Advertisement for orthosearch.org.uk
Results 1 - 20 of 29
Results per page:
The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 33 - 39
1 Jul 2019
Lachiewicz PF O’Dell JA

Aims. There is insufficient evidence to recommend the use of alternative polyethylene bearings in modular, fixed-bearing total knee arthroplasty (TKA). The purpose of this study was to compare standard polyethylene (SP) and highly crosslinked polyethylene (XLP) tibial liners in posterior-stabilized TKA, with osteolysis as the primary outcome and clinical results and the rate of re-operation as the secondary outcomes. Patients and Methods. This is a single-surgeon, prospective randomized study involving one design of modular posterior-stabilized TKA. An analysis of 122 TKAs with an SP compression moulded liner and 123 with an XLP liner was performed, with a mean follow-up of six years (2 to 11). Patients were evaluated clinically using the Knee Society score, Lower Extremity Activity Score (LEAS), and the presence of an effusion, and standard radiographs were assessed for radiolucent lines and osteolytic lesions. Results. Osteolysis was present in four TKAs (3.3%) in the SP group, and no knees in the XLP group (p = 0.06). There were no significant differences between the Knee Society total score, change in total score, knee function score, change in function score, LEAS, and change in LEAS in the two groups. There was a significant difference in the presence of an effusion (10/122 with SP liners, 1/123 with XLP liners; p = 0.02). There was no significant difference in the rate of re-operation between the two groups (p = 0.36). There were no complications related to the XLP liner. Conclusion. At this length of follow-up, there were no advantages and no complications related to the use of this XLP tibial liner. The presence of effusion and small osteolytic lesions was more frequent with SP than XLP liners, but of unknown clinical significance. Cite this article: Bone Joint J 2019;101-B(7 Supple C):33–39


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 277 - 283
1 Mar 2023
Gausden EB Puri S Chiu Y Figgie MP Sculco TP Westrich G Sculco PK Chalmers BP

Aims

The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution.

Methods

We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims

Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance.

Methods

An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1502 - 1506
1 Nov 2005
Arora J Ogden AC

We analysed at a mean follow-up of 7.25 years the clinical and radiological outcome of 117 patients (125 knees) who had undergone a primary, cemented, modular Freeman-Samuelson total knee replacement. While the tibial and femoral components were cemented, the patellar component was uncemented. A surface-cementing technique was used to secure the tibial components. A total of 82 knees was available for radiological assessment. Radiolucent lines were seen in 41 knees (50%) and osteolytic lesions were seen in 13 knees (16%). Asymptomatic, rotational loosening of the patellar implant was seen in four patients and osteolysis was more common in patients with a patellar resurfacing. Functional outcome scores were available for 41 patients (41 knees, 35%) and the mean Western Ontario McMasters Universities score was 77.5 (sd 19.5) and the cumulative survival was 93.4% at ten years with revision for aseptic loosening as an endpoint. Increased polyethylene wear from modular components, a rotationally-loose patella, and the surface-cementing technique may have contributed to the high rate of osteolysis seen in our study.


Bone & Joint Open
Vol. 4, Issue 5 | Pages 393 - 398
25 May 2023
Roof MA Lygrisse K Shichman I Marwin SE Meftah M Schwarzkopf R

Aims

Revision total knee arthroplasty (rTKA) is a technically challenging and costly procedure. It is well-documented that primary TKA (pTKA) have better survivorship than rTKA; however, we were unable to identify any studies explicitly investigating previous rTKA as a risk factor for failure following rTKA. The purpose of this study is to compare the outcomes following rTKA between patients undergoing index rTKA and those who had been previously revised.

Methods

This retrospective, observational study reviewed patients who underwent unilateral, aseptic rTKA at an academic orthopaedic speciality hospital between June 2011 and April 2020 with > one-year of follow-up. Patients were dichotomized based on whether this was their first revision procedure or not. Patient demographics, surgical factors, postoperative outcomes, and re-revision rates were compared between the groups.


