Polyethylene liners of modular acetabular components wear and sometimes need to be replaced, despite the metal shell being well fixed. Replacing the liner is a relatively simple procedure, but very little is known of the outcome of liner revision. We prospectively followed up 1126 Harris-Gallante 1 metal backed, uncemented cups for between 9 and 19 years. 38 (3.4%) liners out of 1126 acetabular components wore and required revision. These revisions were then followed up for a mean of 4.8 years. The rate of dislocation was 28.9%. Nine of the dislocations were single dislocations and 2 were recurrent. The overall re-revision rate was 3 out of 38 total hip replacements (7.9%) at a mean follow up of 4.8 years. This gives a 92.1% survivorship at just under 5 years. In isolated liner revision we had a complication rate of 23%. In liner revision combined with stem revision we had a complication rate of 48%. Possible reasons for high dislocation rates are discussed. Leaving the well fixed acetabular shell in-situ leads to an increased risk of instability. However, this needs to be balanced against the otherwise low complication rate for liner revision. Patients should be consented accordingly


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1103 - 1110
1 Jun 2021
Tetreault MW Hines JT Berry DJ Pagnano MW Trousdale RT Abdel MP

Aims

This study aimed to determine outcomes of isolated tibial insert exchange (ITIE) during revision total knee arthroplasty (TKA).

Methods

From 1985 to 2016, 270 ITIEs were performed at one institution for instability (55%, n = 148), polyethylene wear (39%, n = 105), insert fracture/dissociation (5%, n = 14), or stiffness (1%, n = 3). Patients with component loosening, implant malposition, infection, and extensor mechanism problems were excluded.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_III | Pages 534 - 534
1 Aug 2008
Bailie AG Howell JR Hubble MJ Timperley AJ Gie GA
Full Access

Introduction: Recurrent dislocation can be a significant problem after total hip replacement. The use of a constrained tripolar liner is an option in the surgical treatment of dislocation or instability.

Methods: A retrospective review was carried out of patients identified from a prospective database. All patients had a constrained liner cemented onto a satisfactory pre-existing cement mantle, cemented into a reconstruction ring, or cemented into a well fixed cementless shell. The Osteonics Tripolar Liner was used in all cases; the outer aspect of the liner was prepared with a burr to create grooves and thus improve cement interlock. Data collected included demographics, reason for revision, re-revision rate, outcome and survival.

Results: There were 58 cases identified where a cemented constrained liner was inserted at revision hip surgery. Average age at time of surgery was 77years (range 40–94). There were 9 patients who died with less than 2 years follow-up; they were excluded, leaving a study group of 49 cases. No cases were lost to follow-up. Average duration of follow-up was 46months (range 24–76). There have been 4 infections, one of which required removal of prostheses and a 2-stage revision. There was one case of fall post-operatively and fracture of the contra-lateral femoral neck. There have been 4 implant failures requiring re-revision. All failures were due to disarticulation of the liner, 2 of which occurred in the same patient on separate occasions. There have been no revisions for loosening, and there have been no cases of failure at the bone-cement interface or at the cement-cement interface with the cement-in-cement technique. Overall survival of the cemented constrained liner was 90% at average 3.8years.

Conclusion: This study demonstrates that cementing a constrained liner into the acetabulum is a viable option in revision hip surgery, particularly in the management of instability.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 53 - 53
1 Mar 2009
Bailie G Howell J Hubble M Timperley J Gie G
Full Access

Introduction: Recurrent dislocation is a significant problem after total hip replacement. Aetiology is multifactorial and treatment should address the reason for dislocation. The use of a constrained tripolar liner is an option in the surgical treatment of dislocation.

Methods: A retrospective review was carried out of patients who have undergone revision hip surgery and had a constrained liner cemented into the acetabulum. Patients were identified from a computer database. All patients had a constrained liner cemented onto a satisfactory pre-existing cement mantle, cemented into a reconstruction ring, or cemented into a well fixed cementless shell. The Osteonics Tripolar Liner was used in all cases and the outer aspect of the tripolar liner was prepared with a burr to create grooves and thus improve cement interlock. Data collected included demographics, reason for revision, components used, re-revision rate, outcome and survival.

Results: There were 58 cases identified where a cemented constrained liner was inserted at revision hip surgery. Average age at time of surgery was 77years (range 40–94). Reason for use of a constrained liner was recurrent dislocation in over 95% of cases. There were 9 patients who died with less than 2 years follow-up; they were excluded, leaving a study group of 49 cases. Average duration of follow-up was 46months (range 24–76).

