Advertisement for orthosearch.org.uk
Results 1 - 8 of 8
Results per page:
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 222 - 222
1 Sep 2012
Li C Hussain A Pamu J Kamali A
Full Access

INTRODUCTION. Hip wear simulator test results could be affected by many non-bearing related factors such as fixation surface conditions, equipment calibration and component set-up. In an effort to improve the accuracy, reliability and repeatability of hip simulator test, a quality management system has been established at the IDC hip tribology laboratory, which has been accredited by UKAS (United Kingdom Accreditation Service) in accordance with the recognised international standard ISO17025. This study demonstrates that under well-controlled laboratory and testing conditions, satisfactory repeatability can be achieved during hip simulator studies. METHODS. Between 2008 to 2010, ten 50 mm Birmingham Hip Resurfacing (BHR) devices were tested by the IDC tribology laboratory using two ProSim hip wear simulators in three different tests (T1, T2 and T3). All tests were performed following the same IDC testing protocols at 1 Hz frequency for 5 million cycles (Mc) or until after a steady state was reached. Paul type stance phase loadings with a maximum load of 3 kN and a swing phase load of 0.3 kN was used. The flexion and extension angles were 30 and 15 degree. The internal/external rotation angel was ±10 degree. Wear was measured gravimetrically using an analytical balance (Mettler, Toledo xp504) with an accuracy of 0.1 mg. RESULTS. Results showed that wear of all the components tested followed the typical biphasic trend of wear for MoM hip joint device with a high “running-in” wear during the initial stage followed by a low “steady state” wear in the later stage. The data points scattered in a narrow range following the same trend line throughout the tests. The maximum difference in accumulated volume loss between any two tests at any given cycle was less in 0.2 mm3. The average running-in wear rates (±SD) were 0.72 ±0.17, 0.86 ±0.01 and 0.71 ±0.19 mm3/Mc respectively for test T1, T2 and T3. The running-in wear rates for test T1 and T3 were identical, whilst that for T2 was slightly higher. In spite of this, the wear rates for the bearings tested in T2 were still in the same range as those tested in T1 and T3 during the running-in stage. The steady state (>1.0Mc) wear rate of three different tests was practically the same, being 0.18 ±0.02, 0.18 ±0.06 and 0.16 ±0.04 mm3/Mc respectively for test T1, T2 and T3. Overall, the mean running-in wear rate was 0.74 ±0.16 mm3/Mc, and the steady state wear rate was 0.17 ±0.04 mm3/Mc for the 10 BHR devices tested in three different studies. DISCUSSIONS. This study shows that satisfactory repeatability can be achieved during hip simulator test of metal on metal hip joint devices. Well controlled test procedures including calibration of testing and measurement equipment, correct and consistence test set-up and attention to non-bearing surfaces are some of the most important prerequisites for reliable and accurate hip simulator test results


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 323 - 323
1 Sep 2012
Li C Hussain A Joseph P Kamali A
Full Access

INTRODUCTION. One of the recent advances in the hard-on-hard hip arthroplasty is the development of a new material of diffusion hardened oxidised zirconium (DHOxZr). The DHOxZr material consists of a ceramic layer on the top surface which is supported by a thick oxygen diffusion hardened (DH) zone underneath. With the desired properties of metal substrate, ceramic surface and a gradient structure of the oxygen diffusion zone, the DHOxZr-on-DHOxZr bearing combination is expected to produce low wear and minimal metal ions. This can possibly address the concerns associated with metal hypersensitivity associated with metal on metal bearings and fracture risk associated with ceramics. The aim of this study was to evaluate the wear of DHOxZr-on-DHOxZr as a possible hard on hard bearing combination in hips. METHODS. Three pairs of 50 mm DHOxZr prototype hip joint devices, each consisting of a DHOxZr modular head and a DHOxZr liner were wear tested in a ProSim hip joint simulator under standard testing conditions used by the Implant Development Centre (IDC), Smith & Nephew, Leamington Spa for 5 million cycles (Mc). The flexion/extension was 30° and 15°. The internal/external rotation was ± 10°. The force was Paul-type stance phase loading, with a maximum load of 3 kN and a standard ISO swing phase load of 0.3 kN. The test frequency was 1 Hz. Gravimetric analysis was carried out at 0, 0.5, 1, 2, 3, 4 & 5 million cycles. The lubricant was new born calf serum with 2 g/l sodium azide concentration diluted with de-ionised water to achieve average protein concentration of 20 g/l. Lubricant was changed every 0.25Mc during the first million cycles of the test and at every 0.33 Mc from 1 to 5Mc. RESULTS. A biphasic wear pattern was observed for the DHOxZr on DHOxZr devices during the test, with a running in phase from 0–1 Mc and a steady state phase from 1–5 Mc. At a confidence level of 95%, the mean wear rate was 0.21 ±0.06 mm3/Mc during the running-in stage, and the wear rate was reduced to 0.01 ±0.03 mm3/Mc during the steady state for the device tested. The wear volume loss of the DHOxZr on DHOxZr devices was significantly lower than that generated by CoCrMo metal on metal (MoM) devices (p < 0.05) under identical simulator test conditions


