Introduction

While TKA procedures have demonstrated clinical success, medial/lateral overhang of the femoral component in total knee arthroplasty (TKA) of ≥3mm may be associated with an increased risk of knee pain, and distal femoral size may vary across ethnic populations. The aim of this study was to determine and compare the prevalence of femoral component overhang among an inclusive (non-segmented) and Asian-identified (Asian-segmented) population, using a flexible intramedullary-rod, posterior referencing method.

Methods

CT Scans from bilateral lower limbs of skeletally mature subjects (981 inclusive, 267 Asian-identified), without bone pathology were prospectively acquired. Bones were segmented and landmarks were modeled using a flexible intramedullary-rod, posterior referencing method. Femoral components were virtually positioned by aligning the lateral implant edge with the lateral bone edge, where the anterior flange meets the anterior chamfer. Medial and lateral component overhang was measured at three zones: (1) intersection of the anterior flange and anterior chamfer (medial only), (2) anterior chamfer mid-line, and (3) distal face mid-line. The central tendency of the samples was determined using the observed mean and median and the 95% confidence interval.