There have been 4 infections, one of which required removal of prostheses and 2 stage revision. There was one case of fall post-operatively and fracture of the contra-lateral femoral neck. There have been 3 implant failures requiring re-revision. All failures were due to disarticulation of the liner, 2 of which occurred in the same patient on separate occasions. There have been no revisions for loosening, and there have been no cases of failure at the bone-cement interface or at the cement-cement interface with the cement-in-cement technique. Overall survival of the cemented constrained liner was 91.8% at average 3.8years.

Conclusion: This study demonstrates that a cemented constrained tripolar liner is a viable option in revision hip surgery, particularly in the treatment of recurrent dislocation.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 90 - 90
10 Feb 2023
Burn P
Full Access

Polyimide (MP-1, MMATech, Haifa, Israel), is a high performance aerospace thermoplastic used for its lubricity, stability, inertness and radiation resistance. A wear resistant thin robust bearing is needed for total hip arthroplasty (THR). After independent laboratory testing, in 2006, the author used the material as a bearing in two Reflection (Smith and Nephew, USA) hip surgeries. The first, a revision for polyethylene wear, survives with no evidence of wear, noise, new osteolysis or complications related to the MP-1 bearing after 16 yrs. The second donated his asymptomatic MP-1 hip at 6.5yrs for post-mortem examination. There were no osteoclasts, cellular reaction bland in contrast to that of polyethylene. In 2013 a clinical study with ethical committee approval was started using a Biolox Delta (Ceramtec, Germany) head against a polyimide liner in 97 patients. MMATech sold all liners, irradiated: steam 52:45. Sixteen were re-machined in New Zealand. Acetabular shells were Delta PF (LIMA, Italy). The liner locked by taper. The cohort consisted of 46:51 M:F, and ages 43 to 85, mean 65. Ten received cemented stems. For contralateral surgery, a ceramic or polyethylene liner was used. Initial patients were lower demand, later, more active patients, mountain-biking and running. All patients have on-going follow up, including MP-1 liner revision cases. There has been no measurable wear, or osteolysis around the acetabular components using weight-bearing radiographs. Squeaking within the first 6 weeks was noted in 39 number of cases and subtle increase in palpable friction, (passive rotation at 50 degrees flexion), but then disappeared. There were 6 revisions, four of which were related to cementless Stemsys implants (Evolutis, Italy) fixed distally with proximal linear lucencies in Gruen zones 1 and 7, and 2 and 6. No shells were revised and MP-1 liners were routinely changed to ceramic or polyethylene. The liners showed no head contact at the apex, with highly polished contact areas. There were no deep or superficial infections, but one traumatic anterior dislocation at 7 years associated with 5 mm subsidence of a non-collared stem. The initial squeaking and increased friction was due to the engineering of the liner / shell composite as implanted, not allowing adequate clearance for fluid film lubrication and contributed to by shell distortion during impaction. The revised bearings were “equatorial” rather than polar, and with lack of wear or creep this never fully resolved. Where the clearance was better, function was normal. The “slow” utilization was due to my ongoing concern with clearances not being correct. The revision of 4 Stemsys stems, tribology issues may have contributed, but non “MP-1” / Stemsys combinations outside this study have shown the same response, thought to be due to de-bonding of the hydroxyapatite coating. With correct engineering and clearances, a 3.6 mm thick MP-1 bearing, a surface Ra<0.5, steam sterilized, shows no appreciable wear, and with confidence, can be used as a high performance THR bearing


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 89 - 97
1 May 2024
Scholz J Perka C Hipfl C