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 200 - 200
1 Sep 2012
Williams S Isaac G Fisher J
Full Access

INTRODUCTION. Ceramic-on-metal hip replacements (COM, where the head is a Biolox Delta ceramic and liner is Co Cr alloy), have demonstrated reduced wear under standard conditions in vitro compared to metal-on-metal (MOM) [1]. Early clinical results are also encouraging [2]. Recently concerns have been raised regarding the poor clinical performance of MOM hip resurfacings [3], particularly when cups are steeply inclined. Laboratory hip simulator testing has been used to replicate edge loading, also demonstrating elevated wear [4]. Therefore, a range of conditions to replicate sub-optimal use clinically to better predict in vivo performance should be used. The aim of this study was to compare the wear rates of MOM and COM under adverse edge loading conditions in an in vitro hip simulator test. METHODS. Ceramic-on-metal (n=3) and metal-on-metal (n=3) 36mm hip prostheses (supplied by DePuy International Ltd, UK) were tested in the Leeds Physiological Anatomical Hip Joint Simulator. Liners were mounted to provide a clinical angle of 45o, and stems positioned anatomically. A simplified gait cycle and microseparation was applied as previously described [5] for two million cycles in 25% new born calf serum. Gravimetric analysis was completed every million cycles and wear volumes calculated. RESULTS. The overall mean volumetric wear rate of COM bearings was 0.36 ± 0.55mm3 per million cycles, this was significantly less than the MOM bearing wear (1.32 ± 0.91mm3 per million cycles). For both COM and MOM bearings wear under these edge loading conditions was significantly greater if compared to previously reported wear under standard conditions [1]. DISCUSSION. The reduced wear of COM has been attributed to the differential hardness decreasing adhesive wear and reduced corrosive wear [6]. Wear under the harsh edge-loading conditions in this study is also reported to be significantly less in COM bearings compared to MOM. In MOM bearings in edge contact conditions, the wear zone becomes starved of lubrication, this elevates wear and increases damage at the edge of the cup. In COM bearings the harder head does not become damaged when there is lubricant starvation and hence wear does not accelerate in the same way. In conclusion, COM bearings show reduced wear compared to MOM bearings under standard and adverse conditions and there is some early evidence to support this finding clinically. ACKNOWLEDGEMENTS. Supported by DePuy International Ltd. SW is supported by a Royal Academy of Engineering/EPSRC (UK) fellowship


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 229 - 229
1 Sep 2012
Masson B Pandorf T
Full Access

