header advert
Results 1 - 1 of 1
Results per page:
Applied filters
Knee

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 64 - 64
1 Oct 2020
Moskal JT
Full Access

Introduction

Malalignment of total knee arthroplasty components may affect implant function and lead to decreased survival, regardless of preferred alignment philosophy – neural mechanical axis restoration or kinematic alignment. A common technique is to set coronal alignment prior to adjusting slope. If the guide is not maintained in a neutral position, adjustment of the slope may alter coronal alignment. Different implant systems recommend varying degrees of slope for ideal function of the implant, from 0–7°. The purpose of this study was to quantify the change in coronal alignment with increasing posterior tibial slope comparing two methods of jig fixation.

Methods

Prospective consecutive series of 100 patients undergoing total knee arthroplasty using computer navigation. First cohort of 50 patients had extramedullary cutting jig secured distally with ankle clamp and proximally with one pin and a second cohort of 50 patients with the jig secured distally with ankle clamp and proximally with two pins. The change in coronal alignment was recorded with each degree of increasing posterior slope from 0–7° using computer navigation. Mean coronal alignment and change in coronal alignment was compared between the two cohorts.