header advert
Results 1 - 3 of 3
Results per page:
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 38 - 38
2 Jan 2024
Frese J Schulz A Kowald B Gerlach U Frosch K Schoop R
Full Access

In a consecutive retrospective analysis of 190 patients treated with the Masquelet technique at the BG Klinikum Hamburg from January 2012 to January 2022, defect-specific features such as the extent and morphology of the defect were recorded, and their influence on the time to reach full weight-bearing of the affected limb was investigated.

A total of 217 defects were treated in 190 patients using the Masquelet technique. 70% of all defects were located in the tibia, followed by 22% in the femur and only about 7% in the upper extremity. The average length of all defects was 58 mm (+/−31 mm), with the largest defect measuring 180 mm and the smallest measuring 20 mm. 89% of the patients achieved full weight-bearing at the end of therapy. The average time from initiation of therapy to reaching safe full weight-bearing was 589 days. There was a significant correlation between defect length and time to reach full weight-bearing (p = 0.0134). These results could serve as a basis for creating a score for prognostics and evaluation of bone healing after treatment with the Masquelet technique. Additionally, the results could help guide indications for secondary stabilization using internal fixation.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 191 - 197
1 May 2016
Kienast B Kowald B Seide K Aljudaibi M Faschingbauer M Juergens C Gille J

Objectives

The monitoring of fracture healing is a complex process. Typically, successive radiographs are performed and an emerging calcification of the fracture area is evaluated. The aim of this study was to investigate whether different bone healing patterns can be distinguished using a telemetric instrumented femoral internal plate fixator.

Materials and Methods

An electronic telemetric system was developed to assess bone healing mechanically. The system consists of a telemetry module which is applied to an internal locking plate fixator, an external reader device, a sensor for measuring externally applied load and a laptop computer with processing software. By correlation between externally applied load and load measured in the implant, the elasticity of the osteosynthesis is calculated. The elasticity decreases with ongoing consolidation of a fracture or nonunion and is an appropriate parameter for the course of bone healing. At our centre, clinical application has been performed in 56 patients suffering nonunion or fracture of the femur.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 398 - 404
1 Mar 2012
Seide K Aljudaibi M Weinrich N Kowald B Jürgens C Müller J Faschingbauer M

In an interdisciplinary project involving electronic engineers and clinicians, a telemetric system was developed to measure the bending load in a titanium internal femoral fixator. As this was a new device, the main question posed was: what clinically relevant information could be drawn from its application? As a first clinical investigation, 27 patients (24 men, three women) with a mean age of 38.4 years (19 to 66) with femoral nonunions were treated using the system. The mean duration of the nonunion was 15.4 months (5 to 69). The elasticity of the plate-callus system was measured telemetrically until union. Conventional radiographs and a CT scan at 12 weeks were performed routinely, and healing was staged according to the CT scans. All nonunions healed at a mean of 21.5 weeks (13 to 37). Well before any radiological signs of healing could be detected, a substantial decrease in elasticity was recorded. The relative elasticity decreased to 50% at a mean of 7.8 weeks (3.5 to 13) and to 10% at a mean of 19.3 weeks (4.5 to 37). At 12 weeks the mean relative elasticity was 28.1% (0% to 56%). The relative elasticity was significantly different between the different healing stages as determined by the CT scans.

Incorporating load measuring electronics into implants is a promising option for the assessment of bone healing. Future application might lead to a reduction in the need for exposure to ionising radiation to monitor fracture healing.