Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 124 - 124
1 Feb 2017
Li G Dimitriou D Tsai T Park K Kwon Y Freiberg A Rubash H
Full Access

Introduction

An equal knee joint height during flexion and extension is of critical importance in optimizing soft-tissue balancing following total knee arthroplasty (TKA). However, there is a paucity of data regarding the in-vivo knee joint height behavior. This study evaluated in-vivo heights and anterior-posterior (AP) translations of the medial and lateral femoral condyles before and after a cruciate-retaining (CR)-TKA using two flexion axes: surgical transepicondylar axis (sTEA) and geometric center axis (GCA).

Methods

Eleven patient with advanced medial knee osteoarthritis (age: 51–73 years) who scheduled for a CR TKA and 9 knees from 8 healthy subjects (age: 23–49 years) were recruited. 3D models of the tibia and femur were created from their MR images. Dual fluoroscopic images of each knee were acquired during a weight-bearing single leg lunge. The OA knee was imaged again one year after surgery using the fluoroscopy during the same weight-bearing single leg lunge. The in vivo positions of the knee along the flexion path were determined using a 2D/3D matching technique. The GCA and sTEA were determined based on existing methods. Besides the anterior-posterior translation, the femoral condyle heights were determined using the distances from the medial and lateral epicondyle centers on the sTEA and GCA to the tibial plateau surface in coronal plane (Fig. 1). The paired t-test was applied to compare the medial and lateral condyle motion within each group (Healthy, OA, and CR-TKA). Two-way ANOVA followed post hoc Newman–Keuls test was adopted to detect significant differences among the groups. p<0.05 was considered significant.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_22 | Pages 40 - 40
1 Dec 2016
Kwon Y
Full Access

The contemporary metal-on-metal (MoM) bearings were reintroduced due to their lower volumetric wear rates in comparison to conventional metal-on-polyethylene bearings. This has the potential to substantially reduce wear-induced osteolysis as the major cause of failure and greater implant stability with the use of large femoral heads. It has been estimated that since 1996 more than 1,000,000 MoM articular couples have been implanted worldwide. However, with increasing clinical experience, the national joint registries have recently reported the failure rate of THA with MoM bearings to be 2–3 fold higher than contemporary THA with non-metal-on-metal bearings. Moreover, adverse periprosthetic tissue reactions have emerged as an important reason for failure in MoM patients.

A painful MoM hip arthroplasty has various intrinsic and extrinsic causes. As in all painful THA, a thorough clinical history, a detailed physical examination, as well as radiographic and laboratory tests are essential to delineate potential cause(s) of pain in patients with MoM hip arthroplasty. While specialised tests such as metal ion analysis are useful modalities for assessing MoM hip arthroplasty, over-reliance on any single investigative tool in the clinical decision-making process should be avoided. There should be a low threshold to perform a systematic evaluation of patients with MoM hip arthroplasty as early recognition and diagnosis will facilitate the initiation of appropriate treatment prior to significant adverse biological reactions. A systematic risk stratification recommendation, for multiple modes of failure including adverse local tissue reactions, based on the currently available evidence will be presented to optimise management.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_22 | Pages 48 - 48
1 Dec 2016
Kwon Y
Full Access

Since the advent of total hip arthroplasty (THA), there have been many changes in implant design that have been implemented in an effort to improve the outcome of the procedure and enhance the surgeon's ability to reproducibly perform the procedure. Some of these design features have not stood the test of time. However, the introduction of femoral stem head/neck modularity made possible by the Morse taper has now been a mainstay design feature for over two decades. Modularity at the head-neck junction facilitates intraoperative adjustments. ‘Dual Taper’ modular stems in total hip arthroplasty have interchangeable modular necks with additional modularity at the neck and stem junction. This ‘dual taper’ modular femoral stem design facilitates adjustments of the leg length, the femoral neck version and the offset independent of femoral fixation. This has the potential advantage of optimizing hip biomechanical parameters by accurately reproducing the center of rotation of the hip.

