Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 130 - 130
11 Apr 2023
Biddle M Wilson V Miller N Phillips S
Full Access

Our aim was to ascertain if K-wire configuration had any influence on the infection and complication rate for base of 4th and 5th metacarpal fractures. We hypothesised that in individuals whose wires crossed the 4th and 5th carpometacarpal joint (CMCJ), the rate of complications and infection would be higher. Data was retrospectively analysed from a single centre. 106 consecutive patients with a base of 5th (with or without an associated 4th metacarpal fracture) were analysed between October 2016 and May 2021. Patients were split into two groups for comparison; those who did not have K-wires crossing the CMCJ's and those in whose fixation had wires crossing the joints. Confounding factors were accounted for and Statistical analysis was performed using SPSS version 20 software. Of 106 patients, 60 (56.6%) patients did have K-wires crossing the CMCJ. Wire size ranged from 1.2-2.0 with 65 individuals (65.7%) having size 1.6 wires inserted. The majority of patients, 66 (62.9%) underwent fixation with two wires (range 1-4). The majority of infected cases (88.9%) were in patients who had k-wires crossing the CMCJ, this trended towards clinical significance (p=0.09). Infection was associated with delay to theatre (p=0.002) and longer operative time (p=0.002). In patients with a base of 4th and 5th metacarpal fractures, we have demonstrated an increased risk of post-operative infection with a K-wire configuration that crosses the CMCJ. Biomechanical studies would be of use in determining the exact amount of movement across the CMCJ, with the different K-wire configuration in common use, and this will be part of a follow-up study


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 114 - 114
1 Dec 2020
Cullu E Olgun H Tataroğlu C Ozgezmez FT Sarıerler M
Full Access

Thermal osteonecrosis is a side effect when used Kirschner (K) wires and drills in orthopaedic surgeries. This osteonecrosis may endanger the fixation. Orthopaedic surgeons sometimes have to use unsharpened K-wires in emergent surgery. The thermal effect of used and unsharpened K wire is ambiguous to the bone. This experimental study aims to assess the thermal osteonecrosis while drilling bone with three different types of K-wires especially a previously used unsharpened wire and its thermographic measurements correlation. Two different speeds of rotation were chosen to investigate the effect of speed on thermal necrosis to the bone. A total of 16 New Zealand white rabbits weighing a mean of 2.90 kg (2.70 – 3.30 kg) were used. All rabbits were operated under general anaesthesia in a sterile operating room. Firstly, 4 cm longitudinal lateral approach was used to the right femur and then the femur was drilled with 1.0 mm trochar tip, spade tip and previously used unsharpened K-wires and 1.0 mm drill bit at 1450 rpm speed. Left femur was drilled with same three type K-wires and drill bit at 330 rpm speed. One cm distance was left among four penetrations on the femur. The thermal changes were recorded by Flir® E6 Thermal Camera from 50 cm distance and 30-degree angle. Thermographic measurements saved for every drilling process and recorded for the highest temperature (°C) during the drilling. All subjects were sacrificed post-operatively on the eighth day and specimens were prepared for the histological examination. The results of osteonecrosis assessment score and thermographic correlation were evaluated statistically. Histological specimens were evaluated by the scoring of osteonecrosis, osteoblastic activity, haemorrhage, microfracture and inflammation. Results were graded semi-quantitatively as none, moderate or severe for osteonecrosis, haemorrhage and inflammation. The microfracture and osteoblastic activity were evaluated as present or absent. There was no meaningful correlation between osteonecrosis and the drilling speed (p=0.108). There was less microfracture zone which was drilled with trochar tip K-wires at 1450 rpm speed (p=0.017). And the drilling temperature of trochar tip K-wires was higher than the others(p=0.001). Despite this evaluation, osteonecrosis zone of spade and unsharpened tip K-wires were more than trochar tip K-wires (p=0.039). The drill bit at 330 rpm caused the least osteonecrosis and haemorrhage and respectfully the lowest drilling temperature (p=0,001). The osteoblastic activity shows no difference between the groups. (p=0,122; 0,636;0.289). On the contrary to the literature, our experiment showed that there is no meaningful correlation between osteonecrosis score and temperature produced by drilling. The histological assessment showed the osteonecrosis during short drilling time but, not clarify the relation with drilling temperature. Eventually, the osteonecrosis showed a positive correlation with drilling time independently of drilling temperature at 330 rpm. (p=0,042) These results show that we need more studies to understand about osteonecrosis and its relationship with drilling heat temperature. Trochar tip K-wires creates higher drilling temperature but less osteonecrosis than a spade and unsharpened cut tip K-wires. Using unsharpened tip K-wire causes more osteonecrosis. Previously used and, unsharpened K-wires should be discarded


