The Bone & Joint Journal
Vol. 99-B, Issue 9 | Pages 1153 - 1156
1 Sep 2017
Harrison PL Harrison T Stockley I Smith TJ

Aims. Tantalum (Ta) trabecular metal components are increasingly used to reconstruct major bone defects in revision arthroplasty surgery. It is known that some metals such as silver have antibacterial properties. Recent reports have raised the question regarding whether Ta components are protective against infection in revision surgery. This laboratory study aimed to establish whether Ta has intrinsic antibacterial properties against planktonic bacteria, or the ability to inhibit biofilm formation. Materials and Methods. Equal-sized pieces of Ta and titanium (Ti) acetabular components were sterilised and incubated with a low dose inoculum of either Staphylococcus (S.) aureus or S. epidermidis for 24 hours. After serial dilution, colony forming units (cfu) were quantified on Mueller-Hinton agar plates. In order to establish whether biofilms formed to a greater extent on one material than the other, these Ta and Ti pieces were then washed twice, sonicated and washed again to remove loosely adhered planktonic bacteria. They were then re-incubated for 24 hours prior to quantifying the number of cfu. All experiments were performed in triplicate. Results. More than 1x10. 8. cfu/ml were observed in both the Ta and Ti experiments. After washing and sonication, more than 2x10. 7. cfu/ml were observed for both Ta and Ti groups. The results were the same for both S. aureus and S. epidermidis. Conclusion. Compared with Ti controls, Ta did not demonstrate any intrinsic antibacterial activity or ability to inhibit biofilm formation. Hence, intrinsic antimicrobial properties of Ta do not account for the previously observed reduction in the frequency of subsequent infections when Ta was used in revision procedures. . Cite this article: Bone Joint J 2017;99-B:1153–6


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 173 - 178
1 Feb 2012
Malizos KN Papasoulis E Dailiana ZH Papatheodorou LK Varitimidis SE

The introduction of a trabecular tantalum rod has been proposed for the management of early-stage osteonecrosis of the femoral head but serves as a single-point of support of the necrotic lesion. We describe a technique using two or three 4.2 mm (or later 4.7 mm) tantalum pegs for the prevention of collapse of the necrotic lesion. We prospectively studied 21 patients (26 hips) with non-traumatic osteonecrosis of the femoral head treated in this manner. Of these, 21 patients (24 hips) were available for radiological and clinical evaluation at a mean follow-up of 46 months (18 to 67). Radiological assessment showed that only eight hips deteriorated according to the Association Research Circulation Osseous classification, and four hips according to the Classification of the Japanese Investigation Committee of Health and Welfare. Functional improvement was obtained with an improvement in the mean Harris hip score from 65.2 (33.67 to 95) to 88.1 (51.72 to 100), the mean Merle D’Aubigné-Postel score from 13 (6 to 18) to 16 (11 to 18), a mean visual analogue score for pain from 5.2 (0 to 9.5) to 2.6 (0 to 7), and the mean Short-Form 36 score from 80.4 (56.8 to 107.1) to 92.4 (67.5 to 115.7). Of these 24 hips followed for a minimum of 18 months, three were considered as failures at the final follow-up, having required total hip replacement. One of the hips without full follow-up was also considered to be a failure. In more than two-thirds of the surviving hips a satisfactory clinical outcome was achieved with promising radiological findings. The estimated mean implant survival was 60 months (95% confidence interval 53.7 to 66.3)


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1175 - 1179
1 Sep 2016
Olsen M Lewis PM Morrison Z McKee MD Waddell JP Schemitsch EH

