Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_25 | Pages 5 - 5
1 May 2013
Dalgleish S Finlayson D Cochrane L Hince A
Full Access

Radiation exposure is a hazard to orthopaedic surgeons, theatre staff and patients intra-operatively. Obesity is becoming a more prevalent problem worldwide and there is little evidence how a patient's body habitus correlates with the radiation doses required to penetrate the soft tissues for adequate imaging. We aimed to identify if there was a correlation between Body Mass Index (BMI) and radiation exposure intra-operatively. We performed a retrospective review of 75 patients who underwent sliding hip screw fixation for femoral neck fractures in one year. We recorded Body Mass Index (BMI), screening time, dose area product (DAP), American Society of Anesthesiologists (ASA) grade, seniority of surgeon and complexity of the fracture configuration. We analysed the data using statistical tests. We found that there was a statistically significant correlation between dose area product and patient's BMI. There was no statistically significant relationship between screening time and BMI. There was no statistical difference between ASA grade, seniority of surgeon, or complexity of fracture configuration and dose area product. Obese patients are exposed to increased doses of radiation regardless of length of screening time. Surgeons and theatre staff should be aware of the increased radiation exposure during fixation of fractures in obese patients and, along with radiographers, ensure steps are taken to minimise these risks


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_5 | Pages 27 - 27
1 May 2015
Bryant H Dearden P Harwood P Wood T Sharma H
Full Access

Methods:. Total radiation exposure accumulated during circular frame treatment of distal tibial fractures was quantified in 47 patients treated by a single surgeon from March 2011 until Nov 2014. The radiation exposures for all relevant radiology procedures for the distal tibial injury were included to estimate the radiation risk to the patient. Results:. The median time of treatment in the frame was 169 days (range 105 – 368 days). Patients underwent a median of 13 sets of plain radiographs; at least one intra operative exposure and 16 patients underwent CT scanning. The median total effective dose per patient from time of injury to discharge was 0.025 mSv (interquartile range 0.013 – 0.162 and minimum to maximum 0.01–0.53). CT scanning is the only variable shown to be an independent predictor of cumulative radiation dose on multivariate analysis, with a 13 fold increase in overall exposure. Conclusion:. Radiation exposure during treatment of distal tibial fractures with a circular frame in this group was well within reasonably safe limits. CT was the only significant predictor of overall exposure serves as a reminder to individually assess the risk and utility of radiological investigations on an individual basis. This is consistent with the UK legal requirements (Ionising Radiation (Medical Exposure) Regulations 2000. 1. )


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_16 | Pages 50 - 50
1 Apr 2013
Hak D Thornton R Dauer LT Quinn B Miodownik D
Full Access

Introduction. Radiation exposure to the eye causes cataracts. Few orthopaedists wear leaded glasses when using fluoroscopy despite regulatory limits for maximum annual eye exposure. Methods. Using anthropomorphic patient and surgeon phantoms, radiation dose at the surgeon phantom's lens was measured with and without leaded glasses during fluroscopic acquisition of 16 common pelvic and hip views. The magnitude of lens dose reduction was calculated by dividing the unprotected dose by the dose measured behind leaded glasses. Results. The unprotected lens dose varied considerably among the different views, ranging from 0 μRem for a single obturator oblique pelvic view, to 257 μRem for a single lateral sacral view with the image intensifier opposite the surgeon. On average, use of leaded glassess reduced radiation to the surgeon phantom's eye by 90%. The greatest reduction was seen with a crosstable lateral radiograph of the hip (22 μRem unprotected lens dose vs. 1 μRem with leaded glasses). Conclusions. The use of leaded glasses could permit an orthopaedist to perform up to 10 times more cases before reaching the annual limit of 20 mSv (20 mSv = 2×10. 6. mRem). A large number of images are often required during a single operative procedure and varies greatly between cases. The number of cases which a surgeon can safely perform without leaded glasses is dependent on the number and type of images, and location of the surgeon with respect to the image intensifier


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 696 - 702
1 May 2016
Theologis AA Burch S Pekmezci M

Aims

We compared the accuracy, operating time and radiation exposure of the introduction of iliosacral screws using O-arm/Stealth Navigation and standard fluoroscopy.