Introduction. In total hip arthroplasty ceramic on ceramic bearing couples are used more and more frequently and on a wordwide basis. The main reason of this choice is reduction of wear debris and osteolysis. The tribological properties and the mechanical behaviour of the implanted ceramic must remain the same throughout the patient's life. The aim of this study was to evaluate the resistance of Alumina Matrix Composite to environmental degradation. Material and method. The alumina matrix composite or BIOLOX ® delta is manufactured in Germany by CeramTec. It is made up of 80 vol.% Al2O3, 17 vol.% Yttria Stabilized ZrO2 and 3vol.% strontium aluminate platelets. The zirconia grains account for 1.3 mol.% of the Yttria content. Accelerated aging tests in water steam at 142°C, 134°C, 121°C, and 105°C were performed to evaluate the aging kinetics of the composite. X-ray diffraction was used to determine the monoclinic phase content on the material surface. Phase transformation is associated with weakness and increase in roughness of zirconia ceramic implants. Results. The data below shows average monoclinic contents before and after aging in water vapour according to the ISO standard test (134°C, 2 bars water steam, 10 h) on the surface and inside the 28 mm taper(12/14 taper) femoral ball heads manufactured in alumina ceramic composite. There are precisions concerning the roughness and the load to failure before and after aging concerning 28mm diameter heads. Before Aging 13%+/-3% of Monoclinic content. After 10 H at 134°C23%+/-3% of Monoclinic content the roughness of the polished surface remain the same (5nm+/− 2). The load to failure of 28 mm heads before aging is 76 kN +/− 5kN, and 72 kN +/− 5kN after aging. The results show that although a rise in monoclinic content is predictable after long aging duration in vivo, the impact of the transformation is quite different to monolithic zirconia. A zirconia femoral head exhibits an important increase of roughness from 2 nm to more than 50 nm when submitted to the same duration of ageing. Other tests with hip simulators under severe micro separation have been done to analyse the impact of aging on wear performance. The main wear zone on femoral heads underwent a phase transformation from tetragonal to monoclinic (23% monoclinic) at 5 milion cycles duration without any change in roughness after 5Mc duration. Conclusion. This experimental testing program has enabled a prediction for the long-term in vivo environmental resistance of prostheses made out of Alumina Matrix Composite. The substantial improvement in mechanical properties and the excellent wear behaviour, even under severe microseparation conditions has been clinically confirmed. Today more than 960,000 ceramic ball heads and more than 450 000 ceramic inserts made of the alumina matrix composite have been implanted. Additionally, due to enhanced mechanical behaviour, new applications in orthopaedics are possible


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 107 - 107
1 Sep 2012
Hadley M Hadfield F Hardaker C Isaac G Fisher J Wye J Barnett J
Full Access

Introduction. Hip wear simulation is a widely used technique for the pre-clinical evaluation of new bearing designs. However, wear rates seen in vitro can often be significantly different to those seen clinically. This can be attributed to the difference between the optimal conditions in a simulator and wide ranging conditions in real patients. This study aimed to develop more clinically relevant simulator tests, looking specifically at the effects of cup inclination angle (in vivo) and stop-dwell-start (SDS) protocols on a clinically available product. Method. Five tests using a Paul type walking cycle (ISO 14242) were carried out on two ProSim hip simulators: . 28mm MoM, standard walking, cup inclination 45°, (n = 5). 36mm MoM, standard walking, cup 45°, (n = 4). 36mm MOM, SDS: 10 walking cycles and pause of 5s with stance load of 1250N cup 45°, (n = 5). 36mm MOM, SDS: 10 walking cycles and pause of 30s with stance load of 1250N, (n = 5) cup 45°. 36mm MOM, standard walking, cup 55°(n = 5), and 65°(n = 5). All samples had matched clearances, measured using a CMM (Prismo Navigator, Zeiss, Germany). Wear was measured gravimetrically (Sartorius ME235S: 0.01mg). Results and Discussion. Metal-on-metal hip bearings are known to exhibit a bi-phasic wear pattern in-vitro (1), with the majority of wear occurring over the first 1–1.5mc, followed by lower steady state wear as the components conform to each other. Therefore bedding in wear over the first 1 million cycles was compared for each of the sample groups in this study. Bedding in wear for 36mm bearings under standard angle (45 inclination) and walking conditions was 0.16±0.15mm3/mc, while for 28mm bearings, wear was 2.67±2.06mm3/mc. Positioning the acetabular cup at an increased angle led to increased wear. At 55bedding in wear was .25±0.29mm3/mc, while at 65wear was 0.65±0.32mm3/mc. When SDS conditions were applied, no significant increase in wear was observed under either the 5s or 30s pause conditions, with bedding in wear of 0.19±0.31mm3/mc and 0.14±0.13mm3/mc respectively. Conclusion. This study illustrates the robust performance of the 36mm MoM bearing under extreme test conditions when compared with 28mm MoM. No effect was observed with the application of stop-dwell-start conditions, and only a small increase in wear was observed when the cup was inclined to a greater angle. Despite the application of extreme test conditions, the 36mm MoM still demonstrates lower wear than 28mm under all conditions. In all but the 65 inclination conditions, this difference is statistically significant. This study replicates some of the more frequently observed extreme conditions measured in real patients with THRs (2,3). Even under these more aggressive scenarios, wear of the 36mm bearing is still relatively low when compared with a commercially available 28mm THR


