Background. Pseudotumours have been associated with metal-on-metal (MoM) hip replacements. We define it as a solid mass which may have cystic components that is neither neoplastic nor infectious in aetiology. The cause of a pseudotumour is not fully understood but could be due to excessive wear, metal hypersensitivity or due to an as-yet unknown cause. Aim. We present the retrieval analysis of early failure MoM hips revised for pain, loosening or a symptomatic mass. Tissues and implants were examined for the possible causes of failure and pseudotumour formation. Corrosion as a potential new cause for pseudotumour formation will be presented. Methods. A group of 16 MoM hip replacements were collected for retrieval analysis. Six of which had a pseudotumour. An Artificial Hip Profiler (Redlux Ltd) was used to measure wear. Edge loading was determined using the 3D wear data. Tissues were histologically evaluated using a 10-point
Hip resurfacing arthroplasty is an established and effective intervention for osteoarthritis of the hip in the young active patient, relying on the principle of femoral bone-stock preservation. A recognised mode of failure is neck thinning leading to radiological evidence of neck collapse and clinical failure. We report on a series of these slow-neck-failure patients and highlight the increased incidence of this phenomenon in post-menopausal female patients. This is a single operator, single implant series; 172 cases were identified from databases at our institution. 76 were female, mean (SD) age 52 (7) years. 96 were male, mean (SD) age 51(12) years. 15 (8.7%) patients required revision. 12 (80%) were female, 9 (75%) of these were due to slow neck failure. In the men one patient developed
Early failure of metal-on-metal (MoM) total hip replacements (THR) is now well established. We review 93 consecutive patients with CPT¯ stems MoM THR. Our series demonstrates a new mechanism of failure, which may be implant combination specific. Between January 2005 and June 2009, 93 consecutive MoM total hip replacements were preformed using CPT stems by 3 surgeons at our unit. 73 CPT¯ stems, Metasul¯ Large Diameter Heads (LDH) with Durom¯ acetabulae and 20 CPT¯ stems, Metasul¯ 28mm diameter heads in Allofit¯ shells (zimmer). Clinical outcomes were collected prospectively before surgery, at 3 months, 1 year, 2 years, 3 years, and at 5 years post surgery. Revision for any cause was taken as the primary endpoint and the roentgenograms and explanted prostheses were analyzed for failure patterns. In the LDH/Durom¯ group a total of 13 (18%) patients required revision (figs. 1) at a median of 35 months (range 6-44). 6 (8%) for periprosthetic fracture. All 6 periprostethic fractures were associated with minimal or no trauma and all had