Materials and Methods

Iliosacral screws were introduced percutaneously into the first sacral body (S1) of ten human cadavers, four men and six women. The mean age was 77 years (58 to 85). Screws were introduced using a standard technique into the left side of S1 using C-Arm fluoroscopy and then into the right side using O-Arm/Stealth Navigation. The radiation was measured on the surgeon by dosimeters placed under a lead thyroid shield and apron, on a finger, a hat and on the cadavers.


Bone & Joint Open
Vol. 5, Issue 2 | Pages 117 - 122
9 Feb 2024
Chaturvedi A Russell H Farrugia M Roger M Putti A Jenkins PJ Feltbower S

Aims. Occult (clinical) injuries represent 15% of all scaphoid fractures, posing significant challenges to the clinician. MRI has been suggested as the gold standard for diagnosis, but remains expensive, time-consuming, and is in high demand. Conventional management with immobilization and serial radiography typically results in multiple follow-up attendances to clinic, radiation exposure, and delays return to work. Suboptimal management can result in significant disability and, frequently, litigation. Methods. We present a service evaluation report following the introduction of a quality-improvement themed, streamlined, clinical scaphoid pathway. Patients are offered a removable wrist splint with verbal and written instructions to remove it two weeks following injury, for self-assessment. The persistence of pain is the patient’s guide to ‘opt-in’ and to self-refer for a follow-up appointment with a senior emergency physician. On confirmation of ongoing signs of clinical scaphoid injury, an urgent outpatient ‘fast’-wrist protocol MRI scan is ordered, with instructions to maintain wrist immobilization. Patients with positive scan results are referred for specialist orthopaedic assessment via a virtual fracture clinic. Results. From February 2018 to January 2019, there were 442 patients diagnosed as clinical scaphoid fractures. 122 patients (28%) self-referred back to the emergency department at two weeks. Following clinical review, 53 patients were discharged; MRI was booked for 69 patients (16%). Overall, six patients (< 2% of total; 10% of those scanned) had positive scans for a scaphoid fracture. There were no known missed fractures, long-term non-unions or malunions resulting from this pathway. Costs were saved by avoiding face-to-face clinical review and MRI scanning. Conclusion. A patient-focused opt-in approach is safe and effective to managing the suspected occult (clinical) scaphoid fracture. Cite this article: Bone Jt Open 2024;5(2):117–122


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 473 - 473
1 Sep 2012
Kotwal R Rath N Paringe V Lyons K Thomas R
Full Access

Introduction. The assessment of the accuracy of reduction of the ankle syndesmosis has traditionally been made using plain radiographic measurements. Recent studies have shown that computerized tomography (CT) scan is more sensitive than radiographs in detecting diastasis. The ethos has now therefore shifted towards CT scan assessment of the syndesmosis. There is however no validated method to scan the syndesmosis and measure it on the CT scans. This exposes the patient to significant radiation risk and also to anxiety from inappropriate interpretation from these scans. The objectives of this research project are to investigate the current practice of CT scanning the syndesmosis at a University Hospital and to devise a new CT protocol to reduce radiation exposure to patients and to assist surgeons in interpreting the observations. Methods. Research Ethics Committee approval was obtained. Current practice was evaluated. A new 5 cut CT protocol was devised. Starting at the level of the distal tibial plafond, 5 cuts were made proximally 0.5 cm apart. Accuracy of the syndesmosis reduction was assessed just above the distal tibial plafond. Both the injured and the normal sides were scanned 12 weeks post surgery. The normal side served as a control. Results. Current practice revealed that patients had on an average 620 cuts CT scan with radiation exposure of 0.2 mSv. 25 patients were prospectively recruited for the new 5 cut CT protocol. The radiation exposure with the new protocol was only 0.002 mSv. Comparison with the normal side revealed that 5 (20%) of syndesmosis had residual diastasis. The only difficulty with the new protocol was getting both the ankle joint lines at the same level for the first CT cut. Discussion and Conclusion. Our CT scan protocol has insignificant radiation risk, even lower than a single chest radiograph. Comparing the measurements between the fractured and the normal sides provides an accurate assessment of the reduction of the syndesmosis. Significance of the measurements on the CT scan will be correlated with functional scores