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 239 - 239
1 Sep 2012
Hussain A Hussain A Kamali A Li C Pamu J Ashton R
Full Access

INTRODUCTION. Analysis of retrieved ceramic components have shown areas of localized ‘stripe wear’, which have been attributed to joint laxity and/or impingement resulting in subluxation of the head, causing wear on the edge of the cup. Studies have been conducted into the effects of mild subluxation, however few in vitro tests have looked at severe subluxation. The aim of this study was to develop a more clinically relevant subluxation protocol. MATERIALS & METHODS. Seven (Subluxation n=4; standard test n=3) of 36mm Biolox Forte (R3, Smith & Nephew) ceramic devices were tested for 0.5m cycles (mc). Two of the subluxed joints were further tested to 1 Mc. The devices were subjected to subluxation under standard testing conditions. The flex/ext was 30° and 15° respectively, with internal/external rotation of ±10°. The force was Paul type stance phase loading with a maximum load of 3 kN, and a standard ISO swing phase load of 0.3 kN at 1 Hz. The test was conducted on a ProSim hip joint wear simulator (SimSol, UK). The simulator is equipped with a novel mechanism to achieve translation of the head, to achieve subluxation. During the ISO swing phase load of 0.3kN, a controlled lateral force required for the translation of the head is applied by a cam mechanism, head retraction then occurs during heel strike. The lubricant used was new born calf serum diluted with de-ionised water to achieve average protein concentration of 20 g/l, with 0.2 wt % concentration NaN3, and changed every 250k cycles. Measurements have been taken at 0.5 & 1 mc stages. RESULTS. Linear wear measurements conducted on the subluxed joints resulted in stripe wear similar to that reported in vivo. Average length, width and depth dimensions were 25.34±1.96 mm, 8±1.60 mm and 16.95±3.87 μm (± 95% CL) respectively. Linear wear at 0.5 Mc for standard joints, were undistinguishable from the original profile. Gravimetrically, weight loss was undetectable for joints tested under standard conditions. The volume loss of the joints under subluxation was 1.9± 0.7 mm3 at 0.5 mc. Two joints tested to 1mc generated an average volume loss of 3.1±2.3 mm3. The stripe wear length, width and depth at 1 Mc were 25.30±3.33mm, 8±3.92mm and 35±17.07 μm respectiveley. DISCUSSION. The current study presents test results of a hip joint simulator with a novel subluxation mechanism to simulate severe and clinically relevant hip joint. Past techniques have had to reduce the swing phase load to achieve stripe wear patches of varying size and depth. The subluxed joints produced significantly higher volumetric wear than the standard joints. Dimensional measurements in terms of length, width and depth of wear patches of subluxed joints generated similar results to that which have been observed following retrieval analysis. Tests that can simulate different types of activity in hip joint simulators will help to improve the design and understanding of implant behaviour in vivo


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 204 - 204
1 Sep 2012
Zietz C Fritsche A Mittelmeier W Bader R
Full Access