Introduction. Polyethylene wear and osteolysis remain a concern with the use of modular, fixed bearing total knee arthroplasty (TKA). A variety of highly cross-linked polyethylenes (XLPs) have been introduced to address this problem, but there are few data on the results and complications of this polyethylene in posterior-stabilized knee prosthesis. We have previously reported an interim analysis of a study comparing polyethylene tibial liners. Methods. This is a prospective randomized study of one modular posterior-stabilized total knee arthroplasty by a single surgeon. 265 patients (329 knees) were randomized to receive a standard compression molded liner (SP) or a highly cross-linked (6.5 CGy electron-beam irradiated and remelted) polyethylene liner (XLP). Patients were evaluated clinically using the classic Knee Society scores, LEAS score, presence of a knee effusion, and by standard radiographs for radiolucent lines and osteolytic lesions. The analysis was performed at a mean of 6 years (range, 2–11 years). Results. There were no clinical differences between 122 knees with SP and 123 knees with XLP in Knee Society total score; change in total score; knee function score; change in function score; LEAS score; and change in LEAS score. There was a difference in the presence of effusion (one of 123 XLP, and 10 of 122 SP; p=0.02). There was no difference in the frequency of radiolucent lines (21 knees with SP and 22 with XLP). Osteolysis was present in 4 knees (3.3%) with SP, and no knees with XLP (p=0.06). There was no difference in frequency of reoperation between the two groups (3 infection in 123 knees allocated to XLP group and six (3 infection, 1 femoral loosening, 1 instability, 1 fracture plating) in 122 knees allocated to SP group. There were no complications related to the XLP liner. Conclusion. At this length of follow-up time, with the numbers available, there were no complications, but no advantages, related to the use of this XLP tibial liner. The presence of effusion and small osteolytic lesions are more frequent with SP than XLP, but of unknown clinical importance


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 904 - 910
1 Jul 2007
Kim Y Kim D Kim J

We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1). The total knee score, pain score, functional score and range of movement were not statistically different (p > 0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a mobile bearing


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 58 - 63
1 Jan 2009
Hooper G Rothwell A Frampton C

We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p < 0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p < 0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 150 - 157
1 Jun 2021
Anderson LA Christie M Blackburn BE Mahan C Earl C Pelt CE Peters CL Gililland J

Aims

Porous metaphyseal cones can be used for fixation in revision total knee arthroplasty (rTKA) and complex TKAs. This metaphyseal fixation has led to some surgeons using shorter cemented stems instead of diaphyseal engaging cementless stems with a potential benefit of ease of obtaining proper alignment without being beholden to the diaphysis. The purpose of this study was to evaluate short term clinical and radiographic outcomes of a series of TKA cases performed using 3D-printed metaphyseal cones.

Methods

A retrospective review of 86 rTKAs and nine complex primary TKAs, with an average age of 63.2 years (SD 8.2) and BMI of 34.0 kg/m2 (SD 8.7), in which metaphyseal cones were used for both femoral and tibial fixation were compared for their knee alignment based on the type of stem used. Overall, 22 knees had cementless stems on both sides, 52 had cemented stems on both sides, and 15 had mixed stems. Postoperative long-standing radiographs were evaluated for coronal and sagittal plane alignment. Adjusted logistic regression models were run to assess malalignment hip-knee-ankle (HKA) alignment beyond ± 3° and sagittal alignment of the tibial and femoral components ± 3° by stem type.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1331 - 1340
3 Oct 2020
Attard V Li CY Self A Mann DA Borthwick LA O’Connor P Deehan DJ Kalson NS

Aims

Stiffness is a common complication after total knee arthroplasty (TKA). Pathogenesis is not understood, treatment options are limited, and diagnosis is challenging. The aim of this study was to investigate if MRI can be used to visualize intra-articular scarring in patients with stiff, painful knee arthroplasties.

