Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of
Aims. Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for
Background. Adverse local tissue reactions (ALTR) in metal-on-polyethylene (MoP) total hip arthroplasty (THA) with head-neck taper corrosion is likely to be multifactorial involving implant and patient factors. However, there is a paucity of clinical data on implant parameters as predisposing factors in MoP head-neck taper corrosion. The aim of this study was to identify any potential implant factors associated with failed MoP THA due to head-neck taper corrosion. Methods. A total of 67 MoP THA patients in two groups was investigated: 1)
Introduction. Modular dual-mobility constructs reduce the risk of dislocation after total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chrome (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the mid-term serum Co and Cr levels after modular dual-mobilities used in revision and complex primary THAs, and specifically included patients revised for
Introduction. Failure of metal on metal (MOM) total hip arthroplasty (THA) and resurfacing arthroplasty (HRA) due to development of adverse local tissue reaction (ALTRs) is a significant problem. The prevalence of ALTRs in asymptomatic MOM arthroplasty patients is highly variable. The purpose of this prospective, longitudinal study was to: 1) determine MRI
The differential diagnosis of the painful total
hip arthroplasty (resurfacing or total hip) includes infection,
failure of fixation (loosening), tendinitis, bursitis, synovitis,
adverse local tissue reaction (ALTR) to cobalt-chromium alloys,
and non-hip issues, such as spinal disorders, hernia, gynecologic,
and other pelvic pain. Assuming that the hip is the source of pain,
the first level question is prosthetic or non-prosthetic pain generator?
The second level prosthetic question is septic or aseptic? The third
level question (aseptic hips) is well-fixed or loose?
Recently, corrosion at the head-neck junction in metal-on-polyethylene bearing total hip arthroplasty (THA) has been recognized as a cause of adverse local tissue reactions (ALTR). Serum metal levels have been advocated as a tool for the diagnosis of
Aims. Advances in surgical technique and implant design may influence the incidence and mechanism of failure resulting in revision total hip arthroplasty (rTHA). The purpose of the current study was to characterize aetiologies requiring rTHA, and to determine whether temporal changes existed in these aetiologies over a ten-year period. Methods. All rTHAs performed at a single institution from 2009 to 2019 were identified. Demographic information and mode of implant failure was obtained for all patients. Data for rTHA were stratified into two time periods to assess for temporal changes: 2009 to 2013, and 2014 to 2019. Operative reports, radiological imaging, and current procedural terminology (CPT) codes were cross-checked to ensure the accurate classification of revision aetiology for each patient. Results. In all, 2,924 patients with a mean age of 64.6 years (17 to 96) were identified. There were 1,563 (53.5%) female patients, and the majority of patients were Caucasian (n = 2,362, 80.8%). The three most frequent rTHA aetiologies were infection (27.2%), aseptic loosening (25.2%), and wear (15.2%). The frequency of rTHA for adverse local tissue reaction (ALTR) was significantly greater from 2014 to 2019 (4.7% vs 10.0%; p < 0.001), while the frequency of aseptic loosening was significantly greater from 2009 to 2013 (28.6% vs 21.9%; p < 0.001). Conclusion. Periprosthetic joint infection was the most common cause for rTHA in the current cohort of patients. Complications associated with
Introduction. Modular neck (MN) components in total hip arthroplasty (THA) offer intraoperative flexibility, but adverse local tissue reactions (ALTR) due to tribocorrosion at modular junctions are a potential complication of such designs. Serum ion levels and metal artifact reduction sequence (MARS) MRI are used to assess
Introduction. Adverse local tissue reactions (ALTR) can result in devastating soft tissue and osseous destruction, while potentially increasing the risk of concomitant periprosthetic joint infection (PJI). The aims of this study were to evaluate cobalt (Co) and chromium (Cr) levels generated in simulators from metal-on-polyethylene (MoP) and ceramic-on-polyethylene (CoP) constructs, and determine their impact on native tissues and PJI risk through evaluation of human adipose-derived mesenchymal stem cells (AMSCs) and Staphylococcus epidermidis isolates. Methods. Ten hip simulator constructs were assembled with 36-mm high-offset femoral heads, highly cross-linked polyethylene liners, and titanium stems. Five constructs used CoCr femoral heads and five used ceramic. Constructs were submerged in bovine serum (BS) and run for 1,000,000 cycles. Samples of BS were collected and evaluated for CoCr concentration. Various concentrations of CoCr were chosen for further assessment of cytotoxicity and growth impact on AMSCs and S. epidermidis and compared to inert SiO2. Results. After 1,000,000 cycles, mean MoP and CoP Co concentration was 2264 ng/mL and 0.6 ng/mL, respectively (p<0.001). Mean MoP and CoP Cr concentration was 217 ng/mL and 4.3 ng/mL, respectively (p<0.001). Mean MoP Co:Cr ratio was 10. Co nanoparticles were significantly more toxic to human AMSCs than control SiO2 in a dose-response manner (p<0.001). S. epidermidis growth was not significantly impacted by Co concentrations derived from the simulators. Conclusions. MoP constructs built in ideal conditions generated substantial CoCr debris, highlighting a baseline risk with these implants that may be exacerbated by host factors or imperfect surgical technique. Evaluation of impact on AMSCs suggests that debris levels produced under ideal conditions can be cytotoxic. Additionally, these concentrations did not potentiate or inhibit S. epidermidis growth, suggesting elevated PJI rates with