Aims. One method of femoral head preservation following avascular necrosis (AVN) is core decompression and insertion of a tantalum rod. However, there may be a high failure rate associated with this procedure. The purpose of this study was to document the clinical and radiological outcomes following total hip arthroplasty (THA) subsequent to failed tantalum rod insertion. Patients and Methods. A total of 37 failed tantalum rods requiring total hip arthroplasty were identified from a prospective database. There were 21 hips in 21 patients (12 men and nine women, mean age 37 years, 18 to 53) meeting minimum two year clinical and radiographic follow-up whose THAs were carried out between November 2002 and April 2013 (mean time between tantalum rod implantation and conversion to a THA was 26 months, 6 to 72). These were matched by age and gender to individuals (12 men, nine women, mean age 40 years, 18 to 58) receiving THA for AVN without prior tantalum rod insertion. Results. There were no functional outcome differences between the two groups. Tantalum residue was identified on all post-operative radiographs in the tantalum group. Linear wear rates were comparable between groups with no evidence of catastrophic wear in either group. Conclusion. In the short term, tantalum rod implantation does not demonstrate an adverse effect on subsequent total joint replacement surgery. There is however, a high rate of retained tantalum debris on post-operative radiographs and thus there is an unknown risk of accelerated articular wear necessitating longer term study. Cite this article: Bone Joint J 2016;98-B:1175–9


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 767 - 771
1 Jun 2016
Konan S Duncan CP Masri BA Garbuz DS

Aims. Reconstruction of the acetabulum after failed total hip arthroplasty (THA) can be a surgical challenge in the presence of severe bone loss. We report the long-term survival of a porous tantalum revision acetabular component, its radiological appearance and quality of life outcomes. Patients and Methods. We reviewed the results of 46 patients who had undergone revision of a failed acetabular component with a Paprosky II or III bone defect and reconstruction with a hemispherical, tantalum acetabular component, supplementary screws and a cemented polyethylene liner. Results. After a minimum follow-up of ten years (ten to 12), the survivorship of the porous tantalum acetabular component was 96%, with further revision of the acetabular component as the end point. The ten-year survivorship, with hip revision for any reason as the end point, was 92%. We noted excellent pain relief (mean Western Ontario and McMaster Universities Arthritis Index (WOMAC) score pain 92.6, (40 to 100)) and good functional outcomes (mean WOMAC function 90.3 (30.9 to 100), mean University of California Los Angeles activity scale 5 (2 to 10)) and generic quality of life measures (mean Short Form-12 (SF-12) physical component 48.3 (18.1 to 56.8), mean SF-12 mental component 56.7 (32.9 to 70.3)). Patient satisfaction with pain relief, function and return to recreational activities were excellent. Take home message: Uncemented acetabular reconstruction using a tantalum acetabular component gives excellent clinical and quality of life outcomes at a minimum follow-up of ten years. Cite this article: Bone Joint J 2016;98-B:767–71


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 43 - 43
1 Nov 2021
Callary S Abrahams J Zeng Y Clothier R Costi K Campbell D Howie D Solomon L
Full Access

First-time revision acetabular components have a 36% re-revision rate at 10 years in Australia, with subsequent revisions known to have even worse results. Acetabular component migration >1mm at two years following revision THA is a surrogate for long term loosening. This study aimed to measure the migration of porous tantalum components used at revision surgery and investigate the effect of achieving press-fit and/or three-point fixation within acetabular bone. Between May 2011 and March 2018, 55 patients (56 hips; 30 female, 25 male) underwent acetabular revision THR with a porous tantalum component, with a post-operative CT scan to assess implant to host bone contact achieved and Radiostereometric Analysis (RSA) examinations on day 2, 3 months, 1 and 2 years. A porous tantalum component was used because the defects treated (Paprosky IIa:IIb:IIc:IIIa:IIIb; 2:6:8:22:18; 13 with pelvic discontinuity) were either deemed too large or in a position preventing screw fixation of an implant with low coefficient of friction. Press-fit and three-point fixation of the implant was assessed intra-operatively and on postoperative imaging. Three-point acetabular fixation was achieved in 51 hips (92%), 34 (62%) of which were press-fit. The mean implant to host bone contact achieved was 36% (range 9–71%). The majority (52/56, 93%) of components demonstrated acceptable early stability. Four components migrated >1mm proximally at two years (1.1, 3.2, 3.6 and 16.4mm). Three of these were in hips with Paprosky IIIB defects, including 2 with pelvic discontinuity. Neither press-fit nor three-point fixation was achieved for these three components and the cup to host bone contact achieved was low (30, 32 and 59%). The majority of porous tantalum components had acceptable stability at two years following revision surgery despite treating large acetabular defects and poor bone quality. Components without press-fit or three-point fixation were associated with unacceptable amounts of early migration