Aims. There is little information in the literature about the use of dual-mobility (DM) bearings in preventing re-dislocation in revision total hip arthroplasty (THA). The aim of this study was to compare the use of DM bearings, standard bearings, and constrained liners in revision THA for recurrent dislocation, and to identify risk factors for re-dislocation. Methods. We reviewed 86 consecutive revision THAs performed for dislocation between August 2012 and July 2019. A total of 38 revisions (44.2%) involved a DM bearing, while 39 (45.3%) and nine (10.5%) involved a standard bearing and a constrained liner, respectively. Rates of re-dislocation, re-revision for dislocation, and overall re-revision were compared. Radiographs were assessed for the positioning of the acetabular component, the restoration of the centre of rotation, leg length, and offset. Risk factors for re-dislocation were determined by Cox regression analysis. The modified Harris Hip Scores (mHHSs) were recorded. The mean age of the patients at the time of revision was 70 years (43 to 88); 54 were female (62.8%). The mean follow-up was 5.0 years (2.0 to 8.75). Results. DM bearings were used significantly more frequently in elderly patients (p = 0.003) and in hips with abductor deficiency (p < 0.001). The re-dislocation rate was 13.2% for DM bearings compared with 17.9% for standard bearings, and 22.2% for constrained liners (p = 0.432). Re-revision-free survival for DM bearings was 84% (95% confidence interval (CI) 0.77 to 0.91) compared with 74% (95% CI 0.67 to 0.81) for standard articulations, and 67% (95% CI 0.51 to 0.82) for constrained liners (p = 0.361). Younger age (hazard ratio (HR) 0.92 (95% CI 0.85 to 0.99); p = 0.031), lower comorbidity (HR 0.44 (95% CI 0.20 to 0.95); p = 0.037), smaller heads (HR 0.80 (95% CI 0.64 to 0.99); p = 0.046), and retention of the acetabular component (HR 8.26 (95% CI 1.37 to 49.96); p = 0.022) were significantly associated with re-dislocation. All DM bearings which re-dislocated were in patients with abductor muscle deficiency (HR 48.34 (95% CI 0.03 to 7,737.98); p = 0.303). The radiological analysis did not reveal a significant relationship between restoration of the geometry of the hip and re-dislocation. The mean mHHSs significantly improved from 43 points (0 to 88) to 67 points (20 to 91; p < 0.001) at the final follow-up, with no differences between the types of bearing. Conclusion. We found that the use of DM bearings reduced the rates of re-dislocation and re-revision in revision THA for recurrent dislocation, but did not guarantee stability. Abductor deficiency is an important predictor of persistent instability. Cite this article: Bone Joint J 2024;106-B(5 Supple B):89–97


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 34 - 34
1 Oct 2020
Lombardi AV
Full Access

Background. Ultraporous metals have now been used in acetabular reconstruction for two decades with excellent survival. The purpose of this study is to evaluate a newer porous metal made from Ti6Al4V titanium alloy in complex primary and revision hip arthroplasty. Methods. A retrospective review as performed on all total hip arthroplasty (THA) procedures in which a G7 Osseo-Ti (Zimmer Biomet, Warsaw, IN) acetabular component was used between 2015 and 2017. Patients with 2-year minimum follow-up or failure were included, yielding a cohort of 123 patients (126 hips). There were 50 male patients (41%; 51 hips) and 73 females (59%; 75 hips). Mean age was 65 years (range, 43–88) and mean BMI was 30.7 kg/m. 2. (range, 18–56). Indications for ultraporous metal components were in hips with compromised bone stock or severe acetabular deformity. Procedures were 35 complex primary THA and 91 revision THA that included 12 conversions and 24 reimplantations as part of 2-staged exchange for treatment of infection. Results. With an average 3.3-year follow-up (range, 2–5 years), 1 hip in the primary series (2.9%) and 4 hips in the revision series (4.4%) were revised for aseptic loosening of the acetabular component. Three of these re-revisions required custom triflange devices. Five patients (4%) failed for periprosthetic infection, which included 1 primary THA done for rheumatoid arthritis and post-radiation necrosis, and 4 second-stage reimplantation revision THAs for prior infection. Two revision patients, one done for active instability and one multiply revised, subsequently dislocated and required liner revision to constrained constructs. Kaplan-Meier analysis to endpoint of acetabular revision for aseptic loosening was 96.6% (±3.4%) in the primary series and 95.3% (±2.3%) in the revision series. Conclusion. This three-dimensionally printed ultraporous titanium acetabular component demonstrated promising early results in complex primary and revision total hip arthroplasty


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 59 - 65
1 May 2024
Liu WKT Cheung A Fu H Chan PK Chiu KY

Aims

Isolated acetabular liner exchange with a highly crosslinked polyethylene (HXLPE) component is an option to address polyethylene wear and osteolysis following total hip arthroplasty (THA) in the presence of a well-fixed acetabular shell. The liner can be fixed either with the original locking mechanism or by being cemented within the acetabular component. Whether the method used for fixation of the HXLPE liner has any bearing on the long-term outcomes is still unclear.

