Limb salvage in bone tumour patients replaces the bone with massive segmental prostheses where achieving bone integration at the shoulder of the implant through extracortical bone growth has been shown to prevent loosening. This study investigates the effect of multidrug chemotherapy on extracortical bone growth and early radiological signs of aseptic loosening in patients with massive distal femoral prostheses. A retrospective radiological analysis was performed on adult patients with distal femoral arthroplasties. In all, 16 patients were included in the chemotherapy group with 18 patients in the non-chemotherapy control group. Annual radiographs were analyzed for three years postoperatively. Dimensions of the bony pedicle, osseointegration of the hydroxyapatite (HA) collar surface, bone resorption at the implant shoulder, and radiolucent line (RLL) formation around the cemented component were analyzed.Aims
Methods
We investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the distal femur after resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database, review of the design of the implant and case records. In total, 335 patients underwent a
The use of extendible
We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
We have made a retrospective study of 1001 custom-made prostheses used as replacements after surgery for bone tumours. There were 493 distal femoral, 263 proximal femoral and 245 proximal tibial prostheses. Aseptic loosening was shown to be the principal mode of failure of the implants, and 71 patients had revision for aseptic loosening of a cemented intramedullary stem. The probability of a patient surviving aseptic loosening for 120 months was 93.8% for a proximal femoral replacement, 67.4% for a distal femoral prosthesis and 58% for a proximal tibial implant. In patients with
Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.Aims
Methods
Segmental resection of malignant bone disease in the femoral diaphysis with subsequent limb reconstruction is a major undertaking. This is a retrospective review of 23 patients who had undergone limb salvage by endoprosthetic replacement of the femoral diaphysis for a primary bone tumour between 1989 and 2005. There were 16 males and seven females, with a mean age of 41.3 years (10 to 68). The mean overall follow-up was for 97 months (3 to 240), and 120 months (42 to 240) for the living patients. The cumulative patient survival was 77% (95% confidence interval 63% to 95%) at ten years. Survival of the implant, with failure of the endoprosthesis as an endpoint, was 85% at five years and 68% (95% confidence interval 42% to 92%) at ten years. The revision rate was 22% and the overall rate of re-operation was 26%. Complications included deep infection (4%), breakage of the prosthesis (8%), periprosthetic fracture (4%), aseptic loosening (4%), local recurrence (4%) and metastases (17%). The 16 patients who retained their diaphyseal endoprosthesis had a mean Musculoskeletal Tumour Society score of 87% (67% to 93%). They were all able to comfortably perform most activities of daily living. Femoral diaphyseal endoprosthetic replacement is a viable option for reconstruction following segmental resection of malignant bone disease. It allows immediate weight-bearing, is associated with a good long-term functional outcome, has an acceptable complication and revision rate and, most importantly, does not appear to compromise patient survival.
We undertook a retrospective review of 33 patients who underwent total femoral endoprosthetic replacement as limb salvage following excision of a malignant bone tumour. In 22 patients this was performed as a primary procedure following total femoral resection for malignant disease. Revision to a total femoral replacement was required in 11 patients following failed segmental endoprosthetic or allograft reconstruction. There were 33 patients with primary malignant tumours, and three had metastatic lesions. The mean age of the patients was 31 years (5 to 68). The mean follow-up was 4.2 years (9 months to 16.4 years). At five years the survival of the implants was 100%, with removal as the endpoint and 56% where the endpoint was another surgical intervention. At five years the patient survival was 32%. Complications included dislocation of the hip in six patients (18%), local recurrence in three (9%), peri-prosthetic fracture in two and infection in one. One patient subsequently developed pulmonary metastases. There were no cases of aseptic loosening or amputation. Four patients required a change of bushings. The mean Musculoskeletal Tumour Society functional outcome score was 67%, the mean Harris Hip Score was 70, and the mean Oxford Knee Score was 34. Total femoral endoprosthetic replacement can provide good functional outcome without compromising patient survival, and in selected cases provides an effective alternative to amputation.