Methods

Well-functioning primary TKAs (n = 11), failed non-fibrotic TKAs (n = 5), and patients with a clinical diagnosis of fibrosis1 (n = 8) underwent an MRI scan with advanced metal suppression (Slice Encoding for Metal Artefact Correction, SEMAC) with gadolinium contrast. Fibrotic tissue (low intensity on T1 and T2, low-moderate post-contrast enhancement) was quantified (presence and tissue thickness) in six compartments: supra/infrapatella, medial/lateral gutters, and posterior medial/lateral.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 123 - 128
1 Jun 2020
Martin JR Geary MB Ransone M Macknet D Fehring K Fehring T

Aims

Aseptic loosening of the tibial component is a frequent cause of failure in primary total knee arthroplasty (TKA). Management options include an isolated tibial revision or full component revision. A full component revision is frequently selected by surgeons unfamiliar with the existing implant or who simply wish to “start again”. This option adds morbidity compared with an isolated tibial revision. While isolated tibial revision has a lower morbidity, it is technically more challenging due to difficulties with exposure and maintaining prosthetic stability. This study was designed to compare these two reconstructive options.

Methods

Patients undergoing revision TKA for isolated aseptic tibial loosening between 2012 and 2017 were identified. Those with revision implants or revised for infection, instability, osteolysis, or femoral component loosening were excluded. A total of 164 patients were included; 88 had an isolated tibial revision and 76 had revision of both components despite only having a loose tibial component. The demographics and clinical and radiological outcomes were recorded.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 559 - 564
1 May 2019
Takemura S Minoda Y Sugama R Ohta Y Nakamura S Ueyama H Nakamura H

Aims

The use of vitamin E-infused highly crosslinked polyethylene (HXLPE) in total knee prostheses is controversial. In this paper we have compared the clinical and radiological results between conventional polyethylene and vitamin E-infused HXLPE inserts in total knee arthroplasty (TKA).

Patients and Methods

The study included 200 knees (175 patients) that underwent TKA using the same total knee prostheses. In all, 100 knees (77 patients) had a vitamin E-infused HXLPE insert (study group) and 100 knees (98 patients) had a conventional polyethylene insert (control group). There were no significant differences in age, sex, diagnosis, preoperative knee range of movement (ROM), and preoperative Knee Society Score (KSS) between the two groups. Clinical and radiological results were evaluated at two years postoperatively.


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 996 - 1002
1 Aug 2017
Brown TS Van Citters DW Berry DJ Abdel MP

Advances in polyethylene (PE) in total hip arthroplasty have led to interest and increased use of highly crosslinked PE (HXLPE) in total knee arthroplasty (TKA). Biomechanical data suggest improved wear characteristics for HXLPE inserts over conventional PE in TKA. Short-term results from registry data and few clinical trials are promising. Our aim is to present a review of the history of HXLPEs, the use of HXLPE inserts in TKA, concerns regarding potential mechanical complications, and a thorough review of the available biomechanical and clinical data.

Cite this article: Bone Joint J 2017;99-B:996–1002.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 647 - 652
1 May 2017
Abdel MP Ledford CK Kobic A Taunton MJ Hanssen AD

Aims

The number of revision total knee arthroplasties (TKA) that are performed is expected to increase. However, previous reports of the causes of failure after TKA are limited in that they report the causes at specific institutions, which are often dependent on referral patterns. Our aim was to report the most common indications for re-operations and revisions in a large series of posterior-stabilised TKAs undertaken at a single institution, excluding referrals from elsewhere, which may bias the causes of failure.

Patients and Methods

A total of 5098 TKAs which were undertaken between 2000 and 2012 were included in the study. Re-operations, revisions with modular component exchange, and revisions with non-modular component replacement or removal were identified from the medical records. The mean follow-up was five years (two to 12).


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 105 - 112
1 Jan 2016
Khan M Osman K Green G Haddad FS

Total knee arthroplasty (TKA) is a cost effective and extremely successful operation. As longevity increases, the demand for primary TKA will continue to rise. The success and survivorship of TKAs are dependent on the demographics of the patient, surgical technique and implant-related factors.

