Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 116 - 116
1 Jan 2016
Burgett-Moreno M Medina E Burton P Donaldson T Clarke I
Full Access

A 35-year-old female (age 35Yrs) had primary MOM total hip arthroplasty (THA) in 2008. At 8 months this patient postoperatively developed headaches, memory loss, vertigo, and aura-like symptoms that progressed to seizures. At 18 months review, she complained of progressive hip pain, a popping sensation and crepitus with joint motion. This patient weighed 284lbs with BMI of 38.5. Radiographs revealed the cup had 55° inclination, 39° anteversion (Fig. 1). Metal ion concentrations were high (blood: Co=126 mcg/L, Cr= 64mcg/L). Revision was performed in November 2010 A dark, serous fluid was observed, along with synovitis. The implants were well fixed and the femoral head could not be removed; thus the stem was removed by femoral osteotomy. With the head fused on this femoral stem, for the 1st time it was possible to precisely determine the habitual patterns of MOM wear relative to her in-vivo function. We investigated (1) size and location of wear patterns and (2) signs of cup-stem impingement to help explain her symptoms developed over 32 months follow-up.

The retrieved MOM was a Magnum™ with head diameter 50mm and 50×56mm cup (Biomet). This was mounted on a Taperloc™ lateralized porous-coated stem. Components were examined visually and wear damage mapped by stereo-microscopy, interferometry, CMM, SEM, and EDS. Main-wear zone (MWZ) areas were calculated using standard spherical equations1 and centroidal vectors determined.

The head-cup mismatch was 427um with the cup revealing a form factor of 228um. The cup showed wear area of 1275mm² that extended up to the cup rim over 150°arc. The cup rim was worn thin over a 90° arc with loss of cup bevel. The head showed an elliptical wear area of 2200mm2 located centrally on the superior-medial surface (ellipsoidal ratio ×1.2). Compared to the hemispherical surface (50mm: hemi-area = 3927mm2), the worn area represented hemi-area ratio of 56%. The centroidal vectors measured 8° anterior and 24° superior to the head's polar axis (Fig. 2). Stripe wear damage revealed multiple impingement sites. SEM and EDS revealed stripes were contaminated by metal transfer from the stainless-steel instruments used at revision. The main impingement position was identified (Fig. 3) indicating the site of repetitive subluxations whereby the subluxing head thinned the cup, i.e. “edge wear”.

Cup and head wear patterns corresponded well, reinforcing our definition of the MWZ locations in vivo. The femoral MWZ was centrally located superiorly and medially with respect to the polar axis of the femoral neck and head. The noted impingement position indicated this patient had experienced repetitive subclinical subluxations (RSS).2 The taper inside the fused head may also have been a contributory factor that we cannot ignore. Nevertheless her excessive cup thinning was likely a result of a steep cup and considerable anteversion allowing the femoral head to sublux over the cup rim, thus thinning the cup and wearing the rim bevel, and producing MOM wear debris.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 187 - 187
1 Dec 2013
Van Der Straeten C Van Quickenborne D Pennynck S De Smet K Victor J
Full Access

Background:

Potential systemic toxicity of metal ions from metal-on-metal hip arthroplasties (MoMHA) is concerning. High blood cobalt (Co) levels have been associated with neurological, cardiac and thyroid dysfunctions.

Questions/purposes:

The aim of this research was to investigate the prevalence of systemic Co toxicity in a MoMHA population, to identify confounding factors, and to indicate a Co level above which there is a high risk for systemic toxicity.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVII | Pages 50 - 50
1 May 2012
Baker J Byrne D Walsh P Mulhall K
Full Access

Introduction

Local anaesthetic has been reported to have a detrimental effect on human chondrocytes both in vitro and in vivo. Magnesium, an NMDA-receptor antagonist, may be an alternative intra-articular analgesic agent following arthroscopy. We aimed to report the dose response effect of commonly used local anaesthteitc on chondrocyte viability and also report on the effect of adding magnesium to local anaesthetic.

Methods

Human chondrocytes were grown under standard conditions. Cells were exposed to either lignocaine (0.5, 1, 2%), levobupivacaine (0.125, 0.25, 0.5%), bupivacaine (0.125, −.25, 0.5%) or ropivacaine (0.1875, 0.375, 0.75%) for 15 minutes. Cells were also exposed to a local anesthetic agent with the addition of magnesium (10, 20, or 50%). Cells exposed to media or saline served as controls. The MTS assay was used to assess cell viability 24-hours after exposure.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 1 - 1
1 Apr 2018
Jo S Lee S Lim W Kim D Lee J
Full Access