Methods

Data were retrieved for all patients who underwent isolated acetabular component liner exchange surgery with a HXLPE component in our institute between August 2000 and January 2015. Patients were classified according to the fixation method used (original locking mechanism (n = 36) or cemented (n = 50)). Survival and revision rates were compared. A total of 86 revisions were performed and the mean duration of follow-up was 13 years.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_10 | Pages 69 - 69
1 Jun 2018
Rosenberg A
Full Access

Nutritional Status and Short-Term Outcomes Following THA; Initial Metal Ion Levels Predict Risk in MoM THA; THA Bearing Surface Trends in the US ‘07- ’14; Dislocation Following Two-Stage Revision THA; Timing of Primary THA Prior to or After Lumbar Spine Fusion; Failure Rate of Failed Constrained Liner Revision; ESR and CRP vs. Reinfection Risk in Two-Stage Revision?; Mechanical Complications of THA Based on Approach; Impaction Force and Taper-Trunnion Stability in THA; TKA in Patients Less Than 50 Years of Age; Post-operative Mechanical Axis and 20-year TKA Survival and Function; Return to Moderate to High-intensity Sports after UKA; “Running Two Rooms” and Patient Safety in TJA; Varus and Implant Migration and Contact Kinematics after TKA; Quadriceps Snips in 321 Revision TKAs; Tubercle Proximalization for Patella Infera in Revision TKA; Anterior Condylar Height and Flexion in TKA; Compression Bandage Following Primary TKA; Unsupervised Exercise vs. Traditional PT After Primary TKA and UKA


The protective effect of lipped polyethylene uncemented acetabular liners against revision THA for instability has been reported. However, the effect of lip size has not been explored, nor has the effect on revision THA for loosening. We aimed to determine if uncemented acetabular liner geometry, and lip size, influences the risk of revision THA for instability or loosening. 202511 primary THAs with uncemented polyethylene acetabular components were identified from the NJR dataset (2003 – 2017). The effect of acetabular liner geometry and lip size on the risk of revision THA for instability or loosening was investigated using binomial regression and competing risks survival analyses (competing risks were revision for other causes or death) adjusting for age, gender, ASA grade, diagnosis, side, institution type, surgeon grade, surgical approach, head size and polyethylene crosslinking. The distribution of acetabular liners was: neutral – 39.4%, offset neutral – 0.9%, 10-degree – 34.5%, 15-degree – 21.6%, 20-degree – 0.8%, offset reorientating – 2.82%. There were 690 (0.34%) revision THAs for instability and 604 (0.3%) for loosening. Significant subhazard risk ratios were found in revision THA for instability with 10-degree (0.63), 15-degree (0.48) and offset reorientating (1.6) liners, compared to neutral liners. There was no association found between liner geometry and risk of revision THA for loosening. This Registry based study confirms a significantly lower risk of revision THA for instability when a lipped liner is used, compared to neutral liners, and a higher risk with the use of offset reorientating liners. Furthermore, 15degree liners seem to have a lower risk than 10degree liners. We did not find an association between acetabular liner geometry and revision THA for loosening. 10- and 15-degree lipped polyethylene liners seem to offer a lower revision risk over neutral liners, at least at medium term followup. Further studies are required to confirm if this benefit continues into the long-term


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 6 - 6
1 Oct 2019
Masri BA Zamora T Garbuz DS Greidanus NV
Full Access