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1545 - 1550
1 Nov 2007
Koslowsky TC Mader K Dargel J Koebke J Hellmich M Pennig D

We have evaluated four different fixation techniques for the reconstruction of a standard Mason type-III fracture of the radial head in a sawbone model. The outcome measurements were the quality of the reduction, and stability. A total of 96 fractures was created. Six surgeons were involved in the study and each reconstructed 16 fractures with 1.6 mm fine-threaded wires (Fragment Fixation System (FFS)), T-miniplates, 2 mm miniscrews and 2 mm Kirschner (K-) wires; four fractures being allocated to each method using a standard reconstruction procedure. The quality of the reduction was measured after definitive fixation. Biomechanical testing was performed using a transverse plane shear load in two directions to the implants (parallel and perpendicular) with respect to ultimate failure load and displacement at 50 N. A significantly better quality of reduction was achieved using the FFS wires (Tukey’s post hoc tests, p < 0.001) than with the other devices with a mean step in the articular surface and the radial neck of 1.04 mm (. sd. 0.96) for the FFS, 4.25 mm (. sd. 1.29) for the miniplates, 2.21 mm (. sd. 1.06) for the miniscrews and 2.54 mm (. sd. 0.98) for the K-wires. The quality of reduction was similar for K-wires and miniscrews, but poor for miniplates. The ultimate failure load was similar for the FFS wires (parallel, 196.8 N (. sd. 46.8), perpendicular, 212.5 N (. sd. 25.6)), miniscrews (parallel, 211.8 N (. sd. 47.9), perpendicular, 208.0 N (. sd. 65.9)) and K-wires (parallel, 200.4 N (. sd. 54.5), perpendicular, 165.2 N (. sd. 37.9)), but significantly worse (Tukey’s post hoc tests, p < 0.001) for the miniplates (parallel, 101.6 N (. sd. 43.1), perpendicular, 122.7 N (. sd. 40.7)). There was a significant difference in the displacement at 50 N for the miniplate (parallel, 4.8 mm (. sd. 2.8), perpendicular, 4.8 mm (. sd. 1.7)) vs FFS (parallel, 2.1 mm (. sd. 0.8), perpendicular, 1.9 mm (. sd. 0.7)), miniscrews (parallel, 1.8 mm (. sd. 0.5), perpendicular, 2.3 mm (. sd. 0.8)) and K-wires (parallel, 2.2 mm (. sd. 1.8), perpendicular, 2.4 mm (. sd. 0.7; Tukey’s post hoc tests, p < 0.001)). The fixation of a standard Mason type-III fracture in a sawbone model using the FFS system provides a better quality of reduction than that when using conventional techniques. There was a significantly better stability using FFS implants, miniscrews and K-wires than when using miniplates


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 137 - 137
11 Apr 2023
Quinn A Pizzolato C Bindra R Lloyd D Saxby D
Full Access

There is currently no commercially available and clinically successful treatment for scapholunate interosseous ligament rupture, the latter leading to the development of hand-wrist osteoarthritis. We have created a novel biodegradable implant which fixed the dissociated scaphoid and lunate bones and encourages regeneration of the ruptured native ligament. To determine if scaphoid and lunate kinematics in cadaveric specimens were maintained during robotic manipulation, when comparing the native wrist with intact ligament and when the implant was installed. Ten cadaveric experiments were performed with identical conditions, except for implant geometry that was personalised to the anatomy of each cadaveric specimen. Each cadaveric arm was mounted upright in a six degrees of freedom robot using k-wires drilled through the radius, ulna, and metacarpals. Infrared markers were attached to scaphoid, lunate, radius, and 3rd metacarpal. Cadaveric specimens were robotically manipulated through flexion-extension and ulnar-radial deviation by ±40° and ±30°, respectively. The cadaveric scaphoid and lunate kinematics were examined with 1) intact native ligament, 2) severed ligament, 3) and installed implant. Digital wrist models were generated from computed tomography scans and included implant geometry, orientation, and location. Motion data were filtered and aligned relative to neutral wrist in the digital models of each specimen using anatomical landmarks. Implant insertion points in the scaphoid and lunate over time were then calculated using digital models, marker data, and inverse kinematics. Root mean squared distance was compared between severed and implant configurations, relative to intact. Preliminary data from five cadaveric specimens indicate that the implant reduced distance between scaphoid and lunate compared to severed configuration for all but three trials. Preliminary results indicate our novel implant reduced scapho-lunate gap caused by ligament transection. Future analysis will reveal if the implant can achieve wrist kinematics similar to the native intact wrist


