Aims. Following the resection of an extensive amount of bone in the
treatment of a tumour, the residual segment may be insufficient
to accept a standard length intramedullary
Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes. This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.Aims
Methods
For rare cases when a tumour infiltrates into the hip joint, extra-articular resection is required to obtain a safe margin. Endoprosthetic reconstruction following tumour resection can effectively ensure local control and improve postoperative function. However, maximizing bone preservation without compromising surgical margin remains a challenge for surgeons due to the complexity of the procedure. The purpose of the current study was to report clinical outcomes of patients who underwent extra-articular resection of the hip joint using a custom-made osteotomy guide and 3D-printed endoprosthesis. We reviewed 15 patients over a five-year period (January 2017 to December 2022) who had undergone extra-articular resection of the hip joint due to malignant tumour using a custom-made osteotomy guide and 3D-printed endoprosthesis. Each of the 15 patients had a single lesion, with six originating from the acetabulum side and nine from the proximal femur. All patients had their posterior column preserved according to the surgical plan.Aims
Methods
We reviewed 25 patients in whom a MUTARS megaprosthesis with a conical fluted stem had been implanted. There were three types of stem: a standard stem was used in 17 cases (three in the proximal femur, nine in the distal femur and five proximal tibia), a custom-made proximal femoral stem in four cases and a custom-made distal femoral stem in four cases. The mean age of the patients was 40.1 years (17 to 70) and the mean follow-up was for 2.5 years (0.9 to 7.4). At follow-up two patients had died from their disease: one was alive with disease and 22 were disease-free. One of 23 prostheses had been removed for infection and another revised to a
The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection. From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.Aims
Methods
The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases. This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for loosening and failure. Patient and implant survival were assessed using a competing risk analysis. Secondary parameters included primary malignancy, oncological treatment, surgical factors, length of stay in hospital, and postoperative complications.Aims
Methods
Hydroxyapatite (HA)-coated collars have been shown to reduce aseptic loosening of massive endoprostheses following primary surgery. Limited information exists about their effectiveness in revision surgery. The aim of this study was to radiologically assess osteointegration to HA-coated collars of cemented massive endoprostheses following revision surgery. Retrospective review of osseointegration frequency, pattern, and timing to a specific HA-coated collar on massive endoprostheses used in revision surgery at our tertiary referral centre between 2010 to 2017 was undertaken. Osseointegration was radiologically classified on cases with a minimum follow-up of six months.Aims
Methods
We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
Massive endoprostheses using a
Limb salvage in bone tumour patients replaces the bone with massive segmental prostheses where achieving bone integration at the shoulder of the implant through extracortical bone growth has been shown to prevent loosening. This study investigates the effect of multidrug chemotherapy on extracortical bone growth and early radiological signs of aseptic loosening in patients with massive distal femoral prostheses. A retrospective radiological analysis was performed on adult patients with distal femoral arthroplasties. In all, 16 patients were included in the chemotherapy group with 18 patients in the non-chemotherapy control group. Annual radiographs were analyzed for three years postoperatively. Dimensions of the bony pedicle, osseointegration of the hydroxyapatite (HA) collar surface, bone resorption at the implant shoulder, and radiolucent line (RLL) formation around the cemented component were analyzed.Aims
Methods
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).Aims
Patients and Methods
The aim of this study was to report the results of custom-made endoprostheses with extracortical plates plus or minus a short, intramedullary stem aimed at preserving the physis after resection of bone sarcomas in children. Between 2007 and 2017, 18 children aged less than 16 years old who underwent resection of bone sarcomas, leaving ≤ 5 cm of bone from the physis, and reconstruction with a custom-made endoprosthesis were reviewed. Median follow-up was 67 months (interquartile range 45 to 91). The tumours were located in the femur in 11 patients, proximal humerus in six, and proximal tibia in one.Aims
Patients and Methods
The aim of this study was to present the long-term surgical outcomes, complications, implant survival, and causes of implant failure in patients treated with the modified Harrington procedure using antegrade large diameter pins. A cohort of 50 consecutive patients who underwent the modified Harrington procedure for periacetabular metastasis or haematological malignancy between January 1996 and April 2018 were studied. The median follow-up time for all survivors was 3.2 years (interquartile range 0.9 to 7.6 years).Aims
Patients and Methods
Aim. The use of megaprostheses for knee reconstruction after distal femur resection in young bone sarcoma patients has become popular since early ′80. The authors reviewed their experience with different distal femur megaprostheses in children. Method. Clinico-radiographic evolution in a consecutive series of 113 children, that had implanted below age 15 (range 6-14) a distal femur megaprosthesis in the period 1984-2007, was analized. A modular implant was used in 97 cases with uncemented femoral stem (three different models along the period). The implant presented fixed-hinge joint in 78 cases while rotating-hinge knee was utilized in 19 cases. In 39 cases the fixed-hinge joint had a tibial component with a polished stem to allow the residual growth of proximal tibia; in two cases a mechanically extendable prosthesis was used. A custom-made noninvasive extendable prosthesis with
We have made a retrospective study of 1001 custom-made prostheses used as replacements after surgery for bone tumours. There were 493 distal femoral, 263 proximal femoral and 245 proximal tibial prostheses. Aseptic loosening was shown to be the principal mode of failure of the implants, and 71 patients had revision for aseptic loosening of a
Dislocation rates are reportedly lower in patients requiring
proximal femoral hemiarthroplasty than for patients undergoing hip
arthroplasty for neoplasia. Without acetabular replacement, pain
due to acetabular wear necessitating revision surgery has been described.