More recently, however, there is increasing concern regarding the occurrence of adverse local tissue reactions in patients with taper corrosion, which is emerging as an important reason for failure requiring revision surgery. Although adverse tissue reactions or ‘pseudotumor’ were initially described as a complication of metal-on-metal (MoM) bearings, the presence of pseudotumor in patients with taper corrosion is thought to result from corrosion at the neck-stem taper junction, secondary to reciprocating movement at the modular junction leading to fretting corrosion in a process described as mechanically assisted crevice corrosion (MACC). Therefore, the focus of this presentation is to summarise clinical challenges in diagnosis and treatment of patients with adverse tissue reactions due to taper corrosion and review up-to-date evidence.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 107 - 107
1 May 2016
Kirking B El-Gohary M Kwon Y
Full Access

Measurements of shoulder kinematics during activities of daily living (ADL) can be used to evaluate patient function before and after treatment and help define device testing conditions. However, due to the difficulties of making 3D motion measurements outside of laboratory conditions, there are few reports of measured shoulder 3D kinematics during ADL. The purpose of this study was to demonstrate the feasibility of using wearable inertial measurement units (IMUs) to track shoulder joint angles.

A nonrandom sample of 5 subjects with normal shoulders was selected based on occupation. The occupations were: dental hygienist, primary school teacher, mechanical project engineer, administrative assistant, and retail associate. Subjects wore two OPAL IMUs (APDM, Portland OR) as shown in Figure 1 on the sternum and on the upper arm for approximately 4 hours while at their workplace performing their normal work place activities and then up to 4 hours while off-work.

Orientation angles from IMUs have traditionally been estimated by integrating gyroscope data and calculating inclination angles relative to gravity with accelerometers. A significant problem is that inaccuracies inherent in the measurements can degrade accuracy. In this study, we used an Unscented Kalman Filter (UKF) with IMU output to track shoulder angles. The UKF mitigates the effect of random drift by incorporating domain knowledge about the shoulder normal range of motion, and the gyroscope and accelerometer characteristics into the state-space models. Initially, in the horizontal plane, without gravity measurements from the accelerometer to aid the gyroscope data, there were unacceptable errors in transverse rotation. To mitigate this error, additional constraints were applied to model gyroscope drift and a zero velocity update strategy was included. These additions decreased tracker errors in heading by 63%. The resulting accuracy with the modified tracker in all motion planes was about 2° (Figure 2).

Subjects commented that the IMUs were well tolerated and did not interfere with their ability to perform tasks in a normal manner. The overall averaged 95th percentile angles (Figure 3) were: flexion 128.8°, adduction 128.4°, and external rotation 69.5°. These peaks angles are similar to other investigator's reports using laboratory simulations of ADL tasks measured with optical and electromagnetic technologies, though this study's observations did show 17% greater extension and 40% greater adduction. Additionally, in these observations, occurrences of maximal internal rotation were rare compared to maximal external rotation and when maximum external rotation did occur, it was in combination with an average flexion angle of 103°. Finally, by performing a Fourier transform of the arm angles and using the 50th percentile frequency the number of arm cycles in a 10 year period was calculated at over 600,000 cycles.

Application of the UKF with the additional drift correction made substantial improvements in shoulder tracking performance and this feasibility data suggests that IMUs with the UKF are suitable for extended use outside of laboratory settings. The motion data collected provides a novel description of arm motion during ADLs including estimating the cycle count of the upper arm at more than 600,000 cycles over 10 years.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 82 - 82
1 Jan 2016
Nebergall A Greene M Sillesen N Rubash HE Kwon Y Malchau H
Full Access

Introduction

Osteolysis caused by wear of the ultrahigh molecular weight polyethylene (UHMWPE) often leads to failure. Cross-linking improves wear, but also produces residual free radicals that decrease oxidative stability. In vitro studies have shown that the anti-oxidative properties of vitamin E UHMWPE stabilize free radicals while retaining the physical and chemical properties of UHMWPE. The porous surface of the Regenerex™ shell was developed for improved bone in-growth fixation. The increased porosity of the Regenerex™ shell promotes early bony in-growth with the goal of greater long-term stability. The purpose of this study was to evaluate vitamin E infused polyethylene (VEPE) wear and stability of acetabular and femoral components using RSA.

