Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 10 - 10
1 Aug 2018
Hooper G Gilchrist N Maxwell R Frampton C
Full Access

Stress shielding has been a well-recognised problem with uncemented femoral components resulting in proximal bone loss and dysfunction, but less attention has been paid to the preservation of acetabular bone stock. Uncemented acetabular components often demonstrate reduced bone density on plain radiographs in the mid-portion of the cup (zone 2), which may be due to the rigidity of the outer shell. This study compares the change in bone density around three different cups with varying moduli of elasticity at a minimum of 2 years. Our hypothesis was that less rigid cups would be associated with improved bone density and less stress shielding. This prospective randomised controlled trial compared the bone mineral content (BMC) adjacent to three different cups with marked differences in stiffness. Cup A was an all titanium shell, cup B was a titanium coated all polyethylene implant and cup C was a tantalum backed shell. All articulations used a 32mm ceramic femoral head. Cup B used polyethylene modified by treatment with vitamin E whereas cups A and C used a liner made of irradiated cross linked polyethylene. Five regions of interest (ROI) were established adjacent to the cup, regions 2, 3 and 4 where similar to the DeLee and Charnley regions 1, 2 and 3. Bone density was measured using IDXA preoperatively, postoperatively, 6 months, 1 and 2 years and compared for each ROI and implant. Precision measurements showed significant reliability. All areas showed a reduction in BMC following insertion of the acetabular cup. Bone loss was less in ROI 1 and 4 in the area of rim fit for all cups and the maximal bone loss was seen in ROI 2 and 3 at the dome of the cup. The more elastic cup (Cup B) produced the least bone loss in this area (p<0.05). Cup C produced the largest bone loss at ROI 2 (40%) which continued increasing at 2 years. Cup stiffness is related to bone loss adjacent to the acetabulum, presumably due to a similar process of stress shielding as seen in the femur. All cups produced similar changes at the periphery of the cup but the more elastic cup retained bone density beneath the cup which continued past 2 years. This improvement in bone quality is likely to be associated with better acetabular bone stock into the future and more reliable long term cup fixation


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 880 - 886
1 Jul 2017
Mohaddes M Shareghi B Kärrholm J

Aims. The aim of this study was to compare the incidence of aseptic loosening after the use of a cemented acetabular component and a Trabecular Metal (TM) acetabular component (Zimmer Inc., Warsaw, Indiana) at acetabular revision with bone impaction grafting. Patients and Methods. A total of 42 patients were included in the study. Patients were randomised to receive an all- polyethylene cemented acetabular component (n = 19) or a TM component (n = 23). Radiostereometric analysis and conventional radiographic examinations were performed regularly up to two years post-operatively or until further revision. Results. The proximal migration was significantly higher in the cemented group. At two years, the median proximal migration was 1.45 mm and 0.25 mm in the cemented and TM groups, respectively (p = 0.02). One cemented component was revised due to dislocation. There were no revisions in the TM group. Conclusion. Lower proximal migration in the TM group suggests that this design might be associated with a lower risk of aseptic loosening in the long term compared with an all polyethylene cemented component. Longer follow-up is required to confirm the clinical advantages of using this component at acetabular revision. Cite this article: Bone Joint J 2017;99-B:880–6


Bone & Joint Open
Vol. 2, Issue 5 | Pages 278 - 292
3 May 2021
Miyamoto S Iida S Suzuki C Nakatani T Kawarai Y Nakamura J Orita S Ohtori S

Aims

The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA.

Methods

A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 414 - 422
1 Apr 2020
Gwynne-Jones DP Gray AR

Aims

To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem.

Methods

Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or uncemented acetabular component (454) were collected from local records and the New Zealand Joint Registry. A competing risks regression survival analysis was performed with death as the competing risk with adjustments made for age, sex, approach, and bearing.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 25 - 30
1 Jan 2017
Waddell BS Della Valle AG

This review summarises the technique of impaction grafting with mesh augmentation for the treatment of uncontained acetabular defects in revision hip arthroplasty.

The ideal acetabular revision should restore bone stock, use a small socket in the near-anatomic position, and provide durable fixation. Impaction bone grafting, which has been in use for over 40 years, offers the ability to achieve these goals in uncontained defects. The precepts of modern, revision impaction grafting are that the segmental or cavitary defects must be supported with a mesh; the contained cavity is filled with vigorously impacted morselised fresh-frozen allograft; and finally, acrylic cement is used to stabilise the graft and provide rigid, long-lasting fixation of the revised acetabular component.

Favourable results have been published with this technique. While having its limitations, it is a viable option to address large acetabular defects in revision arthroplasty.

Cite this article: Bone Joint J 2017;99-B(1 Supple A):25–30.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 307 - 312
1 Mar 2016
Maggs JL Smeatham A Whitehouse SL Charity J Timperley AJ Gie GA

Aims

We report on the outcome of the Exeter Contemporary flanged cemented all-polyethylene acetabular component with a mean follow-up of 12 years (10 to 13.9). This study reviewed 203 hips in 194 patients. 129 hips in 122 patients are still in situ; 66 hips in 64 patients were in patients who died before ten years, and eight hips (eight patients) were revised. Clinical outcome scores were available for 108 hips (104 patients) and radiographs for 103 hips (100 patients).

Patients and Methods

A retrospective review was undertaken of a consecutive series of 203 routine primary cemented total hip arthroplasties (THA) in 194 patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 730 - 737
1 Jun 2009
Carrington NC Sierra RJ Gie GA Hubble MJW Timperley AJ Howell JR

The first 325 Exeter Universal stems (309 patients) implanted at the originating centre were inserted between March 1988 and February 1990 by a group of surgeons with differing experience. In this report we describe the clinical and radiological results at a mean of 15.7 years (14.7 to 17.3) after operation with no loss to follow-up. There were 97 patients (108 hips) with replacements still in situ and 31 (31 hips) who had undergone a further procedure. With an endpoint of revision for aseptic loosening, the survivorship at 17 years was 100% and 90.4% for the femoral and acetabular component, respectively. The mean Merle D’Aubigné and Postel scores at review were 5.4 (sd 0.97) for pain and 4.5 (sd 1.72) for function. The mean Oxford score was 38.4 (sd 9.8) (0 to 48 worst-to-best scale) and the mean combined Harris pain and function score was 73.2 (sd 16.9). Radiological review showed excellent preservation of bone stock in the proximal femur and no failures of the femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 556 - 561
1 May 2008
Glyn-Jones S McLardy-Smith P Gill HS Murray DW

The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis.

We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years.

After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial.

We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.