The aim of this study was to establish what happens to patients
in the long term after endoprosthetic replacement for a primary
malignant tumour of bone. We conducted a retrospective analysis of a prospectively maintained
database to identify all patients who had undergone an endoprosthetic
replacement more than 25 years ago and who were still alive. Their
outcomes were investigated with reference to their complications and
need for further surgery. A total of 230 patients were identified.
Their mean age at diagnosis was 20.7 years (five to 62). The most
common diagnosis was osteosarcoma (132). The most common site was
the distal femur (102). Aims
Patients and Methods
Massive endoprostheses using a cemented intramedullary stem are widely used to allow early resumption of activity after surgery for tumours. The survival of the prosthesis varies with the anatomical site, the type of prosthesis and the mode of fixation. Revision surgery is required in many cases because of aseptic loosening. Insertion of a second cemented endoprosthesis may be difficult because of the poor quality of the remaining bone, and loosening recurs quickly. We describe a series of 14 patients with triplate fixation in difficult revision or joint-sparing tumour surgery with a minimum follow-up of four years. The triplate design incorporated well within a remodelled cortex to achieve osseomechanical integration with all patients regaining their original level of function within five months. Our preliminary results suggest that this technique may provide an easy, biomechanically friendly alternative to insertion of a further device with an intramedullary stem, which has a shorter lifespan in revision or joint-sparing tumour surgery. A short segment of bone remaining after resection of a tumour will not accept an intramedullary stem, but may be soundly fixed using this method.
We reviewed retrospectively the results in 211 consecutive patients who had undergone limb salvage for bone neoplasia with endoprosthetic reconstruction of the proximal femur (96), distal femur (78), proximal tibia (30) and total femur (7). Their mean age was 50 years (11 to 86) and the mean follow-up period was 37.3 months (1 to 204). A total of 35 (16.6%) prostheses failed. Overall, implant survival was 78% (95% confidence interval (CI) 0.29 to 0.54) at five years, 60% (95% CI 0.93 to 2.35) at ten years and 60% (95% CI 1.27 to 3.88) at 15 years. Survivorship of the limb was 97.6% (95% CI 1.73 to 3.35) at ten years. The gender, age, diagnosis and location of the tumour were not prognostic variables for failure. Modular endoprosthetic replacement in the lower limb is a durable long-term reconstructive option, with the implants generally outlasting the patient.
We used a knee-sparing distal femoral endoprosthesis in young patients with malignant bone tumours of the distal femur in whom it was possible to resect the tumour and to preserve the distal femoral condyles. The proximal shaft of the endoprosthesis had a coated hydroxyapatite collar, while the distal end had hydroxyapatite-coated extracortical plates to secure it to the small residual femoral condylar fragment. We reviewed the preliminary results of this endoprosthesis in eight patients with primary bone tumours of the distal femur. Their mean age at surgery was 17.years (14 to 21). The mean follow-up was 24 months (20 to 31). At final follow-up the mean flexion at the knee was 102° (20° to 120°) and the mean Musculoskeletal Tumour Society score was 80% (57% to 96.7%). There was excellent osteointegration at the prosthesis-proximal bone interface with formation of new bone around the hydroxyapatite collar. The prosthesis allowed preservation of the knee and achieved a good functional result. Formation of new bone and remodelling at the interface make the implant more secure. Further follow-up is required to determine the long-term structural integrity of the prosthesis.
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
We reviewed our initial seven-year experience
with a non-invasive extendible prosthesis in 34 children with primary bone
tumours. The distal femur was replaced in 25 cases, total femur
in five, proximal femur in one and proximal tibia in three. The
mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain
alive. The prostheses were lengthened by an electromagnetic induction
mechanism in an outpatient setting and a mean extension of 32 mm
(4 to 80) was achieved without anaesthesia. There were lengthening
complications in two children: failed lengthening in one and the
formation of scar tissue in the other. Deep infection developed
in six patients (18%) and local recurrence in three. A total of
11 patients required further surgery to the leg. Amputation was
necessary in five patients (20%) and a two-stage revision in another.
