostly limited to patients with known diagnoses of FAI. Therefore, little is known of the native articular wear patterns encountered in other disease states. Therefore, we aimed to define common osteochondral wear patterns for a cohort of patients managed via hip arthroscopy. We retrospectively analysed intraoperative data for 1127 patients managed via hip arthroscopy between 2008 and 2013, for either therapeutic or diagnostic purposes. Intraoperative data was categorized by location (A-E as defined by Fontana et al. 2016) and chondral damage (0-4 scale as defined by Beck et al. 2005) with respect to both acetabulum and femoral head. Data for 1127 patients were included. Location of acetabular chondral pathology was variable with locations C. 1. and D. 1. representing the most common regions of damage. Labral tears predominated in locations C and D. Femoral chondral pathology was evenly distributed. The degree of femoral chondral injury was predominantly grade 1, whilst acetabular wear was evenly distributed. Large proportions of wear were observed at the peripheral superior and anterior regions of the lunate surface of the acetabulum in keeping with prior works. However, we observed higher rates of central wear and lower rates of grade 4 acetabular damage extending into superior/posterior zones, in our cohort. Our work characterises common articular wear patterns encountered at the time of hip arthroscopy. Further inquiry into the natural history of osteochondral lesions is needed to better understand and manage these conditions


Full Access

We report the outcome of 320 primary Total Hip Arthroplasties (THA) with minimum 10-year follow-up (range 10–17 years, mean 12.6 years), performed by a single surgeon in Tauranga New Zealand, with the Exeter Contemporary Flanged all-polyethylene cup and Exeter femoral stem via a posterior approach. The aim of the study is to compare the results with the published results from the design centre and create a baseline cohort for further outcomes research in this centre. All patients were prospectively followed at 6 weeks, 1 year, 5 years, 10 years, (and 15 years when available). Of 333 cases that matched the inclusion criteria, 13 procedures in 12 patents were excluded because of concomitant bone grafting and/or supplementary cage fixation, leaving 320 primary THA procedures in 280 patients, including 26 bilateral procedures in 13 patients. Mean follow-up of the surviving cases was 12.6 (range 5.0-17.1) years. There were 12 revisions – 2 for fracture, 5 for instability, 1 for impingement pain and 4 for infection. There were no revisions for aseptic cup loosening. Kaplan-Meier survivorship with revision for aseptic loosening as the endpoint was 100% at 15.0 years (with minimum 40 cases remaining at risk). All-cause acetabular revision in 12 cases result in a Kaplan-Meier survival of 95.9% (95% CI: 93.5 to 98.3%). Cemented THA with the Exeter Contemporary Flanged cup and the Exeter stem is a durable combination with results that can be replicated outside of the design centre. The Exeter Contemporary Flanged cup has excellent survivorship at 15 years when used with the Exeter stem. Cemented THA with well-proven components should be considered the benchmark against which newer designs and materials should be compared


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_11 | Pages 42 - 42
1 Nov 2022
Kumar K Van Damme F Audenaert E Khanduja V Malviya A
Full Access

Abstract. Introduction. Recurrent groin pain following periacetabular osteotomy (PAO) is a challenging problem. The purpose of our study was to evaluate the position and dynamics of the psoas tendon as a potential cause for recurrent groin pain following PAO. Methods. Patients with recurrent groin pain following PAO were identified from a single surgeon series. A total of 13 patients with 18 hips (4.7%) out of a 386 PAO, had recurrent groin pain. Muscle path of the psoas tendon was accurately represented using 3D models from CT data were created with Mimics software. A validated discrete element model using rigid body springs was used to predict psoas tendon movement during hip circumduction and walking. Results. Five out of the 18 hips did not show any malformations at the osteotomy site. Thirteen hips (72%) showed malformation secondary to callus at the superior pubic ramus. These were classified into: osteophytes at the osteotomy site, hypertrophic callus or non-union and malunion at the osteotomy. Mean minimal distance of the psoas tendon to osteophytes was found to be 6.24 mm (n=6) and to the osteotomy site was 14.18 mm (n=18). Conclusions. Recurrent groin pain after PAO needs a thorough assessment. One need to have a high suspicion of psoas issues as a cause. 3D CT scan may be necessary to identify causes related to healing of the pubic osteotomy. Dynamic ultrasound of the psoas psoas tendon may help in evaluating for psoas impingement as a cause of recurrent groin pain in these cases


