Metallic contacts in hip replacements are susceptible to wear and corrosion processes which lead to the release of particles and metal ions. Adverse local tissue reactions (ALTRs) and systemic manifestations to solid and soluble debris can be debilitating for the patients. It is believed that particles originating from CoCrMo taper junctions trigger more severe body reactions compared to debris from MoM hip bearings. The body's reaction is highly dependent on particle characteristics, such as size, morphology, composition and aggregation state, which can reflect the specific wear and corrosion conditions at the site of release. Here we proposed to investigate wear and corrosion flakes collected from around CoCrMo tapers at the time of revision. The particles were initially characterised with scanning electron microscopy (SEM) and energy dispersive X-ray analysis (EDX). This revealed the microstructure of the corrosion products, which appeared to be made of smaller metallic aggregates, entrapped in a biological matrix. The in depth characterisation of the particles released from the organo-metallic composite, was performed with transmission electron microscopy (TEM) and scanning transmission electron microscopy (STEM), both fitted with EDX. The investigation revealed clusters and individual nanoparticles, as small as 3 nm, which represent the building blocks of the large corrosion flakes, reported and characterised in the past mainly with low resolution microscopy techniques. The majority of the particles consisted of Cr and O, potentially in the form of chromium oxides, with little evidence of Co and Mo. Particles size distribution (PSD) provided by STEM and TEM characterisation showed statistically different results. The STEM technique was able to resolve tiny particles found in close proximity and provided a PSD shift towards the smaller end of the size range. The study is the first to show microscopy evidence of Cr rich nanoparticles (3–60 nm) released in vivo from the modular taper interface, which can have important health implications caused by their increased potential to disseminate and corrode within the body.
The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions. Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour.Aims
Methods
Trunnionosis, due to mechanical wear and/or corrosion at the head stem taper junction, can occur in metal on polyethylene (MOP) hip implants. In some patients this results in severe soft tissue destruction or Adverse Reaction to Metal Debris (ARMD). The amount of material required to cause ARMD is unknown but analyses of retrieved hips may provide the answer to this clinically important question. We collected implants from 20 patients with failed hips with MOP bearings, revised due to ARMD. We collected clinical, imaging and blood test data. We graded the severity of taper corrosion (1 to 4), and quantified the volume of material loss from this junction. We compared our results with previous data collected for metal-on-metal (MOM) hips. The median time to revision of the MOP hips was 51.3 (23.1–56.4) months. All head tapers were moderately to severely corroded with a median corrosion score of 4. The median (range) of total material loss at the taper of the MOP hips was 3.9 mm3 (2.96 – 7.85 mm3) and the material loss rate was 1.4 mm3 / year (0.56 – 1.82). Comparison with MOM hips revealed no significant difference in taper material loss (p=0.7344) with a median rate of 0.81 mm3 / year (0.01–3.45). We are the first to quantify the volume of material loss at the head taper of hip implants with MOP bearings that were revised due to trunnionosis. This data indicates that a clinically significant dose of cobalt and chromium to induce ARMD is approximately 1.4 mm3 / year. We have identified a clinically significant volume of taper material loss in MOP hips.
Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.Aims
Methods