Management of pseudotumours associated with MoM THA can be difficult and complications are frequent. The functional outcome of patients after revision surgery may be suboptimal. The objective of this study was to assess our experience with revisions of failed MoM THA due to pseudotumours. 78 hips were diagnosed with pseudotumours in 70 patients following metal-on-metal hip replacements. Of these, 68 MoM THA were revised in 62 patients. Pre operative symptoms, radiographic analysis, metal ion levels, MRI results, intra-operative findings, WOMAC scores, the satisfaction level and the complication rate were recorded. Five patients had a resurfacing arthroplasty as their primary implants while the remaining 63 hips in 57 patients had MoM THA of different brands. The average time between the primary and revision surgery was 69 months (range 15–120). The average age at revision was 59 years (43–87). The mean follow-up was 24 months (range 2–73). 36 patients had minimal one year follow-up. Most lesions consisted of cystic changes and solid lesions were observed in 19 patients. In 57 hips, the pseudotumours were located posteriorly or postero-laterally around the greater trochanter. Intra operatively, muscle necrosis was observed in 15(22%) patients. Most THA cases demonstrated wear and corrosion at the head neck junction of the femoral implants. Thirty-five patients (44.9%) had greater than 50 degrees of cup abduction, including 10 patients (12.8%) with an abduction angle greater than 60 degrees. The average pre operative and postoperative Co ion levels were 27.46 ug/L (range 0.36–145.6) and 2.46 (range 0.4–12.48), respectively. Post revision, a total of 10 hips (14.7%) sustained a dislocation, with seven (10.3%) of them experiencing recurrent dislocations. In 8/10 hips, the femoral head size was 36mm or greater. Revision for dislocation occurred in seven(10.3%) patients. Three(4.4%) deep and one(1.47%) superficial infections occurred and deep infections were re-operated. One(1.47%) fracture of the greater trochanter and one (1.47%) psoas tendinitis did not need revision. Therefore, a total of 10 patients (14.7%) were reoperated. 6 revisions for instability were performed in the first 34 patients, while 1 were done in the last 34 patients. At one year post revision surgery, the mean WOMAC score was 19.68 (range 0–48). In comparison, the mean WOMAC score of the same patients one year after their primary surgery was 8.1 (0–63). Patient satisfaction level of patients one year post revision surgery was 7.61 (range 5–10) compared to 4.15 (range 0–7) pre-revision one. The complication rate after revision of pseudotumours is high. Most re revisions occurred secondary to instability despite the use of larger femoral heads. The functional outcome at one year post revision seems to be lower than that seen after primary THA but similar to other revisions in the literature. Experience in the management of these patients may reduce the complication rate.
Introduction. Adverse reaction to metal debris (ARMD) is an increasingly recognised complication of metal-on-metal hip arthroplasty. A previous study described poor results following revision and recommended early intervention. 1. We determined the outcome of revision for
Introduction. Metal on metal hip resurfacing (MoM HR) is attractive for young active patients. Patients with osteonecrosis of the femoral head (ONFH) are relatively young. HR can be an option of treatment, however, long-term stability of the femoral component is a concern because of the necrotic lesion in the femoral head. There is also a concern of
Background. The failure and subsequent withdrawal of the ASR device in both its resurfacing and THR form has been well documented. The National Joint Registry report of 2010 quoted figures of 12–13% failure at five years. Adverse reaction to metal debris (ARMD) is a poorly understood condition and patients developing severe metal reactions may go unrecognised for sometime. Patients and Methods. In 2004 a single surgeons prospective study of the ASR bearing surface was undertaken. We present the
A-70-year old woman underwent uncomplicated total hip arthroplasty using a titanium modular stem with a 46mm CoCr femoral head, a titanium shell, and a metal linear (Wright Medical Technology). Eight years after implantation, she presented with a painful left hip. A pelvic radiograph revealed adequate positioning of both hip implants without any signs of wear of loosening. CT scanning confirmed the presence of a 5 × 5 cm soft tissue mass in the ilium above the cup component accompanied by the iliac fracture. The patient was diagnosed as having an adverse reaction to metal debris (ARMD) after a metal-on-metal THA and revision was performed. Perioperatively?tissue necrosis and partial destruction of the abductor mechanism were found in the absence of any macroscopic infection. Both the neck trunnion and bore of the head showed slight signs of corrosion. The modular neck was revised with a ceramic 28mm head and a new dual-mobility liner(Zimmer Biomet). The iliac fracture was fixed with a porous trabecular metal augment(Zimmer Biomet). The histopathology of tissue sample revealed extensively necrotic material with focal cellular areas of inflammatory cells containing macrophages and neutrophilas. Metalic debris was also scattered in the necrotic materials. After the revision, the patient was recovered without pain or dislocation, and iliac fracture was well fixed. Instability is a substantial problem in the revision of