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 10 - 10
1 Jun 2016
Harrison T Harrison P Smith T Stockley I
Full Access

Introduction. Tantalum trabecular metal components are increasingly used to reconstruct major bone defects in revision arthroplasty surgery. It is known that some metals such as silver have antibacterial properties. Recent reports have raised the question as to whether Tantalum components are protective against infection in revision surgery. This is based on a retrospective, single institution review, of revision cases comparing tantalum with titanium acetabular implants, which reported a lower incidence of subsequent infection in the tantalum group. This laboratory study aimed to establish if tantalum had any intrinsic antibacterial properties against planktonic bacteria or ability to inhibit biofilm formation. Materials and methods. Equal sized pieces of tantalum (Trabecular metal, Zimmer UK) and titanium (Trilogy, Zimmer UK) were sterilised and then incubated with a low dose inoculum of either Staphylococcus aureus or Staphylococcus epidermidis for 24 hours. After serial dilution, colony forming units were quantified on MH agar plates. To establish the ability to inhibit biofilm formation these tantalum and titanium pieces were then washed twice, sonicated and washed again to remove loosely adhered planktonic bacteria. They were then re-incubated for 24 hours prior to quantifying colony forming units. All experiments were performed in triplicate. Results. More than 1×10. 8. cfu/ml were observed in both the titanium and tantalum experiments. After washing and sonication more than 2×10. 7. cfu/ml were observed for both tantalum and titanium groups. The results were the same for both Staph Aureus and Staph Epidermidis. Discussion. Compared with titanium controls tantalum did not demonstrate any intrinsic antibacterial activity or ability to inhibit biofilm formation. The intrinsic properties of tantalum do not account for the previously observed reduction in subsequent infection when tantalum was used in the revision procedure. Conclusion. Tantalum does not have any intrinsic antimicrobial properties or ability to inhibit biofilm formation


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 23 - 23
1 Aug 2018
Sousa P Abdel M Francois E Hanssen A Lewallen D
Full Access

Highly porous tantalum cups have been used in complex acetabular revisions for nearly 20 years but reports of long term results are limited. This study was designed to report ten year results of revision using a single porous tantalum cup design with special attention to re-operation for any reason, all-cause revision, and revision for aseptic loosening. Retrospective review of all revision THA cases performed from 1999–2006 using a highly porous tantalum acetabular component design with multiple screw holes and a cemented polyethylene liner (Zimmer Biomet, Warsaw, IN). Our institutional medical record and total joint registry were used to assess follow-up and xrays were reviewed. The Paprosky classification system was used to rate acetabular bone loss. Radiographic loosening was defined as new/progressive radiolucencies in all 3 acetabular zones, or cup migration (>2mm). Kaplan-Meier survivorship was used to assess survivorship free of cup revision/removal for any reason, and free of revision for aseptic loosening. Between 1999 and 2006 this tantalum cup was used in 916 revisions. Mean age: 66 (±6), BMI: 29 (±6), and male: 42%. Indications for revision: aseptic loosening 346 (38%), osteolysis 240 (26%), and infected arthroplasty 168 (18%). Large (3A or 3B) bone defects were present in 260, and pelvic discontinuity in 61. Reoperation for any reason: 133 (15%), but 84 of 133 cases did not require cup revision for instability (38) or femoral failure (24). Tantalum cup removal/revision was required in 49 (5.3%) for deep infection (39) and recurrent dislocation (6), and aseptic loosening (4). 10 year survivorship free of cup revision for any reason: 95% and for aseptic loosening: 99%. Radiographic review (mean 10 years): suspicious for aseptic loosening in another 4 cups. A highly porous tantalum acetabular component with multiple screws and a cemented polyethylene insert provided durable long term fixation for an array of acetabular revision problems. Long term aseptic loosening was very rare (<1%) and cup removal was mainly related to deep infection, and rarely dislocation