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_15 | Pages 26 - 26
7 Nov 2023
de Wet J Gray J Verwey L Dey R du Plessis J Vrettos B Roche S
Full Access

The musculoskeletal (MSK) profiles of water polo players and other overhead athletes has been shown to relate to injury and throwing performance (TP). There have been no robust studies conducted on the MSK profiles and the variables affecting TP amongst female, adolescent, elite water polo players. A prospective quantitative cohort design was conducted amongst eighty-three female adolescent, elite water polo players (range 14–19 years). All participants filled out the Kerlan-Jobe Orthopaedic Clinic questionnaire, followed by a battery of screening tests aimed to identify possible MSK factors affecting TP. Pain provocation tests, range of motion (ROM), upward scapula rotation (USR), strength and pectoralis minor length measurements were all included. Participants also performed throwing speed (TS) and throwing accuracy (TA) tests. All the data collected were grouped together and analysed using SPSS 28.0. The condition for statistical significance was set as p <0.05. Multi-collinearity was tested for among variables to find out inter-variable correlations. Finally, a multiple regression analysis was performed. The mean KJOC score was 82.55 ± 14.96. 26.5% tested positive for at least one of the impingement tests. The MSK profile revealed decreased internal rotation ROM, increased external rotation ROM, a downwardly rotated scapula, weak external rotators, weak serratus anterior strength, strong lower trapezius and gluteus medius strength and a shorter pectoralis minor length all on the dominant side. Age, pectoralis minor length, upper trapezius and serratus anterior strength as well as upward scapula rotation were all positively correlated with TS, while sitting height, upper trapezius and serratus anterior strength and glenohumeral internal rotation ROM were positively correlated with TA. Multiple MSK parameters were found to be related to TS and TA in elite, adolescent water polo players


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_4 | Pages 9 - 9
3 Mar 2023
Zahid A Mohammed R
Full Access

Anterior cervical discectomy and fusion (ACDF) is a well-established spinal operation for cervical disc degeneration disease with neurological compromise. The procedure involves an anterior approach to the cervical spine with discectomy to relieve the pressure on the impinged spinal cord to slow disease progression. The prosthetic cage replaces the disc and can be inserted stand-alone or with an anterior plate that provides additional stability. The literature demonstrates that the cage-alone (CA) is given preference over the cage-plate (CP) technique due to better clinical outcomes, reduced operation time and resultant morbidity. This retrospective case-controlled study compared CA versus CP fixation used in single and multilevel anterior cervical discectomy and fusion for myelopathy in a tertiary centre in Wales. A retrospective clinico-radiological analysis was undertaken, following ACDF procedures over seven years in a single tertiary centre. Inclusion criteria were patients over 18 years of age with cervical myelopathy who had at least six-month follow-up data. SPSS was used to identify any statistically significant difference between both groups. The data were analysed to evaluate the consistency of our findings in comparison to published literature. Eighty-six patients formed the study cohort; 28 [33%] underwent ACDF with CA and 58 [67%] with CP. The patient demographics were similar in both groups, and fusion was observed in all individuals. There was no statistical difference between the two constructs when assessing subsidence, clinical complication (dysphagia, dysphonia, infection), radiological parameters and reoperations. However, a more significant percentage [43% v 61%] of patients improved their cervical lordosis angle with CP treatment. Furthermore, the study yielded that surgery to upper cervical levels results in a higher incidence of dysphagia [65% v 35%]. Finally, bony growth across the cage was observed on X-ray in 12[43%] patients, a unique finding not mentioned in the literature previously. Our study demonstrates no overall difference between the two groups, and we recommend careful consideration of individual patient factors when deciding what construct to choose