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_8 | Pages 36 - 36
1 May 2018
Fawdington R Beaven A Fenton P Lofti N
Full Access

Background. In a systematic review of 1125 distal tibia fractures treated with an intramedullary nail, the reported incidence of malalignment was 14%. The purpose of our study is to assess whether the addition of blocking screws during intramedullary nailing of a distal tibia fracture improved radiological outcomes. As a secondary outcome, the time to radiographic union was compared to see if a more rigid bone-implant construct had an effect on healing. Methods. We searched computerised records at a UK level 1 major trauma centre. The joint alignment was measured on the immediate post-operative radiograph and compared to the most recent radiograph. We used a difference of 2 degrees to indicate a progressive deformity and a RUST score greater or equal to 10, to indicate radiographic fracture union. Results. Twenty-seven patients were included. Nineteen patients had no blocking screw and 8 patients had a blocking screw. Five patients had a difference in their coronal plane alignment of 2 degrees or more (3/5 had no blocking screw). The results were analysed and found to be not statistically significant (p=0.88). The addition of a blocking screw has also been shown not to have an effect on the time to union. Conclusion. We have changed our surgical practice. We use a 2.5mm blocking wire to aid in fracture reduction prior to reaming / nail insertion and then remove the wire when the nail has been adequately locked. Implications. This could save surgical procedure time, radiation exposure for the patient, implant costs and potential complications


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 506 - 506
1 Sep 2012
Baxter J Emmett S Barlow T Costa M
Full Access

Introduction. The Patient Archiving and Communication System (PACS) has revolutionised the way that radiographs are stored and viewed in orthopaedic surgery. A recent advance has been the ability to upload images directly from the image intensifier to PACS. We postulated that this facility may reduce the need for post operative ‘check’ radiographs following many orthopaedic trauma procedures. Patients and Method. We performed an audit of post-operative radiographs requested in our University Hospital over three time periods: 31 days immediately before the direct upload facility was introduced, 31 days immediately after and a 31 day period two months later. Details from the operating lists were cross-referenced with image intensifier records to identify cases where it had been used. PACS records were then checked to determine if these images were available to view and if a formal ‘check’ radiograph was performed in the period prior to discharge. Results. A total of 624 orthopaedic trauma operations were reviewed. In the period before direct-upload of images began, 62% of all trauma procedures utilised the image intensifier and in 26% post-operative ‘check’ radiographs were taken. In the period immediately after uploading began, the image intensifier was used in 66% of cases with 84% of these images being uploaded. Fifteen percent of the patients had a check radiograph during this period. Two months later the image intensifier was used in 67% of cases. Although 96% of these images were uploaded, 14% still had a formal post-operative radiograph. Conclusion. The facility to upload images from the image intensifier in the operating theatre directly onto the PACS system has reduced the number of post-operative ‘check’ radiographs requested in our trauma department by 11%. This has reduced both the cost and radiation exposure involved in orthopaedic trauma surgery but has not eliminated the need for post-operative radiographs completely


Bone & Joint Open
Vol. 5, Issue 2 | Pages 147 - 153
19 Feb 2024
Hazra S Saha N Mallick SK Saraf A Kumar S Ghosh S Chandra M

Aims

Posterior column plating through the single anterior approach reduces the morbidity in acetabular fractures that require stabilization of both the columns. The aim of this study is to assess the effectiveness of posterior column plating through the anterior intrapelvic approach (AIP) in the management of acetabular fractures.