Introduction. Cobalt chrome femoral head has been used widely in total hip arthroplasty and has shown favorable outcome. However, there is still of concern of potential metal toxicity from the wear debris. In the other hand, titanium is well known for its biocompatibility but it is not used in bearing surface of arthroplasty due to its brittleness. Recently, coating of the prosthesis using plasma electrolytic oxidation (PEO) has shown favorable surface protection. Thus, in this study, we tried to find out whether the PEO coating on the titanium surface would provide surface protection. Materials and methods. Five Titanium alloy (Ti-6Al-4V) ball mimicking femoral head was manufactured and was coted using plasma electrolytic oxidation. Wear rate was tested using validated wear tester with 10N compression force at 80rpm. The amount of wear was detected by measuring change of weight after wear test was completed. This was compared with femoral head manufactured with titanium alloy without PEO coating. Toxicity of the debris was also tested using MTT assay with human osteoblast cell line. Results. Compare to the base titanium metal, PEO coated metal head has shown to provide surface protection. The wear rate has significantly decreased with PEO coating (median value : 0.00015g/mm. 2. vs 0.00006 g/mm. 2. ). MTT assay revealed no cytotoxicity with the amount of debris generated from the wear test. Conclusion. The result of the current study indicate that the PEO coating on the titanium femoral head can significantly decrease the wear rate and is non cytotoxic. This indicates that the femoral head manufactured with titanium alloy and PEO coating maybe a potential alternative to be used in total hip arthroplasty


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 56 - 56
1 Apr 2019
Goswami K Cho JE Manrique J Tan T Higuera C Della Valle C Parvizi J
Full Access

Introduction. The use of irrigation solution during surgical procedures is a common and effective practice in reduction of bioburden and the risk of subsequent infection. The optimal irrigation solution to accomplish this feat remains unknown. Many surgeons commonly add topical antibiotics to irrigation solutions assuming this has topical effect and eliminates bacteria. The latter reasoning has never been proven. In fact a few prior studies suggest addition of antibiotics to irrigation solution confers no added benefit. Furthermore, this practice adds to cost, has the potential for anaphylactic reactions, and may also contribute to the emergence of antimicrobial resistance. We therefore sought to compare the antimicrobial efficacy and cytotoxicity of irrigation solution containing polymyxin-bacitracin versus other commonly used irrigation solutions. Methods. Using two in vitro breakpoint assays of Staphylococcus aureus (ATCC#25923) and Escherichia coli (ATCC#25922), we examined the efficacy of a panel of irrigation solutions containing topical antibiotics (500,000U/L Polymyxin-Bacitracin 50,000U/L; Vancomycin 1g/L; Gentamicin 80mg/L), as well as commonly used irrigation solutions (Normal saline 0.9%; Povidone-iodine 0.3%; Chlorhexidine 0.05%; Castile soap 0.45%; and Sodium hypochlorite 0.125%) following 1 minute and 3 minutes of exposure. Surviving bacteria were counted in triplicate experiments. Failure to eradicate all bacteria was considered to be “not effective” for that respective solution and exposure time. Cytotoxicity analysis in human fibroblast, osteoblast, and chrondrocyte cells exposed to each of the respective irrigation solutions was performed by visualization of cell structure, lactate dehydrogenase (LDH) activity and evaluation of vital cells. Toxicity was quantified by determination of LDH release (ELISA % absorbance; with higher percentage considered a surrogate for cytotoxicity). Descriptive statistics were used to present means and standard deviation of triplicate experimental runs. Results. Polymyxin-Bacitracin, Saline and Castile soap irrigation at both exposure times were not effective at eradicating S aureus or E coli (Figure 1). In contrast, Povidone-iodine, Chlorhexidine, and Sodium hypochlorite irrigation were effective at eradicating both S aureus and E coli. Vancomycin irrigation was effective at S aureus eradication but not against E coli, whereas Gentamicin irrigation showed partial efficacy against E coli eradication but none against S aureus. The greatest cytotoxicity was seen with Chlorhexidine (49.4% ± 1.9). This was followed by Castile soap (33.2% ±3.9), Vancomycin (9.01% ±5.1), Polymyxin-Bacitracin (8.45% ±1.5), and Gentamicin irrigation (4.72% ±2.3) (Figure 2 and Figure 3 microscopy images). Povidone-iodine and Sodium hypochlorite showed least cytotoxicity (0.05% ±0.08 and 0.11%±0.19, respectively). Similar trends were seen at both exposure times and across fibroblasts, osteoblasts and chondrocytes. Discussion. This in vitro study suggests that addition of polymyxin-bacitracin to saline irrigation solution is a futile exercise. Taken within the context of its associated expense, risk of hypersensitivity and impact upon antimicrobial resistance, our findings bring its widespread clinical usage into question. Povidone-iodine may be a more effective option, with a more favorable cytotoxicity profile than the other commonly used irrigation solutions. Clinical outcomes should be studied to determine the most effective agent, concentration, and exposure time for intraoperative irrigation


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 582 - 589
1 May 2015
Brennan SA Ní Fhoghlú C Devitt BM O’Mahony FJ Brabazon D Walsh A

Implant-associated infection is a major source of morbidity in orthopaedic surgery. There has been extensive research into the development of materials that prevent biofilm formation, and hence, reduce the risk of infection. Silver nanoparticle technology is receiving much interest in the field of orthopaedics for its antimicrobial properties, and the results of studies to date are encouraging. Antimicrobial effects have been seen when silver nanoparticles are used in trauma implants, tumour prostheses, bone cement, and also when combined with hydroxyapatite coatings. Although there are promising results with in vitro and in vivo studies, the number of clinical studies remains small. Future studies will be required to explore further the possible side effects associated with silver nanoparticles, to ensure their use in an effective and biocompatible manner. Here we present a review of the current literature relating to the production of nanosilver for medical use, and its orthopaedic applications.

Cite this article: Bone Joint J 2015; 97-B:582–9.