Methods

58 patients (64 observed hips), all with osteoarthritis, gave informed consent to participate in a 5 year RSA study. Each patient received a VEPE liner, a Regenerex™ acetabular shell, and an uncemented stem with either a 32mm or 36 mm cobalt chrome femoral head. Tantalum beads were inserted into the VEPE, the pelvic and the femoral bone to measure head penetration into the polyethylene, and shell and stem stability over time, using RSA. RSA radiographs were scheduled immediately postoperatively (up to 6 weeks) and 6 months, 1, 2, 3, and 5 years after surgery. The Wilcoxon signed-ranks nonparametric test was used to determine if changes in penetration or migration were significant over time at p≤0.05.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_16 | Pages 22 - 22
1 Oct 2014
Li G Tsai T Dimitriou D Kwon Y
Full Access

Combined acetabular and femoral anteversion (CA) of the hip following total hip arthroplasty (THA) is critical to the hip function and longevity of the components. However, no study has been reported on the accuracy in restoration of CA of the hip after operation using robotic assistance and conventional free-hand techniques. The purpose of this study was to evaluate if using robotic assistance in THA can better restore native CA than a free-hand technique.

Twenty three unilateral THA patients participated in this study. Twelve of them underwent a robotic-arm assisted THA (RIO® Robotic Arm Interactive Orthopedic System, Stryker Mako., Fort Lauderdale, FL, USA) and eleven received a free-hand THA. Subject specific 3D models of both implanted and non-implanted hips were reconstructed using post-operative CT scans. The anteversion and inclination of the native acetabulum and implanted cup were measured and compared. To determine the differences of the femoral anteversion between sides, the non-implanted native femur was mirrored and aligned with the remaining femur of the implanted side using an iterative closest point algorithm. The angle between the native femoral neck axis and the prosthesis neck axis in transverse plane was measured as the change in femoral anteversion following THA. The sum of the changes of the acetabular and femoral anteversion was defined as the change of CA after THA. A Wilcoxon signed rank test was performed to test if the anteversion of the navigation and free-hand THAs were different from the contralateral native hips (α = 0.05).

The acetabular anteversion were 22.0°±7.4°, 35.9°±6.5° and 32.6°±22.6° for the native hips, robotic assisted THAs and free-hand THAs, respectively, and the corresponding values of the acetabular inclinations were 52.0°±2.9°, 35.4°±4.4° and 43.1°±7.1°. The acetabular anteversion was increased by 12.2°±11.1° (p=0.005) and 12.5°±20.0° (p=0.102) for the robotic assisted and the free-hand THAs. The femoral anteversion was increased by 6.3°±10.5° (p=0.077) and 11.0°±13.4° (p=0.014) for the robotic assisted and free-hand THAs, respectively. The CA were significantly increased by 18.5°±11.7° (p<0.001) and 23.5°±26.5° (p=0.019) for the robotic assisted and the free-hand THAs. The changes of the CA of the free-hand THAs varied in a larger range than those of the robotic assisted THAs.

This study is the first to evaluate the changes in acetabular and femoral anteversions of the hips after robotic assisted and free-hand THAs using the contralateral native hip as a control. The results demonstrate that both the navigation and free-hand THAs significantly increased the CA compared to the contralateral native hips, but the changes of the robotic assisted THAs (18.5°±11.7°) were smaller and varied less than those of the free-hand THAs (23.5°±26.5°). These data suggest that the robotic assisted THA can better restore the native hip CAs with higher repeatability than the free-hand technique. Further studies are needed to investigate the effects of the hip anteversion changes on the in-vivo function of the hip and the long-term outcomes in THA patients.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 118 - 118
1 Dec 2013
Li G Li J Hosseini A Kwon Y Rubash HE
Full Access

Due to technology advancement, many studies have reported on in-vivo human knee kinematics recently (Dannis, 2005; Moro-oka, 2008; Tashman, 2003; Koo, 2008). This abstract summarized the joint kinematics during three motions usually seen in our daily living, i.e. gait, step-up (stair ascending) and single-legged lunge that was measured using a combined dual-fluoroscopic imaging system and MRI based modeling technique (Li, 2008). Cartilage contacts or condylar motion using transepicondylar axis (TEA)/geometric center axis (GCA) were used to describe the motion characters of the knee during these motions.