Introduction. The number of medial unicompartmental knee replacements (UKR) performed for arthritis has increased and as such, revisions to total knee replacement (TKR) is increasing. Previous studies have investigated survivorship of UKR to TKR revision and functional outcomes compared to TKR to TKR revision, but have failed to detail the surgical considerations involved in these revisions. Our objectives are to investigate the detailed surgical considerations involved in UKR to TKR revisions. Methods. This study is a retrospective comparative analysis of a prospectively collected database. From 2005 to 2017, 61 revisions of UKR to TKR were completed at a single center. Our inclusion criteria included: revision of UKR to TKR or TKR to TKR with minimum 1 year follow-up. Our exclusion criteria include: single component and liner revisions and revision for infection. The 61 UKR to TKR revisions were matched 2:1 with respect to age, ASA and BMI to a group of 122 TKR to TKR revisions. The following data was collected: indication for and time to revision, operative skin to skin surgical time, the use of specialized equipment (augment size/location, stem use), intraoperative and postoperative complications, re-operations and outcome scores (WOMAC, Oxford 12, SF 12, satisfaction score). Results. There were no statistical differences between the demographic data from either group (age, BMI, ASA, sex and follow-up range). Progression of arthritis was the most common reason for revision in the UKR to TKR group (30/61, 49%, p < 0.001). Aseptic loosening was the most common reason for revision in the TKR to TKR group (73/122, 60%,) and was encountered more often than aseptic loosening in the UKR to TKR group (21/61, 35%, p=0.002). The operative time was longer in the TKR to TKR group (77 vs 112 min, p< 0.001). Femoral augmentation was required for one 1/61 (1.64%) UNI and 92/122 (75%) TKR revisions, respectively (p <0.001). Medial tibial augments were required in 9/61 (14.8%) of the UKR to TKR group while 12/122 (10%) and 10/122 (8%) of the TKR to TKR group required medial and full tibial augments, respectively (p=0.7). UKR to TKR revisions never required femoral stems while 120/122 (98%) of the TKR to TKR group did (p<0.001). Tibial stems were required in 19/61 (31%) and 122/122 (100%) of UKR to TKR and TKR to TKR groups, respectively (p<0.001). There was no statistical difference in the overall complication rate of either group (15% in the UKR to TKR group and 13% in the TKR to TKR group, p = 0.9). Stiffness was a common complication of UKR to TKR and TKR to TKR re-revisions at 2/61 (3%), and 6/122 (5%), respectively (P = 0.6). Aseptic loosening was also a common complication of in both groups at 2/61 (3%) and 4/122 (3%) in the UKR to TKR and TKR to TKR groups, respectively (p = 0.7). There was no statistical difference in the re-operation rate of either group (10% in the UKR to TKR group and 7% in the TKR to TKR group, P = 1). Stiffness was the most common indication for re-operation in the UKR to TKR group (2/61, 3%, p = 0.11) while aseptic loosening was the most common in the TKR to TKR group (4/122, 3.2%, p = 0.7). The survivorship in the UKR to TKR was 93% and 90% at 5 and 9 years, respectively. The survivorship in the TKR to TKR group was 95% and 94% at 5 and 9 years, respectively, which was not statistically different from the UKR group. Discussion. The most common reason for revision was different between the two groups (p < 0.001) while the skin to skin time was longer in the TKR to TKR group. In terms of revision components, femoral stems were never required in the UKR to TKR group while tibial stems were only required in 31%. Similarly, medial tibial augments were only required in 15% of the UKR to TKR group. While the surgeon must be prepared to use augmentation and stems in UKR to TKR revisions, they can often be completed with primary components and therefor will have an overall lower cost to the health care system. Furthermore, the survivorship and re-operation between the two groups was similar which supports previous literature. The results of this study will allow for a more in-depth cost-effectiveness analysis of UKR to TKR vs TKR to TKR in arthroplasty decision making. Unicompartmental knee replacements should be considered in appropriate patients to decrease the lifetime cost of arthroplasty intervention and potentially decrease the burden on the health care system. For figures, tables, or references, please contact authors directly


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1045 - 1051
1 Oct 2023
Turgeon TR Righolt CH Burnell CD Gascoyne TC Hedden DR Bohm ER

Aims

The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components.

Methods

Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).


Bone & Joint Open
Vol. 4, Issue 10 | Pages 782 - 790
18 Oct 2023
Hamilton DF Gaston P Macpherson GJ Simpson P Clement ND

Aims

The primary aim of this study is to assess the survival of the uncemented hydroxyapatite (HA) coated Trident II acetabular component as part of a hybrid total hip arthroplasty (THA) using a cemented Exeter stem. The secondary aims are to assess the complications, joint-specific function, health-related quality of life, and radiological signs of loosening of the acetabular component.

Methods

A single-centre, prospective cohort study of 125 implants will be undertaken. Patients undergoing hybrid THA at the study centre will be recruited. Inclusion criteria are patients suitable for the use of the uncemented acetabular component, aged 18 to 75 years, willing and able to comply with the study protocol, and provide informed consent. Exclusion criteria includes patients not meeting study inclusion criteria, inadequate bone stock to support fixation of the prosthesis, a BMI > 40 kg/m2, or THA performed for pain relief in those with severely restricted mobility.


Bone & Joint Open
Vol. 4, Issue 5 | Pages 393 - 398
25 May 2023
Roof MA Lygrisse K Shichman I Marwin SE Meftah M Schwarzkopf R

Aims

Revision total knee arthroplasty (rTKA) is a technically challenging and costly procedure. It is well-documented that primary TKA (pTKA) have better survivorship than rTKA; however, we were unable to identify any studies explicitly investigating previous rTKA as a risk factor for failure following rTKA. The purpose of this study is to compare the outcomes following rTKA between patients undergoing index rTKA and those who had been previously revised.

