Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 19 - 19
1 May 2016
Halloran J Zadzilka J Colbrunn R Bonner T Anderson C Klika A Barsoum W
Full Access

Introduction

Improper soft-tissue balancing can result in postoperative complications after total knee arthroplasty (TKA) and may lead to early revision. A single-use tibial insert trial with embedded sensor technology (VERASENSE from OrthoSensor Inc., Dania Beach, FL) was designed to provide feedback to the surgeon intraoperatively, with the goal to achieve a “well-balanced” knee throughout the range of motion (Roche et al. 2014). The purpose of this study was to quantify the effects of common soft-tissue releases as they related to sensor measured joint reactions and kinematics.

Methods

Robotic testing was performed using four fresh-frozen cadaveric knee specimens implanted with appropriately sized instrumented trial implants (geometry based on a currently available TKA system). Sensor outputs included the locations and magnitudes of medial and lateral reaction forces. As a measure of tibiofemoral joint kinematics, medial and lateral reaction locations were resolved to femoral anterior-posterior displacement and internal-external tibial rotation (Fig 1.). Laxity style joint loading included discrete applications of ± 100 N A-P, ± 3 N/m I-E and ± 5 N/m varus-valgus (V-V) loads, each applied at 10, 45, and 90° of flexion. All tests included 20 N of compressive force. Laxity tests were performed before and after a specified series of soft-tissue releases, which included complete transection of the posterior cruciate ligament (PCL), superficial medial collateral ligament (sMCL), and the popliteus ligament (Table 1). Sensor outputs were recorded for each quasi-static test. Statistical results were quantified using regression formulas that related sensor outputs (reaction loads and kinematics) as a function of tissue release across all loading conditions. Significance was set for p-values ≤ 0.05.