In the treadmill gait, the movement of the medial femoral condyle along the anteroposterior direction was significantly greater than that of the lateral femoral condyle during the stance phase using either TEA (9.7 ± 0.7 mm vs. 4.0 ± 1.7 mm, respectively; p < 0.01; Fig. 1A) or GCA (17.4 ± 2.0 mm vs. 7.4 ± 6.1 mm, respectively; p < 0.01; Fig. 1B). A “lateral-pivoting” of the knee was observed (Kozanek, 2009).

In the step-up motion, both medial and lateral contact points moved anteriorly on the tibial articular surfaces along the step-up motion path. The contact points on the medial and lateral tibial plateau moved anteriorly (13.5 ± 3.2 mm vs. 10.7 ± 5.0 mm, respectively; p > 0.05; Fig. 2A) with knee extension. Using the TEA (Fig. 2B), the femoral condylar motions presented a similar pattern as the contact points; nonetheless, using the GCA (Fig. 2C), the femoral condylar motion pattern was dramatically different. The medial condyle moved anteriorly, while the lateral condyle shifted posteriorly. However, none of them showed a significant pivoting phenomenon (Li, 2013).

In the single-legged lunge, both medial and lateral contact points moved similarly before 120° of knee flexion, but the lateral contact moved posteriorly and significantly more than the medial compartment in high flexion (1.9 ± 2.1 mm vs. 4.8 ± 2 mm, respectively; p < 0.05). The single-legged lunge didn't show a single motion pattern (Fig. 3) (Qi, 2013).

These data provide baseline knowledge for the understanding of normal physiological function of the knee during gait, step-up and lunge activities. The findings of these studies demonstrated that knee joint kinematics is activity-dependent and indicated that the knee joint motions could not be described using a single motion character such as “medial-pivoting” that has recently been popularized in total knee arthroplasty design areas.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 234 - 234
1 Dec 2013
Barr C Nebergall A Scarborough D Braithwaite G Kwon Y Rubash HE Muratoglu O Malchau H
Full Access

Introduction:

Acetabular cup position is an important factor in successful total hip arthroplasty (THA). Optimal cup placement requires surgeons to possess an accurate perception of pelvic orientation during cup impaction, however, varying pelvic anatomy and limited visual cues in the surgical field may interfere with this process. The purpose of this study was to evaluate the utility of an inertial measurement unit (IMU) in monitoring pelvic position during THA.

Materials & Methods:

Ten patients scheduled to undergo THA were IRB-approved and consented by four surgeons. A small IMU was placed over the patient's sacrum pre-operatively and zeroed in standing position. Pelvic orientation data was streamed and captured wirelessly throughout the procedure. Surgeons were blinded to all data throughout the study period. Prior to cup impaction, the surgeon indicated his intended cup abduction angle and the degree to which the cup impactor was manipulated to compensate for perceived AP pelvic tilt. The degree of pelvic tilt as determined by the IMU (angle β) was then recorded (Figure 1). AP-pelvis radiographs were measured in Martell Hip Analysis Suite post-operatively to calculate the cup abduction angle, which was then compared to the surgeon's intended abduction angle to determine surgeon accuracy. To predict the final cup abduction angle, the degree of pelvic tilt recorded by the IMU (angle β) was subtracted from the abduction angle of the cup impactor (angle α) that was positioned using the OR table as a reference (Figure 1). This value was then compared to the measured post-operative cup abduction angle in order to assess the accuracy of the IMU in measuring pelvic tilt. Surgeon accuracy and IMU accuracy were compared to determine if the IMU was more or less effective than surgeon perception at determining pelvic tilt.