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 184 - 184
1 Jul 2014
Hydorn C Nathe K Kanwisher M DesJardins J Rogers M Bertram A
Full Access

Summary Statement. This study examined the fixation stiffness of 13 tibial and 12 femoral Salter-Harris fracture fixation methods, and determined that screws and screws+ k-wires methods provided the highest stability. In situations where k-wire use is unavoidable, threaded k-wires are preferable. Introduction. Salter-Harris fractures of the proximal tibia and distal femur are common in pediatric patients that present to orthopedic surgeons. Salter-Harris type I fractures are characterised by breaks that extend only through the physis while Salter-Harris II fractures are the most common, accounting for 85% of Salter-Harris fractures, and extend past the growth plate, exiting through the metaphyseal bone. Fixation of these fracture types can be accomplished using a variety of methods including the use of Kirschner wires, cannulated screws, and a combination of both materials. Stability of fracture fixation is of utmost importance as persistent motion at the fracture margin leads to deformity. The purpose of this study is to analyze the biomechanical efficacy of various fixation methods used to stabilise Salter-Harris I and II fracture patterns in both the proximal tibia and distal femur. Stiffness, the primary gauge of efficacy, will be tested in flexion and extension, varus and valgus movement, and internal and external rotation and will be compared to determine the optimal fixation method. Materials and Methods. This study utilised 39 tibia and 36 femur 4. th. generation synthetic bones (Model 3401 and 3403, Pacific Research Laboratories Inc.) The synthetic bones were fractured and fixated to model Salter-Harris fractures and common fixation methods. Fixation methods used employed 6.5mm cannulated screws, 4.5mm cannulated screws, 2mm smooth K-wires, and 2mm threaded K-wires. Tibias were fractured according to Salter-Harris I, valgus Salter-Harris II, and flexion Salter-Harris II patterns with 13 different fixation methods. Femurs were fractured according to Salter-Harris I and Salter-Harris II patterns with 12 different fixation methods. Testing was performed in three orientations, flexion/extension, varus/valgus, and internal/external rotation, on a materials testing machine (Model 8874, Instron, Norwood, MA) and cyclic displacement tests were performed using Wavematrix software. These displacement tests recorded the torque required to reach an angulation of ±5° for 10 cycles. From this data, the rotational stiffness of the loading phases for each cycle was determined. Statistical analysis was performed to compare construct stiffness and differences between groups using analysis of variance. Results. Results show superior fixation for threaded k-wires in both femoral and tibial Salter-Harris I fractures. Methods utilizing transverse screws were least optimal for the fixation of femoral Salter-Harris II fractures, while a combination of k-wires and screws or the use of oblique screws was more effective. Fixation utilizing a combination of k-wires and screws yielded greater stiffness in valgus and flexion tibial Salter-Harris II fractures. Internal and external rotational stiffness values were low for all fixation methods and no significant variance existed for internal and external rotational stiffnesses in most fracture patterns. Discussion/Conclusion. Based on the results and statistical analysis, we believe that significant variance exists between most of the studied fixation methods for each fracture type. Fixation methods utilizing screws and a combination of screws and k-wires would provide optimal stability. In situations where the use of k-wires is unavoidable, threaded k-wires are preferable


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 86 - 86
1 Mar 2021
Hope N Arif T Stagl A Fawzy E
Full Access

