This study describes the Osseointegration Group of Australia’s
Accelerated Protocol two-stage strategy (OGAAP-1) for the osseointegrated
reconstruction of amputated limbs. We report clinical outcomes in 50 unilateral trans-femoral amputees
with a mean age of 49.4 years (24 to 73), with a minimum one-year
follow-up. Outcome measures included the Questionnaire for persons
with a Trans-Femoral Amputation, the health assessment questionnaire
Short-Form-36 Health Survey, the Amputation Mobility Predictor scores
presented as K-levels, 6 Minute Walk Test and timed up and go tests.
Adverse events included soft-tissue problems, infection, fractures
and failure of the implant.Aims
Patients and Methods
A silver-containing hydroxyapatite (Ag-HA) coating has been developed using thermal spraying technology. We evaluated the osteoconductivity of this coating on titanium (Ti) implants in rat tibiae in relation to bacterial infection in joint replacement. At 12 weeks, the mean affinity indices of bone formation of a Ti, an HA, a 3%Ag-HA and a 50%Ag-HA coating were 97.3%, 84.9%, 81.0% and 40.5%, respectively. The mean affinity indices of bone contact of these four coatings were 18.8%, 83.7%, 77.2% and 40.5%, respectively. The indices of bone formation and bone contact around the implant of the 3%Ag-HA coating were similar to those of the HA coating, and no significant differences were found between them (bone formation, p = 0.99; bone contact, p = 0.957). However, inhibition of bone formation was observed with the 50%Ag-HA coating. These results indicate that the 3%Ag-HA coating has low toxicity and good osteoconductivity, and that the effect of silver toxicity on osteoconductivity depends on the dose.