Aims. Intraoperative pressure sensors allow surgeons to quantify
Aims. Total knee arthroplasty (TKA) using functional alignment aims to implant the components with minimal compromise of the soft-tissue envelope by restoring the plane and obliquity of the non-arthritic joint. The objective of this study was to determine the effect of TKA with functional alignment on mediolateral
Complications after total knee arthroplasty (TKR) such as malalignment, instability, subluxation, excessive wear, and loosening have been attributed to poor
Introduction/Aim. Mid-flexion instability is a well-documented, but often poorly understood cause of failure of TKA. NAVIO robotic-assisted TKA (RA-TKA) offers a novel, integrative approach as a planning, execution as well as an evaluation tool in TKA surgery. RA-TKA provides a hybrid planning technique of measured resection and gap balancing- generating a predictive
Decreasing proprioception of the knee is multifactorial and is a function of age and degenerative joint disease. Soft-tissue release during total knee replacement may have an influence. We have quantified soft-tissue imbalance at the time of knee replacement and attempted to eliminate it at full extension, using established methods. We studied the influence of residual soft-tissue imbalance on postoperative proprioception, assessing this in 38 patients before total knee replacement and at three and six months postoperatively. We found that proprioception improved in varus knees at three and six months after
Smart trials are total knee tibial trial liners
with load bearing and alignment sensors that will graphically show quantitative
compartment load-bearing forces and component track patterns. These
values will demonstrate asymmetrical ligament balancing and misalignments
with the medial retinaculum temporarily closed. Currently surgeons
use feel and visual estimation of imbalance to assess soft-tissue
balancing and tracking with the medial retinaculum open, which results
in lower medial compartment loads and a wider anteroposterior tibial
tracking pattern. The sensor trial will aid the total knee replacement
surgeon in performing
We reviewed two similar groups of patients with medial osteoarthritis of the knee treated by unicompartmental arthroplasty. The group receiving an Oxford meniscal-bearing implant, with no medial release, showed significantly better mechanical alignment than that receiving a fixed-bearing implant. Under-correction, with its ominous mechanical implications, was much more common with the fixed-bearing design. Over-correction was rare and was seen in both designs about equally. Degenerative stenosis of the intercondylar notch was common and appeared to put the anterior cruciate ligament at risk of rupture, especially after correction of the varus deformity. We consider that postoperative leg alignment and
The purpose of this study is to evaluate stiff knees which have a preoperative arc of motion (AOM) < 65 degrees and maximum flexion < 90 degrees under anesthesia for primary TKA. We prospectively evaluated 25 knees, 20 patients, the follow up period was 5±3 years, OA 13, RA 10 and traumatic OA 2 knees. All case were medial para-patella approaches and snip was added in one knee operation, 23 PS-type and 2 constrain-type TKAs.Purpose
Material and Methods
Fifteen-year survivorship studies demonstrate that total knee replacements have excellent survivorship, with reports of 85% to 97%. However, excellent survivorship does not equate to excellent patient reported outcomes. Noble et al. reported that 14% of their patients were dissatisfied with their outcome with more than half expressing problems with routine activities of daily living. There is also a difference in the patient's subjective assessment of outcome and the surgeon's objective assessment. Dickstein et al. reported that a third of total knee patients were dissatisfied, even though the surgeons felt that their results were excellent. Most of the patients who report lower outcome scores due so because their expectations are not being fulfilled by the total knee replacement surgery. Perhaps this dissatisfaction is a result of subtle soft tissue imbalance that we have difficulty in assessing intra-operatively and post-operatively. Soft tissue balancing techniques still rely on subjective feel for appropriate ligamentous tension by the surgeon. Surgical experience and case volume play a major role in each surgeon's relative skill in balancing the knee properly. New technology of “smart trials” with embedded microelectronics and accelerometers, used in the knee with the medial retinaculum closed, can provide dynamic, intra-operative feedback regarding knee quantitative compartment pressures and component tracking. After all bone cuts are made using the surgeon's preferred techniques, trial components with the sensored tibial trial are inserted and the knee is taken through a passive range of motion. After visualizing the resultant compartment pressures and tracking data on a graphical interface, the surgeon can decide whether to perform a soft tissue balance or minor bone recuts. If soft tissue balancing is chosen, pressure data can indicate where to perform the release and allow the surgeon to assess the pressure changes as titrated soft tissue releases are performed. A multi-center study using smart trials has demonstrated dramatically better outcomes out to three years.
