The mechanical alignment (MA) for Total Knee Arthroplasty (TKA) with neutral alignment goal has had good overall long-term outcomes. In spite of improvements in implant designs and surgical tools aiming for better accuracy and reproducibility of surgical technique, functional outcomes of MA TKA have remained insufficient. Therefore, alternative, more anatomical options restoring part (adjusted MA (aMA) and adjusted kinematic alignment (aKA) techniques) or the entire constitutional frontal deformity (unicompartment knee arthroplasty (UKA) and kinematic alignment (KA) techniques) have been developed, with promising results. The kinematic alignment for TKA is a new and attractive surgical technique enabling a patient specific treatment. The growing evidence of the kinematic alignment mid-term effectiveness, safety and potential short falls are discussed in this paper. The current review describes the rationale and the evidence behind different surgical options for knee replacement, including
In a prospective study of 14 patients undergoing total hip replacement we have used dual-energy X-ray absorptiometry (DEXA) to investigate remodelling of the bone around two different designs of cementless femoral prosthesis. The bone mineral density (BMD) was measured at 12-weekly intervals for a year. Eight patients (group A) had a stiff, collarless implant and six (group B) a flexible isoelastic implant. Patients in group A showed a decrease in BMD from 14 weeks after operation. By 12 months, the mean loss in BMD was 27%, both medially and laterally to the proximal part of the implant. Those in group B showed an overall increase in BMD which reached a mean of 12.6% on the lateral side of the distal portion of the implant. Our results support the