There were no cases of loosening, but two patients had implant breakage
and required revision. The mean Musculoskeletal Tumor Society functional
score was 85% (60% to 100%) at last known follow-up. These early
results demonstrate that the non-invasive extendible prosthesis
allows successful lengthening without surgical intervention, but
the high incidence of infection is a cause for concern.
This paper describes the preliminary results of a proximal tibial endoprosthesis which spares the knee joint and enables retention of the natural articulation by replacing part of the tibial metaphysis and diaphysis. In eight patients who had a primary malignant bone tumour of the proximal tibia, the distal stem, which had a hydroxyapatite-coated collar to improve fixation, was cemented into the medullary canal. The proximal end had hydroxyapatite-coated extracortical plates which were secured to the remaining proximal tibial metaphysis using cortical screws. The mean age of the patients at operation was 28.9 years (8 to 43) and the mean follow-up was for 35 months (4 to 48). The mean Musculoskeletal Tumour Society score was 79% (57% to 90%), the mean Oxford Knee score was 40 points of 48 (36 to 46) and the mean knee flexion was 112° (100° to 120°). In one patient, revision to a below-knee amputation through the prosthesis was required because of recurrence of the tumour. Another patient sustained a periprosthetic fracture which healed with a painful malunion. This was revised to a further endoprosthesis which replaced the knee. In the remaining six patients the prosthesis allowed preservation of the knee joint with good function and no early evidence of loosening. Further follow-up is required to assess the longevity of these prostheses.
In developing countries locally-made low-cost prostheses are mainly used in limb-salvage surgery to alleviate the economic burden. We retrospectively collected data on 104 patients treated by limb-salvage surgery between July 1997 and July 2005. We used a locally-designed and fabricated stainless-steel endoprosthesis in each case. Oncological and functional outcomes were evaluated at a mean follow-up of 47 months (12 to 118). A total of 73 patients (70.2%) were free from disease, nine (8.7%) were alive with disease, 19 (18.2%) had died from their disease and three (2.9%) from unrelated causes. According to the Musculoskeletal Tumor Society scoring system, the mean functional score was 76.3% (SD 17.8). The five-year survival for the implant was 70.5%. There were nine cases (8.7%) of infection, seven early and two late, seven (6.7%) of breakage of the prosthesis, three (2.9%) of aseptic loosening and two (1.9%) of failure of the polyethylene bushing. Multivariate analysis showed that a proximal tibial prosthesis and a resection length of 14 cm or more were significant negative prognostic factors. Our survival rates and Musculoskeletal Tumor Society functional scores are similar to those reported in the literature. Although longer follow-up is needed to confirm our results, we believe that a low-cost custom-made endoprosthesis is a cost-effective and reliable reconstructive option for limb salvage in developing countries.
We have investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the proximal tibia following resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database. A total of 194 patients underwent a proximal tibial replacement, with 95 having a fixed-hinge design and 99 a rotating-hinge with a hydroxyapatite collar; their median age was 21.5 years (10 to 74). At a mean follow-up of 14.7 years (5 to 29), 115 patients remain alive. The risk of revision for any reason in the fixed-hinge group was 32% at five years, 61% at ten years and 75% at 15 and 20 years, and in the rotating-hinge group 12% at five years, 25% at ten years and 30% at 15 years. Aseptic loosening was the most common reason for revision in the fixed-hinge knees, fracture of the implant in the early design of rotating hinges and infection in the current version. The risk of revision for aseptic loosening in the fixed-hinge knees was 46% at ten years. This was reduced to 3% in the rotating-hinge knee with a hydroxyapatite collar. The cemented, rotating hinge design currently offers the best chance of long-term survival of the prosthesis.