Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 289 - 289
1 Sep 2012
Bragdon C Martell J Jarrett B Clohisy J White R Goldberg V Della Valle C Berry D Johanson P Harris W Malchau H
Full Access

Introduction. Total hip replacements using highly cross-linked polyethylene show excellent clinical outcomes, low wear, and minimal lysis at 5 years follow-up. A recent RSA study reports a significant increase in femoral head penetration between 5 and 7 years. This study is a multi-center radiographic analysis to determine whether the RSA observation is present in a large patient cohort. Methods. Six centers were enrolled for radiographic analysis of primary total hip arthroplasty for standard head sizes (26mm, 28mm, or 32mm). Radiographic inclusion criteria required a minimum of four films per patient at the following time points: 1 year; 2–4.5 years; 4.5–5.5 years; and 5.5–11 years. The Martell Hip Analysis Suite was used to analyze pelvic radiographs resulting in head penetration values. Wear rates were determined in two ways: the longest follow-up radiograph compared to the 1 year film, and individual linear regressions for the early and late periods. For both methods, average wear rates from the early period (1 to 5.5 years) and late period (>5.5 years) were compared using t-tests. Results. We present the completed analysis of 235 of the 250 hips under active analysis with 3160 film comparisons. Average follow-up was 7.2±0.99 years (range 5.4–10.1). Based on latest follow-up, the average wear rate was not significantly different during the early and late periods (10.1±95μm/year and −3.8±235μm/year respectively, p=0.518). Based on the group regression, the average wear rate was not significantly different during the early and late periods (7.5μm/year and −36μm/year respectively, p=0.13). Conclusion. In this large multicenter radiographic study, we found no late increase in femoral head penetration into highly cross-linked polyethylene as suggested by the RSA report. Additional centers and patients are being recruited in order to reduce the variation in the late period


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 232 - 232
1 Sep 2012
Beck M Kohl S
Full Access

INTRODUCTION. The lifetime of UHMWPE implants may be limited by wear and oxidative degeneration. Wear produced particles are in general biologically active, and may induce osteolysis. As threshold of PE wear rate below which osteolysis is rarely observed is postulated to be less tahn 0.1 mm per year. Moreover, PE delamination and breakage are consequences of the embrittlement of the PE due to oxidation. Both demonstrate, that improving the clinical behaviour of UHMWPE means reduction of wear particles. The first can be achieved by cross-linking the second by the anti-oxidative stabiliser vitamin E. The highly cross-linked PE vitamys ® used for the isoelastic monobloc cup RM Pressfit (Mathys AG Bettlach, Bettlach, Switzerland) is mixed with 0.1% of synthetic vitamin E and is the first and only highly cross-linked PE used in total hip replacement that meets all requirements for the best grade UHMWPE in yield strength, ultimate tensile strength and elongation at break. METHODS AND MATERIAL. With the first implantation of RM Pressfit vitamys® a prospective multicentre study was started. So far 256 cases in 7 clinics from Europe and New Zealand are included. This report presents the first clinical experiences of one Swiss clinic from the multicentre study. Prospective data collection includes Harris Hip score (HHS), patient satisfaction and radiographic analysis. Clinical and radiographic follow-up is done after 6 weeks, 6, 12 and 24 months, and thereafter for long-term results. Standardized documentation of surgery and postoperative course is performed. RESULTS. 81 patients were included in the study. Mean age at implantation is 69 years. The indication was primary OA in 57% and secondary OA in 36%, the others were treated either for necrosis, fracture or hip dysplasia. 68% of the patients received a 36mm femoral head, 31% a 32mm and 1% a 28mm. Intraoperatively two minor complications occurred, one femur and one trochanter fissure. No complications occurred during implantation of the cup. Most of the cups (59%) were implanted with an inclination of 40°-50°, 30% with 35°-40°. There were no signs of early loosening and a good Harris Hip Score was achieved. DISCUSSION. The principals of this monobloc cup with its isoelastic property and the thin titanium coating is a proven concept. Ihle et al. (JBJS 2008) reported 91% survival rate with revision for aseptic loosening as endpoint after 20 years. They found an increase of cup revisions after 14 to 16 years after implantation due to osteolysis probably due to PE wear. At short term, we haven't encountered any problems related to the implantation of this cup. So far one patient in the entire multicentre study needed a stem revision after periprosthetic femur fracture. To prove the durability of this novel material a follow-up of 14 years and more will be required


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 243 - 243
1 Sep 2012
Bragdon C Malchau H Greene M Doerner M Emerson R Gebuhr P Huddleston J Cimbrelo E
Full Access

