Pre-operative computerised three-dimensional planning was carried out in 223 patients undergoing total hip replacement with a cementless acetabular component and a cementless modular-neck femoral stem. Components were chosen which best restored leg length and femoral offset. The post-operative restoration of the anatomy was assessed by CT and compared with the
Total joint arthroplasty (TJA) is one of the commonest and most successful orthopaedic procedures used for the management of end-stage arthritis. With the recent introduction of robotic-assisted joint replacement, Computed tomography (CT) has become part of required
We have developed a technique with the underlying principle being that the difference in height between what is removed and what is inserted will determine the leg length correction (Figure 1). The height of the implant to be inserted is determined from the manufacturer's specifications. We have developed a Vertical Measurement Tool to accurately and reproducibly determine the height of the resected bone (Figure 2). Leg length correction = a−b−c+d. Vertical Measurement Tool validation was performed by 4 separate surgeons on 20 resected femoral heads in the laboratory. Inter and intra-observer error was assessed. Fifty patients were assessed clinically and radiologically, to assess if desired leg length correction was achieved. Statistical analysis showed the device to be accurate with high intra and inter observer reliability. Differences between the observers were tested using a general linear model in a repeated measure design. No main effect and interaction effects were found. Intra operatively the resected head was measured and the formula was applied. The range of desired correction was 0mm to 18mm. In all cases the post-operative correction was within 4mm of the
We describe our technique and rationale using
hybrid fixation for primary total hip arthroplasty (THA) at the Hospital
for Special Surgery. Modern uncemented acetabular components have
few screw holes, or no holes, polished inner surfaces, improved
locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented
sockets can be combined with highly cross-linked polyethylene liners,
which have demonstrated very low wear and osteolysis rates after
ten to 15 years of implantation. The results of cement fixation
with a smooth or polished surface finished stem have been excellent,
virtually eliminating complications seen with cementless fixation
like peri-operative femoral fractures and thigh pain. Although mid-term
results of modern cementless stems are encouraging, the long-term
data do not show reduced revision rates for cementless stems compared
with cemented smooth stems. In this paper we review the conduct
of a hybrid THA, with emphasis on
We evaluated the accuracy with which a custom-made
acetabular component could be positioned at revision arthroplasty
of the hip in patients with a Paprosky type 3 acetabular defect. A total of 16 patients with a Paprosky type 3 defect underwent
revision surgery using a custom-made trabecular titanium implant.
There were four men and 12 women with a median age of 67 years (48
to 79). The planned inclination (INCL), anteversion (AV), rotation
and centre of rotation (COR) of the implant were compared with the post-operative
position using CT scans. A total of seven implants were malpositioned in one or more parameters:
one with respect to INCL, three with respect to AV, four with respect
to rotation and five with respect to the COR. To the best of our knowledge, this is the first study in which
CT data acquired for the
Introduction. Patients with abnormal spinopelvic mobility are at increased risk for hip instability. Measuring the change in sacral slope (ΔSS) with standing and seated lateral radiographs is commonly used to determine spinopelvic mobility pre-operatively. Sacral slope should decrease at least 10 degrees to demonstrate adequate accommodation. Accommodation of <10 deg necessitates acetabular component position change or use of a dual mobility implant. There is potential for different ΔSS measurements in the same patient based on sitting posture. Methods. 78 patients who underwent THA were reviewed to quantify the variability in pre-operative spinopelvic mobility when two different seated positions (relaxed sitting v. pre-rise sitting) were used in the same patient. Results. 34 patients had standardized pre-rise sitting x-rays, while 44 patients had standardized relaxed sitting x-rays. Of the 44 patients with relaxed sitting x-rays, the mean ΔSS (ΔrSS) was 20.4 degrees. No patients exhibited an increase in sacral slope when sitting (ie; reverse accommodation). Of the 34 patients with pre-rise sitting x-rays, the mean pre-rise sit-stand change (ΔprSS) was only 1.85 degrees with 47% (16/34) showing reverse accommodation, actually increasing the seated sacral slope compared to standing sacral slope. 18 patients had both pre-rise and relaxed sitting x-rays. In patients with both seated x-rays, the mean relaxed sit-stand change in sacral slope (ΔrSS) was 18.1 ± 6.1 degrees and only 3.0 ± 10.5 degrees for pre-rise sit-stand (ΔprSS), with a mean ΔSS difference of the 15.1 degrees (p <0.0001). Conclusion. A 15 degrees error could be made in
Total hip replacement for developmental hip dysplasia
is challenging. The anatomical deformities on the acetabular and
femoral sides are difficult to predict. The Crowe classification
is usually used to describe these cases – however, it is not a very
helpful tool for
The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle. Prudent
We report our experience of revision total hip
replacement (THR) using the Revitan curved modular titanium fluted revision
stem in patients with a full spectrum of proximal femoral defects.
