header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 287 - 287
1 Mar 2013
Nowakowski A Roesle I Valderrabano V Widmer K
Full Access

Introduction

After total hip arthroplasty, dislocation is one of the most frequent serious early complications. This occurs in part due to impingement (catching and leverage of the neck-cup on the inlay/cup border). Impingement may also negatively impact long-term outcomes.

Materials and Methods

A preliminary model for an optimised hip endoprosthesis system was developed to offer a mechanical solution to avoid impingement and dislocation. A computer-supported range of motion simulation using parameters of cup anteversion and inclination as well as torsion and CCD shaft angle was then performed to localise areas of anterior and posterior impingement of typical acetabular cups.