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 45 - 49
1 Jan 2015
Tokarski AT Novack TA Parvizi J

We hypothesised that the use of tantalum (Ta) acetabular components in revision total hip arthroplasty (THA) was protective against subsequent failure due to infection. We identified 966 patients (421 men, 545 women and 990 hips) who had undergone revision THA between 2000 and 2013. The mean follow up was 40.2 months (3 months to 13.1 years). The mean age of the men and women was 62.3 years (31 to 90) and 65.1 years (25 to 92), respectively. . Titanium (Ti) acetabular components were used in 536 hips while Ta components were used in 454 hips. In total, 73 (7.3%) hips experienced subsequent acetabular failure. The incidence of failure was lower in the Ta group at 4.4% (20/454) compared with 9.9% (53/536) in the Ti group (p < 0.001, odds ratio 2.38; 95% CI 1.37 to 4.27). Among the 144 hips (64 Ta, 80 Ti) for which revision had been performed because of infection, failure due to a subsequent infection was lower in the Ta group at 3.1% (2/64) compared with 17.5% (14/80) for the Ti group (p = 0.006). . Thus, the use of Ta acetabular components during revision THA was associated with a lower incidence of failure from all causes and Ta components were associated with a lower incidence of subsequent infection when used in patients with periprosthetic joint infection. . Cite this article: Bone Joint J 2015;97-B:45–9


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 7 - 7
1 Jan 2018
Solomon L Abrahams J Callary S Howie D
Full Access

Acetabular components used to treat large defects are at greater risk of loosening. Porous tantalum acetabular components have reported the most promising early to midterm revision rates. Early stability of acetabular components used at revision THR was shown to be a good predictor of later loosening. The primary aim was to assess the migration of porous acetabular component used to reconstruct severe acetabular defects. Secondarily, we investigated the effect of acetabular defect severity and type of component fixation on migration. Radiosterometric analysis was used to measure migration at a mean follow-up of four years, (range 2–10) in 59 reconstructions of severe acetabular defects with porous tantalum components. Acetabular component fixation was classified as superior if augmented with screws through cup, augments or cage in the ilium only. Fixation was classified as combined, superior and inferior, if flanges and/or screws were also placed in the ischium and or pubis. Acceptable limits of proximal migration were defined as ≤1mm within 2 years and ≤2.5mm at any time point. Eight hips had reconstruction of Paprosky II defects with superior fixation only. The mean proximal migration of the eight acetabular components was 0.25mm (0.08–0.40) at 2 years and 0.29mm (0.10–0.81) at last follow-up. Fifty-one hips had reconstruction of Paprosky III defects. Seven of these reconstructions exceeded the migration thresholds. Five reconstructions (four with superior fixation and one cup cage construct with no inferior screw fixation) of hips with pelvic discontinuity developed pain and were re-revised for loosening. Two reconstructions are asymptomatic and migrated 2.68mm (cup-cage construct with superior screws) and 2.86mm (no pelvic discontinuity, superior fixation) at final follow-up. The mean proximal migration of the 51 Paprosky III reconstructions was 0.99mm (0.03 to 16.4) at 2 years and 1.92mm (0.01 to 29.4) at last follow-up. The mean proximal translation at 2 years of the 11 reconstructions with inferior screw fixation was 0.2mm (−0.6 to 0.7mm), compared with 0.9mm (−0.6 to 16.4mm) for the reconstructions without inferior screw fixation. In conclusion, when used to reconstruct Paprosky II defects, porous tantalum acetabular components provide component stability similar to a good performing primary THR. These implants achieve adequate stability when used to treat Paprosky III defects, including those with pelvic discontinuity. For the most severe defects, combined fixation with inferior screws is recommended, particularly when reconstructing hips with pelvic discontinuity


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 414 - 416
1 May 1998
Eldridge JDJ Avramidis K Lee M Learmonth ID