Methods

We retrospectively reviewed the data from R G Kar Medical College, Kolkata, India, from June 2018 to April 2023. Overall, there were 34 acetabulum fractures involving both columns managed by medial buttress plating of posterior column. The posterior column of the acetabular fracture was fixed through the AIP approach with buttress plate on medial surface of posterior column. Mean follow-up was 25 months (13 to 58). Accuracy of reduction and effectiveness of this technique were measured by assessing the Merle d’Aubigné score and Matta’s radiological grading at one year and at latest follow-up.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 418 - 418
1 Sep 2012
Goost H Karius T Deborre C Kabir K Randau T Burger C Wirtz D Pflugmacher R
Full Access

Introduction. Pedicle screw pullout or loosening is increased in the osteoporotic spine. Recent studies showed a significant increase of pullout forces especially for PMMA-augmentation. With application of conventional viscosity PMMA the risk of cement extravasation is associated. This risk can be reduced by using radiofrequency-responsive, ultrahigh viscosity bone cement. Method. 11 fresh-frozen lumbar vertebral bodies (VB) from 5 cadavers were collected and freed from soft-tissue and ligaments. By DEXA scan (Siemens QDR 2000) 8 VB were identified as severely osteoporotic (BMD 0.8 g/cm3), 3 VB were above this level. Two screws (6×45 mm, WSI-Expertise Inject, Peter Brehm, Weisendorf, Germany) were placed in the pedicles. Through the right screw 3ml of radiofrequency-responsive bone cement (StabiliT® ER2 Bone Cement, DFine, Germany) were injected via hydraulic cement delivery system (StabiliT® Vertebral Augmentation System, DFine, Germany). As control group, left pedicle screws remained uncemented. After potting the whole VB in technical PMMA (Technovit 4004, Heraeus Kulzer, Germany) axial pullout test was performed by a material testing device (Zwick-Roell, Zmart-Pro, Ulm, Germany). Results. The mean BMD of all specimen was 0.771 g/cm3 (min./max. 0.615/1.116, SD ± 0.170). Due to the definition of osteoporosis 8 specimens had a BMD lower than 0.8 g/cm3 (mean 0.677, min./max. 0.615/0,730, SD ±0.045). The non-osteoporotic group consisted of 3 specimens with a mean BMD of 1.020 g/cm3 (min/max 0.928/1.116, SD ±0.094). Overall we observed an increase in the mean axial pullout strength of 284% when using cement augmentated screws (non-cemented 385 N vs. cemented 1029 N, p 0,001). In the osteoporotic group the mean pullout force of the non-cemented screws was 407 N vs. 1022 N for the cemented screws (p 0.001). Similarly the pullout force rose in the non-osteoporotic group from 325 N for the non-cemented screws to 1048 N for the cemented screws p 0,001). All surgical procedures could be performed without technical problems. Conclusion. This cadaver study demonstrates the efficacy and effectiveness of pedicle-screw augmentation with ultra-high viscosity cement. Pullout forces are significantly increased, especially in osteoporotic bone. No complications like clogged in cannulated pedicle-screws or extravasation of bone cement were observed. In daily clinical routine radiation exposure to operator during cement delivery is reduced due to remote-controlled, automated delivery of radiofrequency-responsive bone cement. Furthermore availability of longer time to work with the cement (up to 30 min) is achieved; hectically injection or multiple-cement-mixing is not necessary anymore. The WSI-Expertise cannulated pedicle screws can be inserted and also augmented in a minimal-invasive technique. The PMMA then is injected through an extension adapter. PMMA augmentation of pedicle screws is already well established in spine surgery. With the combination of radiofrequency-responsive, ultra-high viscosity bone cement and cannulated pedicle screws this established procedure could become more controlled and safer for both patient and physician


Bone & Joint Open
Vol. 3, Issue 9 | Pages 726 - 732
16 Sep 2022
Hutchison A Bodger O Whelan R Russell ID Man W Williams P Bebbington A

Aims

We introduced a self-care pathway for minimally displaced distal radius fractures, which involved the patient being discharged from a Virtual Fracture Clinic (VFC) without a physical review and being provided with written instructions on how to remove their own cast or splint at home, plus advice on exercises and return to function.