Introduction. Proper cup positioning is a critical component in the success of total hip arthroplasty surgery. A multicenter study has been initiated to study a new type of highly cross-linked polyethylene. This study provides a unique opportunity to a review the acetabular cup placement of over 500 patients implanted in the past 2 years from 13 medical centers from the U.S., Mexico, and Europe. Methods. 482 patients have received primary total hip arthroplasty using components from a single manufacturer in 5 centers in the US and Mexico and 7 centers in Europe. The acetabular anteversion and inclination were measured in post-operative radiographs. An acceptable window of cup position is defined at 5–25° of anteversion and 30–45° of inclination. Results. The measured cup anteversion and inclination averaged 15.89° ± 8.91° (0.00–42.25°) and 43.27° ± 7.17° (23.46–67.79°), respectively. Of the patient radiographs read, 71% were within the acceptable range of anteversion, 55% were in the acceptable range of inclination, and 41% satisfied both criteria. The best performing center had 86% of patients within the acceptable range of anteversion, 63% in the acceptable range of inclination, and 57% satisfied both criteria. The worst performing center had 54% within the acceptable range of anteversion, 29% in the acceptable range of inclination, and 17% satisfied both criteria. Conclusion. A significant variation in acetabular cup anteversion and inclination exists in this study both within and between the participating high volume centers. Correlation to mid- and long-term clinical outcome will show the clinical relevance of the finding, but liner designs with unsupported polyethylene should be used with caution


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 142 - 142
1 Sep 2012
Traynor A Simpson D Ellison P Collins S
Full Access

Introduction. Cobalt chrome on polyethylene remains a widely used bearing combination in total joint replacement. However wear induced osteolysis, bulk material property degradation of highly cross-linked polyethylene (HXLPE) [1], and oxidation after implantation (thought to be as a result of lipid absorption or cyclic loading [2]) remains a concern. ECIMA is a cold-irradiated, mechanically annealed, vitamin E blended next generation HXLPE developed to maintain mechanical properties, minimise wear and to improve the oxidation resistance in the long-term. The aim of this study was to compare the in-vitro wear rate and mechanical properties of three different acetabular liners; conventional UHMWPE, HXLPE and ECIMA. Methods. Twelve liners (Corin, UK) underwent a 3 million cycle (mc) hip simulation. Three conventional UHMWPE liners (GUR1050, Ø32 mm, 30 kGy sterilised in Nitrogen), three HXLPE liners (GUR1020, Ø40 mm, 75 kGy cross-linking and EtO sterilised) and six ECIMA liners (0.1 wt% vitamin E GUR1020, Ø40 mm, 120 kGy cross-linking, mechanically deformed and annealed, and EtO sterilised) articulated against CoCrMo alloy femoral heads to ASTM F75 (Corin, UK). Wear testing was performed in accordance with ISO 14242 parts 1 and 2, with a maximum force of 3.0 kN and at a frequency of 1 Hz. The test lubricant used was calf serum with a protein content of 30 g/l and 1% (v/v) patricin added as an antibacterial agent. Volumetric wear rate was determined gravimetrically after the first 0.5 mc and every 1 mc thereafter. ASTM D638 type V specimens (3.2 mm thick) were machined from ECIMA material for uniaxial tension testing to ASTM D638. Ultimate tensile strength (UTS), yield strength and elongation values were measured. These values were compared to mechanical data available for the other material types. Results. There was a 94% and a 68% reduction in the wear rate for the ECIMA liners compared to the conventional UHMWPE and HXLPE liners respectively. There was an increase in UTS, yield strength and elongation of 11%, 11% and 15% respectively, for ECIMA compared to HXLPE. Discussion. The wear results reported in this study indicate that ECIMA is a very low wearing material which has the potential to reduce wear related osteolysis in-vivo. Importantly, the mechanical properties were generally maintained unlike the degradation found in many modified polyethylene materials and were more comparable to conventional UHMWPE than HXLPE. The reduced wear rate during in-vitro hip simulation of ECIMA compared to conventional UHMWPE, coupled with improved mechanical properties in comparison to HXLPE, makes ECIMA a promising next generation, advanced bearing material


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 95 - 100
1 Jan 2013
Chémaly O Hebert-Davies J Rouleau DM Benoit B Laflamme GY

Early total hip replacement (THR) for acetabular fractures offers accelerated rehabilitation, but a high risk of heterotopic ossification (HO) has been reported. The purpose of this study was to evaluate the incidence of HO, its associated risk factors and functional impact. A total of 40 patients with acetabular fractures treated with a THR weres retrospectively reviewed. The incidence and severity of HO were evaluated using the modified Brooker classification, and the functional outcome assessed. The overall incidence of HO was 38% (n = 15), with nine severe grade III cases. Patients who underwent surgery early after injury had a fourfold increased chance of developing HO. The mean blood loss and operating time were more than twice that of those whose surgery was delayed (p = 0.002 and p < 0.001, respectively). In those undergoing early THR, the incidence of grade III HO was eight times higher than in those in whom THR was delayed (p = 0.01). Only three of the seven patients with severe HO showed good or excellent Harris hip scores compared with eight of nine with class 0, I or II HO (p = 0.049). Associated musculoskeletal injuries, high-energy trauma and head injuries were associated with the development of grade III HO.

The incidence of HO was significantly higher in patients with a displaced acetabular fracture undergoing THR early compared with those undergoing THR later and this had an adverse effect on the functional outcome.

Cite this article: Bone Joint J 2013;95-B:95–100.