There are several techniques for the accurate measurement of the migration of components after arthroplasty some of which require the operative placement of tantalum balls. We have reviewed the position and migration of these markers in 64 patients after total hip arthroplasty. In 40% of cases, one or more balls was seen to be outside the proximal femur on the postoperative radiograph, although all were considered to be within the bone at operation. In two hips, one ball appeared to have migrated towards the joint, although none was seen within the joint. Misplacement was not related to the experience of the surgeon or the operative approach. Migration analysis which necessitates the insertion of tantalum balls requires careful technique to avoid a potential source of third-body wear. It should probably be used only for research in small series of patients


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 2 - 2
23 Jun 2023
Hube R Zimmerer A Nonnenmacher L Najfeld M Fischer M Wassilew GI
Full Access

The use of trabecular metal (TM.) shells supported by two TM augments in the footing technique has been described as a potential option for the treatment of Paprosky 3B acetabular defects. The aim of this study was to assess the mid implant survivorship and radiological and clinical outcomes after acetabular revision using this technique.

We undertook a retrospective, double-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the footing technique using a TM shell supported by two TM augments, for severe acetabular bone loss between 2007 and 2020. The mean age at the time of surgery was 62,9 (28 to 86) years. The mean follow-up was 5,4 (1,5 to 15) years.

The cumulative mid survivorship of the implant with revision for any cause was 89%. 3 hips (7,6%) required further revision due to aseptic loosening, and 1 (2,8%) required revision for infection. The mean Harris Hip Score improved significantly from 48 (29 to 65) preoperatively to 79 points (62 to 98) at the latest follow-up (p < 0.001).

The reconstruction of Paprosky 3B acetabular defects with TM shells and two augments in footing-technique showed excellent mid-term results. This technique appears to be a viable option for treating these defects.


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 311 - 316
1 Mar 2019
Löchel J Janz V Hipfl C Perka C Wassilew GI

Aims

The use of trabecular metal (TM) shells supported by augments has provided good mid-term results after revision total hip arthroplasty (THA) in patients with a bony defect of the acetabulum. The aim of this study was to assess the long-term implant survivorship and radiological and clinical outcomes after acetabular revision using this technique.

Patients and Methods

Between 2006 and 2010, 60 patients (62 hips) underwent acetabular revision using a combination of a TM shell and augment. A total of 51 patients (53 hips) had complete follow-up at a minimum of seven years and were included in the study. Of these patients, 15 were men (29.4%) and 36 were women (70.6%). Their mean age at the time of revision THA was 64.6 years (28 to 85). Three patients (5.2%) had a Paprosky IIA defect, 13 (24.5%) had a type IIB defect, six (11.3%) had a type IIC defect, 22 (41.5%) had a type IIIA defect, and nine (17%) had a type IIIB defect. Five patients (9.4%) also had pelvic discontinuity.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 304 - 309
1 Mar 2006
Macheras GA Papagelopoulos PJ Kateros K Kostakos AT Baltas D Karachalios TS

Between January 1998 and December 1998, 82 consecutive patients (86 hips) underwent total hip arthroplasty using a trabecular metal monoblock acetabular component. All patients had a clinical and radiological follow-up evaluation at six, 12 and 24 weeks, 12 months, and then annually thereafter. On the initial post-operative radiograph 25 hips had a gap between the outer surface of the component and the acetabular host bed which ranged from 1 to 5 mm. All patients were followed up clinically and radiologically for a mean of 7.3 years (7 to 7.5). The 25 hips with the 1 to 5 mm gaps were studied for component migration at two years using the Einzel-Bild-Roentgen-Analyse (EBRA) digital measurement method. At 24 weeks all the post-operative gaps were filled with bone and no acetabular component had migrated. The radiographic outcome of all 86 components showed no radiolucent lines and no evidence of lysis. No acetabular implant was revised. There were no dislocations or other complications. The bridging of the interface gaps (up to 5 mm) by the trabecular metal monoblock acetabular component indicates the strong osteoconductive, and possibly osteoinductive, properties of trabecular metal.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 74 - 74
23 Jun 2023
Wilson JM Maradit-Kremers H Abdel MP Berry DJ Mabry TM Pagnano MW Perry KI Sierra RJ Taunton MJ Trousdale RT Lewallen DG
Full Access