Methods

This retrospective, observational study reviewed patients who underwent unilateral, aseptic rTKA at an academic orthopaedic speciality hospital between June 2011 and April 2020 with > one-year of follow-up. Patients were dichotomized based on whether this was their first revision procedure or not. Patient demographics, surgical factors, postoperative outcomes, and re-revision rates were compared between the groups.


Bone & Joint Open
Vol. 5, Issue 6 | Pages 514 - 523
24 Jun 2024
Fishley W Nandra R Carluke I Partington PF Reed MR Kramer DJ Wilson MJ Hubble MJW Howell JR Whitehouse SL Petheram TG Kassam AM

Aims

In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component.

Methods

Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims

Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up.

Methods

We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_14 | Pages 66 - 66
1 Mar 2013
Grobler G Dower B Nortje M Reid C
Full Access

Purpose of Study. To assess the results of Revision Hip Surgery in which a less invasive technique was utilized in situations where a number of different options was available. Method. The authors rely on an experience of 3,445 hip arthroplasties by a single surgeon over a period of 20 years, of which approximately 20% were revision cases. Of these 617 cases, we report on 175 in which a minimally invasive option was taken. This does not apply to the skin incision, as all cases were adequately exposed. We have adopted this term to describe cases in which a surgical options was taken that resulted in the least morbidity and the shortest surgical time. We postulated that would lead to the best outcomes with the least complications. Acetabular revisions: 1) Isolated polyethylene exchange. 2) Liner revision with cement technique in cases of cup malposition or poor locking mechanism. 3) Revision of cup with a primary prosthesis with significant medial bone loss. Stem revisions: 1) Cement on cement technique. 2) Strut graft and primary stem. Results. We found a very low complication rate utilizing these methods: Fatal pulmonary emboli: 0 Sepsis: 2 Dislocations 3 Repeat revisions 3. Conclusion. Revision surgery offers many challenges that tend to be compounded with successive operations. We believe that good results can be achieved when a philosophy of minimally invasive surgery is adopted. NO DISCLOSURES


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1774 - 1782
1 Dec 2021
Divecha HM O'Neill TW Lunt M Board TN

Aims

The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening.

Methods

A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk regression analyses adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, surgeon grade, surgical approach, head size, and polyethylene crosslinking. Stratified analyses by surgical approach were performed, including pairwise comparisons of liner geometries.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 833 - 843
1 Jul 2022
Kayani B Baawa-Ameyaw J Fontalis A Tahmassebi J Wardle N Middleton R Stephen A Hutchinson J Haddad FS

Aims

This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA).

Methods

Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 37 - 37
1 Aug 2012
Al-Najjim M Chambers I
Full Access

Aims. To assess the early subsidence rate of the femoral stem for patients who had collarless Corail total hip replacement. Methods. Consecutive data was collected retrospectively between August 2007 and December 2009 for patients who had collarless Corail total hip replacement. Radiographic assessment of the degree of subsidence, calcar resorption, stem angulation, canal fill ratio and loosening of the stem were measured. Post operative pain, dislocation and stem revision surgery were also evaluated. Results. 48 patients were identified, providing 51 hips for the study. There were 22 male and 26 female. The mean age 64.2 years (range 38-77). Post-operative radiographs were taken at day 1, 6 weeks and one year post-operatively (range 10-18 months, mean 12.7). Significant subsidence was defined as 3 or more millimetres, we identified two patients with subsidence between 3-5mm, one patient with 6mm and two patients with 10mm subsidence at one year post-operatively. In the 5 patients with subsidence post-operatively, all significant subsidence occurred within the first 6 weeks. Canal fill ratio was measured in all patients; in the non-subsided group the ratio was an average of 72% in the lower third of the stem and 84% in the middle third. In the subsided group 75% in the lower third of the stem and 81% in the middle third, which we felt was clinically insignificant. There was no dislocation or revision for septic loosening. One patient with 10mm subsidence had liner revision for dissociation but the stem was well fixed and not revised. All patients who had significant subsidence still had functioning implants with no pain, revisions for subsidence or features of loosening. Discussion. There were no radiological features predictive of subsidence in our patients. Subsidence may be due to lack of compliance in some patients with partial weight bearing and increased Body Mass Index