We aimed to determine whether wear of the native acetabulum following
hemiarthroplasty necessitates revision surgery with secondary replacement
of the acetabulum after proximal femoral replacement (PFR) for tumour
reconstruction. We reviewed 100 consecutive PFRs performed between January 2003
and January 2013 without acetabular resurfacing. The procedure was
undertaken in 74 patients with metastases, for a primary bone tumour
in 20 and for myeloma in six. There were 48 male and 52 female patients,
with a mean age of 61.4 years (19 to 85) and median follow-up of
two years (interquartile range (IQR) 0.5 to 3.7 years). In total,
52 patients presented with a pathological fracture and six presented
with failed fixation of a previously instrumented pathological fracture.Aims
Patients and Methods
Massive endoprostheses rely on extra-cortical bone bridging (ECBB)
to enhance fixation. The aim of this study was to investigate the
role of selective laser sintered (SLS) porous collars in augmenting
the osseointegration of these prostheses. The two novel designs of porous SLS collars, one with small pores
(Ø700 μm, SP) and one with large pores (Ø1500 μm, LP), were compared
in an ovine tibial diaphyseal model. Osseointegration of these collars
was compared with that of a clinically used solid, grooved design
(G). At six months post-operatively, the ovine tibias were retrieved and
underwent radiological and histological analysis.Aims
Materials and Methods
Excision of the proximal femur for tumour with
prosthetic reconstruction using a bipolar femoral head places a considerable
load on the unreplaced acetabulum. We retrospectively reviewed the changes which occur around the
affected hip joint by evaluating the post-operative radiographs
of 65 consecutive patients who underwent proximal prosthetic arthroplasty
of the femur, and in whom an acetabular component had not been used.
There were 37 men and 28 women with a mean age of 57.3 years (17
to 93). Radiological assessment included the extent of degenerative
change in the acetabulum, heterotopic ossification, and protrusio
acetabuli. The mean follow-up was 9.1 years (2 to 11.8). Degenerative changes
in the acetabulum were seen in three patients (4.6%), Brooker grade
1 or 2 heterotopic ossification in 17 (26%) and protrusion of the
prosthetic head in nine (13.8%). A total of eight patients (12.3%) needed a revision. Five were
revised to the same type of prosthesis and three (4.6%) were converted
to a total hip arthroplasty. We conclude that radiological evidence of degenerative change,
heterotopic ossification and protrusion occur in a few patients
who undergo prosthetic arthroplasty of the proximal femur for tumour.
The limited extent of these changes and the lack of associated symptoms
do not justify the routine arthroplasty of the acetabulum in these patients. Cite this article:
We determined the efficacy of a devitalised autograft
(n = 13) and allograft (n = 16) cortical strut bone graft combined
with long-stem endoprosthetic reconstruction in the treatment of
massive tumours of the lower limb. A total of 29 patients (18 men:11
women, mean age 20.1 years (12 to 45) with a ratio of length of
resection to that of the whole prosthesis of >
50% were treated
between May 2003 and May 2012. The mean follow-up was 47 months
(15 to 132). The stem of the prosthesis was introduced through bone
graft struts filled with cement, then cemented into the residual
bone. Bone healing was achieved in 23 patients (86%). The mean Musculoskeletal
Tumour Society functional score was 85% (57 to 97). The five-year
survival rate of the endoprostheses was 81% (95% confidence intervals
67.3 to 92.3). The mean length of devitalised autografts and allografts
was 8.6 cm (5 to 15), which increased the ratio of the the length
of the stem of the prosthesis to that of the whole length of the
prosthesis from a theoretical 35% to an actual 55%. Cortical strut bone grafting and long-stem endoprosthetic reconstruction
is an option for treating massive segmental defects following resection
of a tumour in the lower limb. Patients can regain good function
with a low incidence of aseptic loosening. The strut graft and the
residual bone together serve as a satisfactory bony environment
for a revision prosthesis, if required, once union is achieved. Cite this article:
Giant cell tumour is the most common aggressive
benign tumour of the musculoskeletal system and has a high rate of
local recurrence. When it occurs in proximity to the hip, reconstruction
of the joint is a challenge. Options for reconstruction after wide
resection include the use of a megaprosthesis or an allograft-prosthesis
composite. We performed a clinical and radiological study to evaluate
the functional results of a proximal femoral allograft-prosthesis
composite in the treatment of proximal femoral giant cell tumour
after wide resection. This was an observational study, between 2006
and 2012, of 18 patients with a mean age of 32 years (28 to 42)
and a mean follow-up of 54 months (18 to 79). We achieved excellent
outcomes using Harris Hip Score in 13 patients and a good outcome
in five. All allografts united. There were no complications such
as infection, failure, fracture or resorption of the graft, or recurrent
tumour. Resection and reconstruction of giant cell tumours with
proximal femoral allograft–prosthesis composite is a better option
than using a prosthesis considering preservation of bone stock and excellent
restoration of function. A good result requires demanding bone banking techniques, effective
measures to prevent infection and stability at the allograft-host
junction. Cite this article: