Despite advancements, revision rates following total ankle replacement (TAR) are high in comparison to other total joint replacements. This explant analysis study aimed to investigate whether there was appreciable metal particulate debris release from various contemporary TARs by describing patterns of material loss. Twenty-eight explanted TARs (9 designs: 3 fixed and 6 mobile bearing), revised for any reason, were studied. The articulating surfaces of the metal tibial and talar components as well as the polyethylene insert were assessed for damage features using light microscopy. Based on the results of the microscopic analysis, scanning electron microscopy with energy dispersive X-ray spectroscopy was performed to determine the composition of embedded debris identified, as well as non-contacting 3D profilometry. Pitting, indicative of material loss, was identified on the articulating surfaces of 54% of tibial components and 96% of talar components. Bearing constraint was not found to be a factor, with similar proportions of fixed and mobile bearing metal components showing pitting. More cobalt-chromium than titanium alloy tibial components exhibited pitting (63% versus 20%). Significantly higher average surface roughness (Sa) values were measured for pitted areas in comparison to unpitted areas of these metal components (p<0.05). Additionally, metallic embedded debris (cobalt-chromium likely due to pitting of the tibial and talar components or
The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
We report the medium-term outcomes of a consecutive series of 118 Zenith total ankle arthroplasties (TAAs) from a single, non-designer centre. Between December 2010 and May 2016, 118 consecutive Zenith prostheses were implanted in 114 patients. Demographic, clinical, and patient-reported outcome measures (PROMs) data were collected. The endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan-Meier survival curves were generated with 95% confidence intervals (CIs) and the rate of failure calculated for each year.Aims
Methods
The Zenith. TM. total ankle replacement (Corin, Cirencester) is a mobile-bearing implant based on the Buechal Pappas design. Key features are the simple fully-jigged instrumentation aiming to improve accuracy and reproducibility of implant positioning, cementless calcium phosphate coated surfaces for improved early osseointegration, and
We report the long-term clinical and radiological outcomes of a consecutive series of 200 total ankle arthroplasties (TAAs, 184 patients) at a single centre using the Scandinavian Total Ankle Replacement (STAR) implants. Between November 1993 and February 2000, 200 consecutive STAR prostheses were implanted in 184 patients by a single surgeon. Demographic and clinical data were collected prospectively and the last available status was recorded for further survival analysis. All surviving patients underwent regular clinical and radiological review. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot scoring system. The principal endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan–Meier survival curves were generated with 95% confidence intervals and the rate of failure calculated for each year.Aims
Patients and Methods
Between 2002 and 2008, 130 consecutive ankles were replaced with an hydroxyapatite (HA) and titanium-HA-coated Ankle Evolutive System total ankle prosthesis. Plain radiographs were analysed by two independent observers. Osteolytic lesions were classified by their size and location, with cavities >
10 mm in diameter considered to be ‘marked’. CT scanning was undertaken in all patients with marked osteolysis seen on the plain radiographs. Osteolytic lesions were seen on the plain films in 48 (37%) and marked lesions in 27 (21%) ankles. The risk for osteolysis was found to be 3.1 (95% confidence interval 1.6 to 5.9) times higher with implants with Ti-HA porous coating. Care should be taken with ankle arthroplasty until more is known about the reasons for these severe osteolyses.
We describe the results of a randomised, prospective study of 200 ankle replacements carried out between March 2000 and July 2003 at a single centre to compare the Buechel-Pappas (BP) and the Scandinavian Total Ankle Replacement (STAR) implant with a minimum follow-up of 36 months. The two prostheses were similar in design consisting of three components with a meniscal polyethylene bearing which was highly congruent on its planar tibial surface and on its curved talar surface. However, the designs were markedly different with respect to the geometry of the articular surface of the talus and its overall shape. A total of 16 ankles (18%) was revised, of which 12 were from the BP group and four of the STAR group. The six-year survivorship of the BP design was 79% (95% confidence interval (CI) 63.4 to 88.5 and of the STAR 95% (95% CI 87.2 to 98.1). The difference did not reach statistical significance (p = 0.09). However, varus or valgus deformity before surgery did have a significant effect) (p = 0.02) on survivorship in both groups, with the likelihood of revision being directly proportional to the size of the angular deformity. Our findings support previous studies which suggested that total ankle replacement should be undertaken with extreme caution in the presence of marked varus or valgus deformity.
We have evaluated the clinical effectiveness
of a metal resurfacing inlay implant for osteochondral defects of
the medial talar dome after failed previous surgical treatment.
We prospectively studied 20 consecutive patients with a mean age
of 38 years (20 to 60), for a mean of three years (2 to 5) post-surgery.
There was statistically significant reduction of pain in each of
four situations (i.e., rest, walking, stair climbing and running;
p ≤ 0.01). The median American Orthopaedic Foot and Ankle Society
ankle-hindfoot score improved from 62 (interquartile range (IQR)
46 to 72) pre-operatively to 87 (IQR 75 to 95) at final follow-up
(p <
0.001). The Foot and Ankle Outcome Score improved on all
subscales (p ≤ 0.03). The mean Short-Form 36 physical component
scale improved from 36 (23 to 50) pre-operatively to 45 (29 to 55)
at final follow-up (p = 0.001); the mental component scale did not
change significantly. On radiographs, progressive degenerative changes
of the opposing tibial plafond were observed in two patients. One
patient required additional surgery for the osteochondral defect.
This study shows that a metal implant is a promising treatment for
osteochondral defects of the medial talar dome after failed previous
surgery. Cite this article:
We performed a retrospective review of a consecutive
series of 178 Mobility total ankle replacements (TARs) performed
by three surgeons between January 2004 and June 2009, and analysed
radiological parameters and clinical outcomes in a subgroup of 129
patients. The mean follow-up was 4 years (2 to 6.3). A total of
ten revision procedures (5.6%) were undertaken. The mean Ankle Osteoarthritis
Scale (AOS) pain score was 17 (0 to 88) and 86% of patients were
clinically improved at follow-up. However, 18 patients (18 TARs,
14%) had a poor outcome with an AOS pain score of >
30. A worse
outcome was associated with a pre-operative diagnosis of post-traumatic degenerative
arthritis. However, no pre- or post-operative radiological parameters
were significantly associated with a poor outcome. Of the patients
with persistent pain, eight had predominantly medial-sided pain.
Thirty TARs (29%) had a radiolucency in at least one zone. The outcome of the Mobility TAR at a mean of four years is satisfactory
in >
85% of patients. However, there is a significant incidence
of persistent pain, particularly on the medial side, for which we
were unable to establish a cause. Cite this article:
We studied a cohort of 26 diabetic patients with chronic ulceration under the first metatarsal head treated by a modified Jones extensor hallucis longus and a flexor hallucis longus transfer. If the first metatarsal was still plantar flexed following these two transfers, a peroneus longus to the peroneus brevis tendon transfer was also performed. Finally, if ankle dorsiflexion was <
5° with the knee extended, a Strayer-type gastrocnemius recession was performed. The mean duration of chronic ulceration despite a minimum of six months’ conservative care was 16.2 months (6 to 31). A total of 23 of the 26 patients were available for follow-up at a mean of 39.6 months (12 to 61) after surgery. All except one achieved complete ulcer healing at a mean of 4.4 weeks (2 to 8) after surgery, and there was no recurrence of ulceration under the first metatarsal. We believe that tendon balancing using modified Jones extensor hallucis longus and flexor hallucis longus transfers, associated in selected cases with a peroneus longus to brevis transfer and/or Strayer procedure, can promote rapid and sustained healing of chronic diabetic ulcers under the first metatarsal head.
The Bologna–Oxford (BOX) total ankle replacement
(TAR) was developed with the aim of achieving satisfactory pain-free
movement of the ankle. To date, only one single multicentre study
has reported its clinical results. The aim of this study was to
conduct an independent review of its mid-term results. We retrospectively reviewed a total of 60 prospectively followed
patients in whom 62 BOX TARs had been implanted between 2004 and
2008. We used the American Orthopedic Foot and Ankle Society (AOFAS)
score to assess the clinical results. Standardised radiographs taken
at the time of final follow-up were analysed by two observers. The
overall survival was 91.9% at a mean follow-up of 42.5 months (24
to 71). The mean AOFAS score had improved from 35.1 points (
We investigated the clinical and radiological outcome after unilateral fracture of the lateral process of the talus in 23 snowboarders with a mean follow-up of 3.5 years (12 to 76 months). In this consecutive cohort study both operative and non-operative cases were considered. The mean American Orthopaedic Foot and Ankle Society hindfoot score was 94 (82 to 100). The non-operative group of seven with a minimally-displaced fracture scored higher (98 points) than the operative group of 16 with displaced or unstable fractures (93 points). In 88% of operative cases, significant concomitant hindfoot injuries were found at operation. All but eight (35%) patients (six operative and two non-operative) regained their pre-injury level of sporting activity. Subtalar osteoarthritis was present in nine (45%) of the 20 patients available for radiological review, including one late-diagnosed non-operative case and eight operative cases with associated injuries or fracture comminution. The outcome after fracture of the lateral process of the talus in snowboarders is favourable provided an early diagnosis is made and adequate treatment, which is related to the degree of displacement and associated injuries, is undertaken.
We carried out 123 consecutive total ankle replacements in 111 patients with a mean follow-up of four years (2 to 8). Patients with a hindfoot deformity of up to 10° (group A, 91 ankles) were compared with those with a deformity of 11° to 30° (group B, 32 ankles). There were 18 failures (14.6%), with no significant difference in survival between groups A and B. The clinical outcome as measured by the post-operative American Orthopaedic Foot and Ankle Surgeons score was significantly better in group B (p = 0.036). There was no difference between the groups regarding the post-operative range of movement and complications. Correction of the hindfoot deformity was achieved to within 5° of neutral in 27 ankles (84%) of group B patients. However, gross instability was the most common mode of failure in group B. This was not adequately corrected by reconstruction of the lateral ligament. Total ankle replacement can safely be performed in patients with a hindfoot deformity of up to 30°. The importance of adequate correction of alignment and instability is highlighted.