The prevalent cause of implant failure after total joint replacement is aseptic loosening caused by wear debris. Improvement of the wear behaviour of the articulating bearing between the cup and femoral head is essential for increased survival rate of artificial hip joints. Cross-linking of the polyethylene (PE) material is one attempt to reduce wear particle release at the articulating surface. Various cross-linked polyethylenes (X-PE) are used in orthopaedics since several years. In total hip arthroplasty (THA) the use of larger femoral head sizes has specific reasons. Larger heads lead to a decreased risk of total hip dislocation and impingement as well as an improved range of motion in comparison to smaller head sizes like 28mm or less. However, the increasing diameter of femoral head can be associated with lower thickness of the PE liner and increased wear rate. Cross-linking of PE can improve the wear rate of the liner and hence supports the use of larger femoral heads. The aim of this experimental study was to evaluate the wear of standard vs. sequential X-PE (X3-PE) liner in combination with different ceramic femoral head sizes. Wear testing was performed for 5 million load cycles using standard UHMW-PE liners (N2Vac) and X3-PE liners (each Stryker GmbH & Co. KG, Duisburg, Germany) combined with 28mm ceramic ball heads and the Trident PSL acetabular cup (Stryker). Furthermore, X3-PE liners with an internal diameter of 36mm and 44mm and decreased wall thickness (5.9mm and 3.8mm) were combined with corresponding ceramic heads. An eight station hip wear simulator according to ISO 14242 (EndoLab GmbH, Rosenheim, Germany) was used to carry out the standard wear tests. The tests were realised in temperature-controlled chambers at 37°C containing calf serum (protein content 20g/l). The average gravimetrical wear rates of the standard UHMW-PE (N2Vac) liners combined with 28mm ceramic heads amounted to 12.6 ± 0.8mg/million cycles. Wear of X3-PE liners in combination with 28 mm ceramic heads was not detectable. The average gravimetrical wear rates of the X3-PE liners in combination with 36mm and 44mm ceramic heads amounted to 2.0 ± 0.5mg and 3.1 ± 0.3mg/million cycles, respectively. The purpose of this study was to evaluate the effect of femoral head size at THA on standard and sequential X-PE liner. The wear simulator tests showed that the wear rate of PE liners with small heads (28mm) decreased by cross-linking of the PE significantly. The amount of wear at X-PE increased slightly with larger head size (36mm and 44mm). However, by sequential cross-linking, the wear rate using thinner liners and larger femoral heads is reduced to a fractional amount of wear at conventional UHMW-PE. Hence, the above-mentioned advantages of larger femoral head diameters can be realised by improved wear behaviour of sequential X-PE


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 223 - 223
1 Sep 2012
Herrera L Loving L Essner A Nevelos J
Full Access

Osteolysis induced by UHMWPE debris has historically been one of the major causes of long term failure of TJR. An increase in concentration of polyethylene particles in the peri-prostheic tissue has been linked to an increased incidence of osteolysis. The dual mobility hip bearing concept mates a femoral head into a polyethylene liner which has an unconstrained articulation into a metal shell. The wear mechanism of the dual mobility hip bearing is distinct from a constrained single articulation design, which may result in a difference in wear debris particles. The aim of this study is to evaluate wear debris generated from a dual mobility hip and compare it to a conventional single articulation design when both are manufactured from sequentially crosslinked and annealed polyethylene. The dual mobility hip (Restoration ADM) incorporated a 28mm CoCr femoral head into a polyethylene liner that articulates against a metal shell (48mm ID). The conventional hip (Trident®) mated a 28mm CoCr femoral head against a polyethylene liner. The polyethylene for all liners was sequentially crosslinked and annealed (X3). A hip joint simulator was used for testing at a rate of 1 Hz with cyclic Paul curve physiologic loading. A serum sample from each testing group was collected. Serum samples were protein digested following the published process by Scott et al. The digested serum was then filtered through a series of polycarbonate filter papers of decreasing size and sputter coated with gold for analysis using SEM. Image fields were randomized and wear debris was compared in terms of its length, width, aspect ration, and equivalent circular diameter (ECD). A total of 149 conventional hip particles and 114 dual mobility hip particles were imaged. Results show a majority of particles are of spherical nature and images do not indicate the presence of fibrillar or larger elongated polyethylene debris. Particle length between designs is not statistically different, while all other comparisons show statistical significance (p<0.05). It is hypothesized that the dual mobility hip system reduces the total amount of cross-shear motion on any one articulation, which aids in the reduction in wear. This design feature may be responsible for the slight difference in morphology of dual mobility wear debris when compared to the constrained hip design. The length of the particles was similar, simply indicating a different shape rather than a marked reduction in overall size. The debris generated is this study was from highly crosslinked polyethylene in two different designs, which produced a very significant decrease in quantity of particles when compared to the quantity of debris from conventional polyethylene. The wear debris was of similar length in both designs and so we do not expect any difference in biological response to debris from either device. The dual mobility design has also shown no effect of cup abduction angle on wear demonstrating forgiveness to implant positioning. This advantage, combined with the low wear rate and similar length wear particles, should lead to good clinical performance of dual mobility cups with sequentially irradiated and annealed polyethylene