A total of 112 patients (116 revisions) with a mean age of 73.4
years (39 to 90) were included in the study. The mean follow-up
was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died
but their data were included in the survival analysis, and four
patients (4 hips) were lost to follow-up. The clinical outcome,
proximal bone regeneration and subsidence were assessed for 101
hips. The mean Harris Hip Score was 88.2 (45.8 to 100) after five years
and there was an increase of the mean Barnett and Nordin-Score,
a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7).
Five stems had to be revised (4.3%), three (2.9%) showed subsidence,
five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%)
a periprosthetic infection. . At the latest follow-up, the survival with revision of the stem
as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%)
and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic
fractures were not observed. We report excellent results with respect to subsidence, the risk
of fracture, and loosening after femoral revision using a modular
curved revision stem with distal cone-in-cone fixation. A successful
outcome depends on careful
Dislocation is one of the most common causes
of patient and surgeon dissatisfaction following hip replacement
and to treat it, the causes must first be understood. Patient factors
include age greater than 70 years, medical comorbidities, female
gender, ligamentous laxity, revision surgery, issues with the abductors,
and patient education. Surgeon factors include the annual quantity
of procedures and experience, the surgical approach, adequate restoration
of femoral offset and leg length, component position, and soft-tissue
or bony impingement. Implant factors include the design of the head
and neck region, and so-called skirts on longer neck lengths. There
should be offset choices available in order to restore soft-tissue
tension. Lipped liners aid in gaining stability, yet if improperly placed
may result in impingement and dislocation. Late dislocation may
result from polyethylene wear, soft-tissue destruction, trochanteric
or abductor disruption and weakness, or infection. Understanding
the causes of hip dislocation facilitates prevention in a majority
of instances. Proper
Clinical graphics allows creation of three dimensional simulation based on CT or MRI that allows
Introduction. The prevalence of the various patterns of spinopelvic abnormalities that increase the risk for prosthetic impingement is unknown. While prior surgery or lumbar fusion are recognized as a risk factors for postoperative dislocation, many patients presenting for THA do not have obvious radiographic abnormalities. The purpose of this study is to determine the prevalence of large posterior pelvic tilt (PPT) when standing, stiff lumbar-spine (SLL) and spino-pelvic sagittal imbalance (SSI) in patients undergoing primary THA. Methods. A consecutive series of 1592 patients (56% female) over 2 years underwent functional analysis of spinopelvic mobility using CT, standing, and flexed seated lateral radiographs as part of
Introduction. The achieved anteversion of uncemented stems is to a large extent limited by the internal anatomy of the bone. A better understanding of this has recently become an unmet need because of the increased use of uncemented stems. We aimed to assess plan compliance in six degrees of freedom to evaluate the accuracy of PSI and guides for stem positioning in primary THAs. Materials and Methods. We prospectively collected 3D plans generated from preoperative CTs of 30 consecutive THAs (17 left and 13 right hips), in 29 patients with OA, consisting of 16 males and 13 females (median age 68 years, range 46–83 years). A single CT-based planning system and cementless type of implant were used. Post operatively, all patients had a CT scan which was reconstructed using state-of-the-art software solution: the plan and CT reconstruction models were. Outcome measures: 1) discrepancy between planned and achieved stem orientation angles Fig.2&3; 2) clinical outcome. Results. 1) The mean (±SD) discrepancy was low for: Varus-valgus −1.1 ± 1.4 deg (IQR −2.2 – 0.3 deg); Anterior-posterior 0.1 ± 1.6 deg (IQR −0.7 – 1.3 deg). The discrepancy was higher for femoral version −1.4 ± 8.2 deg (IQR −8.3 – 7.2 deg). 3D-CT planning correctly predicted sizes in 93% of the femoral components. 2) There was no intra-operative fracture, no case showed evidence of early periprosthetic osseous injury. Discussion. Surgeons and engineers should be cautious with their expectation of achieving the planned femoral stem version of an uncemented femoral stem from the
