header advert
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 234 - 234
1 Dec 2013
Barr C Nebergall A Scarborough D Braithwaite G Kwon Y Rubash HE Muratoglu O Malchau H
Full Access

Introduction:

Acetabular cup position is an important factor in successful total hip arthroplasty (THA). Optimal cup placement requires surgeons to possess an accurate perception of pelvic orientation during cup impaction, however, varying pelvic anatomy and limited visual cues in the surgical field may interfere with this process. The purpose of this study was to evaluate the utility of an inertial measurement unit (IMU) in monitoring pelvic position during THA.

Materials & Methods:

Ten patients scheduled to undergo THA were IRB-approved and consented by four surgeons. A small IMU was placed over the patient's sacrum pre-operatively and zeroed in standing position. Pelvic orientation data was streamed and captured wirelessly throughout the procedure. Surgeons were blinded to all data throughout the study period. Prior to cup impaction, the surgeon indicated his intended cup abduction angle and the degree to which the cup impactor was manipulated to compensate for perceived AP pelvic tilt. The degree of pelvic tilt as determined by the IMU (angle β) was then recorded (Figure 1). AP-pelvis radiographs were measured in Martell Hip Analysis Suite post-operatively to calculate the cup abduction angle, which was then compared to the surgeon's intended abduction angle to determine surgeon accuracy. To predict the final cup abduction angle, the degree of pelvic tilt recorded by the IMU (angle β) was subtracted from the abduction angle of the cup impactor (angle α) that was positioned using the OR table as a reference (Figure 1). This value was then compared to the measured post-operative cup abduction angle in order to assess the accuracy of the IMU in measuring pelvic tilt. Surgeon accuracy and IMU accuracy were compared to determine if the IMU was more or less effective than surgeon perception at determining pelvic tilt.