header advert
Results 1 - 3 of 3
Results per page:
Applied filters
Hip

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 35 - 35
23 Jun 2023
Lavernia C Patron LP Lavernia CJ Gibian J Hong T Bendich I Cook SD
Full Access

Fracture of contemporary femoral stems is a rare occurrence. Earlier THR stems failed due to design issues or post manufacturing heat treatments that weakened the core metal. Our group identified and analyzed 4 contemporary fractured femoral stems after revision surgery in which electrochemical welds contributed to the failure.

All four stems were proximally porous coated titanium alloy components. All failures occurred in the neck region post revision surgery in an acetabular cup exchange. All were men and obese. The fractures occurred at an average of 3.6 years post THR redo (range, 1.0–6.5 years) and 8.3 years post index surgery (range, 5.5–12.0 years). To demonstrate the effect of electrocautery on retained femoral stems following revision surgery, we applied intermittent electrosurgical currents at three intensities (30, 60, 90 watts) to the polished neck surface of a titanium alloy stem under dry conditions.

At all power settings, visible discoloration and damage to the polished neck surface was observed. The localized patterns and altered metal surface features exhibited were like the electrosurgically-induced damage priorly reported.

The neck regions of all components studied displayed extensive mechanical and/or electrocautery damage in the area of fracture initiation. The use of mechanical instruments and electrocautery was documented to remove tissues in all 4 cases.

The combination of mechanical and electrocautery damage to the femoral neck and stem served as an initiation point and stress riser for subsequent fractures. The electrocautery and mechanical damage across the fracture site observed occurred iatrogenically during revision surgery. The notch effect, particularly in titanium alloys, due to mechanical and/or electrocautery damage, further reduced the fatigue strength at the fractured femoral necks. While electrocautery and mechanical dissection is often required during revision THA, these failures highlight the need for caution during this step of the procedure in cases where the femoral stem is retained.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 19 - 19
1 Oct 2019
Lavernia CJ
Full Access

Introduction

Inpatient rehabilitation services following joint replacement have been estimated to cost over $3 billion/ yr. A shift in reimbursement strategies to bundle payments with the goal of decreasing cost and improving quality has given discharge disposition after joint replacement a front row seat. Our objectives were (1) to establish a correlation between the accuracy of current tools utilized to predict discharge location and (2) compare preoperative and postoperative patient oriented outcomes (POO's) according to discharge disposition.

Methods

188 consecutive total hip arthroplasty (THA) surgeries performed by a single surgeon were prospectively studied. Pre-intervention assessment of the probable and preferred discharge disposition was performed using one of 5 tools (1) experienced surgical coordinator evaluation (2) Risk Assessment and Prediction Tool (RAPT); (3) Charlson Score; (4) ASA Score; and (5) Patient Self-Reported Health Status. Demographic characteristics, Visual Analogue Scale, QWB-7, SF-36, WOMAC, clinical scores were recorded before and after surgery. Correlation between final discharge disposition (home vs rehab) and its predictors was performed. Preoperative and postoperative outcomes were compared, p<0.05 was considered significant.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 59 - 59
1 Oct 2018
Lavernia CJ Alcerro JC
Full Access

Introduction

Deaths due to overdose involving opioids have nearly quadrupled in USA in the last 20 years. Several published studies have shown that preoperative opioid use independently predicts perioperative complications in total joint arthroplasty. Our objectives were to (1) assess preoperative opioid consumption in patients undergoing total hip arthroplasty (THA) and (2) to determine how preoperative opioid use affects patient oriented outcomes (POO's) and hip scores before and after THA.

Methods

54 primary THA by a single surgeon were reviewed. A new patient questionnaire was administered to document preoperative opioid consumption (type, dosage and length of use); preoperatively and postoperatively POO's measurements [visual analog scale (VAS), QWB-7, SF-36, and WOMAC] and hip scores were recorded. Patients were stratified into two groups: (1) Opioid Users and (2) Non-Opioid Users. Oral morphine equivalents (OME) were calculated for opioid using patients using standard methodology.