Currently the risk of failure of a TKA requiring revision surgery ten years post-operatively is 5%.

The most common indications for revision include aseptic loosening (29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses considerable clinical burdens on patients and financial burdens on healthcare systems.

We present a current concepts review on the epidemiology of failed TKAs using data from worldwide National Joint Registries.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):105–12.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1057 - 1063
1 Aug 2013
Zeng Y Shen B Yang J Zhou ZK Kang PD Pei FX

The purpose of this study was to undertake a meta-analysis to determine whether there is lower polyethylene wear and longer survival when using mobile-bearing implants in total knee replacement when compared with fixed-bearing implants. Of 975 papers identified, 34 trials were eligible for data extraction and meta-analysis comprising 4754 patients (6861 knees). We found no statistically significant differences between the two designs in terms of the incidence of radiolucent lines, osteolysis, aseptic loosening or survival. There is thus currently no evidence to suggest that the use of mobile-bearing designs reduce polyethylene wear and prolong survival after total knee replacement.

Cite this article: Bone Joint J 2013;95-B:1057–63.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1657 - 1662
1 Dec 2014
Stambough JB Clohisy JC Barrack RL Nunley RM Keeney JA

The aims of this retrospective study were to compare the mid-term outcomes following revision total knee replacement (TKR) in 76 patients (81 knees) < 55 years of age with those of a matched group of primary TKRs based on age, BMI, gender and comorbid conditions. We report the activity levels, functional scores, rates of revision and complications. Compared with patients undergoing primary TKR, those undergoing revision TKR had less improvement in the mean Knee Society function scores (8.14 (–55 to +60) vs 23.3 points (–40 to +80), p < 0.001), a similar improvement in UCLA activity level (p = 0.52), and similar minor complication rates (16% vs 13%, p = 0.83) at a mean follow-up of 4.6 years (2 to 13.4). Further revision surgery was more common among revised TKRs (17% vs 5%, p = 0.02), with deep infection and instability being the most common reasons for failure. As many as one-third of patients aged < 55 years in the revision group had a complication or failure requiring further surgery.

Young patients undergoing revision TKR should be counselled that they can expect somewhat less improvement and a higher risk of complications than occur after primary TKR.

Cite this article: Bone Joint J 2014; 96-B:1657–62


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 917 - 922
1 Jul 2013
Sabouret P Lavoie F Cloutier J

We report on the long-term results of 163 bicruciate-retaining Hermes 2C total knee replacements in 130 patients at a mean follow-up of 22.4 years (20.3 to 23.5). Even when the anterior cruciate ligament had a partially degenerative appearance it was preserved as long as the knee had a normal anterior drawer and Lachman’s test pre-operatively. The description and surgical technique of this minimally constrained prosthesis were published in 1983 and the ten-year clinical results in 1999.

A total of 12% of the knees (20 of 163) in this study were revised because of wear of the polyethylene tibial insert. Excellent stability was achieved and the incidence of aseptic component loosening was 4.3% (seven of 163). The survival rate using revision for any reason as the endpoint was 82% (95% confidence interval 76.2 to 88.0).

Although this series included a relatively small number of replacements, it demonstrated that the anterior cruciate ligament, even when partially degenerated at the time of TKR, remained functional and provided adequate stability at a long-term follow-up.

Cite this article: Bone Joint J 2013;95-B:917–22.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 134 - 136
1 Nov 2012
Scott RD

PCL retaining fixed-bearing TKA is a highly successful operation with the need for more surgery occurring at the rate of approximately 0.4% per year over the first 27 years. The most common cause for revision surgery is related to polyethylene insert failure and accounts for approximately 50% of re-operations. Late metastatic infection is the next most frequent cause followed by patellar problems, late instability and component loosening in decreasing frequency. A myriad of rare miscellaneous problems can also occur.