We reviewed 44 consecutive revision hip replacements in 38 patients performed using the
Aims. We report the incidence of radiolucent lines (RLLs) using two
flanged acetabular components at total hip arthroplasty (THA) and
the effect of the Rim Cutter. Patients and Methods. We performed a retrospective review of 300 hips in 292 patients
who underwent primary cemented THA. A contemporary flanged acetabular
component was used with (group 1) and without (group 2) the use
of the Rim Cutter and the Rimfit acetabular component was used with
the Rim Cutter (group 3). RLLs and clinical outcomes were evaluated
immediately post-operatively and at five years post-operatively. Results. There was no significant difference in the incidence of RLLs
on the immediate post-operative radiographs (p = 0.241) or at five
years post-operatively (p = 0.463). RLLs were seen on the immediate
post-operative radiograph in 2% of hips in group 1, in 5% in group
2 and in 7% in group 3. Five years post-operatively, there were
RLLs in 42% of hips in group 1, 41% in group 2 and in 49% in group
3. In the vast majority of hips, in each group, the RLL was present
in DeLee and Charnley zone 1 only (86%, 83%, 67% respectively).
Oxford and Harris Hip scores improved significantly in all groups.
There was no significant difference in these scores or in the change
in scores between the groups, with follow-up. Conclusion. Despite the Rim Cutter showing promising results in early laboratory
and clinical studies, this analysis of the radiological and clinical
outcome five years post-operatively does not show any advantage
over and above
Revision of a cemented hemiarthroplasty of the
hip may be a hazardous procedure with high rates of intra-operative complications.
Removing well-fixed cement is time consuming and risks damaging
already weak bone or perforating the femoral shaft. The cement-in-cement
method avoids removal of intact cement and has shown good results
when used for revision total hip arthroplasty (THA). The use of
this technique for the revision of a hemiarthroplasty to THA has
not been previously reported. A total of 28 consecutive hemiarthroplasties (in 28 patients)
were revised to a THA using an Exeter stem and the cement-in-cement
technique. There were four men and 24 women; their mean age was
80 years (35 to 93). Clinical and radiographic data, as well as
operative notes, were collected prospectively and no patient was
lost to follow-up. Four patients died within two years of surgery. The mean follow
up of the remainder was 70 months (25 to 124). Intra-operatively
there was one proximal perforation, one crack of the
femoral calcar and one acetabular fracture. No femoral components
have required subsequent revision for aseptic loosening or are radiologically loose. . Four patients with late complications (14%) have since undergone
surgery (two for a peri-prosthetic fracture, and one each for deep
infection and recurrent dislocation) resulting in an overall major
rate of complication of 35.7%. The
The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility. We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening. A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.
The technique of femoral cement-in-cement revision
is well established, but there are no previous series reporting its
use on the acetabular side at the time of revision total hip replacement.
We describe the technique and report the outcome of 60 consecutive
acetabular cement-in-cement revisions in 59 patients at a mean follow-up
of 8.5 years (5 to 12). All had a radiologically and clinically
well-fixed acetabular cement mantle at the time of revision. During
the follow-up 29 patients died, but no hips were lost to follow-up.
The two most common indications for acetabular revision were recurrent
dislocation (46, 77%) and to complement femoral revision (12, 20%). Of the 60 hips, there were two cases of aseptic loosening of
the acetabular component (3.3%) requiring re-revision. No other
hip was clinically or radiologically loose (96.7%) at the latest
follow-up. One hip was re-revised for infection, four for recurrent
dislocation and one for disarticulation of a constrained component.
At five years the Kaplan-Meier survival rate was 100% for aseptic
loosening and 92.2% (95% CI 84.8 to 99.6), with revision for any cause
as the endpoint. These results support the use of cement-in-cement revision on
the acetabular side in appropriate cases. Theoretical advantages
include preservation of bone stock, reduced operating time, reduced
risk of complications and durable fixation.
Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.Aims
Methods
Aims. Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a
Aims. The aim of this study was to report the initial results of the
Exeter V40 stem, which became available in 2000. Patients and Methods. A total of 540 total hip arthroplasties (THAs) were performed
in our unit using this stem between December 2000 and May 2002.
Our routine protocol is to review patients postoperatively and at
one, five, and ten years following surgery. Results. A total of 145 patients (26.9%) died before ten years and of
the remaining 395 stems, 374 (94.7%) remain in situ.