Distal radius fractures (DRF) are very common injuries. National recommendations (British Orthopaedic Association, National Institute for Health and Care Excellence (NICE)) exist in the UK to guide the management of these injuries. These guidelines provide recommendations about several aspects of care including which type of injuries to treat non-operatively and surgically, timing of surgery and routine follow-up. In particular, current recommendations include considering immobilizing patients for 4 weeks in plaster for those managed conservatively, and operating on fractures within 72 hours for intra-articular injuries and 7 days for extra-articular fractures. With increased demands for services and an ageing population, prompt surgery for those presenting with distal radius fractures is not always possible. A key factor is the need for prompt surgery for hip fracture patients. This study is an audit of the current standard of care at a busy level 2 trauma unit against national guidelines for the management of DRFs. This retrospective audit includes all patients presenting to our emergency department from June to September 2018. Patients over 18 years of age with a diagnosis of a closed distal radius fracture and follow-up in our department were included in the study. Those with open fractures were excluded. Data was retrieved from clinical coding, electronic patient records, and IMPAX Client (Picture archiving and communication system). The following data was collected on patients treated conservatively and those managed surgically:- (1)Time to surgery for surgical management; (2)Period of immobilization for both conservative and operative groups. 45 patients (13 male, 32 female) with 49 distal radius fractures (2 patients had bilateral injuries) were included. Patients had mean age 63 years (range 19 to 92 years) 30 wrists were treated non-operatively and 19 wrists treated surgically (8 K-wires, 10 ORIF, 1 MUA). Mean time to surgery in the operative group was 8 days (range 1 – 21 days, median 7 days). Mean time to surgery for intra-articular fractures was 7 days (range 1 – 21) and 12 days for extra-articular fractures (range 4 – 20). Mean immobilization period in those treated in plaster is 6 weeks (range 4 – 13 weeks, median 5.6 weeks). At busy level 2 trauma units with limited theatre capacity and a high volume of hip fracture admissions, time to surgery for less urgent injuries such as wrist fractures is often delayed. National guidelines are useful in helping to guide management however their standards are often difficult to achieve in the context of increasing populations in urban areas and an ageing population


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 95 - 95
1 May 2012
Molony DC Kennedy J Gheiti AC Mullett JH
Full Access

Background. The treatment of olecranon fractures frequently involves the use of tension band fixation. Although associated with high union rates, this method has a high incidence of morbidity associated with soft tissue compromise and limitation of range of movement requiring frequent re-operation for removal of metal. Objectives. We describe the use of a simple jig to ensure intramedullary placement of longitudinal K-wires and compare the accuracy of placement of wires using this device with the traditional free hand method. Methods & Materials. 10 orthopaedic surgical trainees passed 2 longitudinal k wires into synthetic ulnae. This was done using the free hand method and then plastic jig. The ulnae were then sectioned to identify the position of the wires relative to the intramedullary canal. Results. Of the free hand wires, only 11 of 20 were found to be intramedullary versus 20 of 20 in the group passed using the jig. (p<0.001). The mean distance from the center of the ulnar canal was 6.5 mm in the freehand group (range 1-18mm, SD 6.1mm) and 1.6 mm in the O-Jig group (range 1-2mm, SD 0.5mm). The difference in the mean distance from the centre was 4.9mm which was significant (p< 0.001). Conclusions. We found the distance from the centre of the medullary canal, the range and standard deviations of wire positions to be significantly more precise when the jig was used


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XI | Pages 7 - 7
1 Apr 2012
Mullen M Pillai A Fogg Q Kumar CS
Full Access

The extended lateral approach offers a safe surgical approach in the fixation of calcaneal fractures. Lateral plating of the calcaneum could put structures on the medial side at risk. The aim was to identify structures at risk on the medial side of the calcaneum from wires, drills or screws passed from lateral to medial. Ten embalmed cadaveric feet were dissected. A standard extended lateral approach was performed. The DePuy perimeter plate was first applied and 2mm K-wires were drilled through each of the holes. The medial side was now examined to determine the structures at risk through each hole. The process was repeated with the Stryker plate. The calcaneum was divided into 6 zones, by two vertical lines, from the margins of the posterior facet and a transverse line along the axis of the bone through the highest point of the peroneal tubercle. The DePuy and the Stryker plates have 12 screw positions, 5 of which are common. With both systems, screw positions in zone 1 risk injury to the medial plantar nerve and zone 3 the lateral plantar nerve. A screw through zone 2 compromises the medial plantar in both. Screws through zone 4 risk the lateral plantar nerve with the DePuy plate. Screws through zone 5 of the DePuy plate risk the medial calcaneal nerve. Zone 5 of the Stryker plate and Zone 6 of both are safe. There is significant risk to medial structures from laterally placed wires, drills or screws. Subtalar screws have the highest risk and have to be carefully measured and placed. The Stryker plating system is relatively safer than the DePuy perimeter plate with three safe zones out of six


