header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 224 - 224
1 Sep 2012
Ebreo D Khan A El-Meligy M Armstrong C Peter V
Full Access

INTRODUCTION. The advantages of large diameter metal on metal total hip arthroplasty (MoM THA) and hip resurfacing arthroplasty are decreased wear rate, preservation of bone stock, anatomical restoration and enhanced stability. Large amounts of metal wear particles and metal ions are released which may induce adverse reactions including local soft tissue toxicity, hypersensitivity reactions, bone loss and risk of carcinogenesis. Aseptic loosening can be the result of a peri-prosthetic osteolysis generated as a result of a biological response to particulate wear debris. No reports in the literature exist as to whether circulating levels of Chromium (Cr) and Cobalt (Co) decrease upon removal of a symptomatic large diameter MoM implant or whether levels remain high due to the effect of metal ions debris left behind in the soft tissues after revision surgery. PATIENTS AND METHODS. Between June 2006 and June 2009 we undertook 44 revision surgeries of both large head MoM THAs (femoral head diameter 38mm) and metal-on-metal hip resurfacings for suspected metallosis. Mean time from original implant to revision was 4 years, 8 months (1yr 4mo–7yr 9mo). The mean follow up evaluation was 2 years and 2 months (1yr 2mo–4 years). Blood samples were taken for whole blood Cr and serum Co according to a recognised protocol and compared with reference levels indicated by the Medicines and Healthcare Regulatory Agency recommendation of less than 7ppb for Cr (130nM/L) and Co (119nM/L). RESULTS. 42 patients were found to have histological evidence of either metal allergy, metal toxicity or foreign body reaction. 2 patients had evidence of infection with no features of metal reaction. 3 patients suffered early dislocation requiring closed reduction. 1 patient had infective complications necessitating Girdlestones. 11 patients were lost to follow up, 8 patients were diagnosed pre operatively on Co and Cr levels in urine or synovial fluid aspirate alone. 23 patients had pre revision blood or serum metal ion level results available for direct comparison. Median serum Co level pre revision was 176.6nM/L, falling post revision to 5.1nM/L (p=<0.001∗). The median whole blood Cr level pre revision was 117nM/L and 19nm/L post revision (p=<0.001∗). Mean Oxford Hip Score was 23.7. DISCUSSION. This study demonstrates that at greater than one year post removal of a large diameter MoM hip implant for the indication of symptomatic metallosis or metal hypersensitivity, metal ion levels fall to almost normal levels and that outcome of revision surgery in terms of patient satisfaction is not adversely affected