Advertisement for orthosearch.org.uk
Results 1 - 6 of 6
Results per page:
Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_2 | Pages 38 - 38
1 Jan 2014
Gadd R Barwick T Paling E Davies M Blundell C
Full Access

Introduction:

Prompted by the success of hip and knee arthroplasty, total ankle replacement (TAR) has become increasingly popular as a treatment for end stage arthritic complaints of the ankle. Glazebrook et al proposed a three grade classification of complications to assist prediction of early implant failure. We have compared the experience of a tertiary referral centre in the UK to Glazebrook's proposed system.

Method:

A retrospective review of the Sheffield Foot and Ankle Unit TAR database was performed from 1995 to 2010. All complications were recorded and categorised using Glazebrook's proposed system. Glazebrook described eight main complications of increasing severity. Low grade complications; Post operative bone fracture, Intra-operative bone fracture and wound healing problems were very unlikely to lead to revision. Medium grade complications; technical error and subsidence, lead to failure < 50% of the time. High grade complications; deep infection, aseptic loosening and implant failure lead to revision > 50% of the time.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 985 - 992
1 Sep 2023
Arshad Z Haq II Bhatia M

Aims

This scoping review aims to identify patient-related factors associated with a poorer outcome following total ankle arthroplasty (TAA).

Methods

A scoping review was performed using the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. A computer-based literature search was performed in PubMed, Embase, Cochrane trials, and Web of Science. Two reviewers independently performed title/abstract and full-text screening according to predetermined selection criteria. English-language original research studies reporting patient-related factors associated with a poorer outcome following TAA were included. Outcomes were defined as patient-reported outcome measures (PROMs), perioperative complications, and failure.


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 475 - 479
1 Apr 2018
Ali AA Forrester RA O’Connor P Harris NJ

Aims. The aim of this study was to present a series of patients with aseptic failure of a total ankle arthroplasty (TAA) who were treated with fusion of the hindfoot using a nail. Patients and Methods. A total of 23 TAAs, in 22 patients, were revised for aseptic loosening and balloon osteolysis to a hindfoot fusion by a single surgeon (NH) between January 2012 and August 2014. The procedure was carried out without bone graft using the Phoenix, Biomet Hindfoot Arthrodesis Nail. Preoperative investigations included full blood count, CRP and ESR, and radiological investigations including plain radiographs and CT scans. Postoperative plain radiographs were assessed for fusion. When there was any doubt, CT scans were performed. Results. The mean follow-up was 13.9 months (4.3 to 37.2). Union occurred at the tibiotalar joint in 22 ankles (95.6%) with one partial union. Union occurred at the subtalar joint in 20 ankles (87%) of cases with two nonunions. The nail broke in one patient with a subtalar nonunion and revision was undertaken. The only other noted complication was one patient who suffered a stress fracture at the proximal aspect of the nail, which was satisfactorily treated conservatively. Conclusion. This study represents the largest group of patients reported to have undergone revision TAA to fusion of the hindfoot with good results. Cite this article: Bone Joint J 2018;100-B:475–9


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 472 - 478
1 Apr 2022
Maccario C Paoli T Romano F D’Ambrosi R Indino C Federico UG

Aims

This study reports updates the previously published two-year clinical, functional, and radiological results of a group of patients who underwent transfibular total ankle arthroplasty (TAA), with follow-up extended to a minimum of five years.

Methods

We prospectively evaluated 89 patients who underwent transfibular TAA for end-stage osteoarthritis. Patients’ clinical and radiological examinations were collected pre- and postoperatively at six months and then annually for up to five years of follow-up. Three patients were lost at the final follow-up with a total of 86 patients at the final follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1540 - 1547
1 Nov 2010
Kim BS Knupp M Zwicky L Lee JW Hintermann B

We report the clinical and radiological outcome of total ankle replacement performed in conjunction with hindfoot fusion or in isolation. Between May 2003 and June 2008, 60 ankles were treated with total ankle replacement with either subtalar or triple fusion, and the results were compared with a control group of 288 ankles treated with total ankle replacement alone.

After the mean follow-up of 39.5 months (12 to 73), the ankles with hindfoot fusion showed significant improvement in the mean visual analogue score for pain (p < 0.001), the mean American Orthopaedic Foot and Ankle Society score (p < 0.001), and the mean of a modified version of this score (p < 0.001). The mean visual analogue pain score (p = 0.304) and mean modified American Orthopaedic Foot and Ankle Society score (p = 0.119) were not significantly different between the hindfoot fusion and the control groups. However, the hindfoot fusion group had a significantly lower mean range of movement (p = 0.009) and a higher rate of posterior focal osteolysis (p = 0.04). Both groups showed various complications (p = 0.131) and failure occurring at a similar rate (p = 0.685).

Subtalar or triple fusion is feasible and has minimal adverse effects on ankles treated with total ankle replacement up to midterm follow-up. The clinical outcome of total ankle replacement when combined with hindfoot fusion is comparable to that of ankle replacement alone. Thus, hindfoot fusion should be performed in conjunction with total ankle replacement when indicated.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 69 - 74
1 Jan 2009
Wood PLR Sutton C Mishra V Suneja R

We describe the results of a randomised, prospective study of 200 ankle replacements carried out between March 2000 and July 2003 at a single centre to compare the Buechel-Pappas (BP) and the Scandinavian Total Ankle Replacement (STAR) implant with a minimum follow-up of 36 months. The two prostheses were similar in design consisting of three components with a meniscal polyethylene bearing which was highly congruent on its planar tibial surface and on its curved talar surface. However, the designs were markedly different with respect to the geometry of the articular surface of the talus and its overall shape.

A total of 16 ankles (18%) was revised, of which 12 were from the BP group and four of the STAR group. The six-year survivorship of the BP design was 79% (95% confidence interval (CI) 63.4 to 88.5 and of the STAR 95% (95% CI 87.2 to 98.1). The difference did not reach statistical significance (p = 0.09). However, varus or valgus deformity before surgery did have a significant effect) (p = 0.02) on survivorship in both groups, with the likelihood of revision being directly proportional to the size of the angular deformity. Our findings support previous studies which suggested that total ankle replacement should be undertaken with extreme caution in the presence of marked varus or valgus deformity.