Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_16 | Pages 22 - 22
17 Nov 2023
van Duren B Firth A Berber R Matar H Bloch B
Full Access

Abstract. Objectives. Obesity is prevalent with nearly one third of the world's population being classified as obese. Total knee arthroplasty (TKA) is an effective treatment option for high BMI patients achieving similar outcomes to non-obese patients. However, increased rates of aseptic loosening in patients with a high BMI have been reported. In patients with high BMI/body mass there is an increase in strain placed on the implant fixation interfaces. As such component fixation is a potential concern when performing TKA in the obese patient. To address this concern the use of extended tibial stems in cemented implants or cementless fixation have been advocated. Extend tibial stems are thought to improve implant stability reducing the micromotion between interfaces and consequently the risk of aseptic loosening. Cementless implants, once biologic fixation is achieved, effectively integrate into bone eliminating an interface. This retrospective study compared the use of extended tibial stems and cementless implants to conventional cemented implants in high BMI patients. Methods. From a prospectively maintained database of 3239 primary Attune TKA (Depuy, Warsaw, Indiana), obese patients (body mass index (BMI) >30 kg/m²) were retrospectively reviewed. Two groups of patients 1) using a tibial stem extension [n=162] and 2) cementless fixation [n=163] were compared to 3) a control group (n=1426) with a standard tibial stem cemented implant. All operations were performed by or under the direct supervision of specialist arthroplasty surgeons. Analysis compared the groups with respect to class I, II, and III (BMI >30kg/m², >35 kg/m², >40 kg/m²) obesity. The primary outcome measures were all-cause revision, revision for aseptic loosening, and revision for tibial loosening. Kaplan-Meier survival analysis and Cox regression models were used to compare the primary outcomes between groups. Where radiographic images at greater than 3 months post-operatively were available, radiographs were examined to compare the presence of peri-implant radiolucent lines. Results. The mean follow-up of 4.8, 3.4, and 2.5 years for cemented, stemmed, and cementless groups respectively. In total there were 34 all-cause revisions across all the groups with revision rates of 4.55, 5.50, and 0.00 per 1000-implant-years for cemented, stemmed, and cementless groups respectively. Survival Analysis did not show any significant differences between the three groups for all-all cause revision. There were 6 revisions for aseptic loosening (5 tibial and 1 femoral); all of which were in the standard cemented implant group. In contrast there were no revisions in the stemmed or cementless implant groups, however, this was not significant on survival analysis. Analysis looking at class I, II, and III obesity also did not show any significant differences in survival for all cause revision or aseptic loosening. Conclusion. This retrospective analysis showed that there were no revisions required for aseptic loosening when either a cemented stemmed or cementless implant were used in obese patients. These findings are in line with other studies showing that cementless fixation or extended stem implants are a reasonable option in obese patients who represent an increasing cohort of patients requiring TKR. Declaration of Interest. (b) declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research reported:I declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research project


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 741 - 744
1 May 2005
Beaulé PE Krismer M Mayrhofer P Wanner S Le Duff M Mattesich M Stoeckl B Amstutz HC Biedermann R

Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse - femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers.

The accuracy of the method was found to be ± 1.6 mm for the x-direction and ± 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively.

EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.