Appropriate soft tissue balance is an important factor for postoperative function and long survival of total knee arthroplasty(TKA). Soft tissue balance is affected by ligament release, osteophyte removal, order of soft tissue release, cutting angle of tibial surface and rotational alignment of femoral components. The purpose of this study is to know the characteristics of soft tissue balance in ACL deficient osteoarthritis(OA) knee and warning points during procedures for TKA. We evaluated 139 knees, underwent TKA (NexGen LPS-Flex, fixed surface, Zimmer) by one surgeon (S.A.) for OA. All procedures were performed through a medial parapatellar approach. There were 49 ACL deficient knees. A balanced gap technique was used in 26 ACL deficient knees, and anatomical measured technique based on pre-operative CT was used in 23 ACL deficient knees. To compare flexion-extension gaps and medial- lateral balance during operations between the two techniques, we measured each using an original two paddles tensor (figure 1) at 20lb, 30lb and 40lb, for each knee at a 0 degree extension and 90 degree flexion. We measured bone gaps after removal of all osteophytes and cutting of the tibial surface, then we measured component gaps after insertion of femoral components. Statistical analysis was performed by t-test with significant difference defined as P<0.05.INTRODUCION
METHODS
It is unknown whether kinematic alignment (KA) objectively improves knee balance in total knee arthroplasty (TKA), despite this being the biomechanical rationale for its use. This study aimed to determine whether restoring the constitutional alignment using a restrictive KA protocol resulted in better quantitative knee balance than mechanical alignment (MA). We conducted a randomized superiority trial comparing patients undergoing TKA assigned to KA within a restrictive safe zone or MA. Optimal knee balance was defined as an intercompartmental pressure difference (ICPD) of 15 psi or less using a pressure sensor. The primary endpoint was the mean intraoperative ICPD at 10° of flexion prior to knee balancing. Secondary outcomes included balance at 45° and 90°, requirements for balancing procedures, and presence of tibiofemoral lift-off.Aims
Methods
Soft tissue balance is important for good clinical outcome and good stability after TKA. Ligament balancer is one of the devices to measure the soft tissue balance. The objective of this study is to clarify the effect of the difference in the rotational position of the TKA balancer on medial and lateral soft tissue balance. This study included with 50 knees of the 43 patients (6 males, 37 females) who had undergone TKA with ADLER GENUS system from March 2015 to January 2017. The mean age was 71.1±8.1 years. All patients were diagnosed with medial osteoarthritis of the knee. All implants was cruciate substituted type (CS type) and mobile bearing insert. We developed a new ligament balancer that could be fixed to the tibia with keel and insert trial could be rotated on the paddle. We measured the medial and lateral soft tissue balance during TKA with the new balancer. The A-P position of the balancer was fixed on tibia in parallel with the Akagi line (A-P axis 0 group) and 20 degrees internal rotation (IR group) and 20 degrees external rotation (ER group). Soft tissue balance was measured in extension and 90 degrees of knee flexion on each rotational position. The mean angle of valgus and varus in IR group, 0 group and ER group were 4.6±2.2 degrees varus, 1.9±1.6 degrees varus and 0.4±2.4 degrees varus respectively in extension, and 5.5±3.0 degrees varus, 2.1±2.2 degrees varus and 0.7±3.2 degrees varus respectively in 90 degrees of knee flexion. There were significant differences between three groups in extension (p<0.0001) and flexion (p<0.0001). In other words, when the balancer was fixed on tibia with internal rotation against the Akagi line, the soft tissue balance indicated medial tightness. Conversely, when the balancer was fixed on tibia with external rotation against the Akagi line, the soft tissue balance showed lateral tightness. The insert trial significantly rotated to opposite side against the position of balancer fixed.Materials and Methods
Results
Bone preservation is desired for future revision in any knee arthroplasty. There is no study comparing the difference in the amount of bone resection when soft tissue balance is performed with or without computer navigation. To determine the effect on bony cuts when soft tissue balance is performed with or without use of computer software by standard manual technique in total knee arthroplasty. One hundred patients aged 50 to 88 years underwent navigated TKR for primary osteoarthritis. In group A, 50 patients had both soft tissue release and bone cuts done using computer-assisted navigation. In group B, 50 patients had soft tissue release by standard manual technique first and then bone cuts were guided by computer-assisted navigation. In group A the mean medial tibial resection was 5 ± 2.3 mm and lateral was 8 ± 1 mm compared to 5 ± 2 mm ( Our results show that performing soft tissue release and bone cuts using computer- assisted navigation is more bone conserving as compared to manual soft tissue release and bone cuts using computer navigation for TKR, thus preserving bone for possible future revision surgery.Background