This study reports the mid-term results of a large bearing uncemented metal on metal total hip replacement (MOMHTHR) matched series using the Synergy stem and Birmingham modular head in 36 hips (mean follow up 61 months). All patients underwent clinical, metal ion and MRI assessment. Wear analysis was performed on retrieved heads using Redlux non-contact optical profilometry. Seven patients (19%) have undergone revision surgery. All revisions had two or more of either symptoms, high metal ions or an MRI suggestive of an adverse reaction to metal debris (ARMD). There was no evidence of component malposition or impingement. Frank staining of tissues together with high volume dark brown fluid collections were found in all cases. All stems and cups were well fixed. In 4 cases pubic and ischial lysis (adjacent to the inferior fins) was observed. All 7 cases had radiological, intraoperative and histological evidence of
Introduction. We conducted independent wear analysis of retrieved metal on metal (MoM) hip components from around the world. All patients with resurfaced hips who developed adverse reactions to metal debris (ARMD) were found to have increased wear of the bearing surfaces. This was untrue in patients with large diameter (?36mm) MoM total hip replacements. This led us to search for other factors leading to
Aims. Metal-on-metal total hip replacements (MoM THRs) are frequently revised. However, there is a paucity of data on clinical outcomes following revision surgery in this cohort. We report on outcomes from the largest consecutive series of revisions from MoM THRs and consider pre-revision factors which were prognostic for functional outcome. Materials and Methods. A single-centre consecutive series of revisions from MoM THRs performed during 2006–2015 was identified through a prospectively maintained, purpose-built joint registry. The cohort was subsequently divided by the presence or absence of symptoms prior to revision. The primary outcome was functional outcome (Oxford Hip Score (OHS)). Secondary outcomes were complication data, pre- and post-revision serum metal ions and modified Oxford classification of pre-revision magnetic resonance imaging (MRI). In addition, the study data along with demographic data was interrogated for prognostic factors informing on post-revision functional outcome. Results. 180 revisions in 163 patients were identified at a median follow-up of 5.48 (2–11.7) years. There were 152 (84.4%) in the symptomatic subgroup and 28 (15.6%) in the asymptomatic group. Overall median OHS improved from 29 to 37 with revision (P<0.001). Symptomatic patients experienced greater functional benefit (DOHS 6.5 vs. 1.4, p=0.012) compared to asymptomatic patients, though they continued to report inferior outcomes (OHS 36.5 vs 43, p=0.004). The functional outcome of asymptomatic patients was unaffected by revision surgery (pre-revision OHS 41, post-revision OHS 43, p=0.4). Linear regression analysis confirmed use of a cobalt-chrome (CoCr)-containing bearing surface (MoM or metal-on- polyethylene) at revision and increasing BMI were predictive of poor functional outcome (R. 2. 0.032, p=0.0224 and R. 2. 0.039, p=0.015 respectively). Pre- and post-revision serum metal ions and pre-revision MRI findings were not predictive of outcome. The overall complication rate was 36% (n=65) with a re-revision rate of 6.7%. The most common complication was ongoing adverse reaction to metal debris (ARMD, defined as positive post-revision MRI) in 21.1%. The incidence of ongoing