The last two decades have seen remarkable technological advances in total hip arthroplasty (THA) implant design. Porous ingrowth surfaces and highly crosslinked polyethylene (HXLPE) have been expected to dramatically improve implant survivorship. The purpose of the present study was to evaluate survival of contemporary cementless acetabular components following primary THA. 16,421 primary THAs performed for osteoarthritis between 2000 and 2019 were identified from our institutional total joint registry. Patients received one of 12 contemporary cementless acetabular designs with HXLPE liners. Components were grouped based on ingrowth surface into 4 categories: porous titanium (n=10,952, mean follow-up 5 years), porous tantalum (n=1223, mean follow-up 5 years), metal mesh (n=2680, mean follow-up 6.5 years), and hydroxyapatite (HA) coated (n=1566, mean follow-up 2.4 years). Kaplan-Meier analyses were performed to assess the survivorship free of acetabular revision. A historical series of 182 Harris-Galante-1 (HG-1) acetabular components was used as reference. The 15-year survivorship free of acetabular revision was >97% for all 4 contemporary cohorts. Compared to historical control, porous titanium (HR 0.06, 95% CI 0.02–0.17, p<0.001), porous tantalum (HR 0.09, 95%CI 0.03–0.29, p<0.001), metal mesh (HR 0.11, 95%CI 0.04–0.31, p<0.001), and HA-coated (HR 0.14, 95%CI 0.04–0.48, p=0.002) ingrowth surfaces had significantly lower risk of any acetabular revision. There were 16 cases (0.1%) of acetabular aseptic loosening that occurred in 8 (0.07%) porous titanium, 5 (0.2%) metal mesh, and 3 (0.2%) HA-coated acetabular components. 7 of the 8 porous titanium aseptic loosening cases occurred in one known problematic design. There were no cases of aseptic loosening in the porous tantalum group. Modern acetabular ingrowth surfaces and HXLPE liners have improved on historical results at the mid-term. Contemporary designs have extraordinarily high revision-free survivorship, and aseptic loosening is now a rare complication. At mid-term follow-up, survivorship of contemporary uncemented acetabular components is excellent and aseptic loosening occurs in a very small minority of patients


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 56 - 56
1 Jan 2018
Macheras G Lepetsos P Anastasopoulos P Tzefronis D Galanakos S Poultsides L
Full Access

Porous tantalum is a highly osteoinductive biomaterial, initially introduced in orthopedics in 1997, with a subsequent rapid evolution of orthopedic applications. The use of porous tantalum for the acetabular component in primary total hip arthroplasty (THA) has demonstrated excellent short-term and mid-term results. However, long term data are scarce. The purpose of this prospective study is to report the long-term clinical and radiological outcome following use of an uncemented porous tantalum acetabular component in primary THA with a minimum follow-up of 17.5 years, in a previously studied cohort of patients. We prospectively followed 128 consecutive primary THAs in 140 patients, between November 1997 and June 1999. A press-fit porous tantalum monoblock acetabular component was used in all cases. The presence of initial gaps in the polar region, as sign of incomplete seating of the monoblock cup, was assessed on the immediate postoperative radiographs. All patients were followed clinically and radiographically at 6, 12, and 24 weeks and 12 months and then at 2, 5, 8, 10, and 19 years, for a mean of 18.1 years (range 17.5 – 19 years). Periacetabular dome gap filling, acetabular cup migration and polyethylene wear were assessed by the EBRA digital measurement system, until 2 years postoperatively. Mean age of patients at the time of operation was 60.4 years old (range 24 – 72). Harris hip score, Oxford Hip Score and range of motion (ROM) were dramatically improved in all cases (p < 0.001). In the initial postoperative radiographs, periacetabular dome gaps were observed in the 15% of cases, and were progressively filled within 6 months. In 2 years postoperatively, the mean component migration, as shown in EBRA study, was 0.67 mm. At last follow-up, all cups were radiographically stable with no evidence of migration, gross polyethylene wear, progressive radiolucencies, osteolytic lesions or acetabular fractures. The survivorship with re-operation for any reason as end point was 92.8%, whereas the survivorship for aseptic loosening as an end point was 100%. Upon visual inspection, two removed acetabular components due to recurrent dislocation and infection, respectively, showed extensive bone osseointegration. In our primary THA series, the porous tantalum monoblock cup demonstrated excellent clinical and radiographic outcomes with no failures because of aseptic loosening at a mean follow-up of 18.1 years. Porous tantalum acetabular components showed excellent initial stability, produced less wear debris and revealed a great potential for bone ingrowth. Due to its unique osteoinductive properties and elliptical shape, porous tantalum monoblock cups have demonstrated superior short and long-term survivorship compared to other press fit prostheses in the market