Complex total hip arthroplasty (THA) with subtrochanteric shortening osteotomy is necessary in conditions other than developmental dysplasia of the hip (DDH) and septic arthritis sequelae with significant proximal femur migration. Our aim was to evaluate the hip centre restoration with THAs in these hips. In all, 27 THAs in 25 patients requiring THA with femoral shortening between 2012 and 2019 were assessed. Bilateral shortening was required in two patients. Subtrochanteric shortening was required in 14 out of 27 hips (51.9%) with aetiology other than DDH or septic arthritis. Vertical centre of rotation (VCOR), horizontal centre of rotation, offset, and functional outcome was calculated. The mean followup was 24.4 months (5 to 92 months).Aims
Methods
There is no consensus about the best method of achieving equal
leg lengths at total hip arthroplasty (THA) in patients with Crowe
type-IV developmental dysplasia of the hip (DDH). We reviewed our
experience of a consecutive series of patients who underwent THA
for this indication. We retrospectively reviewed 78 patients (86 THAs) with Crowe
type-IV DDH, including 64 women and 14 men, with a minimum follow-up
of two years. The mean age at the time of surgery was 52.2 years
(34 to 82). We subdivided Crowe type-IV DDH into two major types
according to the number of dislocated hips, and further categorised
them into three groups according to the occurrence of pelvic obliquity
or spinal curvature. Leg length discrepancy (LLD) and functional
scores were analysed.Aims
Patients and Methods
The aim of this study was to evaluate the long-term clinical
and radiographic outcomes of the Birmingham Interlocking Pelvic
Osteotomy (BIPO). In this prospective study, we report the mid- to long-term clinical
outcomes of the first 100 consecutive patients (116 hips; 88 in
women, 28 in men) undergoing BIPO, reflecting the surgeon’s learning
curve. Failure was defined as conversion to hip arthroplasty. The
mean age at operation was 31 years (7 to 57). Three patients (three
hips) were lost to follow-up.Aims
Patients and Methods
The Corail stem has good long-term results. After four years
of using this stem, we have detected a small group of patients who
have presented with symptomatic metaphyseal debonding. The aim of
this study was to quantify the incidence of this complication, to
delineate the characteristics of patients presenting with this complication
and to compare these patients with asymptomatic controls to determine
any important predisposing factors. Of 855 Corail collarless cementless stems implanted for osteoarthritis,
18 presented with symptomatic metaphyseal debonding. A control group
of 74 randomly selected patients was assembled. Clinical and radiological
parameters were measured and a logistic regression model was created
to evaluate factors associated with metaphyseal debonding.Aims
Patients and Methods
Our aim was to evaluate the radiographic characteristics of patients
undergoing total hip arthroplasty (THA) for the potential of posterior
bony impingement using CT simulations. Virtual CT data from 112 patients who underwent THA were analysed.
There were 40 men and 72 women. Their mean age was 59.1 years (41
to 76). Associations between radiographic characteristics and posterior
bony impingement and the range of external rotation of the hip were
evaluated. In addition, we investigated the effects of pelvic tilt
and the neck/shaft angle and femoral offset on posterior bony impingement.Aims
Patients and Methods
The stability of cementless acetabular components is an important
factor for surgical planning in the treatment of patients with pelvic
osteolysis after total hip arthroplasty (THA). However, the methods
for determining the stability of the acetabular component from pre-operative
radiographs remain controversial. Our aim was to develop a scoring
system to help in the assessment of the stability of the acetabular
component under these circumstances. The new scoring system is based on the mechanism of failure of
these components and the location of the osteolytic lesion, according
to the DeLee and Charnley classification. Each zone is evaluated
and scored separately. The sum of the individual scores from the
three zones is reported as a total score with a maximum of 10 points.
The study involved 96 revision procedures which were undertaken
for wear or osteolysis in 91 patients between July 2002 and December
2012. Pre-operative anteroposterior pelvic radiographs and Judet
views were reviewed. The stability of the acetabular component was
confirmed intra-operatively.Aims
Patients and Methods