The goal of total knee arthroplasty (TKA) is to achieve a stable and well-aligned tibiofemoral and patello-femoral (PF) joint, aiming at long-term clinical patient satisfaction. The surgical principles of both cruciate retaining (CR) and posterior stabilized (PS) TKA are accurate osteotomy and proper soft tissue balancing. We have developed an offset-type tensor, and measured intra-operative soft tissue balance under more physiological joint conditions with femoral component in place and reduced PF joint. In this study, we measured intra-operative soft tissue balance and assessed the post-operative knee joint stability quantitatively at one month, six months and one year after surgery, and compared these parameters between CR and PS TKAs. Sixty patients with varus osteoarthritis of the knee underwent TKAs (30 CR TKAs: CR and 30 PS TKAs: PS). Mean varus deformity in standing position was 11.1 degrees in CR, and 12.6 degrees in PS. All TKAs were performed by a single surgeon with measured resection technique. The external rotation of posterior femoral condyle osteotomy was performed according to surgical epicondylar axis in pre-operative CT. We measured intra-operative soft tissue balance using an offset-type tensor with 40 lbs of joint distraction force at 0, 10, 30, 45, 60, 90, 120 and 135 degrees of flexion. The joint component gap (mm) and varus angle (degrees) were measured at each flexion angles. One month, six months and one year after surgery, we evaluated the knee stability at extension by varus and valgus stress radiography using Telos (10kg) and at flexion by epicondylar view with 1.5kg weight at the ankle. We measured joint separation distance at medial as medial joint looseness (MJL) and at lateral as lateral joint looseness (LJL). Intra-operative measurements and post-operative joint stabilities were compared between CR and PS using unpaired t-test. The change of joint looseness in each group was analyzed using repeated measures ANOVA.Objective
Material and Method
The aim of this prospective multicentre study
was to report the patient satisfaction after total knee replacement (TKR),
undertaken with the aid of intra-operative sensors, and to compare
these results with previous studies. A total of 135 patients undergoing
TKR were included in the study. The
The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of
Aims. The purpose of this multicentre observational study was to investigate the association between intraoperative component positioning and
Aims. It is unknown whether gap laxities measured in robotic arm-assisted total knee arthroplasty (TKA) correlate to load sensor measurements. The aim of this study was to determine whether symmetry of the maximum medial and lateral gaps in extension and flexion was predictive of knee balance in extension and flexion respectively using different maximum thresholds of intercompartmental load difference (ICLD) to define balance. Methods. A prospective cohort study of 165 patients undergoing functionally-aligned TKA was performed (176 TKAs). With trial components in situ, medial and lateral extension and flexion gaps were measured using robotic navigation while applying valgus and varus forces. The ICLD between medial and lateral compartments was measured in extension and flexion with the load sensor. The null hypothesis was that stressed gap symmetry would not correlate directly with sensor-defined soft tissue balance. Results. In TKAs with a stressed medial-lateral gap difference of ≤1 mm, 147 (89%) had an ICLD of ≤15 lb in extension, and 112 (84%) had an ICLD of ≤ 15 lb in flexion; 157 (95%) had an ICLD ≤ 30 lb in extension, and 126 (94%) had an ICLD ≤ 30 lb in flexion; and 165 (100%) had an ICLD ≤ 60 lb in extension, and 133 (99%) had an ICLD ≤ 60 lb in flexion. With a 0 mm difference between the medial and lateral stressed gaps, 103 (91%) of TKA had an ICLD ≤ 15 lb in extension, decreasing to 155 (88%) when the difference between the medial and lateral stressed extension gaps increased to ± 3 mm. In flexion, 47 (77%) had an ICLD ≤ 15 lb with a medial-lateral gap difference of 0 mm, increasing to 147 (84%) at ± 3 mm. Conclusion. This study found a strong relationship between intercompartmental loads and gap symmetry in extension and flexion measured with prostheses in situ. The results suggest that ICLD and medial-lateral gap difference provide similar assessment of
Computer assisted total knee arthroplasty helps in accurate and reproducible implant positioning, bony alignment, and
Introduction. Proper