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 14 - 14
7 Jun 2023
Smeeton M Wilcox R Isaac G Anderson J Board T Van Citters DW Williams S
Full Access

Dual Mobility (DM) Total Hip Replacements (THRs) were introduced to reduce dislocation risk, which is the most common cause of early revision. The in-vivo mechanics of these implants is not well understood, despite their increased use in both elective and trauma settings. Therefore, the aim of this study was to comprehensively assess retrieved DM polyethylene liners for signs of damage using visual inspection and semi-quantitative geometric assessment techniques. Retrieved DM liners (n=20) were visually inspected for the presence of seven established modes of polyethylene damage. If embedded debris was identified on the external surface, its material composition was characterised using energy-dispersive x-ray analysis (EDX). Additionally, each liner was geometrically assessed for signs of wear/deformation using a validated methodology. Visual inspection of the liners revealed that scratching and pitting were the most common damage modes on either surface. Burnishing was observed on 50% and 15% of the internal and external surfaces, respectively. In addition, embedded debris was identified on 25% of the internal and 65% of the external surfaces. EDX analysis of the debris identified several materials including iron, titanium, cobalt-chrome, and tantalum. Geometric analysis demonstrated highly variable damage patterns across the liners. The results of this study provide insight into the in-vivo mechanics of DM bearings. For example, the results suggest that the internal bearing (i.e., between the head and liner) acts as the primary articulation site for DM-THRs as evidenced by a higher incidence of burnishing and larger, more concentrated regions of penetration across the liners’ internal surfaces. Furthermore, circumferential, and crescent-shaped damage patterns were identified on the articulating surfaces of the liners thus providing evidence that these components can rotate within the acetabular shell with varying degrees of mobility. The mechanics of DM bearings are complex and may be influenced by several factors (e.g., soft tissue fibrosis, patient activities) and thus further investigation is warranted. Finally, the results of this study suggest that DM liners may be susceptible to ex-vivo surface damage and thus caution is advised when handling and/or assessing these types of components


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 22 - 30
1 Jan 2018
Brown TS Salib CG Rose PS Sim FH Lewallen DG Abdel MP

Aims. Reconstruction of the acetabulum after resection of a periacetabular malignancy is technically challenging and many different techniques have been used with varying success. Our aim was to prepare a systematic review of the literature dealing with these techniques in order to clarify the management, the rate of complications and the outcomes. Patients and Methods. A search of PubMed and MEDLINE was conducted for English language articles published between January 1990 and February 2017 with combinations of key search terms to identify studies dealing with periacetabular resection with reconstruction in patients with a malignancy. Studies in English that reported radiographic or clinical outcomes were included. Data collected from each study included: the number and type of reconstructions, the pathological diagnosis of the lesions, the mean age and follow-up, gender distribution, implant survivorship, complications, functional outcome, and mortality. The results from individual studies were combined for the general analysis, and then grouped according to the type of reconstruction. . Results. A total of 57 studies met the inclusion criteria and included 1700 patients. Most lesions were metastatic (41%), followed by chondrosarcoma (29%), osteosarcoma (10%), Ewing’s sarcoma (7%), and multiple myeloma (2%). The techniques of reconstruction were divided into seven types for analysis: those involving a Harrington reconstruction, a saddle prosthesis, an allograft and allograft prosthesis composite, a pasteurised autograft, a porous tantalum implant, a custom-made prosthesis and a modular hemipelvic reconstruction. The rate of complications was 50%, with infection (14%) and instability (8%) being the most common. Mortality data were available for 1427 patients (84%); 50% had died of disease progression, 23% were alive with disease, and 27% had no evidence of disease at a mean follow-up of 3.4 years (0 to 34). . Conclusion. Both the rate of complications and mortality are high following resection of oncological periacetabular lesions and reconstruction. Many types of reconstruction have been used with unique challenges and complications for each technique. Newer prostheses, including custom-made prostheses and porous tantalum implants and augments, have shown promising early functional and radiographic outcomes. . Cite this article: Bone Joint J 2018;100-B(1 Supple A):22–30