A total of 21 well-fixed stems (5.3%) were revised. Ten were exchanged
using a
Aims. Compared with primary total hip arthroplasty (THA), revision
surgery can be challenging. The cement-in-cement femoral revision
technique involves removing a femoral component from a well-fixed
femoral cement mantle and cementing a new stem into the original
mantle. This technique is widely used and when carried out for the
correct indications, is fast, relatively inexpensive and carries
a reduced short-term risk for the patient compared with the alternative
of removing well-fixed cement. We report the outcomes of this procedure
when two commonly used femoral stems are used. Patients and Methods. We identified 1179 cement-in-cement stem revisions involving
an Exeter or a Lubinus stem reported to the Swedish Hip Arthroplasty
Register (SHAR) between January 1999 and December 2015. Kaplan-Meier
survival analysis was performed. Results. Survivorship is reported up to six years and was better in the
Exeter group (91% standard deviation (. sd). 2.8% versus 85% . sd. 5.0%)
(p = 0.02). There was, however, no significant difference in the
survival of the stem and risk of re-revision for any reason (p =
0.58) and for aseptic loosening (p = 0.97), between revisions in
which the Exeter stem (94% . sd. 2.2%; 98% . sd. 1.6%)
was used compared with those in which the Lubinus stem (95% . sd. 3.2%;
98% . sd. 2.2%) was used. The database did not allow identification
of whether a further revision was indicated for loosening of the acetabular
or femoral component or both. Conclusion. The
Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision necessitate removal of all bone cement from the femur. In our two institutions, we have been using a
Using a
Objectives. Previous studies have evidenced
The aim of this study was to obtain detailed long-term data on the cement-bone interface in patients with cemented stems, implanted using the constrained fixation technique. A total of eight stems were removed together with adjacent bone during post-mortem examinations of patients with well-functioning prostheses. Specimens were cut at four defined levels, contact radiographs were obtained for each level, and slices were prepared for histological analysis. Clinical data, clinical radiographs, contact radiographs and histological samples were examined for signs of loosening and remodelling. The mean radiological follow-up was 9.6 years and all stems were well-fixed, based on clinical and radiological criteria. Contact radiographs revealed an incomplete cement mantle but a complete filling of the medullary canal for all implants. Various amounts of polyethylene particles were evident at the cement-bone interface of seven stems, with no accompanying inflammatory reaction. Cortical atrophy and the formation of an ‘inner cortex’ were confirmed in six of eight stems by contact radiographs and histology, but were only visible on two clinical radiographs. Our results confirm that a complete cement mantle is not essential for the survival of Müller straight stems into the mid term, and support our hypothesis that no benefit to long-term survival can be expected from
The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement. We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a
When Radiolucent lines (RLL) are observed around cemented acetabular components, they may progress and be associated with loosening. We reviewed the incidence and progression of RLLs around the Exeter Contemporary flanged acetabular component and compared our results with other published series. We reviewed a consecutive series of 203 sockets with a minimum 10-year follow-up. Up to date radiographs were reviewed by 2 independent assessors for the presence, location and thickness of RLL and an assessment for loosening/migration was made. Initial post-operative radiographs were examined for any case with RLLs at review. There were no revisions for aseptic loosening. 103 hips remained in situ with a minimum follow up of 10 years. Lucent lines were seen on 37/103 (36%) of hips with a mean follow up of 12.1 years (10.0–13.9 years). In these 37 hips, the lucency was present in one zone in 84%, two zones in 8% and all three zones in 8%. Of the 37 hips with a RLL at minimum 10 years follow up, five exhibited a RLL immediately post-operatively. All 5 of these lines were initially isolated to zone 1 and progressed over the 10 years around at least 1 more zone. Only one line became circumferential, although the cup did not migrate. Compared to previous papers (DeLee & Charnley, Hodgkinson and Garcia-Cimbrelo) the presence of RLL at 10 years is reduced in our series (table 2) and no cup migrated. All RLL seen in cups at both 10 years and immediate post-operatively in our series progressed (table 3), unlike in the previous studies. This reduction in lucent lines may be down to
Numerous studies have evidenced
Introduction. Stem revision with retention of the old cement mantle (cement-in-cement revision) in cases with an intact cement/bone interface is an appealing option. There has been an increasing use of this technique. In 2014 this cement-in- cement technique was used in 10% of all stem revisions reported to the Swedish Hip Arthroplasty Register (SHAR). We analysed the outcome of cement-in-cement stem revisions reported to the SHAR during years 2001–2014. Patients/Materials & Methods. Since 2001 1292 cement-in-cement revisions (study group), performed with a short (≤150 mm) Exeter stem (n=973) or a Lubinus stem (n=319) were reported to the SHAR. Stem revisions, using short Exeter and Lubinus stems without the use of the
Introduction. The Exeter Hip femoral component remains largely unchanged from the original design, introduced in 1970. It is a highly polished, modular, double tapered stem and has undergone various minor modifications to surface, modularity and most recently the taper; changed to the current V40. TM. design in 2000. The effect of any design modification cannot easily be foreseen and greater emphasis is now placed on ensuring appropriate monitoring for such implants. Methods. We present the results of the first 540 V40. TM. Exeter THAs performed in our Centre between December 2000 and May 2002. All patients were reviewed prospectively at 1, 5 and 10 years following surgery. Results. Ten years after surgery there are no revisions for aseptic loosening and no evidence of aseptic loosening in any hip. The fate of every implant is known and all patients remain under review. One hundred and forty-five patients (26.9%) died before 10 years and of the remaining 395 stems, 374 (94.7%) remain in situ, 11 were revised (2.7%). Three were revised for infection, one for recurrent dislocation and one implant failed at the neck of the prosthesis. Ten were exchanged using
Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.Aims
Methods
The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA. A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.Aims
Methods