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 315 - 319
1 Feb 2010
Lalliss SJ Branstetter JG

Using an osteotomy of the olecranon as a model of a transverse fracture in 22 cadaver elbows we determined the ability of three different types of suture and stainless steel wire to maintain reduction when using a tension-band technique to stabilise the bone. Physiological cyclical loading simulating passive elbow movement (15 N) and using the arms to push up from a chair (450 N) were applied using an Instron materials testing machine whilst monitoring the osteotomy site with a video extensometer. Each osteotomy was repaired by one of four materials, namely, Stainless Steel Wire (7), No 2 Ethibond (3), No 5 Ethibond (5), or No 2 FiberWire (7).

There were no failures (movement of > 2 mm) with stainless steel wire or FiberWire and no significant difference in the movements measured across the site of the osteotomy (p = 0.99). The No. 2 Ethibond failed at 450 N and two of the five of No. 5 Ethibond sutures had a separation of > 2 mm at 450 N.

FiberWire as the tension band in this model held the reduction as effectively as stainless steel wire and may reduce the incidence of discomfort from the hardware. On the basis of our findings we suggest that a clinical trial should be undertaken


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 823 - 829
1 Jun 2009
Adachi N Motoyama M Deie M Ishikawa M Arihiro K Ochi M

We evaluated the histological changes before and after fixation in ten knees of ten patients with osteochondritis dissecans who had undergone fixation of the unstable lesions. There were seven males and three females with a mean age of 15 years (11 to 22). The procedure was performed either using bio-absorbable pins only or in combination with an autologous osteochondral plug. A needle biopsy was done at the time of fixation and at the time of a second-look arthroscopy at a mean of 7.8 months (6 to 9) after surgery.

The biopsy specimens at the second-look arthroscopy showed significant improvement in the histological grading score compared with the pre-fixation scores (p < 0.01). In the specimens at the second-look arthroscopy, the extracellular matrix was stained more densely than at the time of fixation, especially in the middle to deep layers of the articular cartilage.

Our findings show that articular cartilage regenerates after fixation of an unstable lesion in osteochondritis dissecans.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 726 - 736
1 May 2010
Hee HT Ismail HD Lim CT Goh JCH Wong HK

Although success has been achieved with implantation of bone marrow mesenchymal stem cells (bMSCs) in degenerative discs, its full potential may not be achieved if the harsh environment of the degenerative disc remains. Axial distraction has been shown to increase hydration and nutrition. Combining both therapies may have a synergistic effect in reversing degenerative disc disease. In order to evaluate the effect of bMSC implantation, axial distraction and combination therapy in stimulating regeneration and retarding degeneration in degenerative discs, we first induced disc degeneration by axial loading in a rabbit model.

The rabbits in the intervention groups performed better with respect to disc height, morphological grading, histological scoring and average dead cell count. The groups with distraction performed better than those without on all criteria except the average dead cell count.

Our findings suggest that bMSC implantation and distraction stimulate regenerative changes in degenerative discs in a rabbit model.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 413 - 416
1 Mar 2007
van Riet RP van Glabbeek F de Weerdt W Oemar J Bortier H

We undertook a study on eight arms from fresh cadavers to define the clinical usefulness of the lesser sigmoid notch as a landmark when reconstructing the length of the neck of the radius in replacement of the head with a prosthesis. The head was resected and its height measured, along with several control measurements. This was compared with in situ measurements from the stump of the neck to the proximal edge of the lesser sigmoid notch of the ulna. All the measurements were performed three times by three observers acting independently.

The results were highly reproducible with intra- and interclass correlations of > 0.99. The mean difference between the measurement on the excised head and the distance from the stump of the neck and the lesser sigmoid notch was −0.02 mm (−1.24 to +0.97). This difference was not statistically significant (p = 0.78).

The proximal edge of the lesser sigmoid notch provides a reliable landmark for positioning a replacement of the radial head and may have clinical application.