Advertisement for orthosearch.org.uk
Results 1 - 5 of 5
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 252 - 257
1 Feb 2015
Wafa H Grimer RJ Reddy K Jeys L Abudu A Carter SR Tillman RM

We conducted a case-control study to examine the merit of silver-coated tumour prostheses. We reviewed 85 patients with Agluna-treated (silver-coated) tumour implants treated between 2006 and 2011 and matched them with 85 control patients treated between 2001 and 2011 with identical, but uncoated, tumour prostheses.

In all, 106 men and 64 women with a mean age of 42.2 years (18.4 to 90.4) were included in the study. There were 50 primary reconstructions (29.4%); 79 one-stage revisions (46.5%) and 41 two-stage revisions for infection (24.1%).

The overall post-operative infection rate of the silver-coated group was 11.8% compared with 22.4% for the control group (p = 0.033, chi-square test). A total of seven of the ten infected prostheses in the silver-coated group were treated successfully with debridement, antibiotics, and implant retention compared with only six of the 19 patients (31.6%) in the control group (p = 0.048, chi-square test). Three patients in the silver-coated group (3.5%) and 13 controls (15.3%) had chronic periprosthetic infection (p = 0.009, chi-square test).

The overall success rates in controlling infection by two-stage revision in the silver-coated group was 85% (17/20) compared with 57.1% (12/21) in the control group (p = 0.05, chi-square test). The Agluna-treated endoprostheses were associated with a lower rate of early periprosthetic infection. These silver-treated implants were particularly useful in two-stage revisions for infection and in those patients with incidental positive cultures at the time of implantation of the prosthesis.

Debridement with antibiotic treatment and retention of the implant appeared to be more successful with silver-coated implants.

Cite this article: Bone Joint J 2015;97-B:252–7.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 988 - 992
1 Jul 2013
Glehr M Leithner A Friesenbichler J Goessler W Avian A Andreou D Maurer-Ertl W Windhager R Tunn P

The aims of this study were to evaluate the incidence of local argyria in patients with silver-coated megaprostheses and to identify a possible association between argyria and elevated levels of silver both locally and in the blood. Between 2004 and 2011, 32 megaprostheses with silver coatings were implanted in 20 female and 12 male patients following revision arthroplasty for infection or resection of a malignant tumour, and the levels of silver locally in drains and seromas and in the blood were determined. The mean age of the patients was 46 years (10 to 81); one patient died in the immediate post-operative period and was excluded.

Seven patients (23%) developed local argyria after a median of 25.7 months (interquartile range 2 to 44.5). Patients with and without local argyria had comparable levels of silver in the blood and aspiration fluids. The length of the implant did not influence the development of local argyria. Patients with clinical evidence of local argyria had no neurological symptoms and no evidence of renal or hepatic failure. Thus, we conclude that the short-term surveillance of blood silver levels in these patients is not required.

Cite this article: Bone Joint J 2013;95-B:988–92.


Bone & Joint Research
Vol. 3, Issue 4 | Pages 101 - 107
1 Apr 2014
Edmondson MC Day R Wood D

Objectives

The most concerning infection of allografts and operative procedures is methicillin resistant Staphylococcus aureus (MRSA) and no current iontophoresed antibiotics effectively combat this microbe. It was initially hypothesised that iontophoresis of vancomycin through bone would not be effective due to its large molecular size and lack of charge. The aim of this study was to determine whether this was a viable procedure and to find the optimum conditions for its use.

Methods

An iontophoresis cell was set up with varying concentrations of Vancomycin within the medulla of a section of sheep tibia, sealed from an external saline solution. The cell was run for varying times, Vancomycin concentrations and voltages, to gain information on optimisation of conditions for impregnating the graft. Each graft was then sectioned and dust ground from the exposed surface. The dust was serially washed to extract the Vancomycin and concentrations measured and plotted for all variables tested.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1425 - 1431
1 Oct 2013
Hardes J Henrichs MP Gosheger G Gebert C Höll S Dieckmann R Hauschild G Streitbürger A

We evaluated the clinical results and complications after extra-articular resection of the distal femur and/or proximal tibia and reconstruction with a tumour endoprosthesis (MUTARS) in 59 patients (mean age 33 years (11 to 74)) with malignant bone or soft-tissue tumours. According to a Kaplan–Meier analysis, limb survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection was the most common indication for subsequent amputation (eight patients). Survival of the prosthesis without revision was 48% (95% CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years post-operatively. Failure of the prosthesis was due to deep infection in 22 patients (37%), aseptic loosening in ten patients (17%), and peri-prosthetic fracture in six patients (10%). Wear of the bearings made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal Tumor Society score was 23 (10 to 29). An extensor lag > 10° was noted in ten patients (17%).

These results suggest that limb salvage after extra-articular resection with a tumour prosthesis can achieve good functional results in most patients, although the rates of complications and subsequent amputation are higher than in patients treated with intra-articular resection.

Cite this article: Bone Joint J 2013;95-B:1425–31.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 924 - 928
1 Jul 2008
Tsuchiya H Shirai T Morsy AF Sakayama K Wada T Kusuzaki K Sugita T Tomita K

We studied the safety of external fixation during post-operative chemotherapy in 28 patients who had undergone distraction osteogenesis (17, group A) or vascularised fibular grafting (11, group B) after resection of a tumour. Four cycles of multi-agent post-operative chemotherapy were administered over a mean period of 14 weeks (6 to 27). The mean duration of external fixation for all patients was 350 days (91 to 828). In total 204 wires and 240 half pins were used.

During the period of post-operative chemotherapy, 14 patients (11 in group A, 3 in group B) developed wire- and pin-track infection. A total of ten wires (4.9%) and 11 half pins (4.6%) became infected. Seven of the ten infected wires were in periarticular locations.

External fixation during post-operative chemotherapy was used safely and successfully for fixation of a vascularised fibular graft and distraction osteogenesis in 27 of 28 patients. Post-operative chemotherapy for malignant bone tumours did not adversely affect the ability to achieve union or cause hypertrophy of the vascularised fibular graft and had a minimal effect on distraction osteogenesis. Only one patient developed osteomyelitis which required further surgery.