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Using the database of
a single academic centre, 126 revision THAs in 117 patients using a single
design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome.Aims
Methods
Serum and blood cobalt (Co) and chromium (Cr) ion levels are used to monitor patients at risk for adverse reaction to metal debris (ARMD) following metal-on-metal (MoM) total hip arthroplasty (THA). However, these levels often do not correlate with the degree of local soft-tissue reaction and damage observed at the time of revision. The purpose of this study was to analyze synovial fluid metal ion concentrations in patients with a failed THA in the setting of an ARMD and determine if these levels can be more predictive of soft-tissue destruction than serum or whole blood levels. Synovial, blood and serum samples were prospectively collected from patients undergoing revision THA with ARMD (n=29) and those undergoing aseptic revision without ARMD (n=29). There was no difference in mean age (P=0.50), BMI (P=0.18), sex distribution (P=0.18), serum creatinine (P=0.74), or time to revision THA (P=0.13) between the cohorts. In the AMRD cohort, the components included MoM THA (n=18), hip resurfacing (n=5), dual-modular taper THA (n=4) and MoM and dual-modular taper THA (n=2). At the time of revision THA, 26 (90%) patients in the metal reaction cohort had gross evidence of metallosis in the soft-tissues, the remaining 3 (10%) had evidence of corrosion of the dual taper neck or MoM bearing. In the non ARMD cohort the bearing surfaces included metal-on-polyethylene (n=19) and ceramic-on-polyethylene (n=10). The indications for revision included isolated acetabular loosening (n=11), isolated femoral component loosening (n=11), polyethylene wear (n=5), recurrent dislocation (n=1) and combined femoral and acetabular component loosening (n=1). None had a clinical diagnosis or gross evidence of taper corrosion. Pre-revision, 21 (72%) patients in the metal reaction group had periarticular fluid collections or a mass on MRI. Mean cyst size was 202.9±71.6 cm3 and masses were grouped into Type I (cyst wall <3 mm, n=10), Type II (cyst wall ≥3 mm, n=8) and Type III (mainly solid, n=3). At the time of revision THA, the mean Co levels were elevated in patients with ARMD compared to those without in synovial fluid (1,833 ppb vs. 12.3 ppb, P=0.008), whole blood (22.6 ppb vs 0.5 ppb, P=0.005)) and serum (19.6 vs. 0.6, P=0.001). Likewise, mean Cr levels were significantly elevated in patients with an ARMD compared to those undergoing revision without in synovial fluid (3,128 ppb vs. 10.3 ppb, P=0.01), whole blood (8.9 ppb vs. 0.5 ppb, P=0.009) and serum (14.1 ppb vs. 0.5 ppb, P=0.005). The synovial fluid Co levels were the most accurate test for detecting pseudotumor (AUC 0.951) and adverse local tissue reaction (AUC 0.826). At a synovial fluid Cr level of 110 ppb, the synovial fluid metal ion analysis was 94% sensitive and 86% specific for pseudotumor formation. In this prospective study, synovial fluid analysis of metal ion levels was more accurate in predicting the presence and extent of pseudo-tumor or
This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection. A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.Aims
Methods
Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA. This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m2, and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported.Aims
Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
This study aimed to determine if macrophages can attach and directly affect the oxide layers of 316L stainless steel, titanium alloy (Ti6Al4V), and cobalt-chromium-molybdenum alloy (CoCrMo) by releasing components of these alloys. Murine peritoneal macrophages were cultured and placed on stainless steel, CoCrMo, and Ti6Al4V discs into a 96-well plate. Cells were activated with interferon gamma and lipopolysaccharide. Macrophages on stainless steel discs produced significantly more nitric oxide (NO) compared to their control counterparts after eight to ten days and remained elevated for the duration of the experiment.Aims
Methods
Leucocyte esterase (LE) has been shown to be an accurate marker of prosthetic joint infection (PJI), and has been proposed as an alternative to frozen section (FS) histology for intraoperative diagnosis. In this study, the intraoperative assessment of LE was compared with FS histology for the diagnosis of prosthetic hip infection. A total of 119 patients undergoing revision total hip arthroplasty (THA) between June 2015 and December 2017 were included in the study. There were 56 men and 63 women with a mean age of 66.2 years (27 to 88). Synovial fluid was collected before arthrotomy for the assessment of LE using enzymatic colourimetric strips. Between five and six samples were stained with haematoxylin and eosin for FS histology, and considered suggestive of infection when at least five polymorphonuclear leucocytes were found in five high-power fields.Aims
Patients and Methods
Dual mobility (DM) bearings are an attractive treatment option to obtain hip stability during challenging primary and revision total hip arthroplasty (THA) cases. The purpose of this study was to analyze data submitted to the American Joint Replacement Registry (AJRR) to characterize utilization trends of DM bearings in the USA. All primary and revision THA procedures reported to AJRR from 2012 to 2018 were analyzed. Patients of all ages were included and subdivided into DM and traditional bearing surface cohorts. Patient demographics, geographical region, hospital size, and teaching affiliation were assessed. Associations were determined by chi-squared analysis and logistic regression was performed to assess outcome variables.Aims
Methods
This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%).Aims
Patients and Methods