Abstract. Introduction. The aims of our systematic review were to assess the Nickel sensitising potential of total knee arthroplasty (TKA), explore the relationship between Nickel hypersensitivity and clinical outcomes, and evaluate the utility of skin patch testing pre- and/or post-operatively. Methodology. A literature search was performed through EMBASE, Medline and PubMed databases. Articles were screened independently by two investigators. Levels of Evidence were assessed using Oxford CEBM criteria and quality was assessed using MINORS and Cochrane risk-of-bias tools. Results. Twenty studies met the eligibility criteria, reporting on 1354 TKAs. Prevalence of Nickel hypersensitivity ranged from 0% to 87.5%. Only one study which compared prevalence of hypersensitivity in the same patient group before and after surgery noted newly positive patch test reactions in 4.1%. Three studies reported lower prevalence of hypersensitivity in post-operative patients compared to pre-operative patients. Seven studies suggested hypersensitivity might cause complications; six studies did not support any relationship. Seven studies recommended pre-operative patch testing in patients with history of
Introduction. The utilization of lymphocyte transformation testing (LTT) has increased for diagnosing metal sensitivity associated with TKA, but its validity for the diagnosis of TKA failure due to an immune reaction has not been established. Methods. We performed a retrospective study of 27 well-fixed, aseptic primary TKAs with persistent pain and/or stiffness, revised by a single, experienced surgeon for suspected
Revision total knee arthroplasty (rTKA) is a technically challenging and costly procedure. It is well-documented that primary TKA (pTKA) have better survivorship than rTKA; however, we were unable to identify any studies explicitly investigating previous rTKA as a risk factor for failure following rTKA. The purpose of this study is to compare the outcomes following rTKA between patients undergoing index rTKA and those who had been previously revised. This retrospective, observational study reviewed patients who underwent unilateral, aseptic rTKA at an academic orthopaedic speciality hospital between June 2011 and April 2020 with > one-year of follow-up. Patients were dichotomized based on whether this was their first revision procedure or not. Patient demographics, surgical factors, postoperative outcomes, and re-revision rates were compared between the groups.Aims
Methods
A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship. A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.Aims
Methods
Aims. The treatment of patients with
We explored the literature surrounding whether
allergy and hypersensitivity has a clinical basis for implant selection
in total knee arthroplasty (TKA). In error, the terms hypersensitivity
and allergy are often used synonymously. Although a relationship
is present, we could not find any evidence of implant failure due
to allergy. There is however increasing basic science that suggests
a link between loosening and metal ion production. This is not an
allergic response but is a potential problem. With a lack of evidence
logically there can be no justification to use ‘hypoallergenic’
implants in patients who have pre-existing skin sensitivity to the
metals used in TKA. Cite this article:
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.