Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_7 | Pages 30 - 30
1 Jul 2020
Faizan A Zhang J Scholl L
Full Access

Iliopsoas tendonitis after total hip arthroplasty (THA) can be a considerable cause of pain and patient dissatisfaction. The optimal cup position to avoid iliopsoas tendonitis has not been clearly established. Implant designs have also been developed with an anterior recess to avoid iliopsoas impingement. The purpose of this cadaveric study was to determine the effect of cup position and implant design on iliopsoas impingement.

Bilateral THA was performed on three fresh frozen cadavers using oversized (jumbo) offset head center revision acetabular cups with an anterior recess (60, 62 and 66 mm diameter) and tapered wedge primary stems through a posterior approach. A 2mm diameter flexible stainless steel cable was inserted into the psoas tendon sheath between the muscle and the surrounding membrane to identify the location of the psoas muscle radiographically. CT scans of each cadaver were imported in an imaging software. The acetabular shells, cables as well as pelvis were segmented to create separate solid models of each. The offset head center shell was virtually replaced with an equivalent diameter hemispherical shell by overlaying the outer shell surfaces of both designs and keeping the faces of shells parallel. The shortest distance between each shell and cable was measured. To determine the influence of cup inclination and anteversion on psoas impingement, we virtually varied the inclination (30°/40°/50°) and anteversion (10°/20°/30°) angles for both shell designs.

The CT analysis revealed that the original orientation (inclination/anteversion) of the shells implanted in 3 cadavers were as follows: Left1: 44.7°/23.3°, Right1: 41.7°/33.8°, Left2: 40/17, Right2: 31.7/23.5, Left3: 33/2908, Right3: 46.7/6.3. For the offset center shells, the shell to cable distance in all the above cases were positive indicating that there was clearance between the shells and psoas. For the hemispherical shells, in 3 out of 6 cases, the distance was negative indicating impingement of psoas. With the virtual implantation of both shell designs at orientations 40°/10°, 40°/20°, 40°/30° we found that greater anteversion helped decrease psoas impingement in both shell designs. When we analyzed the influence of inclination angle on psoas impingement by comparing wire distances for three orientations (30°/20°, 40°/20°, 50°/20°), we found that the effect was less pronounced. Further analysis comparing the offset head center shell to the conventional hemispherical shell revealed that the offset design was favored (greater clearance between the shell and the wire) in 17 out of 18 cases when the effect of anteversion was considered and in 15 out of 18 cases when the effect of inclinations was considered.

Our results indicate that psoas impingement is related to both cup position and implant geometry. For an oversized jumbo cup, psoas impingement is reduced by greater anteversion while cup inclination has little effect. An offset head center cup with an anterior recess was effective in reducing psoas impingement in comparison to a conventional hemispherical geometry. In conclusion, adequate anteversion is important to avoid psoas impingement with jumbo acetabular shells and an implant with an anterior recess may further mitigate the risk of psoas impingement.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 9 - 9
1 Apr 2018
Shao H Chen A Chen C Faizan A Scholl D
Full Access

Purpose

Tibial and femoral component overhang in total knee arthroplasty (TKA) is a source of pain, thus is it important to understand anatomic differences between races to minimize overhang by matching the tibial and femoral shaft axis to the knee articular surface. Thus, this study compared knee morphology between Caucasian and East Asian individuals to determine the optimal placement of tibial and femoral stems.

Methods

A retrospective study was conducted on a matched cohort of 50 East Asians (21F, 29M) and 50 Caucasians (21F, 29M) by age and gender. CT scans were obtained in healthy volunteers using <2mm slices. The distance from the proximal tibial diaphysis axis to the tibial plateau center was measured, and the distance from the distal femoral diaphysis axis to the center of distal femoral articular surface was measured. Tibial measurements were made using Akagi's AP axis and the widest ML diameter, and femoral measurements were based on Whiteside's line and the surgical epicondylar axis.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 106 - 106
1 Mar 2017
Yanoso-Scholl L Pierre D Lee R Ambrosi M Swaminathan V Faizan A TenHuisen K
Full Access

Introduction

In hip arthroplasty, it has been shown that assembly of the femoral head onto the stem remains a non-standardized practice and differs between surgeons [1]. Pennock et al. determined by altering mechanical conditions during seating there was a direct effect on the taper strength [2]. Furthermore, Mali et al. demonstrated that components assembled with a lower assembly load had increased fretting currents and micromotion at the taper junction during cyclic testing [3]. This suggests overall performance may be affected by head assembly method. The purpose of this test was to perform controlled bench top studies to determine the influence of impaction force and compliance of support structure (or damping) on the initial stability of the taper junction.

Materials and Methods


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 112 - 112
1 Feb 2017
Faizan A Chuang P Aponte C Sharkey P
Full Access

Introduction

Various 2D and 3D surfaces are available for cementless fixation of acetabular cups. The goal of these surface modifications is to improve fixation between the metallic cups and surrounding bone. Radiographs have historically been used to evaluate the implant-to-bone fixation around the acetabular cups. In general, a well fixed cup shows no gaps or radiolucency around the cup's outer diameter. In post-operative radiographs, the presence of progressive radiolucent zones of 2mm or more around the implant in the three radiographic zones is indicative of aseptic loosening, as described by DeLee and Charnley [1]. In this cadaveric study, we investigated the X-ray image characteristics of two different types of acetabular shell surfaces (2D and 3D) to evaluate the implant-to-bone interface in the two designs.

Methods

Six human cadavers were bilaterally implanted with acetabular cups by an orthopaedic surgeon. 2D surface cups (Trident, Stryker, Mahwah, NJ) and 3D surface cups (Tritanium, Stryker, Mahwah, NJ) were randomized between the left and right acetabula. The surgeon used his regular surgical technique (1 mm under reaming) to implant the acetabular cups. The cadavers were sent for X-ray imaging after the operation, Figure 1A. Following the X-ray imaging, the acetabular cups were carefully resected from the cadavers. Enough bone around the cups was retained for analysis of the implant-to-bone interface by contact X-ray. The acetabular cups with the surrounding bone were fixed in 70% isopropyl alcohol for about a week and subsequently embedded in polymethyl methacrylate. The embedded cups were sectioned at 30° intervals using a diamond saw in the coronal plane, as recommended by Engh et al [2], Figure 1B. The sectioning of the samples produced 6 slices of each cup where the implant-bone interface could easily be visualized for evaluation with contact X-ray.