Introduction. The bearing surfaces of ceramic-on-ceramic (CoC) total hip replacements (THR) show a substantially lower wear rate than metal-on-polyethylene (MoP) THR in-vitro. However, revision rates for CoC THR are comparable with MoP. Our hypothesis that an explanation could be adverse reaction to metal debris (ARMD) from the trunnion led us to investigate the wear at both the bearing surfaces and the taper-trunnion interface of a contemporary CoC THR in an in-vitro study. Methods. Three 36mm CoC hips were tested in a hip simulator for 5 million cycles (Mc). BIOLOX. ®. delta ceramic femoral heads were mounted on 12/14 titanium (Ti6Al4V) trunnions. Wear of femoral heads, acetabular liners and trunnions was determined gravimetrically using the analytical balance. Roughness measurements (Sa) were taken on the articulating surfaces (pre and post-test) and on the trunnion surfaces (worn and unworn). Furthermore, Energy Dispersive X-ray Spectroscopy (EDX) was used to identify and quantify the wear debris present in the lubricant using scanning electron microscope (SEM). Results and Discussion. The total volumetric wear was 0.25 mm. 3. for CoC joints and 0.29 mm. 3. for titanium trunnions. The total wear volume of the titanium trunnions was in agreement with an explant study (Kocagoz et al, 2016, CORR) which quantified the volumetric material loss from retrieved trunnions with the total wear ranging from 0.0–0.74 mm. 3. The Sa values, pre-and post-test, for heads were 0.003 ± 0.002 and 0.004 ± 0.001 µm and for liners were 0.005 ± 0.001 and 0.005 ± 0.001 µm. Pre-and post-test measurements for Sa of heads (p = 0.184) and liners (p = 0.184) did not show a statistically significant change. The Sa of the trunnions on the unworn and worn areas showed a statistically significant decrease from 0.558 ± 0.060 to 0.312 ± 0.028 µm respectively (p < 0.001). Analysis of wear debris within the lubricant confirmed the presence of titanium. A recent clinical study (Matharu et al, 2016, BMC Musc Dis) found more
Introduction. Metal-on-polyethylene (MoP) is the most commonly used bearing couple in total hip replacements (THRs). Retrieval studies (Cooper et al, 2012, JBJS, Lindgren et al, 2011, JBJS) report adverse reactions to metal debris (ARMD) due to debris produced from the taper-trunnion junction of the modular MoP THRs. A recent retrospective observational study (Matharu et al, 2016, BMC Musc Dis) showed that the risk of
There have been numerous reports regarding “pseudotumor” associated with hip arthroplasty. We present two reports in which main etiology in the pseudotumor formation was titanium (Ti), but not cobalt-chromium (Co-Cr). We should keep in mind that Ti analysis is essential in some cases. (Case 1) A 68-year-old male presented to our institution because of right hip pain and lower extremity swelling four years after a bipolar hemiarthroplasty. MRI predicted a cystic pseudotumor. However, revision surgical findings showed no apparent cause of
All patients with a MOM THR implanted at our unit were recalled for clinical examination, Oxford hip score, radiographs, MARS MRI and cobalt/chromium ion levels. Patients were sub-classified, determining further follow up: a regular screening programme or referral to a revision arthroplasty surgeon. We identified 299 MOM THR's. 59 patients reported a recent deterioration in symptoms. 67 had an abnormal x-ray. 232 hips underwent MARS MRI: 66 showed mild MOM disease, 31 moderate MOM disease, 17 severe MOM disease. 117 had a substantial fluid collection around the hip. Serum cobalt/chromium ions were raised in 135 patients. 46 had one or both above 7 ppb. 148 THR's were referred to a revision surgeon. 21 have been revised and 14 are awaiting revision. In total 64 hips have been revised. 13 were histologically proven ALVAL, 19 had clinical findings of ALVAL (not histologically confirmed), and 1 non-specific reaction to metal wear. 53 asymptomatic patients had evidence of soft tissue reaction on MRI and/or rising metal ions. Our screening programme identified asymptomatic patients with evidence of a suspicious reaction to their MOM THR. We recommend all patients with a MOM THR are fully screened for
We retrospectively reviewed the mid term outcome of 88 MoM THA in 84 patients and 21Hip Resurfacing using Recap Magnum bearing surface performed during 2006 – 2009. There were 41 males and 47 females in the THA group and 17 males and 4 females in the Hip Resurfacing group. All procedures were performed through a posterior approach. The average head size for the THR group was 46mm and the cup size was 52mm and the average head size for the resurfacing was 50mm and cup size was 56mm respectively. Median age for the THA group was 60 yrs. (28–73) and for the Resurfacing it was 51.5 yrs. (32–62). Average follow up was 76 months for the THA group and 78 months for the Resurfacing group. Average serum cobalt for the THA and the Resurfacing groups were 53.2nmol/l (119) and 30.85 and the Chromium levels were 82.44nmol/l(134.5) and 67.49 respectively. Eight MRI scans showed abnormal fluid collections suspicious of
Introduction. In recent years, an increasing number of reports related to adverse reactions to metal debris (ARMD) following metal-on-metal (MOM) total hip arthroplasty (THA) have been published. Some patients who experience