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 73 - 77
1 Nov 2014
Brown NM Hellman M Haughom BH Shah RP Sporer SM Paprosky WG

A pelvic discontinuity occurs when the superior and inferior parts of the hemi-pelvis are no longer connected, which is difficult to manage when associated with a failed total hip replacement. Chronic pelvic discontinuity is found in 0.9% to 2.1% of hip revision cases with risk factors including severe pelvic bone loss, female gender, prior pelvic radiation and rheumatoid arthritis. Common treatment options include: pelvic plating with allograft, cage reconstruction, custom triflange implants, and porous tantalum implants with modular augments. The optimal technique is dependent upon the degree of the discontinuity, the amount of available bone stock and the likelihood of achieving stable healing between the two segments. A method of treating pelvic discontinuity using porous tantalum components with a distraction technique that achieves both initial stability and subsequent long-term biological fixation is described. Cite this article: Bone Joint J 2014;96-B(11 Suppl A):73–7


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 87 - 92
1 Nov 2014
Kwong LM Nielsen ESN Ruiz DR Hsu AH Dines MD Mellano CM

A retrospective review was performed of patients undergoing primary cementless total knee replacement (TKR) using porous tantalum performed by a group of surgical trainees. Clinical and radiological follow-up involved 79 females and 26 males encompassing 115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up was 7 years (2 to 11). Tibial and patellar components were porous tantalum monoblock implants, and femoral components were posterior stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological assessments were made for implant positioning, alignment, radiolucencies, lysis, and loosening. There was 95.7% survival of implants. There was no radiological evidence of loosening and no osteolysis found. No revisions were performed for aseptic loosening. Average tibial component alignment was 1.4° of varus (4°of valgus to 9° varus), and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient satisfaction with improvement in pain was 91%. Cementless TKR incorporating porous tantalum yielded good clinical and radiological outcomes at a mean of follow-up of seven-years. Cite this article: Bone Joint J 2014;96-B(11 Suppl A):87–92


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 624 - 628
1 May 2010
Macheras GA Kateros K Koutsostathis SD Tsakotos G Galanakos S Papadakis SA

Between November 1997 and December 2000 we performed 27 total hip replacements in 22 patients with high congenital dislocation of the hip using porous tantalum monoblock acetabular components implanted in the true acetabular bed. Clinical and radiological evaluation was performed at regular intervals for a mean of 10.2 years (8.5 to 12). The mean Harris Hip Score improved from 48.3 (15 to 65) pre-operatively to 89.5 (56 to 100) at the final follow-up. The mean Oxford Hip Score was 49.5 (35 to 59) pre-operatively and decreased to 21.2 (12 to 48) at one year and 15.2 (10 to 28) at final follow-up. Migration of the acetabular component was assessed with the EBRA software system. There was a mean migration of 0.68 mm (0.49 to 0.8) in the first year and a mean 0.89 mm (0.6 to 0.98) in the second year, after excluding one initial excessive migration. No revision was necessary for any reason, no acetabular component became loose, and no radiolucent lines were observed at the final follow-up. The porous tantalum monoblock acetabular component is an implant offering adequate initial stability in conjunction with a modulus of elasticity and porosity close to that of cancellous bone. It favours bone ingrowth, leading to good mid-term results