Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 3 - 3
7 Aug 2023
Fennelly J Santini A Papalexandris S Pope J Yorke J Davidson J
Full Access

Abstract. Background. Oxidized zirconium (OxZr) has been introduced as an alternative bearing for femoral components in Total Knee Arthroplasty (TKA). It has a ceramic-like zirconium oxide outer layer with a low coefficient of friction. Early studies have found OxZr TKA to have a low incidence of early failure in young high demand patients. Currently no study has reported on the outcome of these implants beyond ten years. Objectives. The purpose of our study was to present an in-depth 15-year survival analysis of cemented Profix II OxZr TKA. Study Design & Methods. Data was collected prospectively and survival analysis undertaken with multiple strict end points. Complication rates were recorded and patient reported outcomes were measured. Results. 617 Profix II OxZr TKAs were performed over four years. Forty-nine patients underwent reoperation. Aseptic tibial loosening was the most common cause of failure (32.7%) on average occurring 2.8 years post primary procedure. There was one recorded failure due to loosening of the zirconium femoral component. Revision rate at 15-years was 6.38%. Cumulative survivorship was 91.52% with failure considered to be reoperation for any reason. WOMAC score improved in 86% of patients by year 1. The average score improved by 21.2 points and met the standard for minimum clinically important difference. Conclusions. This study presents the first 15-year survival analysis of cemented Profix II OxZr TKA. Our data supports current literature on the long-term survivorship of oxidised zirconium total knee replacements


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 42 - 42
1 Oct 2018
McCalden RW Clout A Naudie DD MacDonald SJ Somerville LE
Full Access

Purpose

The success of total knee replacement (TKR) surgery can be attributed to improvements in TKR design, instrumentation, and surgical technique. Over a decade ago oxidized zirconium (OxZr) femoral components were introduced as an alternative bearing surface to cobalt-chromium (CoCr), based on strong in-vitro evidence, to improve the longevity of TKR implants. Early reports have demonstrated the clinical success of this material however no long-term comparative studies have demonstrated the superiority of OxZr implants compared to a more traditional CoCr implant. This study aims to compare long-term survivorship and outcomes in OxZr and CoCr femoral components in a single total knee design.

Methods

We reviewed our institutional database to identify all patients whom underwent a TKA with a posterior stabilized OxZr femoral component with a minimum of 10 years of follow-up. These were then matched to patients whom underwent a TKA with the identical design posterior stabilized CoCr femoral component during the same time period by gender, age and BMI. All patients had their patella resurfaced. All patients were prospectively evaluated preoperatively and postoperatively at 6 weeks, 3 months, 12 months, 2 years and every 1 to 2 years thereafter. Prospectively collected clinical outcome measures included, Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Short-Form 12 (SF-12) and Knee Society clinical rating scores (KSCRS). Charts and radiographs were reviewed to determine the revision rates and survivorship (both all cause and aseptic) at 10 years allowing comparison between the two cohorts. Paired analysis was performed to determine if differences existed in patient reported outcomes.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1221 - 1227
1 Sep 2012
Kim Y Park J Kim J

Despite many claims of good wear properties following total knee replacement (TKR) with an oxidised zirconium (OxZr) femoral component, there are conflicting clinical results. We hypothesised that there would be no difference in either the mid-term clinical and radiological outcomes or the characteristics of the polyethylene wear particles (weight, size and shape) in patients using an OxZr or cobalt-chrome (CoCr) femoral component. In all 331 patients underwent bilateral TKR, receiving an OxZr femoral component in one knee and a CoCr femoral component in the other. The mean follow-up was 7.5 years (6 to 8). Following aspiration, polyethylene wear particles were analysed using thermogravimetric methods and scanning electron microscopy. At the most recent follow-up, the mean Knee Society score, Western Ontario and McMaster Universities Osteoarthritis Index score, range of movement and satisfaction score were not significantly different in the two groups. The mean weight, size, aspect ratio and roundness of the aspirated wear particles were similar for each femoral component. Survivorship of the femoral, tibial and patellar components was 100% in both groups.

In the absence of evidence of an advantage in the medium term we cannot justify the additional expense of an OxZr femoral component.