Methods

All patients managed via this protocol between March and October 2020 were contacted by a medical secretary at a minimum of six months post-injury. The patients were asked to complete the Patient-Rated Wrist Evaluation (PRWE), a satisfaction questionnaire, advise if they had required surgery and/or contacted any health professional, and were also asked for any recommendations on how to improve the service. A review with a hand surgeon was organized if required, and a cost analysis was also conducted.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1668 - 1673
1 Dec 2016
Konda SR Goch AM Leucht P Christiano A Gyftopoulos S Yoeli G Egol KA

Aims

To evaluate whether an ultra-low-dose CT protocol can diagnose selected limb fractures as well as conventional CT (C-CT).

Patients and Methods

We prospectively studied 40 consecutive patients with a limb fracture in whom a CT scan was indicated. These were scanned using an ultra-low-dose CT Reduced Effective Dose Using Computed Tomography In Orthopaedic Injury (REDUCTION) protocol. Studies from 16 selected cases were compared with 16 C-CT scans matched for age, gender and type of fracture. Studies were assessed for diagnosis and image quality. Descriptive and reliability statistics were calculated. The total effective radiation dose for each scanned site was compared.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 191 - 197
1 May 2016
Kienast B Kowald B Seide K Aljudaibi M Faschingbauer M Juergens C Gille J

Objectives

The monitoring of fracture healing is a complex process. Typically, successive radiographs are performed and an emerging calcification of the fracture area is evaluated. The aim of this study was to investigate whether different bone healing patterns can be distinguished using a telemetric instrumented femoral internal plate fixator.

Materials and Methods

An electronic telemetric system was developed to assess bone healing mechanically. The system consists of a telemetry module which is applied to an internal locking plate fixator, an external reader device, a sensor for measuring externally applied load and a laptop computer with processing software. By correlation between externally applied load and load measured in the implant, the elasticity of the osteosynthesis is calculated. The elasticity decreases with ongoing consolidation of a fracture or nonunion and is an appropriate parameter for the course of bone healing. At our centre, clinical application has been performed in 56 patients suffering nonunion or fracture of the femur.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 116 - 121
1 Apr 2016
Leow JM Clement ND Tawonsawatruk T Simpson CJ Simpson AHRW

Objectives

The radiographic union score for tibial (RUST) fractures was developed by Whelan et al to assess the healing of tibial fractures following intramedullary nailing. In the current study, the repeatability and reliability of the RUST score was evaluated in an independent centre (a) using the original description, (b) after further interpretation of the description of the score, and (c) with the immediate post-operative radiograph available for comparison.

Methods

A total of 15 radiographs of tibial shaft fractures treated by intramedullary nailing (IM) were scored by three observers using the RUST system. Following discussion on how the criteria of the RUST system should be implemented, 45 sets (i.e. AP and lateral) of radiographs of IM nailed tibial fractures were scored by five observers. Finally, these 45 sets of radiographs were rescored with the baseline post-operative radiograph available for comparison.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 646 - 651
1 May 2014
Mutch J Laflamme GY Hagemeister N Cikes A Rouleau DM

In this study, we describe a morphological classification for greater tuberosity fractures of the proximal humerus. We divided these fractures into three types: avulsion, split and depression. We retrospectively reviewed all shoulder radiographs showing isolated greater tuberosity fractures in a Level I trauma centre between July 2007 and July 2012. We identified 199 cases where records and radiographs were reviewed and included 79 men and 120 women with a mean age of 58 years (23 to 96). The morphological classification was applied to the first 139 cases by three reviewers on two occasions using the Kappa statistic and compared with the AO and Neer classifications. The inter- and intra-observer reliability of the morphological classification was 0.73 to 0.77 and 0.69 to 0.86, respectively. This was superior to the Neer (0.31 to 0.35/0.54 to 0.63) and AO (0.30 to 0.32/0.59 to 0.65) classifications. The distribution of avulsion, split and depression type fractures was 39%, 41%, and 20%, respectively. This classification of greater tuberosity fractures is more reliable than the Neer or AO classifications. These distinct fracture morphologies are likely to have implications in terms of pathophysiology and surgical technique.

Cite this article: Bone Joint J 2014;96-B:646–51.