It's easy to say that hip resurfacing is a failed technology. Journals and lay press are replete with negative reports concerning metal-on-metal bearing failures, destructive pseudotumors, withdrawals and recalls. Reviews of national joint registries show revision risks with hip resurfacing exceeding those of traditional total hip replacement, and metal bearings fare worst among all bearing couples. Yet, that misses the point. Modern hip resurfacing was never meant to replace total hip replacement (THR). It was intended to preserve bone in young patients who would be expected to need multiple revisions due to their youth and high-demand activities. The stated goal of the developers of the Birmingham Hip Resurfacing (BHR) was to delay THR by 10 years. In the two decades that followed the release of BHR, this goal has been met and exceeded. Much has been learned about indications, patient selection, and surgical technique. We now know that this highly specialised, challenging procedure is best indicated in the young, active male with osteoarthritis, as a complementary, not competitive procedure, to THR. Resurfacing has many advantages. First and foremost, it saves bone, on the day of surgery, and over the next several years by preventing stress shielding. Dislocations are very rare. Leg length discrepancy and changes in offset are avoided. Post-operative activity, including heavy manual labor and contact sports, is unrestricted. More normal loading of the femur and joint stability has allowed professional athletes to regain their careers. Femoral side revisions, if necessary, are simple total hips, and dual mobility constructs allow one to keep the socket. Adverse reactions to metal debris (ARMD), including pseudotumors, have generated great concern. Initially described only in women, it was unclear whether the etiology was allergy, toxicity, or inflammation. A better understanding of the wear properties of the bearing, and its relation to size, anteversion, hip dysplasia and metallurgy, along with retrieval analysis, allow us to conclude that it is excessive wear due to edge loading which is the fundamental mechanism for the vast majority of
The use of modular systems adds versatility to the implant system, better restoration of hip biomechanics and lower inventory to the hospital. There have been reports of high metal ions,
Introduction. Metal on metal hip arthroplasty continues to be controversial. Emerging evidence suggests that there are multiple modes of failure, and that the results of revision surgery are influenced by host and implant factors. Methods. This study compares a single surgeon series of hip resurfacings (Birmingham Hip Resurfacing {BHR}) and large diameter metal on metal total hip replacements (LDMOMTHR). Primary outcome measures included survival rates, failure secondary to histologically identified Adverse Reaction to Metal Debris (ARMD), and patient reported outcome measures (Oxford Hip Score {OHS}) following revision. Between 1999 and 2005, 458 BHR and 175 LDMOMTHR were performed. At latest review 43 BHR's (9.4%) and 28 LDMOMTHR's (14%) have been revised. Results. Failure secondary to
The histopathology of periprosthetic tissues has been important to understanding the relationship between wear debris and arthroplasty outcome. In a landmark 1977paper, Willert and Semlitsch (1) used a semiquantitative rating to show that tissue reactions largely reflected the extent of particulate debris. Notably, small amounts of debris, including metal, could be eliminated without “overstraining the tissues” but excess debris led to deleterious changes. Currently, a plethora of terms is used to describe tissues from metal-on-metal (M-M) hips and corroded modular connections. We reviewed the evaluation and reporting of local tissue reactions over time, and asked if a dose response has been found between metal and tissue features, and how the use of more standardized terms and quantitative methodologies could reduce the current confusion in terminology. Methods. The PubMed database was searchedbetween 2000 and 2015 for papers using “metal sensitivity /allergy /hypersensitivity, Adverse Local Tissue Reaction (ALTR): osteolysis, metallosis, lymphocytic infiltration, Aseptic Lymphocytic Vasculitis-Associated Lesions (ALVAL), Adverse Reaction to Metal Debris (ARMD) or pseudotumor/ pseudotumour” as well as metal-on-metal / metal-metal AND hip arthroplasty/replacement. Reports lacking soft tissue histological analysis were excluded. Results. 131 articles describing M-M tissue histology were found. In earlier studies, the terms metal sensitivity / hypersensitivity /allergy implied or stated the potential for a Type IV delayed type hypersensitivity response as a reason for revision. More recently those terms have largely been replaced by broader terms such as ALTR, ALVAL and
As there are many reports describing avascular reactions to metal debris (ARMD) after Metal-on-Metal Hip Arthroplasty (MoMHA), the use of MoMHA, especially hip resurfacing, is decreasing worldwide. In cases of